← Product Code [FER](/productcode/FER) · K133687

# THD N-ANO ANOSCOPE (K133687)

_Thd Spa · FER · Dec 11, 2013 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133687

## Device Facts

- **Applicant:** Thd Spa
- **Product Code:** [FER](/productcode/FER.md)
- **Decision Date:** Dec 11, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The THD N-Ano Anoscope, is intended for physician use to examine the anal sphincter, anus, rectum to perform various diagnostic and therapeutic procedures.

## Device Story

Disposable, non-sterile anoscope with integrated LED light source in handle. Used by physicians in clinical settings to visualize anal sphincter, anus, and rectum for diagnostic/therapeutic procedures. Device features modified handle to house LED circuit components; eliminates separate LED light source block found in predicate. Provides illumination for direct visual examination by clinician. Benefits include improved handle integration and simplified design compared to predicate.

## Clinical Evidence

Bench testing only; no clinical data provided. Substantial equivalence supported by design control activities, risk management, and hazard analysis of physical modifications.

## Technological Characteristics

Disposable anoscope; materials include ABS Terluran (handle/cover). Integrated LED light source circuit. Dimensions: 105 mm length, 21.7 mm diameter. Non-sterile. Manual operation.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Family of THD disposable Anoscopes, Proctoscopes, Rectoscopes and Light-scope ([K103647](/device/K103647.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

THD N-ANO ANOSCOPE

K133687 · page 1 of 3

# 510(k) Summary for the

DEC 1 1 2013

# Family of disposable THD N-ano anoscope

This 510(k) Summary is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92.

### 2.1. General Information Submitter:

THD S.p.A. Via dell'Industria. 1 42015 - Correggio (RE) Italy

Establishment Registration Number: 3006680097

Contact Person:

Guido Bonapace ISEMED srl Via A. Altobelli Bonetti 3/A 40026 Imola (BO) Italy Mob.phone: +39-335-5378686 Telephone: +39-0542 683803 Fax: +39-0542 698456 Email: regulatory@isemed.eu

Summary Preparation Date:

- 2.2. Names
Device Name: Classification Name: Product Code: Regulation number:

October 29, 2013

THD N-ano anoscope Endoscope and accessories FER/GCP 876.1500

#### 2.3. Predicate Devices

This Special 510(k) is related to the device modifications of the following devices:

| Applicant  | Device name                                                                          | 510(k) Number |
|------------|--------------------------------------------------------------------------------------|---------------|
| THD S.p.A. | Family of THD disposable Anoscopes,<br>Proctoscopes, Rectoscopes and Light-<br>scope | K103647       |

{1}------------------------------------------------

K133687 page 2 of 3

The THD N-ano anoscope and its predicate device are indicated for the same intended use and have the same technological characteristics. Both families include the anoscopes and similar devices manufactured with the same materials and the same process.

The differences between the families are that:

- The handle of THD N-ano anoscope is modified in order to lodge all the . components of LED light source circuit (even though the LED light source circuit is not modified respect to predicate devices)
- As a consequence of the point above the LED light source block, used in predicate o devices to lodge all the components of LED light source circuit, is eliminated
- The cover of the LED light source circuit of THD N-Ano anoscope is made of . ABS Terluran, while the LED light source block of predicate devices is made of ABS Lustran.
- The THD N-Ano has similar dimensions than its predicate devices, but this . doesn't include any change on safety and performances of the device, since the THD N-Ano measures are not more critical than dimensions of predicate devices. In fact length of THD N-ano anoscope is 105 mm, longer than an anoscope (which is 90 mm) but shorter than proctoscope (which is 130 mm)., while diameter is 21,7 mm, which is tighter than a Medium Anoscope/proctoscope (which is 22.1 mm. but there are also Large Anoscopes/proctoscopes with a diameter of 26,1 mm) and larger than a Pediatric - small size anoscope (which is 15.1 mm)
- The white polypropylene tip on the introducer, used in predicate devices is . eliminated.

# 2.4. Device Description

The disposable, not sterile THD N-ano anoscope is a disposable, not sterile anoscope with an integrated LED light source.

The device is designed for the examination o the anal sphincter, anus, rectum to perform various diagnostic and therapeutic procedures.

The light source is integrated on the handle.

Differently than predicate devices (Family of THD disposable Anoscopes, Proctoscopes, Rectoscopes and Light-scope - K103647) disposable, not sterile THD N-ano Anoscope does not have any accessory and is available only "not sterile"

{2}------------------------------------------------

## 2.5. Indications for Use

The THD N-Ano Anoscope, is intended for physician use to examine the anal sphincter, anus, rectum to perform various diagnostic and therapeutic procedures.

The Indications for Use of THD N-Ano Anoscope have not been modified with respect to the previous submission (Family of THD disposable Anoscopes, Proctoscopes, Rectoscopes and Light-scope K 103647).

## 2.6. Design Control Activities

As listed in chapter 2.3 the differences between THD N-Ano anoscpe and its predicate devices are:

- . The handle of THD N-ano anoscope is modified in order to lodge all the components of LED light source circuit (even though the LED Light source is not modified respect to predicate devices)
- As a consequence of the point above the LED light source block, used in predicate . devices to lodge all the components of LED light source circuit, is eliminated
- The cover of the LED light source circuit of THD N-Ano anoscope is made of . ABS Terluran, while the LED light source block of predicate devices is made of ABS Lustran.
- The THD N-Ano has similar dimensions than its predicate devices, but this o doesn't include any change on safety and performances of the device, since the THD N-Ano measures are not more critical than dimensions of predicate devices. In fact length of THD N-ano Anoscope is 105 mm, longer than an anoscope (which is 90 mm) but shorter than proctoscope (which is 130 mm)., while diameter is 21,7 mm, which is tighter than a Medium Anoscope/proctoscope (which is 22,1 mm, but there are also Large Anoscopes/proctoscopes with a diameter of 26,1 mm) and larger than a Pediatric - small size anoscope (which is 15.1 mm)
- The white polypropylene tip on the introducer, used in predicate devices is ● eliminated.

The summary about design review is described in the Annex 4.1. The summary about design review activity(Annex 4.1 - summary design review) relates the modification introduced with the risk analysis assessment (Annex 4.2 - Hazard analysis table).

The risk analysis method used to assess the impact of the modifications is described in the Annex 4.3 - Risk management plan.

A declaration of conformity with design controls is included in annex 1.1

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with its wings spread, and three bars below the eagle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 11. 2013

THD S.p.A. % Guido Bonapace Isemed S.R.L. Via A. Altobelli Bonetti, 3/A Imola (BO), Italy 40026

Re: K133687

> Trade/Device Name: THD N-Ano Anoscope Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FER, GCP Dated: October 29, 2013 Received: December 2, 2013

Dear Guido Bonapace,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{4}------------------------------------------------

Page 2 - Guido Bonapace

CFR Part 807): labeling (21 CFR Part 801): medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely vours.

**Herbert** P.Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

THD S.p.A. SPECIAL 510(K) NOTIFICATION THD N-ANO ANOSCOPE

Indications for Use

510(k) Number (if known): K133687

Device Name: THD N-Ano Anoscope

Indications for Use:

The THD N-Ano Anoscopes, are intended for physician use to examine the anal sphincter, anus, rectum to perform various diagnostic and therapeutic procedures.

Prescription Use ਮ (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

ート メイド デジタル アイデント (1) 17,5 × 4,2,111,5,3,6,1,1,1,5,3,6,1,1,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,

Concurrence of CDRH, Office of Device Evaluation (ODE)

Herbert P. Lerner -5
2013.12.11 16:37:51 -05'00'

---

**Source:** [https://fda-staging.innolitics.com/device/K133687](https://fda-staging.innolitics.com/device/K133687)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).