PORTABLE THERAPEUTIX SQUID ACTIVE COLD COMPRESSION DEVICE AND COLD PACK FOR OTC USE
K133483 · Portable Therapeutix · IRP · Mar 11, 2014 · Physical Medicine
Device Facts
Record ID
K133483
Device Name
PORTABLE THERAPEUTIX SQUID ACTIVE COLD COMPRESSION DEVICE AND COLD PACK FOR OTC USE
Applicant
Portable Therapeutix
Product Code
IRP · Physical Medicine
Decision Date
Mar 11, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 890.5650
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Squid Active Cold Compression device and Cold Pack for OTC use is indicated for the temporary relief of minor muscle aches and pains. The compression device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneeding and stroking of tissues using an inflatable garment. The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
Device Story
Device combines intermittent pneumatic compression with cold therapy. Input: user-selected intensity setting (1-30, 2-50, 3-70, or 4-85 mm Hg). Operation: microprocessor-controlled pump inflates garment air bladders to simulate kneading/stroking; exhaust valve manages deflation. Output: mechanical compression and cold application via reusable thermogel pack. Used in home/OTC setting by patient. Healthcare provider/patient uses output to manage minor muscle aches and improve circulation. Benefits: non-invasive relief of minor aches and localized cold therapy.
Clinical Evidence
No clinical data. Evidence consists of bench testing (baseline verification, compression, bladder leak, valve tests, pressure gauge data, shelf life) and a summative usability study validating OTC use by representative users.
Technological Characteristics
Portable pump-controller with pneumatic compressor; AC/DC or lithium battery power. Wrap with air bladder and Velcro attachments; reusable thermogel cold pack. Pressure range 0-85 mm Hg; 4 intensity settings. Microprocessor-based pressure control. Complies with IEC 60601-1 and IEC 60601-1-2.
Indications for Use
Indicated for healthy individuals requiring temporary relief of minor muscle aches and pains, temporary increase in local circulation, and localized cold therapy. Applicable to leg, foot, arm, shoulders, lower back, and hands.
Regulatory Classification
Identification
A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.
Predicate Devices
Portable Therapeutix Squid Active Cold Compression Cold Pack (K123829)
Submission Summary (Full Text)
{0}------------------------------------------------
# MAR 1 1 2014
## Traditional 510(k) Summary as required by 21 CFR 807.92(a) K133483
| A) Submitted by: | Portable Therapeutix<br>4254 Bluebonnet Dr<br>Stafford, TX 77477<br>1-617-331-7524 |
|----------------------------------|-------------------------------------------------------------------------------------------------|
| Contact: | Sharyn Orton, Ph.D.<br>MEDIcept, Inc.<br>200 Homer Ave<br>Ashland, MA 01721<br>401-330-8264 |
| Date prepared: | January 31, 2014 |
| B) Classification Name: | Massager, Powered Inflatable Tube - Product code IRP<br>Pack, Cold, Reusable - Product code IME |
| Common Name: | Powered inflatable tube massager<br>Cold Pack |
| Proprietary Name: | Portable Therapeutix Squid Active Cold Compression device<br>and Cold Pack for OTC Use |
| Device Regulations:<br>and Class | 21 CFR 890.5650, Class II<br>21 CFR 890.5700, Class I 510(k) exempt |
| Product Codes: | IRP; IME |
| C) Predicates: | K123829 Portable Therapeutix Squid Active Cold<br>Compression Cold Pack |
D) Device Description:
The Squid Active Cold Compression device and Cold Pack for OTC Use combines intermittent compression with cold therapy. The Squid simulates kneading and stroking of tissues using an inflatable garment attached to a gel ice pack and connected to a preprogrammed air pump. The Squid may be used for the leg, foot, feet, arm, shoulders, lower back, and hands.
6-1
{1}------------------------------------------------
The device is manufactured with the following components:
- 1. A portable external pump-controller unit containing the pneumatic compressor (air pump) that may be run on AC-DC external power supply or a built-in lithium battery.
- 2. A wrap that contains an air bladder with sequential compression capability and Velcro attachments for the cold pack. The wrap connects to the pump controller via a flexible tube. There may also be an accessory piece to the wrap to provide an additional securing mechanism. Wraps come in two configurations and three sizes.
- 3. Reusable thermogel cold pack
The Squid Active Cold Compression device and Cold Pack was cleared for marketing as a prescription use device in K123829. The Squid Active Cold Compression device and Cold Pack described in this 510(k) Summary will be marketed as an over-the-counter (OTC) use device.
E) Intended Use/Indication for Use:
The Squid Active Cold Compression device and Cold Pack for OTC use is indicated for the temporary relief of minor muscle aches and pains.
The compression device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneeding and stroking of tissues using an inflatable garment.
The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
F) Comparison to Predicate Device(s):
The Portable Therapeutix Squid Active Cold Compression device and Cold Pack for OTC use is the same device as K123829. There were no changes to the K123829 device.
{2}------------------------------------------------
| | Portable Therapeutix Squid<br>Active Cold Compression device<br>and Cold Pack for OTC Use | Portable Therapeutix Squid<br>Active Cold Compression device<br>and Cold Pack<br>K123829 |
|-----------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Product code | IRP; IME | IRP; IME |
| Indication for<br>Use | The Squid Active Cold<br>Compression device and Cold<br>Packfor OTC use is indicated for<br>the temporary relief of minor<br>muscle aches and pains. | The Squid Active Cold<br>Compression device and Cold<br>Pack is indicated for the<br>temporary relief of minor muscle<br>aches and pains. |
| | The compression device is<br>indicated for temporary increase<br>in circulation of the treated areas<br>in people who are in good health,<br>and simulates kneading and<br>stroking of tissues using an<br>inflatable garment. | The compression device is<br>indicated for temporary increase<br>in circulation of the treated areas<br>in people who are in good health,<br>and simulates kneading and<br>stroking of tissues using an<br>inflatable garment. |
| | The cold pack is indicated for<br>localized therapy in situations<br>where cold temperature therapy<br>is necessary or desirable. | The cold pack is indicated for<br>localized therapy in situations<br>where a physician determines<br>that cold temperature therapy is<br>necessary or desirable. |
| Target | Leg, foot, arm, shoulders, lower back, hands | |
| OTC use | Yes | No |
.
.
:
{3}------------------------------------------------
| | Portable Therapeutix Squid<br>Active Cold Compression device<br>and Cold Pack | Portable Therapeutix Squid<br>Active Cold Compression<br>device and Cold Pack<br>K123829 |
|-------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------|
| Principle of<br>Operation/<br>Mechanical<br>characteristics | Intermittent compression<br>4 intensity settings (modes)<br>1 – 30 mm Hg<br>2 – 50 mm Hg<br>3 – 70 mm Hg<br>4 – 85 mm Hg | |
| Pressure<br>range | 0 – 85 mm Hg | |
| Total<br>treatment<br>time | 15 minutes | |
| Pressure<br>control | Microprocessor and pressure sensor | |
| Inflation by | Pressurization pump | |
| Deflation by | Exhaust valve | |
| Power source | 120V 60 Hz, consumption 26W; AC adapter: 120V 60 Hz,<br>consumption 36W<br>Lithium battery | |
| Cold Pack | Yes | |
## Substantial Equivalence Discussion
The Portable Therapeutix Squid Cold Compression System for OTC use is the same device as the FDA cleared Squid. Treatment targets, total treatment time and mechanical intermittent compression are the same. There is no change to the software, materials used, or power source. There is the addition of alignment markers on the wraps and gel pack to assist in the aligning the wrap with the gel pack for Velcro attachment.
The Portable Therapeutix Squid Cold Compression System has the same indications for use as the predicate device, with the addition of OTC use.
## Performance Testing
There were no changes to the K123829 device. The following were described or testing successfully performed and reported in FDA cleared K123829:
- . Biocompatibility
- Electrical Safety/EMC (including batteries) .
- Bench testing including the following tests: Baseline Verification;Compression . Bladder Leak Verification Test; Blow out Valve Test; Gel Pack Seal Leak Test; Low Pressure Verification Test; Max Pressure, -Total Time, -Max Compression Time, -
Portable Therapeutix OTC Squid Traditional 510(k) Premarket Notification
{4}------------------------------------------------
Min Compression Time, -Max Deflation Time, -Min Deflation Time, -Pressure Gauge Data; Solenoid Valve Release Test (High pressure release)
- . Shelf life of batteries and gel pack
- Reprocessing/sterilization the device is reusable and not supplied sterile. . Cleaning instructions are included in the Instructions for Use.
For this application (K133483) a summative Usability Study was successfully conducted on representative users for OTC use. Findings from the study served to validate that the Squid - Active Cold Compression device and Cold Pack is usable by the devices' intended users.
## Conformity to Standards
There is no change in conformance standards from K123829. The Squid complies with:
- IEC 60601-1:1988 + A1:1991 + A2:1995 Medical Electrical Equipment Part 1: . General Requirements for Safety, 1988; Amendment 1, 1991-11; Amendment 2, 1995).
- IEC 60601-1-2:2007 Medical Electrical Equipment Part 1-2: General . requirements for basic safety and essential performance - Collateral standard: Electromagnetic compatibility - Requirements and Tests
## Conclusion
The Portable Therapeutix Squid Cold Compression System device and Cold Pack for OTC use is substantially equivalent to the predicate device.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
March 11. 2014
Portable Therapeutix, LLC c/o MEDIcept, Inc. Attn: Sharyn Orton Ph.D., Senior Consultant 200 Homer Avenue Ashland, MA 01721
Re: K133483
> Trade Name: Portable Therapeutix Squid Active Cold Compression Device and Cold Pack for OTC use Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: Class II Product Code: IRP, IME Dated: February 5, 2014 Received: February 7, 2014
Dear Dr. Orton:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
Page 2 of 3 - Sharyn Orton, Ph.D.
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events} (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Carlos L. Pena -S
Carlos L. Peña, PhD, MS Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
#### 510(k) Number (if known) K133483
#### Device Name
Portable Therapeutix Squid Active Cold Compression device and Cold Pack for OTC use
#### Indications for Use (Describe)
The Squid Active Cold Compression device and Cold Pack for the temporary relief of minor muscle aches and pains.
The compression device is indicated for temporary increase in circulation of the treated areas in people who are in good health, and simulates kneeding and stroking of tissues using an inflatable garment.
The cold pack is indicated for localized therapy in situations where cold temperature therapy is necessary or desirable.
Type of Use (Select one or both, as applicable)
[ ] Prescription Use (Part 21 CFR 801 Subpart D)
2 Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
## Carlos L. Pena -S
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.