K133648 · Siemens Medical Solutions USA, Inc. · JAK · Jul 7, 2014 · Radiology
Device Facts
Record ID
K133648
Device Name
SYNGO, CT DUAL ENERGY
Applicant
Siemens Medical Solutions USA, Inc.
Product Code
JAK · Radiology
Decision Date
Jul 7, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K133648 · Jul 7, 2014
SYNGO, CT DUAL ENERGY
Siemens Medical Solutions USA, Inc.
Retrospective patient data from several hospitals
Retrospective analysis of clinical patient data was used to support the verification and validation of new dual energy application classes.
Retrospective analysis; Clinical patient data; Verification and validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Retrospective analysis of available patient data; Retrospective analysis
Patients from several hospitals; Number of Sites: Several hospitals
Not applicable for this study
Verification and validation of software specifications for Monoenergetic Plus, Virtual Unenhanced, and Bone Marrow application classes
Indications for Use
syngo.CT Dual Energy is designed to operate with CT images which have been acquired with Siemens Dual Source scanners. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. Depending on the region of interest, contrast agents may be used. These differences provide information on the chemical composition of the scanned body materials. syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information. The functionality of the syngo.CT Dual Energy applications is as follows: Monoenergetic, Brain Hemorrhage, Gout Evaluation, Lung Vessels, Heart PBV, Bone Removal, Lung Perfusion, Liver VNC, Monoenergetic Plus, Virtual Unenhanced, Bone Marrow, Kidney Stones. *) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis under consideration of all available information. The accuracy of identification is decreased in obese patients.
Device Story
Post-processing software package for Siemens Dual Source CT scanners; operates on syngo.via client-server platform. Inputs: CT datasets acquired at two different energy spectra (e.g., 80 kV and 140 kV). Processing: Combines low and high energy images to exploit material-specific attenuation differences; enables qualitative and quantitative evaluation of chemical composition. Outputs: Enhanced visualizations and reports for specific clinical applications (e.g., Monoenergetic Plus, Virtual Unenhanced, Bone Marrow, Kidney Stone analysis). Used by radiologists in clinical settings to improve tissue characterization and diagnostic confidence. Benefits: Enhanced visualization of energy-dependent materials; aids in differentiation of specific pathologies like kidney stone composition or bone marrow assessment. Final diagnosis remains the responsibility of the radiologist using all available clinical information.
Clinical Evidence
Bench testing only. Phantom bench tests conducted for Monoenergetic Plus application. Retrospective analysis of patient data from multiple hospitals performed for all three new application classes (Monoenergetic Plus, Virtual Unenhanced, Bone Marrow). Risk management and software verification/validation testing met acceptance criteria.
Indicated for use with CT images acquired via Siemens Dual Source scanners to visualize chemical composition of body materials. Applications include Monoenergetic, Brain Hemorrhage, Gout, Lung Vessels, Heart PBV, Bone Removal, Lung Perfusion, Liver VNC, Virtual Unenhanced, Bone Marrow, and Kidney Stone differentiation (uric vs. non-uric acid). Accuracy for kidney stones is reduced in obese patients; requires clinical correlation by a radiologist.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K133648
Page 1 of 5
## 510(K) Summary
JUL 07 2014
FOR
# syngo.CT Dual Energy Software Package
Submitted by: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway Malvern, PA 19355
## Date Prepared: May 16, 2014
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
- 1. General Information Importer/Distributor: Siemens Medical Solutions USA, Inc. 51 Valley Stream Parkway
Malvern, PA 19355
Establishment Registration Number: 2240869
Manufacturing Site: Siemens AG Medical Solutions Henkestraße 127 D-91052 Erlangen, Germany
Establishment Registration Number: 3002808157
### 2. Contact Person:
Kimberly Manqum Requlatory Affairs Technical Specialist Siemens Medical Solutions, Inc. USA 51 Valley Stream Parkway, D02 Malvern, PA 19355-1406 Phone: (610) 448-1772 Fax: (610) 448-1778 Email: kimberly.mangum@siemens.com
- 3. Device Name and Classification Product Name: syngo.CT Dual Energy Propriety Trade Name: syngo.CT Dual Energy Classification Name: Computed Tomography X-ray System
{1}------------------------------------------------
Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class: Class II Product Code: 90JAK
Legally Marketed Predicate Devices
| Trade Name: | SOMATOM DRI X-Ray Scanner CT System |
|----------------------------|-------------------------------------|
| 510(k)#: | K837107 |
| Clearance Date: | March 09, 1983 |
| Classification Name: | System, X-Ray, Tomography, Computed |
| Classification Panel: | Radiology |
| Classification Regulation: | 21 CFR § 892.1750 |
| Device Class: | II |
| Product Code: | JAK |
syngo® Dual Energy with extended functionality Trade Name: 510(k)#: K083524 Clearance Date: April 01, 2009 Classification Name: System, X-Ray, Tomography, Computed Classification Panel: Radiology Classification Regulation:21 CFR § 892.1750 Device Class: = JAK Product Code:
### Substantial Equivalence: 4.
The subject device syngo. CT Dual Energy software package is substantially equivalent to following medical devices in commercial distribution:
| Predicate Device Name | FDA Clearance Number | FDA Clearance Date |
|----------------------------------------|----------------------|--------------------|
| syngo® Dual Energy Software<br>Package | K083524 | April 1, 2009 |
| SOMATOM DRI X-Ray Scanner<br>CT System | K837107 | March 09, 1983 |
### 5. Device Description:
Dual energy offers functions for qualitative and quantitative evaluations. Dual energy CT can be used to improve the visualization of the chemical composition of various energy dependent materials in the human body when compared to single energy CT.
Depending on the organ of interest, the user can select and modify different application classes or parameters and algorithms. syngo.CT Dual Energy Software Package is a post processing application package consisting of several post processing application classes that can be used to improve visualization of various energy dependent materials in the human body.
{2}------------------------------------------------
syngo.CT Dual Energy is a post processing software package designed to operate on the most recent version syngo via client server platform, which supports preprocessing and loading of datasets by syngo.via depending on configurable rules. Similar to the predicate device (K083524, clearance date April 01, 2009), syngo.CT Dual Energy allows the evaluation of CT data acquired via spiral or sequence scans at two sufficiently different energy spectra (for example 80 kV and 140 kV) achieved with dual source CT scanners.
## 6. Summary of Technical Characteristics of the Subject Device as Compared with the Predicate Device:
syngo.CT Dual Energy Software Package is a post processing application operating on the multi-user syngo.via client server platform. The subject syngo.CT Dual Energy provides similar evaluation, reporting and visualization tools, and functionality as the predicate device syngo® Dual Energy This includes image processing and visualization tools such as basic visualization of various energy dependent materials in the human body and VRT visualization. In addition to the previously cleared dual energy applications, syngo.CT Dual energy supports the following dual energy application classes:
- · Monoenergetic Plus
- Virtual Unenhanced
- · Bone Marrow
syngo.CT Dual Energy does not have significant changes in technological characteristics when compared to the predicate device syngo® Dual Energy. The Indication for Use, operating principle, and the scientific technology are similar; therefore, Siemens believes that syngo.CT Dual Energy Package is substantially equivalent to the predicate devices.
#### 7. Nonclinical Testing:
syngo.CT Dual Energy Software Package is designed to fulfill the requirements of following standards:
- IEC 60601-1-6 : 2007; Medical electrical equipment Part 1-6: General . requirements for basic safety and essential performance - Collateral Standard: Usability
- IEC 60601-1-4:2000; Consol. Ed. 1.1, Medical electrical equipment -● Part 1-4: General requirements for safety -- Collateral standard: Programmable electrical medical systems, edition 1.1
- IEC 62304 Ed. 1.0. "Medical Device Software Software Lifecvcle � Processes"
- ISO 14971:2007; Medical devices Application of risk management to . medical devices
{3}------------------------------------------------
- . Digital Imaging and Communications in Medicine (DICOM) Set Standard: 2008 DICOM conformity is fully covered by syngo.via implementations.
Phantom bench tests have been conducted in the case of Monoenergetic Plus as well as a retrospective analysis of available patient data from several hospitals for all three new application classes.
The Risk analysis was completed and risk control implemented to mitigate identified hazards. The testing results supports that all the software specifications have met the acceptance criteria. Testing for verification and validation of the device was found acceptable to support the claims of substantial equivalence.
Software Documentation for a Moderate Level of Concern software per FDA's Guidance Document "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices" issued on May 11, 2005 is also included as part of this submission.
## 8. Indications for Use:
syngo.CT Dual Energy is designed to operate with CT images which have been acquired with Siemens Dual Source scanners. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. Depending on the region of interest, contrast agents may be used. These differences provide information on the chemical composition of the scanned body materials, syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information.
The functionality of the syngo. CT Dual Energy applications is as follows:
- Monoenergetic .
- Brain Hemorrhage ●
- Gout Evaluation .
- . Lung Vessels
- Heart PBV ●
- Bone Removal .
- Lung Perfusion .
- Liver VNC
- Monoenergetic Plus .
- Virtual Unenhanced
- Bone Marrow .
- Kidney Stones" .
*) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis under consideration of all available information. The accuracy of identification is decreased in obese patients.
{4}------------------------------------------------
K133648
Page 5 of 5
### General Safety and Effectiveness Concerns: 9.
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled during development, verification and validation testing. To minimize electrical, mechanical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards. -
### 10. Conclusion as to Substantial Equivalence
In summary, Siemens is of the opinion that the syngo.CT Dual Energy Software Package does not introduce any new potential safety risk and is substantially equivalent to and performs as well as the predicate devices
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol of an eagle with three wings.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 7, 2014
Siemens Medical Solutions USA, Inc. % Ms. Kimberly Mangum Technical Specialist. Regulatory Submissions 51 Valley Stream Parkway, D02 MALVERN PA 19355
Re: K133648
Trade/Device Name: syngo.CT Dual Energy Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: June 20, 2014 Received: June 23, 2014
Dear Ms. Mangum:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Ms. Mangum
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.

for
Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
SIEMENS Traditional 510(k) Submission: syngo.CT Dual Energy Software Package
510(k) Number (if known): K133648
Device Name: syngo.CT Dual Energy
Indications for Use:
syngo.CT Dual Energy is designed to operate with CT images which have been acquired with Siemens Dual Source scanners. The various materials of an anatomical region of interest have different attenuation coefficients, which depend on the used energy. Depending on the region of interest, contrast agents may be used. These differences provide information on the chemical composition of the scanned body materials. syngo.CT Dual Energy combines images acquired with low and high energy spectra to visualize this information.
The functionality of the syngo.CT Dual Energy applications is as follows:
- Monoenergetic ●
- Brain Hemorrhage ●
- Gout Evaluation .
- Lung Vessels .
- .Heart PBV .
- Bone Removal .
- Lung Perfusion .
- Liver VNC .
- Monoenergetic Plus ●
- Virtual Unenhanced .
- Bone Marrow .
- י Kidney Stones .
*) Kidney Stones is designed to support the visualization of the chemical composition of kidney stones and especially the differentiation between uric acid and non-uric acid stones. For full identification of the kidney stone additional clinical information should be considered such as patient history and urine testing. Only a well-trained radiologist can make the final diagnosis under consideration of all available information. The accuracy of identification is decreased in obese patients.
| Prescription Use | X |
|-----------------------------|---|
| (Part 21 CFR 801 Subpart D) | |
AND/OR
| Over-The-Counter Use | |
|------------------------|--|
| (21 CFR 807 Subpart C) | |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostics and Radiological Health (OIR)
(Division Sign Off)
Division of Radiological Health
Office of In Vitro Diagnostic and Radiological Health
K133648 510(k) -----------------------------------------------------------------------------------------------------------------------------------------------------------------------
Page I of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.