← Product Code [LJS](/productcode/LJS) · K133634

# SPECTRUM TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET (K133634)

_Cook, Inc. · LJS · Dec 23, 2013 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133634

## Device Facts

- **Applicant:** Cook, Inc.
- **Product Code:** [LJS](/productcode/LJS.md)
- **Decision Date:** Dec 23, 2013
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5970
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter (PICC) Sets and Trays are intended for short- or long-term use for venous pressure monitoring, blood sampling, administration of drugs and fluids, and for use with power injectors for delivery of contrast in CT studies. The catheter is impregnated with the antimicrobials minocycline and rifampin to help provide protection against catheter-related bloodstream infections (CRBSIs). The Spectrum Turbo-Ject® PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure limit setting for Power Injectors used with the Turbo-Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated.

## Device Story

Radiopaque polyurethane PICC for venous access; short- or long-term use. Impregnated with minocycline and rifampin to reduce CRBSI risk. Inserted via Peel-Away introducer or over-the-wire. Supports power injection of contrast media for CT studies (max 325 psi). Used in clinical settings by healthcare professionals. Provides venous access for fluid/drug delivery, blood sampling, and pressure monitoring. Benefits patient by providing reliable vascular access while reducing infection risk via antimicrobial coating.

## Clinical Evidence

Bench testing only. Testing included tensile testing (ISO 10555-1:1995), dynamic pressure testing, static failure pressure testing, liquid leakage testing, air leakage testing, and antimicrobial efficacy testing (zone of inhibition). All tests met predetermined acceptance criteria.

## Technological Characteristics

Radiopaque polyurethane catheter; minimally tapered design. Impregnated with minocycline and rifampin (min 15 mm zone of inhibition). 4.0 Fr and 5.0 Fr sizes; single and double lumen configurations. Max pressure 325 psi. Supplied sterile for single use. No software or electronic components.

## Regulatory Identification

A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”

## Predicate Devices

- Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter ([K081690](/device/K081690.md))
- Turbo-Ject® Peripherally Inserted Central Venous Catheter ([K072625](/device/K072625.md))
- [K132334](/device/K132334.md)
- [K132885](/device/K132885.md)

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below "COOK" is the word "MEDICAL" in smaller, white letters.

DEC 2 3 2013

### 510(k) SUMMARY - K133634 Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter Set 21 CFR \$807.92 Date Prepared: December 09, 2013

### Submitted By:

Applicant: Contact: Applicant Address: Cook Incorporated Sean Spence, RAC Cook Incorporated 750 Daniels Way Bloomington, IN 47404 (812) 335-3575 x105127 (812) 332-0281

#### Device Information:

Contact Fax Number:

Contact Phone Number:

| Trade name:          | Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter Set |
|----------------------|------------------------------------------------------------------------|
| Common name:         | PICC Set                                                               |
| Classification Name: | Percutaneous, Implanted, Long-Term Intravascular Catheter              |
| Regulation:          | 21 CFR §880.5970                                                       |
| Product Code:        | LJS                                                                    |

# Intended Use:

Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter (PICC) Sets and Trays are intended for short- or long-term use for venous pressure monitoring, blood sampling, administration of drugs and fluids, and for use with power injectors for delivery of contrast in CT studies. The catheter is impregnated with the antimicrobials minocycline and rifampin to help provide protection against catheter-related bloodstream infections (CRBSIs). The Spectrum Turbo-Ject® PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure limit setting for Power Injectors used with the Turbo-Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated.

#### Predicate Devices:

The Spectrum Turbo-Ject® PICC Sets, subject of this submission, are modifications to the following Cook Incorporated PICCs:

- K081690, Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter �
	- O K072625, Turbo-Ject® Peripherally Inserted Central Venous Catheter
		- o K132334 & K132885, modifications to add minimal taper design

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below the word "COOK" is the word "MEDICAL" in smaller, white letters.

COOK INCORPORATED 750 DANIELS WAY. P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 1. 012.339.2235 rou firer 800.457.4500 WWW.COOKMEDICAL.COM

# Comparison to Predicate Devices:

The scope of this submission is to first add a minimal taper design modification to the 4.0 Fr single and 5.0 Fr single and double lumen antimicrobial (Spectrum) PICCs cleared in the primary predicate, K081690. Second, this submission will add a Spectrum 4.0 Fr double lumen configuration not found in the primary predicate. Complete performance testing, including power injection, was reviewed for a 4.0 Fr double lumen PICC without Spectrum under K072625. Subsequently, the design was modified under K132334 to a minimal taper. Due to market demand, the Spectrum technology from K081690 is being added to the 4.0 Fr double lumen device design cleared under K132334. Lastly, this submission is also meant to standardize the labeling (IFU) format with the non-Spectrum minimally tapered PICCs as amended during K132334 (4.0 Fr) and K132885 (5.0 Fr).

|                                    | Cook Spectrum Turbo-Ject PICC<br>K081690!                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                       | Cook Spectrum Turbo-Ject PICC Set<br>SUBJECT OF THIS SUBMISSION                                                                    |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number/<br>Product Code | 21 CFR 880.5970, LJS                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            | Identical                                                                                                                          |
| Classification Name                | Intravascular Catheter                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          | Identical                                                                                                                          |
| Class                              | II                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Identical                                                                                                                          |
| Intended Use                       | Intended for short- or long-term use for venous<br>pressure monitoring, blood sampling,<br>administration of drugs and fluids, and for use<br>with power injectors for delivery of contrast in<br>CT studies. The catheter is impregnated with<br>the antimicrobials minocycline and rifampin to<br>help provide protection against catheter-related<br>bloodstream infections (CRBSIs). The Spectrum Turbo-Ject PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure limit setting for Power Injectors used with the Spectrum Turbo- Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rated indicated. | Identical                                                                                                                          |
| Catheter Shaft Material.           | Polyurethane                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | Identical                                                                                                                          |
| French Size /<br>Length (cm)       | 4.0 and 5.0 / 60                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                | Identical                                                                                                                          |
| Taper                              | 2 Fr sizes                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 1 Fr size                                                                                                                          |
| Number of Lumens.                  | 4.0 Fr - Single lumen<br>4.0 Fr - Double lumen (K132334)<br>5.0 Fr - Single lumen<br>5.0 Fr - Double lumen                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                      | 4.0 Fr - Single lumen<br>4.0 Fr - Double lumen<br>5.0 Fr - Single lumen<br>5.0 Fr - Double lumen                                   |
| Flow Rate.                         | 4.0 Fr Single lumen - 4 mL/sec<br>4.0 Fr Double lumen - 3 mL/sec (K132334)<br>5.0 Fr Single lumen - 7mL/sec<br>5.0 Fr Double lumen - 5mL/sec                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    | 4.0 Fr Single lumen - 5 mL/sec<br>4.0 Fr Double lumen - 3 mL/sec<br>5.0 Fr Single lumen - 7mL/sec<br>5.0 Fr Double lumen - 5mL/sec |
| Maximum Pressure<br>Rating         | 325 psi                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                         | Identical                                                                                                                          |

# Table 1: Comparison Table

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, sans-serif font at the top. Below it, the word "MEDICAL" is in a smaller, sans-serif font. The background is black, and the text is white.

|                                           | Cook Spectrum Turbo-Ject PICC K081690'                                                                                                                                                                               | Cook Spectrum Turbo-Ject PICC Set<br>SUBJECT OF THIS SUBMISSION                                                                        |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Antimicrobial Agents                      | Minocycline<br>• Synonym: Minocin<br>• Chemical family : Tetracycline<br>• Formula : C23 H27 N3 O7 · HCl<br>Rifampin<br>• Synonym: Rifampicin<br>• Chemical family : Macrolide antibiotic<br>• Formula : C43H58N4O12 | Identical                                                                                                                              |
| Antimicrobial Activity<br>Specification   | Minimum zone of inhibition of 15 mm                                                                                                                                                                                  |                                                                                                                                        |
| Method of<br>antimicrobial<br>application | Impregnation                                                                                                                                                                                                         | Identical                                                                                                                              |
| Inside Diameter (inch)                    | 4.0 Fr Single lumen - 0.030<br>4.0 Fr Double lumen - 0.022/0.038* (K132334)<br>5.0 Fr Single lumen - 0.037<br>5.0 Fr Double lumen - 0.022/0.044*                                                                     | 4.0 Fr Single lumen - 0.037<br>4.0 Fr Double lumen - 0.022/0.038*<br>5.0 Fr Single lumen - 0.048<br>5.0 Fr Double lumen - 0.024/0.048* |
| Outside Diameter (inch)                   | 4.0 Fr Single lumen - 0.053<br>4.0 Fr Double lumen - 0.055 (K132334)<br>5.0 Fr Single lumen - 0.066<br>5.0 Fr Double lumen - 0.066                                                                                   | 4.0 Fr Single lumen - 0.055<br>4.0 Fr Double lumen - 0.055<br>5.0 Fr Single lumen - 0.066<br>5.0 Fr Double lumen - 0.066               |
| Primary set components                    | Obturator, Peel-Away introducer, entry needles,<br>wire guide, injection caps, syringe, scalpel, and<br>securement device                                                                                            | Identical                                                                                                                              |

1K 132334 is referenced where appropriate for the 4.0 Fr double lumen device *height/width

# Device Description:

The proposed Spectrum Turbo-Ject® PICCs are radiopaque polyurethane peripherally inserted central venous catheters for short- or long-term use, and can be inserted through a Peel-Away® introducer, or over-the-wire. The proposed devices are minimally tapered 4.0 Fr and 5.0 Fr single and double lumen catheters. The set components may include the PICC, obturator, Peel-Away® introducer, entry needles, wire guide, and other convenience components. The set is supplied sterile and is intended for one-time use.

# Test Data:

The following tests were performed to demonstrate that the proposed Spectrum Turbo-Ject® PICC Set met applicable design and performance requirements and support a determination of substantial equivalence.

- . Tensile Testing - In conformance with ISO 10555-1:1995, testing demonstrated that the peak load value was greater than 10 N.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is in black and white, with the word "COOK" in large, bold letters on the top line. Below the word "COOK" is the word "MEDICAL" in smaller letters. The logo is simple and professional.

COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489

- Dynamic Pressure Testing Testing demonstrated that the catheters did not fail during . simulated use.
- Static Failure Pressure Testing demonstrated that static failure pressure was at or above the . acceptance criterion.
- 트 Liquid Leakage Testing - Testing demonstrated that the catheters did not leak liquid.
- l Air Leakage Testing – Testing demonstrated that the catheters did not exhibit air leakage.
- . Antimicrobial Testing - Testing demonstrated that the catheters met the predetermined acceptance criteria.

## Conclusions Drawn from the Tests:

The results of these tests provide reasonable assurance that the Spectrum Turbo-Ject® PICC Set is as safe and effective as the predicate devices and support a determination of substantial equivalence.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a symbol that resembles a stylized human figure embracing a globe. The image is black and white and appears to be a scan of a document.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

December 23, 2013

Cook. Incorporated C/O Mr. Sean Spence. RAC 750 Daniels Way BLOOMINGTON IN 47404

Re: K133634

Trade/Device Name: Spectrum Turbo-Jet Peripherally Inserted Central Venous Catheter Set Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: November 25, 2013 Received: November 26, 2013

Dear Mr. Spence:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{5}------------------------------------------------

· Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor YowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Kwame O Ulmer
-s FDA

for

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

*** : : ... . ...

## Indications for Use

510(k) Number (if known) K133634

Device Name

Spectrum (antimicrobial) Turbo-Ject PICCs

#### Indications for Use (Describe)

Spectrum Turbo-Ject Peripherally Inserted Central Venous Catherer (PICC) Sets and Trays are intended for short-or long-term use for venous pressure monitoring, blood sampling, administration of drugs and for use with power injectors for delivery of contrast in CT studies. The catheter is impregnated with the antimicrobials minocycline and rifampin to help protection against catherer-related bloodstream infections (CRBSIs). The Spectrum Turbo-Ject PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure liniectors used with the Spectrum Turbo-Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated.

Type of Use (Select one or both, as applicable)

[ Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

Form Approved: OMB No. 0910-0120

Expiration Date: December 31, 2013

See PRA Statement on last page.

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

CONTRACTORIE FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)

Image /page/6/Picture/13 description: The image shows a logo with the letters FDA in a stylized, geometric design. The letters are formed by a series of parallel lines, giving them a three-dimensional appearance. The FDA logo is a recognizable symbol associated with the Food and Drug Administration.

Digitally signed by Richard C. Chapman Date: 2013.12.23 14:21:50 -05'00' 

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. . . . . . .

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

----------------

---

**Source:** [https://fda-staging.innolitics.com/device/K133634](https://fda-staging.innolitics.com/device/K133634)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
