SPECTRUM TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET
K133634 · Cook, Inc. · LJS · Dec 23, 2013 · General Hospital
Device Facts
Record ID
K133634
Device Name
SPECTRUM TURBO-JECT PERIPHERALLY INSERTED CENTRAL VENOUS CATHETER SET
Applicant
Cook, Inc.
Product Code
LJS · General Hospital
Decision Date
Dec 23, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 880.5970
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter (PICC) Sets and Trays are intended for short- or long-term use for venous pressure monitoring, blood sampling, administration of drugs and fluids, and for use with power injectors for delivery of contrast in CT studies. The catheter is impregnated with the antimicrobials minocycline and rifampin to help provide protection against catheter-related bloodstream infections (CRBSIs). The Spectrum Turbo-Ject® PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure limit setting for Power Injectors used with the Turbo-Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated.
Device Story
Radiopaque polyurethane PICC for venous access; short- or long-term use. Impregnated with minocycline and rifampin to reduce CRBSI risk. Inserted via Peel-Away introducer or over-the-wire. Supports power injection of contrast media for CT studies (max 325 psi). Used in clinical settings by healthcare professionals. Provides venous access for fluid/drug delivery, blood sampling, and pressure monitoring. Benefits patient by providing reliable vascular access while reducing infection risk via antimicrobial coating.
Clinical Evidence
Bench testing only. Testing included tensile testing (ISO 10555-1:1995), dynamic pressure testing, static failure pressure testing, liquid leakage testing, air leakage testing, and antimicrobial efficacy testing (zone of inhibition). All tests met predetermined acceptance criteria.
Technological Characteristics
Radiopaque polyurethane catheter; minimally tapered design. Impregnated with minocycline and rifampin (min 15 mm zone of inhibition). 4.0 Fr and 5.0 Fr sizes; single and double lumen configurations. Max pressure 325 psi. Supplied sterile for single use. No software or electronic components.
Indications for Use
Indicated for patients requiring short- or long-term venous access for pressure monitoring, blood sampling, and administration of drugs, fluids, or contrast media via power injection (max 325 psi) in CT studies. Includes antimicrobial impregnation to reduce risk of catheter-related bloodstream infections.
Regulatory Classification
Identification
A percutaneous, implanted, long-term intravascular catheter is a device that consists of a slender tube and any necessary connecting fittings, such as luer hubs, and accessories that facilitate the placement of the device. The device allows for repeated access to the vascular system for long-term use of 30 days or more, and it is intended for administration of fluids, medications, and nutrients; the sampling of blood; and monitoring blood pressure and temperature. The device may be constructed of metal, rubber, plastic, composite materials, or any combination of these materials and may be of single or multiple lumen design.
Special Controls
*Classification.* Class II (special controls) Guidance Document: “Guidance on Premarket Notification [510(k)] Submission for Short-Term and Long-Term Intravascular Catheters.”
Predicate Devices
Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter (K081690)
Turbo-Ject® Peripherally Inserted Central Venous Catheter (K072625)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below "COOK" is the word "MEDICAL" in smaller, white letters.
DEC 2 3 2013
### 510(k) SUMMARY - K133634 Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter Set 21 CFR \$807.92 Date Prepared: December 09, 2013
### Submitted By:
Applicant: Contact: Applicant Address: Cook Incorporated Sean Spence, RAC Cook Incorporated 750 Daniels Way Bloomington, IN 47404 (812) 335-3575 x105127 (812) 332-0281
#### Device Information:
Contact Fax Number:
Contact Phone Number:
| Trade name: | Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter Set |
|----------------------|------------------------------------------------------------------------|
| Common name: | PICC Set |
| Classification Name: | Percutaneous, Implanted, Long-Term Intravascular Catheter |
| Regulation: | 21 CFR §880.5970 |
| Product Code: | LJS |
# Intended Use:
Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter (PICC) Sets and Trays are intended for short- or long-term use for venous pressure monitoring, blood sampling, administration of drugs and fluids, and for use with power injectors for delivery of contrast in CT studies. The catheter is impregnated with the antimicrobials minocycline and rifampin to help provide protection against catheter-related bloodstream infections (CRBSIs). The Spectrum Turbo-Ject® PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure limit setting for Power Injectors used with the Turbo-Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated.
#### Predicate Devices:
The Spectrum Turbo-Ject® PICC Sets, subject of this submission, are modifications to the following Cook Incorporated PICCs:
- K081690, Spectrum Turbo-Ject® Peripherally Inserted Central Venous Catheter �
- O K072625, Turbo-Ject® Peripherally Inserted Central Venous Catheter
- o K132334 & K132885, modifications to add minimal taper design
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, white letters on a black background. Below the word "COOK" is the word "MEDICAL" in smaller, white letters.
COOK INCORPORATED 750 DANIELS WAY. P.O. BOX 489 BLOOMINGTON, IN 47402-0489 U.S.A. 1. 012.339.2235 rou firer 800.457.4500 WWW.COOKMEDICAL.COM
# Comparison to Predicate Devices:
The scope of this submission is to first add a minimal taper design modification to the 4.0 Fr single and 5.0 Fr single and double lumen antimicrobial (Spectrum) PICCs cleared in the primary predicate, K081690. Second, this submission will add a Spectrum 4.0 Fr double lumen configuration not found in the primary predicate. Complete performance testing, including power injection, was reviewed for a 4.0 Fr double lumen PICC without Spectrum under K072625. Subsequently, the design was modified under K132334 to a minimal taper. Due to market demand, the Spectrum technology from K081690 is being added to the 4.0 Fr double lumen device design cleared under K132334. Lastly, this submission is also meant to standardize the labeling (IFU) format with the non-Spectrum minimally tapered PICCs as amended during K132334 (4.0 Fr) and K132885 (5.0 Fr).
| | Cook Spectrum Turbo-Ject PICC<br>K081690! | Cook Spectrum Turbo-Ject PICC Set<br>SUBJECT OF THIS SUBMISSION |
|------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|
| Regulation Number/<br>Product Code | 21 CFR 880.5970, LJS | Identical |
| Classification Name | Intravascular Catheter | Identical |
| Class | II | Identical |
| Intended Use | Intended for short- or long-term use for venous<br>pressure monitoring, blood sampling,<br>administration of drugs and fluids, and for use<br>with power injectors for delivery of contrast in<br>CT studies. The catheter is impregnated with<br>the antimicrobials minocycline and rifampin to<br>help provide protection against catheter-related<br>bloodstream infections (CRBSIs). The Spectrum Turbo-Ject PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure limit setting for Power Injectors used with the Spectrum Turbo- Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rated indicated. | Identical |
| Catheter Shaft Material. | Polyurethane | Identical |
| French Size /<br>Length (cm) | 4.0 and 5.0 / 60 | Identical |
| Taper | 2 Fr sizes | 1 Fr size |
| Number of Lumens. | 4.0 Fr - Single lumen<br>4.0 Fr - Double lumen (K132334)<br>5.0 Fr - Single lumen<br>5.0 Fr - Double lumen | 4.0 Fr - Single lumen<br>4.0 Fr - Double lumen<br>5.0 Fr - Single lumen<br>5.0 Fr - Double lumen |
| Flow Rate. | 4.0 Fr Single lumen - 4 mL/sec<br>4.0 Fr Double lumen - 3 mL/sec (K132334)<br>5.0 Fr Single lumen - 7mL/sec<br>5.0 Fr Double lumen - 5mL/sec | 4.0 Fr Single lumen - 5 mL/sec<br>4.0 Fr Double lumen - 3 mL/sec<br>5.0 Fr Single lumen - 7mL/sec<br>5.0 Fr Double lumen - 5mL/sec |
| Maximum Pressure<br>Rating | 325 psi | Identical |
# Table 1: Comparison Table
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for Cook Medical. The word "COOK" is in large, bold, sans-serif font at the top. Below it, the word "MEDICAL" is in a smaller, sans-serif font. The background is black, and the text is white.
| | Cook Spectrum Turbo-Ject PICC K081690' | Cook Spectrum Turbo-Ject PICC Set<br>SUBJECT OF THIS SUBMISSION |
|-------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Antimicrobial Agents | Minocycline<br>• Synonym: Minocin<br>• Chemical family : Tetracycline<br>• Formula : C23 H27 N3 O7 · HCl<br>Rifampin<br>• Synonym: Rifampicin<br>• Chemical family : Macrolide antibiotic<br>• Formula : C43H58N4O12 | Identical |
| Antimicrobial Activity<br>Specification | Minimum zone of inhibition of 15 mm | |
| Method of<br>antimicrobial<br>application | Impregnation | Identical |
| Inside Diameter (inch) | 4.0 Fr Single lumen - 0.030<br>4.0 Fr Double lumen - 0.022/0.038* (K132334)<br>5.0 Fr Single lumen - 0.037<br>5.0 Fr Double lumen - 0.022/0.044* | 4.0 Fr Single lumen - 0.037<br>4.0 Fr Double lumen - 0.022/0.038*<br>5.0 Fr Single lumen - 0.048<br>5.0 Fr Double lumen - 0.024/0.048* |
| Outside Diameter (inch) | 4.0 Fr Single lumen - 0.053<br>4.0 Fr Double lumen - 0.055 (K132334)<br>5.0 Fr Single lumen - 0.066<br>5.0 Fr Double lumen - 0.066 | 4.0 Fr Single lumen - 0.055<br>4.0 Fr Double lumen - 0.055<br>5.0 Fr Single lumen - 0.066<br>5.0 Fr Double lumen - 0.066 |
| Primary set components | Obturator, Peel-Away introducer, entry needles,<br>wire guide, injection caps, syringe, scalpel, and<br>securement device | Identical |
1K 132334 is referenced where appropriate for the 4.0 Fr double lumen device *height/width
# Device Description:
The proposed Spectrum Turbo-Ject® PICCs are radiopaque polyurethane peripherally inserted central venous catheters for short- or long-term use, and can be inserted through a Peel-Away® introducer, or over-the-wire. The proposed devices are minimally tapered 4.0 Fr and 5.0 Fr single and double lumen catheters. The set components may include the PICC, obturator, Peel-Away® introducer, entry needles, wire guide, and other convenience components. The set is supplied sterile and is intended for one-time use.
# Test Data:
The following tests were performed to demonstrate that the proposed Spectrum Turbo-Ject® PICC Set met applicable design and performance requirements and support a determination of substantial equivalence.
- . Tensile Testing - In conformance with ISO 10555-1:1995, testing demonstrated that the peak load value was greater than 10 N.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for Cook Medical. The logo is in black and white, with the word "COOK" in large, bold letters on the top line. Below the word "COOK" is the word "MEDICAL" in smaller letters. The logo is simple and professional.
COOK INCORPORATED 750 DANIELS WAY, P.O. BOX 489
- Dynamic Pressure Testing Testing demonstrated that the catheters did not fail during . simulated use.
- Static Failure Pressure Testing demonstrated that static failure pressure was at or above the . acceptance criterion.
- 트 Liquid Leakage Testing - Testing demonstrated that the catheters did not leak liquid.
- l Air Leakage Testing – Testing demonstrated that the catheters did not exhibit air leakage.
- . Antimicrobial Testing - Testing demonstrated that the catheters met the predetermined acceptance criteria.
## Conclusions Drawn from the Tests:
The results of these tests provide reasonable assurance that the Spectrum Turbo-Ject® PICC Set is as safe and effective as the predicate devices and support a determination of substantial equivalence.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a symbol that resembles a stylized human figure embracing a globe. The image is black and white and appears to be a scan of a document.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 23, 2013
Cook. Incorporated C/O Mr. Sean Spence. RAC 750 Daniels Way BLOOMINGTON IN 47404
Re: K133634
Trade/Device Name: Spectrum Turbo-Jet Peripherally Inserted Central Venous Catheter Set Regulation Number: 21 CFR 880.5970 Regulation Name: Percutaneous, Implanted, Long-Term Intravascular Catheter Regulatory Class: II Product Code: LJS Dated: November 25, 2013 Received: November 26, 2013
Dear Mr. Spence:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
· Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor YowIndustry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
Kwame O Ulmer
-s FDA
for
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
*** : : ... . ...
## Indications for Use
510(k) Number (if known) K133634
Device Name
Spectrum (antimicrobial) Turbo-Ject PICCs
#### Indications for Use (Describe)
Spectrum Turbo-Ject Peripherally Inserted Central Venous Catherer (PICC) Sets and Trays are intended for short-or long-term use for venous pressure monitoring, blood sampling, administration of drugs and for use with power injectors for delivery of contrast in CT studies. The catheter is impregnated with the antimicrobials minocycline and rifampin to help protection against catherer-related bloodstream infections (CRBSIs). The Spectrum Turbo-Ject PICC is indicated for multiple injections of contrast media through a power injector. The maximum pressure liniectors used with the Spectrum Turbo-Ject PICC may not exceed 325 psi and the flow rate may not exceed the maximum flow rate indicated.
Type of Use (Select one or both, as applicable)
[ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
See PRA Statement on last page.
# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
CONTRACTORIE FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR Concurrence of Center for Devices and Radiological Health (CDRH) (Signalure)
Image /page/6/Picture/13 description: The image shows a logo with the letters FDA in a stylized, geometric design. The letters are formed by a series of parallel lines, giving them a three-dimensional appearance. The FDA logo is a recognizable symbol associated with the Food and Drug Administration.
Digitally signed by Richard C. Chapman Date: 2013.12.23 14:21:50 -05'00'
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW. . . . . . .
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
----------------
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.