K133635 · Retractable Technologies, Inc. · FMI · Jun 17, 2014 · General Hospital
Device Facts
Record ID
K133635
Device Name
EASYPOINT NEEDLE
Applicant
Retractable Technologies, Inc.
Product Code
FMI · General Hospital
Decision Date
Jun 17, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Indications for Use
The Easy Point™ Needle is intended for use with syringes for general purpose fluid injection and/or aspiration, and venipuncture to obtain blood collection. The EasyPoint™ Needle aids in the prevention of needlestick injuries.
Device Story
EasyPoint™ Needle is a single lumen hypodermic needle designed for percutaneous fluid injection, aspiration, and venipuncture. It features an integrated sharps injury prevention mechanism (chamber) that covers the needle after use. The device is intended for use with luer-fitting syringes (3mL or smaller). It is operated by healthcare professionals in clinical settings. The safety mechanism is retraction-based, similar to the VanishPoint® Syringe. By covering the needle post-use, it reduces the risk of accidental needlestick injuries to the operator. The device is provided as a stand-alone needle.
Clinical Evidence
No clinical trials were performed. Evidence consists of bench testing and a simulated use human factors validation study. Bench testing evaluated liquid/air leakage, dead space, needle puncture force, functionality force, deactivation force, chamber strength, and cannula/hub bond strength per ISO 594-1, ISO 594-2, ISO 7864, ISO 7886, and ISO 9626. Simulated use study demonstrated that the device performs safely and effectively when used by a variety of healthcare workers.
Technological Characteristics
Single lumen hypodermic needle with integrated retraction-based sharps injury prevention chamber. Materials tested for biocompatibility per ISO 10993. Performance standards: ISO 594-1, ISO 594-2, ISO 7864, ISO 7886, ISO 9626. Sterilization validated per ISO 11137 (SAL 10-6). Available in 23G and 25G sizes.
Indications for Use
Indicated for use with syringes for general purpose fluid injection, aspiration, and venipuncture for blood collection in patients requiring such procedures. Aids in the prevention of needlestick injuries.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
Predicate Devices
BD Single Lumen Needle (PrecisionGlide™) (K021475)
{0}------------------------------------------------
## K133635
# - JUN 1-7-2014`
# 6/16/14
.
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
ﺎ
## PREMARKET NOTIFICATION 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS FOR VANISHPOINT® BLOOD COLLECTION SET (21 CFR 807.92)
| Applicant Name: | Retractable Technologies, Inc.<br>511 Lobo Lane<br>Little Elm, TX 75068 |
|---------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Phone: | 972-294-1010 |
| Contact Person: | Rhonda Wells<br>Regulatory Affairs Manager |
| Date of Summary Preparation: | June 16, 2014 |
| Trade Name: | EasyPoint™ Needle |
| Common Name: | Needle, Hypodermic, Single Lumen |
| Regulation Number: | 880.5570 |
| Product Code | FMI: Hypodermic Single Lumen Needle |
| Device Classification: | Class II |
| Legally Marketed Substantially Equivalent Device: | |
| | K021475 - BD Single Lumen Needle (PrecisionGlide™)<br>K970803 - VanishPoint® Syringe |
| Description of Device: | The EasyPoint™ Needle is intended for general purpose<br>percutaneous injection or aspiration of fluids and venipuncture to<br>obtain blood collection.<br>The device contains a sharps injury prevention feature (needlestick<br>prevention feature-chamber) that covers the entire needle after use.<br>Initially available in sizes of 23G and 25G x 1" and 25G x 5/8" for<br>use with 3mL syringes or smaller. |
| Intended Use: | The Easy Point™ Needle is intended for use with syringes<br>for general purpose fluid injection and/or aspiration, and<br>venipuncture to obtain blood collection.<br>The EasyPoint™ Needle aids in the prevention of needlestick<br>injuries. |
| Engineering Testing: | Performance testing on the EasyPoint™ Needle was performed<br>according to applicable design requirements of ISO 594-1, ISO 594-<br>2, ISO 7864, ISO 7886 and ISO 9626. Those tests include, liquid<br>leakage, air leakage, dead space and needle puncture force.<br>Additional performance tests were developed by Retractable<br>Technologies, Inc. to measure the functionality force, deactivation<br>force, chamber strength/ rigidity, cannula/ hub bond strength. |
| Simulated Use Study: | A simulated use pre-market study was performed to demonstrate<br>through human factors validation testing that the EasyPoint™ Needle |
す
{1}------------------------------------------------
device performs safely and effectively when used by a variety of healthcare workers in a simulated use scenario.
#### Comparison of Technical Characteristics:
The subject EasyPoint™ Needle utilizes a retraction based safety mechanism like the predicate device VanishPoint® Syringe. The 2nd predicate device, BD PrecisionGlide® single lumen needle is available as a stand-alone needle but does not have an integrated safety mechanism. The BD single lumen needle is also available as a needle/barrel combination syringe. The EasyPoint™ Needle will initially be sold as a stand-alone needle for use with luer fitting devices. The VanishPoint® Syringe and BD single lumen needle are commercially available in various needle length and gauge sizes. Biocompatibility testing has been performed on the EasyPoint™ Needle in accordance with ISO 10993 with acceptable results.
### Substantial Equivalence:
Conclusion:
The operation, similar design and materials between the subject device and the predicate devices do not raise new issues of safety and effectiveness when used as labeled: The intended use of the subject and predicate devices is virtually identical. In engineering testing, the subject device performed as well or better than the predicate devices. The Simulated Use Study demonstrates that the subject device will perform as intended. The necessary Biocompatibility testing was performed with acceptable results and a Sterilization Validation was performed according to ISO 11137 to ensure sterility of (SAL 106). It is our opinion that the devices are substantially equivalent.
The EasyPoint™ Needle is substantially equivalent to the Vanish Point® Syringe and BD PrecisionGlide® single lumen needle. The intended uses are the same No new concerns regarding safety and effectiveness were raised.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three lines forming its body and wings. The eagle is enclosed in a circle with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 17, 2014
Retractable Technologies, Incorporated Ms. Rhonda Wells Regulatory Affairs Manager 511 Lobo Lane Little Elm, TX 75068
Re: K133635
Trade/Device Name: EasyPoint™ Needle Regulation Number: 21 CFR 880.5570 Regulation Name: Needle, Hypodermic, Single Lumen Regulatory Class: Il Product Code: FMI Dated: March 18, 2014 Received: March 19, 2014
Dear Ms. Wells:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Ms. Wells
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Mary Runner
-S FILA
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
### Indications for Use
Form Approved: OMB No, 0910-0120 Expiration Date: January 31, 2017 Sea PRA Statement below.
510(k) Number (if known) KI33632
Device Name
Easy Point Needle
Indications for Use (Describe)
The Easy Point Needle is intended for use with syringes for general purpose fluid injections and venipuncture to obtain blood collection.
The Easy Point Needle aids the prevention of needlestick injuries.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE CHIP STORE CONTRACTOR FOR FOR FOR FOR FOR FOR CONTRACTOR CONTRACTOR COLLECT
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/4/Picture/15 description: The image shows a logo with the letters "FDA" in a stylized, blocky font. The letters are formed by a series of horizontal and vertical lines, creating a grid-like pattern within each letter. The overall design is geometric and somewhat abstract, with a focus on the shapes of the letters rather than a traditional typeface.
Digitally signed by Richard C. Chapman -S Date: 2014.06.16 09:35:17 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *Do NoT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the rine to review instructions, search existing data sources, gather and maintain the data needed and complete line to review instruction. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.