K133638 · Firstkind Limited · IPF · Aug 21, 2014 · Physical Medicine
Device Facts
Record ID
K133638
Device Name
GEKO
Applicant
Firstkind Limited
Product Code
IPF · Physical Medicine
Decision Date
Aug 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5850
Device Class
Class 2
Attributes
Therapeutic, Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K133638 · Aug 21, 2014
GEKO
Firstkind Limited
Unpublished clinical study (Harbour Hospital, UK)
The study evaluated the safety and effectiveness of the geko™ T-1 for post-surgical DVT prophylaxis in 40 patients undergoing total hip replacement, comparing venous blood velocity outcomes against standard-of-care compression stockings.
DVT prophylaxis; post-surgical; total hip replacement; venous blood velocity
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Unpublished study at Harbour Hospital; Randomized clinical study; Follow-up/Duration: From immediately post-surgery to hospital discharge (approx. 4 days)
40 patients undergoing elective unilateral total hip replacement (THR); Sample Size: 40; Number of Sites: 1
Thromboembolism Deterrent Stockings (TEDS)
Venous blood velocity
Indications for Use
The geko™ T-1 Neuromuscular Stimulator is intended for: - Increasing local blood circulation, and - Immediate post-surgical stimulation of the calf muscles to prevent venous thrombosis.
Device Story
The geko™ T-1 is a disposable, single-channel neuromuscular stimulator. It consists of a constant current pulse generator, battery, and integrated silver electrode with hydrogel, all in an over-molded plastic casing. Applied to the common peroneal nerve behind the knee, it delivers electrical pulses to induce calf muscle contraction, mimicking the cadence and energy of walking to increase blood flow. A single button controls power and intensity (seven pulse-width levels). Used in clinical or home settings, it is operated by patients or clinicians. The device provides a non-invasive method for DVT prophylaxis, offering an alternative to intermittent pneumatic compression or stockings. It benefits patients by augmenting venous and arterial blood flow, potentially reducing post-surgical thrombosis risk. The device is designed for single-patient use, with a 30-hour battery life.
Clinical Evidence
Evidence includes literature and three independent investigations. Tucker et al. established 1 Hz stimulation increases venous/arterial flow. Warwick et al. (10 healthy volunteers) showed significant peak venous velocity increases (p<0.05) with geko™ in various postures and under plaster casts. Jawad et al. (10 healthy volunteers) reported geko™ significantly outperformed (p<0.001) two intermittent pneumatic compression devices in blood flow metrics. A study of 40 total hip replacement patients showed geko™ significantly increased venous blood velocity compared to no treatment and Thromboembolism Deterrent Stockings (TEDS), with no device-related adverse events.
Indicated for increasing local blood circulation and immediate post-surgical stimulation of calf muscles to prevent venous thrombosis. Contraindicated for patients with cardiac demand pacemakers or recently diagnosed deep vein thrombosis.
Regulatory Classification
Identification
A powered muscle stimulator is an electrically powered device intended for medical purposes that repeatedly contracts muscles by passing electrical currents through electrodes contacting the affected body area.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized design of three human profiles facing right, with flowing lines suggesting movement or connection. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the design.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
August 21, 2014
Firstkind Limited c/o Sheila Hemeon-Heyer, JD, RAC Radcliffe Consulting, Inc. 231 Fairbanks Street West Boylston, MA 01583
Re: K133638
Trade/Device Name: geko™ T-1 Neuromuscular Stimulator Regulation Number: 21 CFR 890.5850 Regulation Name: Powered Muscle Stimulator Regulatory Class: Class II Product Code: IPF Dated: July 25, 2014 Received: July 25, 2014
Dear Ms. Hemeon-Heyer:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{1}------------------------------------------------
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aquel -S
for Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K133638
Device Name geko™ T-1 Neuromuscular Stimulator
#### Indications for Use (Describe)
The geko™ T-1 Neuromuscular Stimulator is intended for:
- Increasing local blood circulation, and
- Immediate post-surgical stimulation of the calf muscles to prevent venous thrombosis.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
| Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# te: 2014.08.21 32 -04'00'
FORM FDA 3881 (9/13)
{3}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{4}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
| A. | Submitter: | Firstkind Limited<br>Hawk House<br>Peregrine Business Park<br>High Wycombe, UK<br>HP13 7DL |
|----|------------|--------------------------------------------------------------------------------------------|
|----|------------|--------------------------------------------------------------------------------------------|
Contact: Rachel Fallon Chief Technology Officer
+44 (0) 1494 572045 Tel: Email: rachel.fallon@firstkindmedical.com
#### B. Date Prepared: July 25, 2014
### C. Device Name and Classification Information:
| Trade Name: | geko™ T-1 Neuromuscular Stimulator (geko) |
|----------------------|-------------------------------------------|
| Common/usual Name: | Powered Muscle Stimulator |
| Classification Name: | Stimulator, Muscle, Powered |
| Product Code, CFR: | IPF, 21 CFR 890.5850 |
| Panel code: | 89 |
| Class: | II |
- SYS*STIM ME 208 (K031017) D. Predicate Device:
#### Device Description: ய்
The qeko™ T-1 (qeko™) Neuromuscular Stimulator is a disposable, fully integrated unit composed of a constant current pulse generator with embedded software and battery enclosed in an over-molded plastic casing, and a silver electrode with a hydrogel coating which provides a means of attachment of the device and electrical contact with the patient. A single button controls the On/Off function and the intensity level of the device, which is achieved through changes in the delivered pulse width. The geko™ is applied so that the electrodes lie over the common peroneal nerve behind
{5}------------------------------------------------
the knee. Stimulation of the common peroneal nerve causes contraction of the calf muscles through the direct activation of the motor neurons, resulting in increased blood flow.
The geko stimulus intensity varies with the pulse width, which can be set to one of seven levels to produce the appropriate muscle contraction within the patient comfort zone (70μs, 100μs, 140μs, 200μs, 400μs, and 560μs). The asymmetric biphasic waveform results in a net charge of zero to the patient during each pulse cycle. The pulse rate is fixed at a frequency of 1 Hz and is used to isometrically stimulate the leg and foot muscles with a cadence and energy similar to that of walking.
Electrical contact is made with the patient through a hydrogel layer applied during manufacture to the integrated electrode. The patient contacting materials have been tested per the requirements of ISO 10993-1 and shown to be biocompatible for prolonged (up to 30 days) contact with intact skin. There are no separate electrode leads or electrodes.
## F. Indications for Use:
The geko™ T-1 device is a Neuromuscular Stimulator and is intended for:
- Increasing local blood circulation, and .
- Immediate post-surgical stimulation of the calf muscles to prevent ● venous thrombosis.
# G. Contraindications
Powered muscle stimulators should not be used on patients with cardiac demand pacemakers.
Powered muscle stimulators should not be used on patients with recently diagnosed deep vein thrombosis.
{6}------------------------------------------------
| Parameter | geko™ T-1 | SYS*STIM ME 208<br>(K031017) | Substantial<br>Equivalence |
|---------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use<br>and Indications<br>for Use | The geko T-1 is a<br>neuromuscular<br>stimulation device and is<br>intended for:<br>Increasing local blood<br>circulation Immediate post-<br>surgical stimulation of<br>the calf muscles to<br>prevent venous<br>thrombosis | 1. Symptomatic relief of<br>chronic intractable pain,<br>acute post traumatic pain<br>or acute post-surgical pain<br>2. Temporary relation of<br>muscle spasm<br>3. Prevention of post-<br>surgical phlebo-<br>thrombosis through<br>immediate stimulation of<br>calf muscles<br>4. Increasing local blood<br>circulation<br>5. Prevention or<br>retardation of disuse<br>atrophy<br>6. Muscle re-education<br>7. Maintaining or<br>increasing range of motion | The indications for<br>use of the geko™ T-1<br>are a subset of the<br>indications for use of<br>the predicate device<br>(see #3 and #4 for<br>the SYS*STIM 208) |
| Clinical<br>application | Prescription use only for<br>use in a clinical or home<br>use setting.<br>Single patient use for up<br>to 30 hours (replacement<br>recommended after 24<br>hours) | Prescription use only for<br>use in a clinical or home<br>use setting.<br>The device is reusable but<br>the electrodes are single<br>use, disposable. | Substantially<br>equivalent. Both<br>devices are for<br>prescription use in a<br>clinical or home<br>environment. |
| Anatomical<br>Sites | The electrodes are<br>applied to the posterior<br>aspect of the knee only<br>for stimulation of the<br>peroneal nerve. | Electrodes can be applied<br>to multiple anatomical<br>sites, including the<br>posterior aspect of the<br>knee only for stimulation of<br>the peroneal nerve. | Substantially<br>equivalent. Both<br>devices can be used<br>for stimulation of the<br>peroneal nerve. |
| Parameter | geko™ T-1 | SYS*STIM ME 208<br>(K031017) | Substantial<br>Equivalence |
| Basic Unit Characteristics | | | |
| Power source | One CR2032 primary<br>lithium coin cell. Not<br>replaceable by user | Mains (line) voltage | Substantially<br>equivalent. Both<br>devices meet<br>electrical safety<br>standards. |
| -Method of Line<br>Current<br>Isolation | N/A | Transformer isolation | |
| -Patient<br>Leakage<br>Current | | | Substantially<br>equivalent. |
| -Normal<br>Condition | < 20μA | <100 μA | Leakage current of<br>geko is less than |
| -Single<br>Fault Condition | < 20μA | <300 μA | predicate device. |
| Number of<br>output modes | Single mode with seven<br>discrete stimulation<br>settings corresponding to<br>the seven pulse widths. | Three output modes<br>The "Pulse" mode has a<br>200 μs pulse width | Substantially<br>equivalent to "Pulse"<br>mode of predicate. |
| Number of<br>output<br>channels | Single channel | Single channel | Same |
| -Synchronous<br>or alternating | N/A (single channel) | N/A (single channel) | Same |
| -Method of<br>channel<br>isolation | Capacitor | Transformer | Substantially<br>equivalent. Both<br>methods are effective<br>methods of channel<br>isolation. |
| Regulated<br>current or<br>regulated<br>voltage | Current regulated | Voltage regulated | Substantially<br>equivalent. Both<br>methods comply with<br>electrical safety<br>standards. |
| Microprocessor<br>controlled? | Yes | Yes | Same |
| Parameter | geko™ T-1 | SYS*STIM ME 208<br>(K031017) | Substantial<br>Equivalence |
| Automatic<br>overload trip | Yes | No | Substantially<br>equivalent or better. |
| Automatic no-<br>load trip | Yes | No | Substantially<br>equivalent or better. |
| Automatic<br>shut-off | Yes | Yes | Same |
| Patient over-<br>ride control | Yes | Yes | Same |
| Indicator<br>display | | | |
| - On/Off status | Yes | Yes | Same |
| - Low battery | Yes (device switches off) | N/A | Substantially<br>equivalent |
| -Voltage /<br>current level | N/A (device has fixed<br>constant current) | Yes (intensity control) | Substantially<br>equivalent. |
| -Charge level<br>(pulse width) | Yes (pulse width varies to<br>change delivered<br>charge), The stimulation<br>level (pulse width) is<br>indicated by the flashing<br>of the LED The number<br>of times LED flashes in<br>sequence indicates the<br>level of stimulation, with<br>level 1 (a single flash)<br>equating to 70 μs being<br>the lowest stimulation<br>level and level 7 (7<br>flashes in sequence)<br>equating to 560 μs, the<br>highest. stimulation level | N/A (device has fixed<br>pulse width (200 μs)). | Substantially<br>equivalent. |
| Timer range in<br>minutes | 1800 minutes maximum<br>(device is disabled after<br>30 hours battery run time) | 60 minute timer can be<br>reset indefinitely | Substantially<br>equivalent. |
| Parameter | geko™ T-1 | SYS*STIM ME 208<br>(K031017) | Substantial<br>Equivalence |
| Compliance<br>with voluntary<br>standards | Yes<br>IEC 60601-1:1998 A1, A2<br>IEC 60601-2-10:1987, A1<br>EN 60601-1-2:2007<br>ISO 10993-1 | Yes<br>UL 2601-1<br>CAN/CSA-22.601.2.10-92 | Substantially<br>equivalent |
| Compliance<br>with 21 CFR<br>898 | N/A (electrodes are<br>integral with the device,<br>there are no separate<br>leads) | Yes | Substantially<br>equivalent |
| Weight | 18 g | 1.4 pounds (636g) | Substantially<br>equivalent. Both<br>devices are small<br>and lightweight. |
| Dimensions | 6" x 1.6" x 0.43" | 2.75" x 6.1" x 8" | Substantially<br>equivalent |
| Housing<br>material and<br>construction | Plastic injection molding | Plastic injection molding | Substantially<br>equivalent |
| Output Specifications | | | |
| Waveform<br>-Pulsed<br>monophasic or<br>biphasic<br>- Shape | Biphasic (asymmetrical<br>biphasic with zero net<br>DC)<br>Rectangular, with charge<br>balancing second phase | Biphasic (asymmetrical<br>biphasic with zero net DC)<br>Rectangular, with charge<br>balancing second phase | Same |
| Maximum<br>output voltage | 13.5 V @ 500 Ω<br>54 V @ 2000 Ω<br>110 V @ 10,000 Ω | 92 V @ 500 Ω (measured)<br>144 V @ 2000 Ω<br>166 V @ 10,000 Ω | Substantially<br>equivalent. Geko can<br>operate at lower<br>voltage as intensity is<br>increased by<br>increasing pulse<br>width. |
| Maximum<br>output current | 27 mA @ 500 Ω<br>27 mA @2000 Ω<br>11 mA @ 10,000 Ω | 184 mA @ 500 Ω<br>72 mA @2000 Ω<br>17 mA @ 10,000 Ω | Substantially<br>equivalent for same<br>reason as above. |
| Parameter | geko™ T-1 | SYS*STIM ME 208<br>(K031017) | Substantial<br>Equivalence |
| Pulse width | 70, 100, 140, 200, 280,<br>400, 560 µs | 200 µs | Substantially<br>equivalent. Predicate<br>varies intensity by<br>increasing<br>voltage/current in<br>fixed pulse width. |
| Frequency | 1 Hz | 1 to 80 Hz | Substantially<br>equivalent. Low<br>frequency used to<br>increase circulation. |
| For<br>interferential<br>modes only:<br>-beat<br>Frequency (Hz) | N/A | N/A | Same |
| Multiphasic<br>waveforms | | | |
| -Symmetrical<br>phases | No | No | Same |
| -Phase duration | 70-560 µs for positive<br>phase, second (negative)<br>phase is an exponential<br>decay with a 0.1 s time<br>constant. | 200 µs | Substantially<br>equivalent. Reasons<br>previously discussed. |
| Net charge | 0 µC at 500 Ω | ~0 µC at 100 Ω | Same |
| -How achieved | Capacitor coupling | Transformer coupling | Methods are<br>substantially<br>equivalent. |
| Maximum<br>phase charge | 18.3 µC at 500 Ω | 33.5 µC at 500 Ω | Substantially<br>equivalent or better |
| Maximum<br>current density | 6.67 mA/cm² | 9.12 mA/cm² | Substantially<br>equivalent or better. |
| Maximum<br>power density | 0.000044 W/cm² | 0.012 W/cm² | Substantially<br>equivalent or better. |
| Parameter | geko™ T-1 | SYS*STIM ME 208<br>(K031017) | Substantial<br>Equivalence |
| Burst mode<br>a) Pulses per<br>burst<br>b) Bursts per<br>second<br>c) Burst<br>duration<br>(seconds)<br>d) Duty Cycle<br>[Line (b) x<br>Line (c)] | N/A (single pulse, no<br>pulse train or burst) | N/A when used in PULSE<br>mode (single pulse) | Same. |
| ON Time<br>(seconds) | N/A Stimulation is<br>delivered at 1 Hz, with<br>single pulses of 70µs to<br>560µs | N/A Stimulation is<br>delivered at 1 Hz, with<br>single pulses of 200 µs | Same. |
| OFF Time<br>(seconds) | N/A Stimulation is<br>delivered at 1 Hz, with<br>single pulses of 70µs to<br>560µs | N/A Stimulation is<br>delivered at 1 Hz, with<br>single pulses of 200 µs | Same. |
| Additional Features | | | |
| - On/Off status | Yes | Yes | Same |
| - Low battery | Yes (device switches off) | N/A (mains powered) | Substantially<br>equivalent |
| -Voltage /<br>current level | Yes. Fixed constant<br>current (pulse width<br>varies to change<br>delivered charge) | Yes Continuously variable<br>using un-calibrated dial | Substantially<br>equivalent |
| -Low stim<br>current | Yes. Unit turns off after 4<br>minutes below low<br>current threshold…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.