← Product Code [JAA](/productcode/JAA) · K133553

# KALARE (K133553)

_Toshiba America Medical Systems, In.C · JAA · Jun 13, 2014 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133553

## Device Facts

- **Applicant:** Toshiba America Medical Systems, In.C
- **Product Code:** [JAA](/productcode/JAA.md)
- **Decision Date:** Jun 13, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.1650
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Pediatric

## Indications for Use

Intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations. Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body parts. Applications can be performed with the patient sitting, standing, or lying in prone or supine position.

## Device Story

KALARE (DREX-KL80) is a universal diagnostic imaging system for fluoroscopy and radiography. It captures X-ray images using fluorescent scintillation effects as X-rays pass through the patient. The system supports examinations of the gastrointestinal tract, endoscopy, nonvascular contrast studies, and general skeletal/abdominal imaging. Operated by physicians or technologists in clinical settings, the device allows imaging with patients in various positions (sitting, standing, prone, supine). The system processes X-ray inputs to produce diagnostic images for clinical review. The current modification adds a Flat Panel Detector (FPD) to the existing Image Intensifier (I.I.) configuration. Healthcare providers use these images to inform medical diagnosis and treatment decisions, benefiting patients through improved visualization of anatomical structures.

## Clinical Evidence

Bench testing only. Testing conducted per IEC standards for medical devices and X-ray systems. Bench tests confirmed the FPD installation met manufacturer specifications. Comparative performance testing between the modified device and the predicate device evaluated image quality, artifacts, and motion/dynamic capabilities. No clinical data was used to support the substantial equivalence determination.

## Technological Characteristics

Universal radiographic and fluoroscopic X-ray system. Components: Image Intensifier (I.I.) and Flat Panel Detector (FPD). Energy source: X-ray. Compliance: 21 CFR Subchapter J, 21 CFR 1020, IEC 60601-1, IEC 60601-2-7, IEC 60601-2-28, IEC 60601-2-32. Manufacturing: ISO 13485, 21 CFR 820. Connectivity: System-integrated imaging.

## Regulatory Identification

An image-intensified fluoroscopic x-ray system is a device intended to visualize anatomical structures by converting a pattern of x-radiation into a visible image through electronic amplification. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

## Special Controls

*Classification.* Class II (special controls). An anthrogram tray or radiology dental tray intended for use with an image-intensified fluoroscopic x-ray system only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9. In addition, when intended as an accessory to the device described in paragraph (a) of this section, the fluoroscopic compression device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

## Predicate Devices

- KALARE (DREX-KL80) ([K110785](/device/K110785.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# JUN 1 3 2014

Image /page/0/Picture/1 description: The image shows the text "K1335553 Page 1 of 3" in a handwritten style. The text appears to be a page number or document identifier. The handwriting is somewhat stylized, with distinct loops and curves in the letters and numbers.

### 510(k) - SUMMARY OF SAFETY AND EFFECTIVENESS

# 1. CLASSIFICATION and DEVICE NAME:

| Classification Name:    | System, X-ray, Fluoroscopic, Image-Intensified    |
|-------------------------|---------------------------------------------------|
| Regulation Number:      | 21 CFR 892.1650 (Class II)                        |
| Product Code            | JAA – Image Intensified Fluoroscopic X-ray System |
| Trade Proprietary Name: | KALARE                                            |
| Model Number:           | DREX-KL80                                         |

# 2. ESTABLISHMENT REGISTRATION: 2020563

# 3. CONTACT PERSON, U.S. AGENT and ADDRESS:

Contact Person:

Charlemagne Chua Manager, Regulatory Affairs (714) 669-7896

U.S. Agent: Paul Biggins Director, Regulatory Affairs

(714) 730-5000

#### Establishment Name and Address:

Toshiba America Medical Systems, Inc. (TAMS) 2441 Michelle Drive Tustin, Ca. 92780

### 4. MANUFACTURING SITE

Toshiba Medical Systems Corporation (TMSC) 1385 Shimoishigami Otawara-shi, Tochigi 324-8550, Japan

- 5. Date OF SUBMISSION: November 18, 2013
#### 6. PERFORMANCE STANDARD:

21 CFR Subchapter J, Federal Diagnostic X-ray Equipment Standard

{1}------------------------------------------------

- 7. PREDICATE DEVICE: KALARE (DREX-KL80) , K110785
- 8. REASON FOR SUBMISSION: Modification of a cleared device
- 9. SUBMISSION TYPE:

Traditional 510(k)

#### 10. DEVICE DESCRIPTION:

The main function of the KALARE (DREX-KL80) is to perform fluoroscopy/radiography of the examinations of the gastrointestinal tract examination, support for endoscopy, nonvascular contrast study, general abdominal radiography and general skeletal radiography. Using fluorescent scintillation effects of X-rays that have passed through the patient's body, image information is obtained for medical diagnosis and treatment.

#### 11. SUMMARY OF INTENDED USES:

- a. Intended to be used as a universal diagnostic imaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations
- b. Intended for use by a qualified/trained doctor or technologist on both adult and pediatric subjects taking diagnostic radiographic and fluoroscopic exposures of the whole body, skull, spinal column, chest, abdomen, extremities and other body. parts. Applications can be performed with the patient sitting, standing, or lying in prone or supine position.

#### 12. SUMMARY OF CHANGE(S)

- a. Addition of a previously cleared FPD.
#### 13. SUBSTANTIAL EQUIVALENCE:

This device is substantially equivalent to the KALARE, K110785, marketed by Toshiba America Medical Systems. KALARE includes modifications to the cleared device which adds FPD (in addition to I.I). The basic system configuration, method of operation, base software and manufacturing process remain unchanged from the cleared device.

#### 14. SAFETY:

The device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards. This device is in conformance with the applicable parts of the IEC60601-1 standards, its collateral standards and particular standards; IEC 60601-2-7, IEC60601-2-28 and IEC 60601-2-32. All requirements of the Federal Diagnostic Equipment Standard, as outlined in 21 CFR §1020, that apply to this device, will be met and reported via product report.

{2}------------------------------------------------

K B35553
Page 3 of 3

### 15. TESTING

Testing of the modified system was conducted in accordance with the applicable standards published by the International Electromechanical Commission (IEC) for Medical Devices and XR Systems. Bench testing was done to confirm that the installation of the detector met the stated specifications of the component manufacturer. Additional testing was done to compare the performance between the predicate device and the modified device that included testing directed at image quality, artifacts and motion/dynamic capabilities. The conclusion of this testing demonstrated that substantial equivalence to the predicate device could be proven without the use of clinical images. Additionally, in support of this conclusion the detector component has been the subject of other 510k Premarket Notifications where the clinical aspects of the component have been addressed.

#### 16. CONCLUSION

The modifications incorporated into the KALARE, do not change the indications for use or the intended use of the device. Safety and effectiveness have been verified via risk management and the application of design controls to the modifications.

{3}------------------------------------------------

Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

June 13, 2014

Toshiba Medical Systems Corporation % Mr. Paul Biggins Director Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780

Re: K133553

Trade/Device Name: KALARE (DREX-KL80) with FPD Regulation Number: 21 CFR 892.1650 Regulation Name: Image-intensified fluoroscopic x-ray system Regulatory Class: II Product Code: JAA Dated: May 15, 2014 Received: May 16, 2014

Dear Mr. Biggins:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{4}------------------------------------------------

Page 2-Mr. Biggins

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours.

Janine M. Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

K133553

| AND ANNOLO AND COLLEGIAL AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND AND A<br>The program and the comments of the comments of<br>And Continent on the contract<br>Device Name                                                                                                          | ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A Property of Children Annual Property of | "What" the family of was "the first weeks of the first of the first of the chamble of the |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------|
| KALARE (DREX-KL80)<br>CDD<br>ר ו"ג' 17.                                                                                                                                                                                                                                                                                                                                                    |                                                                                                                                                                                                                                                                                                                                                                                                               |                                                                                           |
| "While the first and and and and a many of the commend of the comments of the comments of the contribution of the contribution of the contribution of the contribution of the<br>CONTRACTOR COLLECT<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------<br>A | STATIS AND AND A BRIDGE OF CHARACT AND THE COLLECTION OF CHEAR<br>THE TEN BEAUTY<br>10-40-4-4-4-4-4-4-4<br>------------------------------------------------------------------------------------------------------------------------------------------------------------------------------                                                                                                                     | .                                                                                         |

#### Indications for Use (Describe)

1. Intended to be used as a universal diaging system for radiographic and fluoroscopic examinations, including general R&F and pediatric examinations.

2. Intended for use by a qualified/trained doctor or technologist on both and pediatric subjects taking diagnostic and fluoroscopic exposures of the whole body, skull, spinal column, extremities and other body parts. Applications can be performed with the patient sitting, standing, or lying in prone or supine position.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

_ Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

# FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Smh7)

---

**Source:** [https://fda-staging.innolitics.com/device/K133553](https://fda-staging.innolitics.com/device/K133553)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
