K133526 · Philips Medical Systems Nederland B.V. · LNH · Feb 12, 2014 · Radiology
Device Facts
Record ID
K133526
Device Name
MDIXON-QUANT
Applicant
Philips Medical Systems Nederland B.V.
Product Code
LNH · Radiology
Decision Date
Feb 12, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1000
Device Class
Class 2
Indications for Use
mDIXON-Quant: · is a software option intended for use on Achieva and Ingenia 1.5T & 3.0T MR Systems · is a non-invasive triglyceride fat fraction calculation · is a 3D, single breath-hold acquisition and reconstruction technique · allows for water-fat separation generating Water-only, Fat-only images as well as Outphase and In-phase images · is indicated for magnetic resonance imaging of the liver
Device Story
mDIXON-Quant is a software option for Philips Achieva and Ingenia 1.5T/3.0T MR systems; performs non-invasive triglyceride fat fraction calculation. Input: gradient echo (FFE) MR signals with 6+ echoes. Processing: reconstruction algorithm accounts for multiple spectral peaks of triglyceride fat, corrects T2* confounding effects, and reduces T1 bias. Output: Water-only, Fat-only, Out-phase, In-phase images, plus triglyceride fat fraction and T2*/R2* maps. Used in clinical routine by radiologists/technicians; protocol selected via ExamCard. Output viewed on console; supports ROI analysis for quantification. Benefits: provides non-invasive assessment of liver fat content, aiding clinical decision-making.
Clinical Evidence
Clinical validation performed on 1.5T and 3.0T systems using 13.6-14.6 second breath-hold protocols. Study confirmed successful generation of Fat Fraction and T2* images and operator ability to perform ROI analysis. Bench testing demonstrated FF accuracy of ±3.5% and T2* reproducibility of 1.5%. No clinical data regarding diagnostic outcomes provided; validation focused on technical performance and user workflow.
Technological Characteristics
Software-only feature for existing MR platforms. Reconstruction algorithm runs on HP Z400/Z420 workstations (Windows 7, 64-bit). Uses gradient echo (FFE) pulse sequences. No hardware changes. Interoperability via existing cleared MR platform features. No off-the-shelf software used for the feature itself.
Indications for Use
Indicated for magnetic resonance imaging of the liver in patients undergoing 1.5T or 3.0T MR examinations on Achieva or Ingenia systems. Used for non-invasive triglyceride fat fraction calculation.
Regulatory Classification
Identification
A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).
Special Controls
*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "PHILIPS" in large, bold, black letters. Above the word, there is handwritten text that appears to be a document or page identifier, possibly "K133526 page 1 of 5". The handwriting is in black ink and contrasts with the clean, printed font of the brand name.
# Philips Medical Systems Nederland B.V.
# 510(k) Summary
FEB 1222014
# mDIXON-Quant
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
#### General Information 【.
21 CFR 807.92 (a)(1), (2)
| Company Name: | Philips Medical Systems Nederland B.V. |
|-----------------------|----------------------------------------------------------------------------------------------|
| Address: | Veenpluis 4-6<br>5684 PC Best<br>The Netherlands |
| Registration No .: | 3003768277 |
| Contact Person: | Susan Quick<br>Tel: (440)-483-2291<br>Fax: (440)-483-4799<br>E-mail: susan.quick@philips.com |
| Prepared (date): | 2013 November 14, Revised 2014 January 24,<br>2014 February 06 |
| Trade Name of Device: | mDIXON-Quant |
| Classification: | Class II |
| Regulatory Section: | Magnetic Resonance Diagnostic Device.<br>892.1000 |
| Product Code: | 90LNH |
{1}------------------------------------------------
## ` 21 CFR 807.92(a)(3):Legally marketed predicate device to which substantial equivalence is claimed:
| Predicate Device: | IDEAL Software Option |
|----------------------|----------------------------------------|
| Manufacturer: | GE Healthcare |
| Predicate Device k#: | K103411 |
| Reference Device: | mDIXON |
| Manufacturer: | Philips Medical Systems Nederland B.V. |
| Reference Device k#: | K102344 |
## 21 CFR 807.92(a)(4): Description of the device that is the subject of this premarket notification:
#### Summary of functions of the device and its major components
mDIXON-Quant acquisition and reconstruction is based on the mDIXON product developed previously. Acquisition relies on a gradient echo acquisition (FFE) including a large number of echoes (6 or more). Furthermore, mDIXON-Quant reconstruction includes the use of multiple spectral peaks of triglyceride fat, correction of the T2* confounding effect and reduction of T1 bias.
mDIXON-Quant allows for water-fat separation and generates Water-only, Fat-only images as well as Out-phase and In-phase images synthesized from the Water and Fat images. Additionally, mDIXON-Ouant produces images representing triglyceride fat fraction as well as images representing transverse magnetization relaxation.
The feature requires:
- · Specific parameter settings for the mDIXON sequence, within cleared parameter limits, to acquire the MR signals
- · A new calculation function to generate the new images for Fat Fraction, T2 *. This function uses a set of MR images as input that is generated in a cleared manner from the acquired MR signals.
- · The new images need to be stored and displayed with the appropriate labels for Fat Fraction, T2 *, applying the facilities provided by the cleared platform.
#### 21 CFR 807.92(a)(5): Intended Use
mDIXON-Quant:
- · is a software option intended for use on Achieva and Ingenia 1.5T & 3.0T MR Systems
- · is a non-invasive triglyceride fat fraction calculation
- · is a 3D, single breath-hold acquisition and reconstruction technique
- · allows for water-fat separation generating Water-only, Fat-only images as well as Outphase and In-phase images
- · is indicated for magnetic resonance imaging of the liver
{2}------------------------------------------------
#### 21 CFR 807.92(a)(6): Technological Characteristics:
The main functional units in the software are:
- (acquisition of MR signals by means of MR pulse sequences) - Methods
- Reconstruction (transforming the MR signals to images)
- Patient Administration (storing of the images in the database and providing access)
- Viewing (display of images and measuring ROI)
The technical impact of the feature mDJXON-Quant comprises:
- Methods: modify an existing sequence to acquire the data required for mDIXON-Quant, within the cleared limits provided by the basic MR system.
- Reconstruction: add a new calculation function that generates the new output images (Fat Fraction, T2*-maps and R2* (=1/T2*).
- Patient Administration and Viewing: enable storage and display of the new output images with appropriate labels (SW-M and SW-P).
No off-the-shelf software is used for the feature mDIXON-Quant. The off-the-shelf software used in the basic MR system for the Ingenia and Achieva 1.5T and 3.0T MR systems is cleared by K110151.
The basic MR system provides the features for interoperability and internet connections. which are already cleared. The feature mDIXON-Quant does not have any additions or modifications related to interoperability and connections to an external network.
## Hardware platform description
mDIXON-Ouant does not require any change of the hardware platform. The main extension introduced by mDIXON-Quant, the new reconstruction algorithm, runs on the Reconstructor computer.
Computer characteristics:
- · Manufacturer: HP; Model: Z400 or Z420; Processor speed: ≥3.6 GHz; Memory: 16-32 GB RAM, 1-2 x quad core processors
- · Operating system: Windows 7, 64 bits
The mDIXON-Ouant algorithm does not add specific requirements to the Ingenia and Achieva 1.5T and 3.0T MR systems as cleared by K110151.
#### Workflow
For the operator the only new element for the mDIXON-Quant-feature in this clinical routine workflow is:
Protocol selection: The operator selects an ExamCard with mD1XON-Quant protocols
All other steps are not changed. Only in the planscan phase the user may want to change the predefined selection of output image-types. The newly generated image types can be viewed, post-processed, printed and archived as any other image type. The display of the images on the console can be switched between grey-scale and color scale. During post processing the operator can do further analysis (such as calculation of average values in a ROI) using the image processing facilities as provided by the standard platform.
{3}------------------------------------------------
#### 21 CFR 807.92(b)(1): Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification:
All non-clinical testing was performed on a 1.5T Ingenia and a 3.0T Achieva TX.
All Integration testing completed with passing results, with regression testing completing without error messages and greater than 90% of all scans present in the batch running successfully.
All System Verification testing completed with passing results. For the System Verification testing, a dS-Torso. Posterior coil and Anterior coil were used on the Ingenia 1.5T and a Sense-XL-Torso coil was used on the Achieva 3.0T. All Acquisition and Reconstruction, Viewing, Calculation, and Performance testing passed. In the Acquisition and Reconstruction testing, there were no error messages, scan time was < 20 seconds and FF and T2* images were available. In the Viewing testing, all images opened in gray scale and the Look-Up Table showed the images in the expected colors. In the Calculation testing, the ROI could be drawn and two values of area and mean were displayed. For the Performance testing, no error messages were displayed during SW startup, phantom positioning, examcard loading, survey scan activation, and images were generated in < 5 minutes.
For the Quantification Performance testing, test results showed that the FF accuracy is ±3.5% for both field strengths and all possible parameter combinations, and the T2* range and reproducibility is 1.5% for both field strengths.
Overall, all tests performed for mDIXON-Quant were successful. While using mDIXON-Quant the system didn't crash or hang-up. Workflow was smooth and no problems occurred.
All defects have been analyzed and are confirmed that they are not safety defects and will not cause any hazardous situation on using this application.
#### 21 CFR 807.92(b)(2): Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:
The clinical user needs were tested as part of the validation testing.
The same systems and coils used in the Verification testing were used for the Validation testing. mDIXON-Quant produced Fat Fraction and T2* images, and operator could create a ROI in the acquired images. All protocol steps as stated in the test plan, completed successfully. The mDIXON-Quant protocols were scanned in 13.6sec for 3T and in 14.6sec for 1.5T. All volunteers were able to hold their breath for that time. As such, the clinical validation of mDIXON-Quant completed successfully. All clinical user needs have passed for Achieva and Ingenia 1.5T and 3T systems.
{4}------------------------------------------------
K133526
Page 5 of 5
#### 21 CFR 807.92(b)(3): The conclusions drawn from the nonclinical and clinical tests that demonstrate that the device is as safe, as effective, and performs as well as or better than the legally marketed device identified in paragraph (a)(3) of this section:
The nonclinical and clinical tests have demonstrated that the device is safe and works according to its intended use.
The mDIXON-Quant software does not introduce new indications for use, nor does the use of the device result in any new potential hazard. Philips Medical Systems considers mDIXON-Quant to be substantially equivalent to the above mentioned predicate device.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol is a stylized representation of an eagle with outstretched wings, with three wavy lines emanating from its body.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-0002
February 12, 2014
Philips Medical Systems Nederland B.V. % Ms. Susan Quick Regulatory Affairs Specialist Philips Medical Systems (Cleveland), Inc. 595 Miner Road CLEVELAND OH 44143
Re: K133526
Trade/Device Name: mDIXON-Quant Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: LNH Dated: November 14, 2013 Received: November 15, 2013
Dear Ms. Quick:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Ms. Quick
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.hum.
Sincerely yours,
Michael D. O'Hara.
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page.
## Indications for Use
510(k) Number (if known) K 133526
#### Device Name mDIXON-Quant
DIXON-Quan
Indications for Use (Describe)
mDIXON-Quant:
- · is a software option intended for use on Achieva and Ingenia 1.5T & 3.0T MR Systems
- · is a non-invasive triglyceride fat fraction calculation
- · is a 3D, single breath-hold acquisition and reconstruction technique
- · allows for water-far separation generating Water-only, Fat-only images as well as Out-phase images
- · is indicated for magnetic resonance imaging of the liver
Type of Use (Select one or both, as applicable)
[x] Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Michael D. O'Hara
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.