RAPIDFRET ORAL FLUID ASSAY FOR OPIATES; RAPIDFRET ORAL FLUID CALIBRATORS, CONTROLS, COLLECTOR; RAPIDFRET INTEGRATED WORK

K133642 · Biophor Diagnostics, Inc. · DJG · Jan 24, 2014 · Clinical Toxicology

Device Facts

Record IDK133642
Device NameRAPIDFRET ORAL FLUID ASSAY FOR OPIATES; RAPIDFRET ORAL FLUID CALIBRATORS, CONTROLS, COLLECTOR; RAPIDFRET INTEGRATED WORK
ApplicantBiophor Diagnostics, Inc.
Product CodeDJG · Clinical Toxicology
Decision DateJan 24, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.3650
Device ClassClass 2

Indications for Use

The RapidFRET Oral Fluid Assay for OPIATES is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Opiates at 40 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay is calibrated against Morphine. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or LC/MS/MS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only. The RapidFRET Oral Fluid Calibrator Set and RapidFRET Oral Fluid Control Set are intended for use only with the RapidFRET Oral Fluid Assay for OPIATES and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only.

Device Story

RapidFRET Oral Fluid Assay for OPIATES is an in vitro diagnostic competitive immunoassay for qualitative opiate detection in neat oral fluid. Input: neat oral fluid collected via RapidEASE Oral Fluid Collector. Operation: homogeneous time-resolved fluorescence (TRF) assay; drug in sample competes with donor-fluorophore-labeled drug for binding sites on acceptor-fluorophore-labeled antibody. Binding brings fluorophores into proximity, enabling energy transfer; signal is inversely proportional to drug concentration. Output: qualitative positive/negative result based on 40 ng/mL cutoff. Used in central laboratories on the RapidFRET Integrated Workstation. Results are preliminary; require confirmation by GC/MS or LC/MS/MS. Assists clinicians in drug screening; professional judgment required for interpretation.

Clinical Evidence

Bench testing only. Performance evaluated using 245 clinical oral fluid samples compared against GC/MS or LC/MS/MS confirmation. Results showed >98% agreement for positive samples and 100% agreement for negative samples. Precision studies (n=279 per level) across 20 days demonstrated 100% frequency of expected results at 100% of cutoff. Cross-reactivity and interference studies evaluated 167 compounds, including common substances and pH variations, confirming assay specificity.

Technological Characteristics

Homogeneous time-resolved fluorescence immunoassay. Components: drug-specific antibody reagent, competitive donor reagent, microtiter plates. Platform: RapidFRET Integrated Workstation. Sample: neat oral fluid. Cutoff: 40 ng/mL. Calibration: Morphine-based.

Indications for Use

Indicated for qualitative screening of opiates in neat human oral fluid at a 40 ng/mL cutoff. For prescription use in central laboratories by professional operators. Requires confirmatory testing via GC/MS or LC/MS/MS for positive results.

Regulatory Classification

Identification

An opiate test system is a device intended to measure any of the addictive narcotic pain-relieving opiate drugs in blood, serum, urine, gastric contents, and saliva. An opiate is any natural or synthetic drug that has morphine-like pharmocological actions. The opiates include drugs such as morphine, morphine glucoronide, heroin, codeine, nalorphine, and meperedine. Measurements obtained by this device are used in the diagnosis and treatment of opiate use or overdose and in monitoring the levels of opiate administration to ensure appropriate therapy.

Special Controls

*Classification.* Class II (special controls). An opiate test system is not exempt if it is intended for any use other than employment or insurance testing or is intended for Federal drug testing programs. The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 862.9, provided the test system is intended for employment and insurance testing and includes a statement in the labeling that the device is intended solely for use in employment and insurance testing, and does not include devices intended for Federal drug testing programs (*e.g.,* programs run by the Substance Abuse and Mental Health Services Administration (SAMHSA), the Department of Transportation (DOT), and the U.S. military).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # JAN 2 4 2014 ## 510(k) Summary for the RapidFRET Oral Fluid Assay for OPIATES This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92 The assigned 510(k) number is: k133642 #### 807.92(a){1}: Contact Information Biophor Diagnostics, Inc. Name: Address: 1201 Douglas Avenue Redwood City, CA 94063 Nathaniel G. Butlin, Ph.D. Contact: 650-367-4954 Phone: Fax: 650-364-4985 #### 807.92(a){2): Device Name, Common Name and Classification RapidFRET Oral Fluid Assay for OPIATES (Enzyme Immunoassay for Opiates) RapidFRET Oral Fluid Calibrator Set (Clinical Toxicology Calibrator) RapidFRET Oral Fluid Control Set (Drug Mixture Control Materials) | Product | Code | Class | Regulation<br>Section | Panel | |----------------------------------------|------|-------|-----------------------|-----------------| | RapidFRET Oral Fluid Assay for OPIATES | DJG | II | 862.3650 | 91 - Toxicology | | RapidFRET Oral Fluid Calibrator Set | DKB | II | 862.3200 | 91 - Toxicology | | RapidFRET Oral Fluid Control Set | DIF | I | 862.3280 | 91 - Toxicology | #### 807.92(a){3}: Identification of Legally Marketed Predicate Devices Thermo Scientific CEDIA® Opiates OFT Assay (k101754). #### 807.92(a)(4): Assay Principle The RapidFRET Oral Fluid Assay for OPIATES is an In Vitro Diagnostic competitive immunoassay used to detect opiates in human oral fluid. This is a ready-to-use homogenous system that involves energy transfer between an acceptor fluorophore labeled to an antibody and a donor fluorophore labeled to drug. The assay is based on competition between drug in the sample and drug labeled with the donor fluorophore for a fixed number of binding sites on the antibody reagent. When acceptor and donor fluorophores are brought into close proximity through a binding event, energy transfer occurs. The fluorescence resonance energy transfer (FRET) signal is measured at the wavelength of the acceptor fluorophore and is inversely proportional to the amount of drug in the sample. A Cutoff Calibrator is used to translate the sample measurement into a positive or negative result. Controls are used to establish and monitor precision and accuracy. {1}------------------------------------------------ ## Device Description The RapidFRET Oral Fluid Assay for Opiates is provided in an all liquid, ready to use format. Two reagents are provided included a drug specific reagent and a second competitive donor reagent. The kit is provided with reagents and microtiter plates. A Cutoff Calibrator is used to translate the sample measurement into a positive or negative result. Controls are used to establish and monitor precision and accuracy. Calibrators and controls are sold separately. ## 807.92(a){5}: Intended Use The RapidFRET Oral Fluid Assay for OPIATES is a homogeneous time-resolved fluorescence assay that is intended for prescription use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Opiates at 40 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay is calibrated against Morphine. This assay provides only a preliminary result. To obtain a confirmed analytical result, a more specific alternate chemical method such as GC/MS or LC/MS/MS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only. The RapidFRET Oral Fluid Calibrator Set and RapidFRET Oral Fluid Control Set are intended for use only with the RapidFRET Oral Fluid Assay for OPIATES and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to determine the cutoff level and translate the assay measurement into a positive or negative result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only. | | Candidate Device<br>(RapidFRET OPIATES) | Predicate Device<br>(Thermo Opiates, K101754) | |--------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for<br>Use | Qualitative determination of opiates in<br>human oral fluid in central labs.<br>Prescription use only. | - Qualitative determination of opiates in<br>human oral fluid in clinical setting. | | Methodology | Competitive homogeneous<br>immunoassay. | Competitive homogeneous<br>immunoassay. | | Kit<br>Components | 1 Drug specific antibody reagent in liquid,<br>ready to use format.<br>1 Drug conjugate reagent in liquid, ready<br>to use format. | 1 Drug specific antibody reagent<br>(marketed in combination as a lyophilized<br>reagent and reconstitution buffer).<br>1 Drug conjugate reagent (marketed in<br>combination as a lyophilized reagent and<br>reconstitution buffer). | | | Candidate Device<br>(RapidFRET OPIATES) | Predicate Device<br>(Thermo Opiates, K101754) | | Performance<br>Characteristics | Precision, accuracy, cross<br>reacting/interfering studies demonstrate<br>equivalence to the predicate device. | Precision, accuracy, cross<br>reacting/interfering studies are similar to<br>the RapidFRET Oral Fluid Assay for<br>OPIATES. | | Safety and<br>Effectiveness | Demonstrated in bench testing and<br>described in PI, equivalent to predicate. | Demonstrated in bench testing and<br>described in PI. | | Neat Oral Fluid<br>Cutoff Level | 40 ng/mL neat oral fluid. | 30 ng/mL neat oral fluid using a 10 ng/mL<br>cutoff calibrator to account for sample<br>dilution by collection device. | | Platform | RapidFRET Integrated Workstation<br>available exclusively from Biophor<br>Diagnostics, Inc. | MGC240 analyzer | | Sample<br>Collection | Neat oral fluid is collected with the<br>RapidEASE Oral Fluid Collector via direct<br>expectoration. No diluent is used and<br>sample is stored in glass sample tube<br>with inert screw cap. | Oral fluid is collected with the Oral-Eze<br>Saliva Collection System. This device uses<br>an absorbent swab and diluent. Sample is<br>stored in plastic tube with snap cap. | | Principle and<br>Procedure | Drugs in the oral fluid sample compete<br>with the drug conjugate donor<br>fluorophore for a fixed number of binding<br>sites on the individual drug antibody<br>acceptor reagents. When acceptor and<br>donor fluorophores are brought into<br>close proximity, through the binding<br>event, fluorescent energy transfer is<br>measured.<br>The amount of drug in the specimen<br>sample is inversely proportional to the<br>assay signal as measured by time<br>resolved fluorescence. | The assay is based on the sample<br>analytes competing with analyte<br>conjugates to one inactive fragment of ß-<br>galactosidase for antibody binding sites.<br>The amount of drug in the specimen is<br>directly proportional to the assay signal<br>as measured by absorbance. | | Controls and<br>Calibrator<br>Levels | Calibrators are available at 0 ng/mL and<br>40 ng/mL. Controls are available at 20<br>ng/mL and 60 ng/mL. | Calibrators are available at 0 ng/mL, 10<br>ng/mL, and 80 ng/mL. Controls are<br>available at additional levels. | ### 807.92(a){6): Technological Similarities and Differences to the Predicate {2}------------------------------------------------ Biophor Diagnostics, Inc. ## Traditional Premarket Notification 510(k) Submission RapidFRET Oral Fluid Assay for Opiates ## 807.92(b)(1): Brief Description of Study Data: A series of studies were performed that evaluated the device performance characteristics including precision and analytical sensitivity, correlation with GC/MS and LC/MS/MS, cross reactivity, and analytical specificity that are summarized below. ## Precision and Analytical Sensitivity . {3}------------------------------------------------ Three lots of the RapidFRET Oral Fluid Assay for OPIATES were analyzed, four times daily, for a minimum of 20 days. Negative oral fluid pools were spiked with Morphine at 0%, 25%, 50%, 75%, 100%, 125%, 150%, 175% and 200% of the cutoff level corresponding to approximately 0, 10, 20, 30, 40, 50, 60, 70 and 80 ng/mL. The aggregate data is summarized in the table below: | Table 18.2.7. All Lots Precision Results Summary by Data Points | | | | | | | | | | |-----------------------------------------------------------------|-----|-----|-----|-----|------|------|------|------|------| | | 0% | 25% | 50% | 75% | 100% | 125% | 150% | 175% | 200% | | POS | 0 | 0 | 0 | 0 | 185 | 278 | 263 | 294 | 278 | | NEG | 279 | 279 | 278 | 279 | 94 | 0 | 0 | 0 | 0 | | N | 279 | 279 | 278 | 279 | 279 | 278 | 263 | 294 | 278 | | Table 18.2.8. All Lots Precision Results Summary by Percent Agreement | | | | | | | | | | |-----------------------------------------------------------------------|------|------|------|------|------|------|------|------|------| | | 0% | 25% | 50% | 75% | 100% | 125% | 150% | 175% | 200% | | POS | 0% | 0% | 0% | 0% | 66% | 100% | 100% | 100% | 100% | | NEG | 100% | 100% | 100% | 100% | 34% | 0% | 0% | 0% | 0% | | N | 279 | 279 | 278 | 279 | 279 | 278 | 263 | 294 | 278 | The data indicate that the analytical sensitivity is between 75% and 125% of cutoff, and expected results were achieved at a 100% frequency. ## Correlation with MS Quantitation Neat oral fluid was collected with the RapidEASE Oral Fluid Collection Device from volunteers potentially positive and negative for opiates. The samples (n=245) were randomized and blinded to the instrument operator and assayed using RapidFRET Opiates reagents. Following screening, positive and negative samples were sent for confirmatory testing. The summarized data are shown below. ## Table of Summary Results | | Negative | Near Cutoff<br>Negative | Near Cutoff<br>Positive | High Positive | |----------|-------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------|--------------------------------------------------------------------|----------------------------------------------------------| | | As determined by the<br>predicate device or less<br>than half the cutoff<br>concentration by<br>GC/MS | Between 50%<br>below the cutoff<br>and the cutoff<br>concentration | Between the<br>cutoff and 50%<br>above the cutoff<br>concentration | Greater than<br>50% above the<br>cutoff<br>concentration | | Positive | 0 | 1 | 10 | 53 | | Negative | 177 | 4 | 0 | 0 | {4}------------------------------------------------ ### Table of Discordant Results | Cutoff Value | Assay<br>(POS / NEG) | Drug / Metabolite LC/MS Value (ng/mL) | |--------------|----------------------|---------------------------------------------------------------------------| | 40 ng/mL | Positive | Total Opiates = 36.5 ng/mL (26.8 ng/mL<br>Morphine and 9.7 ng/mL Codeine) | The data indicate that the RapidFRET Oral Fluid Assay for OPIATES had an agreement of >98% for RapidFRET positive samples and an agreement of 100% for RapidFRET negative samples in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. #### Cross Reactivity and Analytical Specificity A compound library of 167 different structurally related and unrelated compounds including metabolites, OTC and prescription medications and drugs of abuse was used to evaluate the device cross reactivity and specificity. Compounds were spiked at 30,000 ng/mL into neat oral fluid pool aliquots with 0 ng/mL, 20 ng/mL and 60 ng/mL of morphine, processed with the RapidEASE Collector, and tested with the RapidFRET OPIATES assay. Those compounds that gave an unexpected result were further titrated to determine the concentration at which the cross-reacting compound yielded a result approximately equivalent to the cutoff. Twenty nine (29) structurally related compounds were determined to cross-react below 30,000 ng/mL in the absence of morphine with twelve cross-reacting at 1000 ng/mL equivalence or less. | Compound | Level<br>(ng/mL) | 0% Morphinet<br>(0 ng/mL) | 50% Morphinet<br>(20 ng/mL) | 150% Morphinet<br>(60 ng/mL) | |-----------------------------------------------------------------------------------------------|------------------|---------------------------|-----------------------------|------------------------------| | Structurally Related Compounds That Cross React in Neat Oral Fluid Pool with 0 ng/ml Morphine | | | | | | 6-Monoacetylmorphine | 30,000 | 36 [111%] | POS | POS | | Amitriptyline | 30,000 | 12,854 [0.3%] | POS | POS | | Chlorpromazine | 30,000 | 12,215 [0.3%] | POS | POS | | Clomipramine | 30,000 | 2,017 [2.0%] | POS | POS | | Codeine | 30,000 | 31 [129%] | POS | POS | | Cyclizine | 30,000 | 10,179 [0.4%] | POS | POS | | Cyclobenzaprine | 30,000 | 17,887 [0.2%] | POS | POS | | Desipramine | 30,000 | 6,754 [0.6%] | POS | POS | | Diacetylmorphine (Heroin) | 30,000 | 40 [100%] | POS | POS | | Dihydrocodeine | 30,000 | 32 [125%] | POS | POS | | Doxepin | 30,000 | 19,538 [0.2%] | POS | POS | | d-Propoxyphene | 30,000 | 23,593 [0.2%] | POS | POS | | Ethylmorphine | 30,000 | 31 [129%] | POS | POS | | Flurazepam | 30,000 | 5,905 [0.7%] | POS | POS | | Hydrocodone | 30,000 | 41 [98%] | POS | POS | | Hydromorphone | 30,000 | 35 [114%] | POS | POS | {5}------------------------------------------------ | Table 18.8.2. Structurally Related Cross Reactants | | | | | |----------------------------------------------------|------------------|---------------------------|-----------------------------|------------------------------| | Compound | Level<br>(ng/mL) | 0% Morphinet<br>(0 ng/mL) | 50% Morphinet<br>(20 ng/mL) | 150% Morphinet<br>(60 ng/mL) | | Imipramine | 30,000 | 1,368 [2.9%] | POS | POS | | Levorphanol | 30,000 | 348 [11%] | POS | POS | | Meperidine | 30,000 | 12,634 [0.3%] | POS | POS | | Morphine-3βDG | 30,000 | 31 [129%] | POS | POS | | Nalorphine | 30,000 | 37 [108%] | POS | POS | | Naloxone | 30,000 | 429 [9.3%] | POS | POS | | Naltrexone | 30,000 | 2,720 [1.5%] | POS | POS | | Normorphine | 30,000 | 3,318 [1.2%] | POS | POS | | Oxycodone | 30,000 | 533 [7.5%] | POS | POS | | Oxymorphone | 30,000 | 831 [4.8%] | POS | POS | | Rifampin | 30,000 | 8,371[0.5%] | POS | POS | | Thioridazine | 30,000 | 18,104 (0.2%) | POS | * | | Trimipramine | 30,000 | 1,947 (2.1%) | POS | POS | t Results are presented as either the RapidFRET OPIATES screening result (POS / NEG) or the concentration in ng/mL of the cross-reactant that gives a Cutoff equivalent response. *Due to high cross reactivity at 0 ng/mL morphine, a 60 ng/mL morphine spike was not analyzed. A second study evaluated common substances such as foods and dental products as well as pH variations. HSA, ethanol, baking soda, whole blood, hemoglobin, hydrogen peroxide, sodium chloride, cholesterol, denture adhesive, ascorbic acid, bilirubin, lgA, lgG and IgM were spiked into neat oral fluid pool aliquots that contained either 20 ng/mL or 60 ng/ml of morphine. Neat oral fluid pool was titrated to pH values of 5, 6, 7, 8 and 9, spiked with morphine to 20 ng/mL or 60 ng/mL and assayed with the RapidFRET OPIATES Assay. The effects of antiseptic mouthwash, cough syrup, cranberry juice, orange juice, tooth paste, chewing tobacco, cigarettes, chewing gum, hard candy, teeth whitening strips, cola, water, antacid, coffee and tea were evaluated by asking volunteers to use a specific item and provide an oral fluid sample. These samples were then spiked with morphine to 20 ng/mL or 60 ng/mL, processed with a RapidEASE Collector and assayed with the RapidFRET OPIATES device. All compounds at the listed concentrations gave a NEG result when spiked with 20 ng/mL morphine and a POS result when spike with 60 ng/mL morphine. #### 807.92(b)(3): Conclusions The RapidFRET Oral Fluid Assay for OPIATES including the RapidFRET Oral Fluid Negative and Cutoff Calibrators, the RapidFRET Oral Fluid Negative and Positive Controls and the RapidEASE Oral Fluid Collector were determined to be substantially equivalent in safety and effectiveness for their intended use. {6}------------------------------------------------ Image /page/6/Picture/1 description: The image shows the logo for the Department of Health & Human Services (HHS). The logo consists of a stylized depiction of an eagle or bird-like figure with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure. Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 January 24, 2014 BIOPHOR DIAGNOSTICS, INC. NATHANIEL BUTLIN VICE PRESIDENT 1201 DOUGLAS AVE REDWOOD CITY CA 94063 Re: K133642 Trade/Device Name: RapidFRET Oral Fluid Assay For Opiates RapidFRET Oral Fluid Calibrator Set RapidFRET Oral Fluid Control Set Regulation Number: 21 CFR 862.3650 Regulation Name: Opiate test system Regulatory Class: II Product Code: DJG, DKB, DIF Dated: November 26, 2013 Received: November 27, 2013 Dear Mr. Butlin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the {7}------------------------------------------------ Page 2-Mr. Butlin electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulations (21 CER Parts 801 and 809), please contact the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance, You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. Courtney H. Lias-S Courtney H. Lias. Ph.D. . Director Division of Chemistry and Toxicology Devices Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {8}------------------------------------------------ ## DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ## Indications for Use 510(k) Number (if known) k133642 #### Device Name RapidFRET Oral Fluid Assay for OPIATES; RapidFRET Oral Fluid Calibrator Set; RapidFRET Oral Fluid Control Set #### Indications for Use (Describe) The RapidFRET Oral Fluid Assay for OPIATES is a homogeneous ime-resolved fluorescence assay that is intended for pressiption use in central laboratories only on the RapidFRET Integrated Workstation. The assay is used to perform a qualitative screen for Opiates at 40 ng/mL in neat oral fluid samples collected with the RapidEASE Oral Fluid Collector. This assay is calibrated against Morphine. This assay provides only a prelimined analytical result, a more specific alternate chemical method such as GCMS or LCMSMS is required. Professional judgment should be applied to any drug test result, particularly when using preliminary positive results. For In Vitro Diagnostic Use Only. The RapidFRET Oral Fluid Calibrator Set and RapidFRET Oral Fluid Control Set are intended for use only with the RapidFRET Oral Fluid Assay for OPIATES and samples collected with the RapidEASE Oral Fluid Collector. The cutoff calibrator is used to delemine the cutoff level and translate the assay measurement into a positive result. The positive and negative controls are used to monitor laboratory systems, operators, precision, accuracy and assay conditions. For In Vitro Diagnostic Use Only. Type of Use (Select one or both, as applicable) P Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page. ## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # T. Danishefskv - FORM FDA 3881 (9/13) PSC Publishing Services (301) 441-6740 EF
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...