SUZHOU BEINUO SURGICAL STAPLERS - CIRCULAR STAPLERS FOR SINGLE USE, HEMORRHOIDAL CIRCULAR STAPLERS FOR SINGLE USE, LINEA
K133499 · Suzhou Beinuo Medical Equipment Co. , Ltd. · GDW · Jan 14, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133499
Device Name
SUZHOU BEINUO SURGICAL STAPLERS - CIRCULAR STAPLERS FOR SINGLE USE, HEMORRHOIDAL CIRCULAR STAPLERS FOR SINGLE USE, LINEA
Applicant
Suzhou Beinuo Medical Equipment Co. , Ltd.
Product Code
GDW · General, Plastic Surgery
Decision Date
Jan 14, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4750
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Circular Stapler for Single Use The Suzhou Beinuo Circle Stapler for Single Use is used throughout the alimentary tract for the reconstruction of the alimentary tract and the creation of end-to-end, end-to-side and side-to-side anastomoses. Hemorrhoidal Circular Stapler for Single Use The Suzhou Beinuo Hemorrhoidal Circular Stapler for Single Use is a Circular Stapler product, with accessories, that is indicated in anorectal surgical procedures for the treatment of hemorrhoids and anorectal wall defects. Linear Stapler and Reloads for Single Use The Suzhou Beinuo Linear Stapler for Single Use (and reloads) is indicated for the closure of tissue in abdominal, gynecological, and thoracic surgical procedures. Linear Cutter Stapler and Reloads for Single Use The Suzhou Beinuo Linear Cutter Stapler for Single Use (and reloads) is indicated for use in the resection, transection and anastomosis of tissue in gastrointestinal, gynecological, and thoracic surgical procedures.
Device Story
Family of single-use surgical staplers; includes circular, hemorrhoidal, linear, and linear cutter models. Devices function by manual activation (squeezing handle or pushing firing knob) to place double or triple staggered rows of titanium staples into tissue. Circular and hemorrhoidal models feature a central knife blade to resect excess tissue and create anastomoses. Linear and linear cutter models provide mechanical suturing for tissue closure or resection. Used in OR settings by surgeons. Benefits include mechanical replacement of traditional hand-suturing, facilitating tissue sealing and resection during GI, thoracic, and gynecological procedures.
Clinical Evidence
No clinical data. Substantial equivalence supported by bench testing, including physical characteristics (dimensions, compatibility) and performance characteristics (strength, closure performance).
Technological Characteristics
Materials: Unalloyed titanium (ISO 5832-2) staples, stainless steel, plastic, aluminum alloy. Principle: Mechanical stapling and resection. Energy: Manual activation. Form factor: Handheld surgical instruments. Sterilization: Cobalt-60 radiation (ISO 11137-1). Biocompatibility: ISO 10993-1 compliant.
Indications for Use
Indicated for patients undergoing alimentary tract reconstruction (circular stapler); anorectal surgery for hemorrhoids and wall defects (hemorrhoidal stapler); or tissue closure, resection, transection, and anastomosis in abdominal, gynecological, thoracic, or gastrointestinal procedures (linear and linear cutter staplers).
Regulatory Classification
Identification
An implantable staple is a staple-like device intended to connect internal tissues to aid healing. It is not absorbable.
{0}------------------------------------------------
### Section 5 - 510(k) Summary
### 1. Submission Sponsor
Suzhou Beinuo Medical Equipment Co., Ltd. 158-38 Huashan Road Suzhou High - New District China, 215129 Phone: +86 512 66629925 Fax: +86 512 66626238 Contact: Ms Living Huang, Office Manager
#### 2. Submission Correspondent
Emergo Group Suite 1400 816 Congress Ave. Austin, TX 78701 Cell Phone: 940 390 0961 Office Phone: (512) 327.9997 Fax: (512) 327.9998 Contact: Robert Seiple, RAC, Senior Consultant, QA/RA Email: project.management@emergogroup.com
### 3. Date Prepared
13
November 12, 2013
### 4. Device Identification
Suzhou Beinuo family of Surgical Staplers Trade/Proprietary Name: Common/Usual Name: Staple, Implantable Classification Name: Implantable staple Classification Regulation: 21CFR 878.4750 Product Code: GDW Device Class: Class II Classification Panel: General and Plastic Surgery
### 5. Predicate Devices
The Suzhou Frankenman Surgical Staplers (K101378) which includes:
- · Disposable Alimentary Canal Staplers;
- · Single Use Circular Stapler for Rectal Prolapse and Hemorrhoids;
- · Disposable Reloadable Linear Stapler and Reloads; and
- · Disposable Reloadable Linear Cutter Stapler and Reloads;
JAN 2 4 2014
{1}------------------------------------------------
## EMERGO GROUP
The Frankenman Surgical Staplers are substantially to the Suzhou Beinuo family of surgical staplers. The Suzhou Beinuo surgical staplers are virtually identical to the predicate device(s) described above.
### 6. Device Description
The Suzhou Beinuo Staplers were designed in reference to the general principles of surgical staplers. Each stapler/ instrument is activated by squeezing the handle firmly as far as it will go or by pushing the firing knob as far as it will go. Specifics for each stapler include:
- The Suzhou Beinuo Circular Stapler for Single Use places a circular, double staggered row of titanium staples. Immediately after staple formation, the instrument's central knife blade resects the excess tissue, creating a circular anastomosis. The diameter of the staple line is determined by the selection of the 21mm, 28mm, or 32mm stapler. Note that there are two product codes for this stapler (SBW and SBCS L). The total length of SBW stapler is 420mm and the total length of SBCS L version is 520mm. The staplers are identical except for the length.
- The Suzhou Beinuo Hemorrhoidal Circular Staplers for Single Use places two circular peripheral lines of alternating and overlapping staples, thereby sealing off the rectal tissue above the anal canal. The central circular cutting blade cuts the surplus tissue after the sealing to reconstruct the rectal tissue. The diameter of the staple line is determined by the selection of the 32mm or 34mm stapler.
- The Suzhou Beinuo Linear Staplers and Reloads for Single Use places a double(or triple in the case of the SBF 30B) staggered row of titanium staples used for mechanical suturing and closure of tissue. The Linear Stapler is available in 32mm, 60mm, and 88mm line lengths for use in various applications. The instrument may be reloaded during a single procedure.
- The Suzhou Beinuo Linear Cutter Staplers and Reloads for Single Use delivers two doubled staggered rows of titanium staples and anastomoses the internal tissues and incisions during surgical procedures. The Linear Cutter stapler is available in three staple line lengths (61mm, 81mm, or 101mm). The instrument may be reloaded during a single procedure.
### 7. Intended Use
Suzhou Beinuo surgical staplers are indicated for use as follows:
#### . Circular Stapler for Single Use
The Suzhou Beinuo Circle Stapler for Single Use is used throughout the alimentary tract for the reconstruction of the alimentary tract and the creation of end-to-end, end-toside and side-to-side anastomoses.
{2}------------------------------------------------
# K133499 EMERGO GROUP
- . Hemorrhoidal Circular Stapler for Single Use The Suzhou Beinuo Hemorrhoidal Circular Stapler for Single Use is a Circular Stapler product, with accessories, that is indicated in anorectal surgical procedures for the treatment of hemorrhoids and anorectal wall defects.
- . Linear Stapler and Reloads for Single Use
The Suzhou Beinuo Linear Stapler for Single Use (and reloads) is indicated for the closure of tissue in abdominal, gynecological, and thoracic surgical procedures.
- . Linear Cutter Stapler and Reloads for Single Use
The Suzhou Beinuo Linear Cutter Stapler for Single Use (and reloads) is indicated for use in the resection, transection and anastomosis of tissue in gastrointestinal, gynecological, and thoracic surgical procedures.
### 8. Comparison of Technological Characteristics
The following table compares the Suzhou Beinuo family of surgical staplers to the predicate device with respect to intended use, technological characteristics and principles of operation, providing more detailed information regarding the basis for the determination of substantial equivalence.
| Manufacturer | Suzhou Beinuo | Suzhou Frankenman | Significant Differences |
|------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K101378 | n/a |
| Product Code | GDW | GDW | Identical |
| Regulation Number | 21CFR 878.4750 | 21CFR 878.4750 | Identical |
| Regulation Name | Implantable staple | Implantable staple | Identical |
| Indications for Use | Circular Stapler for Single Use<br>The Suzhou Beinuo Circular<br>Stapler for Single Use is used<br>throughout the alimentary<br>tract for the reconstruction of<br>the alimentary tract and the<br>creation of end-to-end, end-<br>to-side and side-to-side<br>anastomoses. | Disposable Alimentary Canal<br>Stapler<br>The Frankenman Disposable<br>Alimentary Canal Stapler is<br>used throughout the<br>alimentary tract for the<br>creation of end-to-end, end-<br>to-side and side-to-side<br>anastomoses in both open<br>and laparoscopic techniques. | Virtually Identical |
| | Hemorrhoidal Circular Stapler<br>for Single Use<br>The Suzhou Beinuo<br>Hemorrhoidal Circular Stapler<br>for Single Use is a Circular<br>Stapler product, with<br>accessories, that is indicated<br>in anorectal surgical<br>procedures for the treatment | Single Use Circular Stapler for<br>Rectal Prolapse and<br>Hemorrhoid<br>The Frankenman Single Use<br>Circular Stapler for Rectal<br>Prolapse and Hemorrhoids is a<br>Circular Stapler product, with<br>accessories, that has | Essentially identical. Suzhou<br>Beinuo indication and the<br>Frankenman indications<br>discuss the same procedures.<br>No safety or efficacy impact to<br>the difference in wording |
| Manufacturer | Suzhou Beinuo | Suzhou Frankenman | Significant Differences |
| | of hemorrhoids and anorectal<br>wall defects. | application for general surgical<br>treatment of haemorrhoids<br>and anorectal wall defects by<br>means of transanal stapling<br>and resection of mucosal and<br>musculo-mucosal tissue<br>resulting in occlusion of<br>haemorrhoidal inflow,<br>restoring the haemorrhoidal<br>tissue to its normal<br>physiological position. | |
| | Linear Stapler and Reloads<br>for Single Use<br>The Suzhou Beinuo Linear<br>Stapler for Single Use (and<br>reloads) is indicated for the<br>closure of tissue in abdominal,<br>gynecological, and thoracic<br>surgical procedures. | Disposable Reloadable Linear<br>Stapler and Reloads<br>The Frankenman Disposable<br>Reloadable Linear Stapler (and<br>Reloads) is used in the<br>resection or transaction of<br>tissue for abdominal,<br>gynecological, pediatric and<br>thoracic surgical procedures. | Essentially identical. Suzhou<br>Beinuo Indication and the<br>Frankenman indication discuss<br>the same procedures. No<br>safety or efficacy impact to<br>the difference in wording |
| | Linear Cutter Stapler and<br>Reloads for Single Use<br>The Suzhou Beinuo Linear<br>Cutter Stapler for Single Use<br>(and reloads) is indicated for<br>use in the resection,<br>transection and anastomosis<br>of tissue in gastrointestinal,<br>gynecological, and thoracic<br>surgical procedures | Disposable Reloadable Linear<br>Cutter Stapler and Reloads<br>The Frankenman Disposable<br>Reloadable Linear Cutter<br>Stapler (and Reloads) has<br>application in abdominal,<br>gynecological, thoracic and<br>pediatric surgery transaction,<br>resection, and the creation of<br>anastomoses. | Essentially identical. Suzhou<br>Beinuo Indication and the<br>Frankenman indication discuss<br>the same procedures. No<br>safety or efficacy impact to<br>the difference in wording |
| Product<br>Descriptions | Circular Staplers for Single<br>Use<br>The Circular Stapler for Single<br>Use is a surgical device for the<br>reconstruction of alimentary<br>tract with mechanical method<br>to replace traditional hand<br>operation.<br>This device is designed on the<br>principal of staplers. It creates<br>side to end, end to end<br>anastomosis in alimentary<br>canal with peripheral double | The Frankenman Disposable<br>Alimentary Canal Stapler (i.e.,<br>CS Stapler, is primarily<br>composed of plastic, titanium,<br>and stainless steel and is used<br>for the reconstruction and<br>anastomosis in the alimentary<br>canal. These disposable<br>staplers place a circular,<br>double staggered row of<br>titanium (ISO 5832-2) staples<br>and then resect the excess<br>tissue, creating a circular | The products are virtually<br>identical. |
| Manufacturer | Suzhou Beinuo | Suzhou Frankenman | Significant Differences |
| | staggered rows of staples, and<br>cut off the residue tissue with<br>the circular cutting blade in<br>the center to ensure a big<br>enough canal for the<br>reconstructed alimentary<br>tract. By squeezing the firing<br>handle, the staples form and<br>the circular knife blade resects<br>the excess tissue, creating a<br>circular anastomosis. The<br>diameter of the staple line is<br>determined by the selection of<br>the 21mm, 25mm, 28mm, or<br>32mm stapler. | anastomosis. The CS Stapler is<br>activated by squeezing the<br>handle firmly.<br>The outer diameter of the<br>staple line is determined by<br>the selection of the 21mm,<br>25mm, 28mm, or 32mm<br>stapler. The stapler is available<br>in one staple diameter,<br>0.28mm and two shaft<br>lengths, 420mm and 520mm. | |
| | The Circular Stapler for Single<br>Use is made of plastic<br>particles, pure titanium,<br>aluminum alloy and stainless<br>steel, packed in specially<br>designed sterile packs and<br>boxes and sterilized by Cobalt-<br>60. The Circular Stapler for<br>Single Use is used for<br>reconstruction and<br>anastomosis in the alimentary<br>canal. The device has a shelf-<br>life of 3 years. | | |
| | Hemorrhoidal Circular Stapler<br>for Single Use<br>Product description<br>Hemorrhoidal Circular Staplers<br>for Single Use is made of<br>plastic particles, pure<br>titanium, aluminum alloy and<br>stainless steel, packed in<br>specially designed sterile<br>packs and boxes and sterilized<br>by Cobalt-60. The product's<br>shelf life is 3 years. | The Frankenman Single Use<br>Circular Stapler for Rectal<br>Prolapse and Hemorrhoids<br>(i.e., stapled<br>Haemorrhoidopexy (CPH), is<br>primarily manufactured from<br>plastic, titanium, aluminum<br>alloy and stainless steel and is<br>used in the treatment of rectal<br>haemorrhoids and anorectal<br>defects of transanal stapling<br>(otherwise known as staples<br>transanal rectal resection or<br>STARR procedure) and | |
| Manufacturer | Suzhou Beinuo | Suzhou Frankenman | Significant Differences |
| | Hemorrhoidal Circular Staplers<br>for Single Use is used in the<br>treatment of rectal wall<br>defects and internal<br>hemorrhoids during anorectal<br>surgery. The Beinuo<br>Hemorrhoidal Circular Staplers<br>for Single Use places a<br>circular, double staggered<br>rows of titanium staples,<br>thereby sealing off the rectal<br>mucosa or rectum above the<br>anal canal. The central circular<br>cutting blade cuts the surplus<br>tissue after the sealing to<br>reconstruct the rectal mucosa<br>or rectum. The diameter of<br>the staple line is determined<br>by the selection of the 32mm<br>or 34mm stapler. | resection of rectal mucosal<br>and musculo-mucosal tissue<br>resulting in occlusion of<br>haemorrhoidal inflow,<br>restoring the haemorrhoidal<br>tissue to its normal<br>physiological position.<br>Specifically, the rectal mucosa<br>above the anal canal is sealed<br>by the placement of two<br>circular peripheral lines of<br>alternating and overlapping of<br>titanium (ISO 5832-2) staples.<br>The central circular cutting<br>blade cuts the surplus tissue<br>after the sealing to<br>reconstruct the rectal mucosa. | |
| | Linear Staplers and Reloads<br>for Single Use<br><br>Product description<br><br>Linear Stapler and Reloads for<br>Single Use are made of plastic<br>particles, titanium and<br>stainless steel, packed in<br>specially designed sterile<br>packs and boxes and sterilized<br>by Cobalt-60. Linear Staplers<br>and Reloads for Single Use is<br>used in the closure of incision<br>and stump of inner organs in<br>general surgery. The shelf life<br>is 3 years.<br><br>The Beinuo Linear Staplers<br>and Reloads for Single Use<br>places a double staggered row<br>of titanium staples (3 rows for<br>white cartridge only) used for<br>mechanical suturing and<br>closure of tissue. The Linear<br>Stapler is available in 32mm. | The Frankenman Disposable<br>Reloadable Linear Stapler is<br>manufactured primarily from<br>plastic, titanium, and stainless<br>steel. Single Use, reloadable<br>linear staplers are used in the<br>process of mechanical<br>suturing and closure of tissue,<br>prior to the removal of excess<br>tissue. Specific surgical<br>procedures where the LS<br>would be used include<br>general, thoracic,<br>gynecological and colorectal<br>surgeries.<br>The LS places a double<br>staggered row of titanium (ISO<br>5832-2) staples, with the<br>exception of model number<br>LS30W which places three<br>staggered rows of staples. This<br>third row provides additional<br>security for closing vessels<br>where bleeding is a significant<br>risk. | |
| Manufacturer | Suzhou Beinuo | Suzhou Frankenman | Significant Differences |
| | 46mm, 60mm, and 88mm line lengths for use in various applications. The instrument may be reloaded during a single procedure. | The LS is available in 30mm, 45mm, 60mm, and 90mm staple line lengths for use in various applications and three stapler sizes (2.5, 3.8mm and 4.5mm) to accommodate various tissue thicknesses. | |
| | <b>Linear Cutter Staplers and Reloads for Single Use</b><br>Product description<br>Linear Cutter Staplers and Reloads for Single Use are made of plastic particles, pure titanium and stainless steel, packed in specially designed sterile packs and boxes and sterilized by Cobalt-60. Linear Cutter Staplers and Reloads for Single Use is used in transection, resection and suture in GI, gynecological, thoracic surgeries. The shelf is three years.<br>The Beinuo Linear Cutter Staplers and Reloads for Single Use delivers two doubled staggered rows of titanium staples and anastomoses the internal tissues and incisions during surgical procedures. The Linear Cutter stapler is available in three staple line lengths (61mm, 81mm, or 101mm). The instrument may be reloaded during a single procedure. | The Frankenman Disposable Reloadable Linear Cutter Stapler (i.e. LC Stapler The Frankenman LC Disposable Reloadable Stapler delivers two doubled staggered rows of titanium staples and is used to resect and/or anastomose the internal tissues during surgical procedures and reloads are manufactured primarily from plastic, titanium, and stainless steel. The Frankenman Disposable Reloadable Linear Cutter stapler is used for abdominal, gynecological, thoracic, and pediatric surgery for transaction, resection, and the creation of anastomoses | |
| Basic Principle of Operation | The stapler places a circular, double-staggered row of staples and then resects any excess tissue, creating a circular anastomosis. | Stapler places a circular, double staggered row of staples and then resects the excess tissue, creating a circular anastomosis. | Identical |
| Manufacturer | Suzhou Beinuo | Suzhou Frankenman | Significant Differences |
| Material | Stainless steel (staplers) &<br>titanium (staples) | Stainless steel (staplers) &<br>titanium (staples) | Identical |
| Sterile | Yes, radiation sterilized | Yes, radiation sterilized | Identical |
| Single-Use | Staplers are single use.<br>Staplers + reloads may be<br>reloaded with additional<br>staples during the procedure<br>(single patient - no re-use or<br>re-sterilization). | Staplers are single use.<br>Staplers + reloads may be<br>reloaded with additional<br>staples during the procedure<br>(single patient - no re-use or<br>re-sterilization). | Identical |
| Shelf Life | 36 months based on the<br>sterilization validation of the<br>packaging. | 24 months based on<br>sterilization validation of the<br>packaging | The Suzhou Beinuo device has<br>12 months additional shelf life<br>based on sterilization<br>validation data. |
| Biocompatibility | Complies with ISO 10993-1<br>and other pertinent standards | Complies with ISO 10993-1<br>and other pertinent standards | Identical |
### Table 5A - Comparison of Characteristics
{3}------------------------------------------------
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
. . . . . . . . .
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
{4}------------------------------------------------
;
# EMERGO GEROUP
{5}------------------------------------------------
### K133499
### Suzhou Beinuo Traditional 510(k) Premarket Submission Surgical Staplers
# EMERGO GGROUP
{6}------------------------------------------------
{7}------------------------------------------------
Suzhou Beinuo . Traditional 510(k) Premarket Submission Surgical Staplers
### EMERGO GROUP
K133499
### Conclusion:
The Suzhou Beinuo surgical staple device shares the same indications for use, device operation, overall technical and functional capabilities, meets the same standards and therefore are substantially equivalent to the predicate device(s). The Suzhou Beinuo device is similar in design and function to the predicate devices for the modes of operation and use.
### 9. Non-Clinical Performance Data
The Suzhou Beinuo surgical stapler complies with the applicable voluntary standards as shown below:
- . Materials of Construction - ISO 5382-2-1999 - Surgical Instruments - Metallic Materials - Part 2 - Unalloyed titanium
- . Sterilization
- o ISO 11737-1: 2006, Sterilization of medical devices -- Microbiological methods -Part 1, Determination of a population of microorganisms on products
- o ISO 11137-1:2006 , Sterilization of healthcare products Radiation Part 1, Requirements for the development, validation and routine control of a sterilization process for medical devices.
- . Biocompatibility - contact materials were tested per the schema in ISO 10993-1. Specifically:
- O AAMI / ANSI / ISO 10993-5:2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity. (Biocompatibility)
- ISO 10993-10:2010, Biological evaluation of medical devices Part 10: Tests for o irritation and skin sensitization. (Biocompatibility)
- . Risk Management activities were carried out as described in ISO 14971: 2007, Medical devices -- Application of risk management to medical devices
- . Packaging for Terminally Sterilized Devices – ISO 11607:2003, Packaging for Terminally Sterilized Medical Products.
{8}------------------------------------------------
Suzhou Beinuo Traditional 510(k) Premarket Submission Surgical Staplers
### EMERGO GROUP
Additionally, the surgical staplers were evaluated to validate physical characteristics (appearance, dimensions, stapler compatibility with the cartridge) and performance characteristics (strength, closure performance). The Suzhou Beinuo staplers were also evaluated for performance testing compared to the analogous Suzhou Frankenman staplers. Evaluation of the two stapler products similar results (in terms of post-operative healing, pain management, anastomotic leakage, and bleeding) as compared to manual suturing and competitors' devices.
Based on this testing, the design and construction of the Suzhou Beinuo and Suzhou Frankenman staplers were determined to be substantially equivalent. The Suzhou Beinuo family of surgical staplers meets all the requirements for overall design, sterilization, biocompatibility, and clinical utility confirming that the output meets the design inputs and specifications. The Suzhou Beinuo surgical staplers passed all testing and supports the claims of substantial equivalence and safe operation.
The Suzhou Beinuo surgical staplers comply with the applicable voluntary standards for biocompatibility and sterilization. The device passed all the testing in accordance with national and international standards.
### 10. Clinical Testing
There was no prospective clinical studies required to support the medical device as the indications for use, technology, materials of construction are virtually identical to the predicate device. These types of devices, including the predicate devices, have been on the market for many years with a proven safety and efficacy for the use of the device. The nonclinical testing detailed in this submission supports the substantial equivalence of the device.
### 11. Statement of Substantial Equivalence
By definition, a device is substantially equivalent to a predicate device when the device has the same intended use and the same technological characteristics as the previously cleared predicate device.
It has been shown in this 510(k) submission that the difference between the Suzhou Beinuo surgical staplers and the Suzhou Frankenmann surgical staplers do not raise any questions regarding safety and effectiveness. Performance testing and compliance with voluntary standards, demonstrate that the Suzhou Beinuo surgical staplers are substantially equivalent to the relevant aspects of the predicate devices in terms of design, components, materials, principals of operation, sterilization, biocompatibility, performance characteristics, and intended use. The Suzhou Beinuo surgical staplers, as designed and manufactured, are determined to be substantially equivalent to the referenced predicate devices.
{9}------------------------------------------------
Image /page/9/Picture/0 description: The image is a black and white logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with its wings spread, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the upper portion of the circle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
January 14, 2014
Suzhou Beinuo Medical Equipment Co., Ltd. C/O Emergo Group Robert Seiple, RAC, Senior Consultant, QA/RA 816 Congress Avenue, Suite 1400 Austin, Texas 78701
Re: K133499
Trade/Device Name: Suzhou Beinuo Family of Surgical Staplers Regulation Number: 21 CFR 878.4750 Regulation Name: Implantable staple Regulatory Class: Class II Product Code: GDW Dated: November 13, 2013 Received: November 14, 2013
Dear Mr. Seiple:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{10}------------------------------------------------
Page 2 - Robert Seiple, RAC, Senior Consultant, QA/RA
.... . ..
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Binita S. Ashar 2014.01.17403.40:56 -05'00'
Binita Ashar, MD, MBA, FACS Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{11}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
510(k) Number (if known) K133499
Device Name
Suzhou Beinuo Surgical Staplers
Indications for Use (Describe)
Suzhou Beinuo surgical staplers are indicated for use as follows:
· Circular Stapler for Single Use
The Suzhou Beinuo Circular Staple Use is used throughout the alimentary tract for the reconstruction of the alimentary tract and the creation of end-to-end, end-to-side and side-to-side anastomoses.
· Hemorrhoidal Circular Stapler for Single Use
The Suzhou Beinuo Hemorthoidal Circular Staple Use is a Circular Stapler product, with accessories, that is indicated in anorectal surgical procedures for the treatment of hemorrhoids and anorectal wall defects.
· Linear Stapler and Reloads for Single Use
The Suzhou Beinuo Linear Stapler and Reloads for Single Use (and reloads) is indicated for the closure of tissue in abdominal, gynecological, and thoracic surgical procedures.
· Linear Cutter Stapler and Reloads for Single Use
The Suzhou Beinuo Linear Cutter Staple Use (and reloads) is indicated for use in the resection, transection, transection and anastomosis of tissue in gastrointestinal, gynecological, and thoracic surgical procedures.
Type of Use (Select one or both, as applicable)
X Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.