← Product Code [MQG](/productcode/MQG) · K133712

# EMBRYOVIEWER SOFTWARE (K133712)

_Unisense Fertilitech A/S · MQG · Aug 8, 2014 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133712

## Device Facts

- **Applicant:** Unisense Fertilitech A/S
- **Product Code:** [MQG](/productcode/MQG.md)
- **Decision Date:** Aug 8, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 884.6120
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The EmbryoScope® time-lapse incubator (version D) provides an environment with controlled temperature, CO2 (and other gasses) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development. The EmbryoViewer™ software is an optional accessory software package for use in displaying, comparing, storing, and transferring images generated by the EmbryoScope® time-lapse incubator (Version D). The software includes a user annotation function for capturing information on embryo development parameters as well as a user-defined modeling function, which allows the user to combine annotated information on embryo development parameters to aid in embryo selection. The EmbryoViewer™ software does not control any hardware components in the EmbryoScope® time-lapse incubator (Version D).

## Device Story

EmbryoScope time-lapse incubator (Version D) provides controlled environment (temperature, CO2/gases) for embryo development; integrates inverted microscope and imaging system for viewing up to 72 embryos simultaneously. EmbryoViewer software acts as accessory for displaying, storing, and transferring time-lapse images; includes annotation tools for developmental parameters (cleavage, fragmentation, multinucleation, pronuclei) and user-defined modeling function to assist embryologists in embryo selection. System used in clinical embryology labs; operated by embryologists. Software does not control hardware; provides database for patient/treatment data and statistical export. Benefits include continuous monitoring of embryo development without removing from incubator, aiding selection for transfer/freezing.

## Clinical Evidence

No clinical testing performed. Bench testing verified software and gas system components (EGS) against requirements and previous version (GB). Electrical safety, electromagnetic compatibility, and environmental performance tested per IEC 60601-1:2005.

## Technological Characteristics

Tri-gas incubator; integrated inverted microscope; IR-cut off filter; Windows 7 OS; optional barcode reader; load door status LEDs. Connectivity: database for image/data storage and export. Software: user-defined modeling and annotation tools. Sterilization: not specified.

## Regulatory Identification

Assisted reproduction accessories are a group of devices used during assisted reproduction procedures, in conjunction with assisted reproduction needles and/or assisted reproduction catheters, to aspirate, incubate, infuse, and/or maintain temperature. This generic type of device may include:(1) Powered aspiration pumps used to provide low flow, intermittent vacuum for the aspiration of eggs (ova). (2) Syringe pumps (powered or manual) used to activate a syringe to infuse or aspirate small volumes of fluid during assisted reproduction procedures. (3) Collection tube warmers, used to maintain the temperature of egg (oocyte) collection tubes at or near body temperature. A dish/plate/microscope stage warmer is a device used to maintain the temperature of the egg (oocyte) during manipulation. (4) Embryo incubators, used to store and preserve gametes and/or embryos at or near body temperature. (5) Cryopreservation instrumentation and devices, used to contain, freeze, and maintain gametes and/or embryos at an appropriate freezing temperature.

## Special Controls

*Classification.* Class II (special controls) (design specifications, labeling requirements, and clinical testing). The device, when it is a simple embryo incubator with only temperature, gas, and humidity control; a syringe pump; a collection tube warmer; a dish/plate/microscope stage warmer; a controlled-rate cryopreservation freezer; or an assisted reproduction laminar flow workstation is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.

## Predicate Devices

- EmbryoScope time-lapse system ([K113075](/device/K113075.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the word "FertiliTech" in bold black font. To the left of the word is a small, circular graphic that appears to be a stylized image of a seed or plant embryo. The text and graphic are the only elements in the image, and they are centered horizontally.

## 510(k) Summary – EmbryoScope time-lapse incubator (Version D) and

### EmbryoViewer software

#### Administrative information

| Name:             | Unisense FertiliTech A/S                 |
|-------------------|------------------------------------------|
| Address:          | Tueager 1<br>DK-8200 Aarhus N<br>Denmark |
| Contact person:   | Henrik Wahlgren, QA Manager              |
| Telephone Number: | +45 29 33 36 85                          |
| Fax Number:       | +45 89 44 95 49                          |
| E-mail:           | hew@fertilitech.com                      |
| Date of summary:  | 8-August-2014                            |

#### Device name and predicate device name

| Trade name                                                               | Common name                   | Device                                                                            | Class | CFR Reference | Product code | Predicate device K number |
|--------------------------------------------------------------------------|-------------------------------|-----------------------------------------------------------------------------------|-------|---------------|--------------|---------------------------|
| EmbryoScope® time-lapse incubator (Version D) and EmbryoViewer™ software | EmbryoScope time-lapse system | Assisted reproduction accessories and Picture archiving and communications system | II    | 884.6120      | MQG          | 113075                    |

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image contains the logo for "FertiliTech". The logo consists of a stylized graphic to the left of the company name. The graphic is a circular shape with an abstract design inside. The text "FertiliTech" is in a bold, sans-serif font.

K133712 Page 2 of 9

#### Indications for Use EmbryoScope time-lapse incubator (Version D) and EmbryoViewer software

The EmbryoScope® time-lapse incubator (version D) provides an environment with controlled temperature, CO2 (and other gasses) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development.

The EmbryoViewer™ software is an optional accessory software package for use in displaying, comparing, storing, and transferring images generated by the EmbryoScope® time-lapse incubator (Version D). The software includes a user annotation function for capturing information on embryo development parameters as well as a user-defined modeling function, which allows the user to combine annotated information on embryo development parameters to aid in embryo selection. The EmbryoViewer™ software does not control any hardware components in the EmbryoScope® time-lapse incubator (Version D).

#### Device description EmbryoScope® time-lapse incubator (Version D)

The EmbryoScope® time-lapse incubator is a tri-gas incubator, which acquires a series of unattended measurements on individual embryos during their development. The measurements include: time-lapse microscopy at multiple focal planes and logging of incubation conditions. Separate processing units control the incubation environment and data acquisition to ensure safe and reliable operation. Up to 72 embryos (6 EmbryoSlide® culture dishes with 12 embryos in each culture dish) can be incubated simultaneously.

|                     | EmbryoScope                                                                                                                                                                                                                                                                                                                                                         | EmbryoScope (K113075)                                                                                                                                                                                                                                                                                                                         |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use | The EmbryoScope® time-lapse incubator (Version D) provides an environment with controlled temperature, CO2 (and other gases) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development. | The EmbryoScope (Version D) provides an environment with controlled temperature, CO2 (and other gases) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development. |

#### Indications for use EmbryoScope® time-lapse incubator (Version D)

Unisense FertiliTech A/S Tueager 1 DK-8200 Aarhus N Denmark

Page 2 of 9

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the word "FertiliTech" in bold, black font. To the left of the word is a small, abstract graphic that appears to be a stylized depiction of a plant or seed. The text is the primary focus of the image, with the graphic serving as a visual element to complement the brand name.

#### Summary of Performance testing EmbryoScope® time-lapse incubator (Version D)

For the EmbryoScope time-lapse incubator software and gas system component (EGS), the test cases that have been performed to verify and validate requirements are an evaluation of tests performed on previous version of the EmbryoScope time-lapse incubator software and gas system component (GB), with addition of tests related to new features described in this submission. Test results have been evaluated and it has been concluded to accept the test results; the EmbryoScope time-lapse incubator is substantially equivalent to the predicate device.

|                                                                                                                                                                                                                                                           | EmbryoScope                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | EmbryoScope (K113075)                                                                                                                                                                                                                                                                                                                         |
|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for use                                                                                                                                                                                                                                       | The EmbryoScope® timelapse incubator (Version D) provides an environment with controlled temperature, CO2 (and other gases) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development.                                                                                                                                                                                                                                                                                                          | The EmbryoScope (Version D) provides an environment with controlled temperature, CO2 (and other gases) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development. |
| Design features                                                                                                                                                                                                                                           | Gas system component (EGS) providing an environment with controlled temperature, CO2 (and other gases) for the development of embryos.<br><br>An integrated inverted microscope and imaging system for embryo viewing.<br><br>IR-cut off filter enabling reduced exposure times and increased image acquisition frequency                                                                                                                                                                                                                                                                                                                                                   | Gas system component (GB) providing an environment with controlled temperature, CO2 (and other gases) for the development of embryos.<br><br>An integrated inverted microscope and imaging system for embryo viewing.                                                                                                                         |
|                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                               |
| Treatment ID should be<br>specified for all EmbryoSlide<br>culture dishes<br><br>Integrated computer<br>running Microsoft Windows<br>7<br><br>Load door LEDs (red/green)<br>indicating the status of the<br>load door lock<br><br>Optional barcode reader | Integrated computer<br>running Microsoft Windows<br>Vista                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                               |
| Non-Clinical Tests<br>performed for<br>Determination of<br>Substantial Equivalence                                                                                                                                                                        | The tests described in the<br>software test plan for the<br>EmbryoScope time-lapse<br>incubator software and gas<br>system component (EGS),<br>derived from the Software<br>Requirements Specification<br>and technical specification,<br>are an evaluation of tests<br>performed on previous<br>version on the EmbryoScope<br>time-lapse incubator<br>software and gas system<br>component (GB), with<br>addition of tests related to<br>new features described in<br>this submission<br><br>Underwent and passed<br>electrical safety<br>electromagnetic<br>compatibility<br>environmental and<br>operating performance<br>testing in accordance with<br>IEC 60601-1:2005 | Underwent and passed<br>electrical safety<br>electromagnetic<br>compatibility<br>environmental and<br>operating performance<br>testing in accordance with<br>IEC 60601-1:1998                                                                                                                                                                 |
| Clinical Tests performed                                                                                                                                                                                                                                  | No clinical testing was<br>performed.<br><br>Tests with and without<br>mouse embryos have been<br>performed to confirm that                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Clinical data show that the<br>EmbryoScope (Version D)<br>performs according to<br>the intended use of the<br>device.                                                                                                                                                                                                                         |
| system component (EGS) is<br>equivalent to the gas system<br>component (GB) in the<br>predicate device.                                                                                                                                                   | No adverse effects of image<br>acquisition were observed.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                               |
| No adverse effects of image<br>acquisition were observed.                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                             |                                                                                                                                                                                                                                                                                                                                               |

#### Comparison to Predicate Device EmbryoScope® time-lapse incubator (Version D)

Unisense FertiliTech A/S Tueager 1 DK-8200 Aarhus N Denmark

Phone: +45 89 44 95 00 Fax: +45 89 44 95 49 sales@fertilitech.com www.fertilitech.com

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for FertiliTech. The logo consists of a stylized symbol to the left of the word "FertiliTech". The symbol appears to be a stylized representation of a seed or a cell, with a swirling design inside a circular shape. The text "FertiliTech" is written in a bold, sans-serif font.

Page 4 of 9

{4}------------------------------------------------

Image /page/4/Picture/0 description: The image shows the logo for "FertiliTech". The logo consists of a stylized plant or seed sprouting icon to the left of the company name. The text "FertiliTech" is written in a bold, sans-serif font, with each letter clearly legible.

The indications for use for the EmbryoScope time-lapse incubator have not changed as a consequence of the change in design features described above. To the name EmbryoScope, the description time-lapse incubator has been added.

The components in the EmbryoScope time-lapse incubator are considered equivalent to the components in the predicate device included in the submission K113075.

Incubation conditions and image quality are not adversely affected by the above changes.

The addition of the features described above to the EmbryoScope time-lapse incubator does not raise new types of safety and effectiveness questions.

#### Device description EmbryoViewer™ software

The purpose of the EmbryoViewer™ software is to assist the embryologist in selecting embryos for transfer or freezing. This is obtained by allowing the embryologist to inspect high-resolution time-lapse images of embryo development, to use detailed annotation tools and to inspect the running conditions of the EmbryoScope® time-lapse system.

The EmbryoViewer™ software does not perform any diagnostics, but only shows data from the EmbryoScope® time-lapse system and data entered by the user.

The EmbryoViewer™ software provides:

- · High-resolution time-lapse images of single embryos
- Embryo annotation tools which assist the user in selecting embryos
- Option to design a set of models that can be used for scoring embryos
- Inspection of incubation details, e.g. temperature and gas conditions
- Export of data for statistical analysis

The EmbryoViewer™ software is an optional accessory package to the EmbryoScope® timelapse system. The time-lapse system provides a controlled culture environment that allows continuous monitoring and observation of up to 72 embryos at a time.

The data from the EmbryoScope® time-lapse system can be seen in the EmbryoViewer™ software. This includes embryo images, incubation details, alarms, log files and other

Page 5 of 9

{5}------------------------------------------------

fech

instrument parameters. The EmbryoViewer™ software can be used to review and compare synchronized time-lapse movies of various embryos.

In addition, the EmbryoViewer™ software contains a database which enables the user to annotate time-lapse images for e.g. cleavage times, fragmentation, multinucleation, pronuclei and other characteristics used for selecting embryos for transfer. Annotations are saved in the database and can be exported to an Excel file for statistical analysis.

The EmbryoViewer™ software also enables input of patient details, treatment data and outcome data. This data is stored in the database and can be exported to an Excel file complete with annotations.

|                     | EmbryoViewer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   | EmbryoViewer (K113075)                                                                                                                                                                                                                                                                                                                                                                                                                                                   |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use | The EmbryoScope® time-<br>lapse system (version D)<br>provides an environment<br>with controlled temperature,<br>CO2 (and other gasses) for<br>the development of embryos.<br>This model has an integrated<br>inverted microscope and<br>imaging system for embryo<br>viewing. Device use is limited<br>to five days (120 hr) covering<br>the time from post-<br>fertilization to day 5 of<br>development.<br><br>The EmbryoViewer™<br>software is an optional<br>accessory software package<br>for use in displaying,<br>comparing, storing, and<br>transferring images<br>generated by the<br>EmbryoScope® time-lapse<br>incubator (Version D). The<br>software includes a user<br>annotation function for<br>capturing information on<br>embryo development<br>parameters as well as a user-<br>defined modeling function,<br>which allows the user to<br>combine annotated | EmbryoViewer is an optional<br>accessory software package<br>for use in displaying,<br>comparing, storing, and<br>transferring EmbryoScope<br>(Version D) generated<br>images. The software<br>includes a user annotation<br>function for capturing<br>information on embryo<br>development parameters,<br>treatment data, and<br>outcome data. The<br>EmbryoViewer software does<br>not control any hardware<br>components in the<br>EmbryoScope (Version D)<br>device. |

#### Indications for use EmbryoViewer™ software

Page 6 of 9

Phone: +45 89 44 95 00 Fax: +45 89 44 95 49 sales@fertilitech.com www.fertilitech.com

{6}------------------------------------------------

Image /page/6/Picture/0 description: The image shows the logo for "FertiliTech". The logo consists of a stylized graphic to the left of the company name. The company name is in a bold, sans-serif font.

#### Summary of Performance testing EmbryoViewer™ software

For the EmbryoViewer software, the test cases that have been performed to verify and validate requirements are an evaluation of tests performed on previous version on the EmbryoViewer software, with addition of tests related to new features described in this submission. Test results have been evaluated and it has been concluded to accept the test results; the EmbryoViewer software is as safe, as effective and performs as well or better than the EmbryoViewer software included in the submission K113075.

#### Comparison to Predicate Device EmbryoViewer™ software

| Indications for Use                                                                | EmbryoViewer                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | EmbryoViewer (K113075)                                                                                                                                                                                                                                                                                                                                                                                     |
|------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                                                                                    | The EmbryoScope® time-lapse system (version D) provides an environment with controlled temperature, CO2 (and other gasses) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development.                                                                                                                                                                                                               | EmbryoViewer is an optional accessory software package for use in displaying, comparing, storing, and transferring EmbryoScope (Version D) generated images. The software includes a user annotation function for capturing information on embryo development parameters, treatment data, and outcome data. The EmbryoViewer software does not control any hardware components in the EmbryoScope (Version |
|                                                                                    | The EmbryoViewer™ software is an optional                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       |                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 |                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                                                    | comparing, storing, and<br>transferring images<br>generated by the<br>EmbryoScope® time-lapse<br>incubator (Version D). The<br>software includes a user<br>annotation function for<br>capturing information on<br>embryo development<br>parameters as well as a user-<br>defined modeling function,<br>which allows the user to<br>combine annotated<br>information on embryo<br>development parameters to<br>aid in embryo selection. The<br>EmbryoViewer™ software<br>does not control any<br>hardware components in the<br>EmbryoScope® time-lapse<br>incubator (Version D). |                                                                                                                                                                                                                                                                                                                                                                                                            |
| Design features                                                                    | Displaying, comparing,<br>storing, and transferring<br>EmbryoScope time-lapse<br>incubator generated images<br>and an annotation function.<br><br>In the Model Designer, the<br>user can define models<br>applied in the Compare &<br>Select view where embryos<br>are listed together with the<br>score from the applied model                                                                                                                                                                                                                                                 | Displaying, comparing,<br>storing, and transferring<br>EmbryoScope (Version D)<br>generated images and an<br>annotation function.                                                                                                                                                                                                                                                                          |
| Non-Clinical Tests<br>performed for<br>Determination of<br>Substantial Equivalence | The EmbryoViewer software<br>is fulfilling the requirements<br>of the IEC 62304 standard<br>according to software testing.<br><br>The tests described in the<br>software test plan for the<br>EmbryoViewer software,<br>derived from the Software<br>Requirements Specification,<br>are an evaluation of tests<br>performed on previous<br>version on the EmbryoViewer                                                                                                                                                                                                          | The EmbryoViewer software<br>is fulfilling the requirements<br>of the IEC 62304 standard<br>according to software<br>testing.                                                                                                                                                                                                                                                                              |
|                                                                                    | software, with addition of<br>tests related to new features<br>described in this submission                                                                                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                            |
| Clinical Tests performed                                                           | This section does not apply.<br>No clinical testing was<br>performed                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | This section does not apply.<br>No clinical testing was<br>performed                                                                                                                                                                                                                                                                                                                                       |

Substantial equivalence is claimed to legally marketed device and accessory covered by K113075, the EmbryoViewer software from Unisense FertiliTech A/S.

Page 7 of 9

{7}------------------------------------------------

Image /page/7/Picture/0 description: The image shows the logo for FertiliTech. The logo consists of a stylized graphic to the left of the word "FertiliTech". The graphic is a circular design with a small extension at the top.

Page 8 of 9

{8}------------------------------------------------

Image /page/8/Picture/0 description: The image contains the word "FertiliTech" in a sans-serif font. The text is dark gray, and there is a light gray graphic to the left of the word. The graphic appears to be a stylized leaf or plant. The background is white.

The addition of the features models and Compare & Select result in a new Indications for Use for the EmbryoViewer software.

The update of the EmbryoViewer software has no impact on incubation conditions in the EmbryoScope time-lapse incubator.The new features models and Compare & Select in the EmbryoViewer software does not introduce or require changes to the EmbryoScope timelapse incubator.

As the EmbryoViewer software does not control any functions on the EmbryoScope timelapse incubator it can not affect the operation of the EmbryoScope time-lapse incubator. The addition of the features models and Compare & Select to the EmbryoViewer software does not raise new types of safety and effectiveness questions.

#### Conclusion

The subject device is substantially equivalent to the predicate device.

Unisense FertiliTech A/S Tueager 1 DK-8200 Aarhus N Denmark

Page 9 of 9

Phone: +45 89 44 95 00 Fax: +45 89 44 95 49 sales@fertilitech.com www.fertilitech.com

{9}------------------------------------------------

Image /page/9/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines forming its body and head. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

August 8, 2014

Unisense FertiliTech A/S Henrik Wahlgren QA & RA Manager Tueager 1 Aarhus N DK-8200 Denmark

Re: K133712

> Trade/Device Name: EmbryoScope time-lapse incubator (Version D) and EmbryoViewer software Regulation Number: 21 CFR§ 884.6120 Regulation Name: Assisted reproduction accessories Regulatory Class: II Product Code: MQG Dated: July 14, 2014 Received: July 17, 2014

Dear Henrik Wahlgren,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading,

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{10}------------------------------------------------

Page 2 - Henrik Wahlgren

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

# Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{11}------------------------------------------------

# Indications for Use

510(k) Number (if known): K133712

Device Name: EmbryoScope time-lapse incubator (Version D) and EmbryoViewer software

Indications For Use:

The EmbryoScope® time-lapse incubator (version D) provides an environment with controlled temperature, CO2 (and other gasses) for the development of embryos. This model has an integrated inverted microscope and imaging system for embryo viewing. Device use is limited to five days (120 hr) covering the time from post-fertilization to day 5 of development.

The EmbryoViewer™ software is an optional accessory software package for use in displaying, comparing, storing, and transferring images generated by the EmbryoScope® time-lapse incubator (Version D). The software includes a user annotation function for capturing information on embryo development parameters as well as a user-defined modeling function, which allows the user to combine annotated information on embryo development parameters to aid in embryo selection. The EmbryoViewer™ software does not control any hardware components in the EmbryoScope® time-lapse incubator (Version D).

× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Herbert P. Lerner -S 2014.08.08 13:09:58 -04'00'

Page 1 of 1 __________________________________________________________________________________________________________________________________________________________________

---

**Source:** [https://fda-staging.innolitics.com/device/K133712](https://fda-staging.innolitics.com/device/K133712)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).