K133482 · Globus Medical, Inc. · OWI · May 21, 2014 · Orthopedic
Device Facts
Record ID
K133482
Device Name
SILC FIXATION SYSTEM
Applicant
Globus Medical, Inc.
Product Code
OWI · Orthopedic
Decision Date
May 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3010
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The SILC™ Fixation System consists of temporary implants for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications: Spinal trauma surgery, used in sublaminar, interspinous or facet wiring techniques: Spinal reconstructive surgery, incorporated into constructs for the purpose of correction of spinal deformities such as idiopathic and neuromuscular scoliosis in patients 8 years of age and older, adult scoliosis, kyphosis, and spondylolisthesis; Spinal degenerative surgery, as an adjunct to spinal fusions. The SILC™ Fixation System may also be used in conjunction with other medical implants made of similar metals whenever "wiring" may help secure the attachment of other implants.
Device Story
SILC™ Fixation System provides temporary bone stabilization during orthopedic surgery. System comprises PET bands, titanium-tipped anchors, and clamps mating with 4.5mm-6.5mm rods; includes manual surgical instruments. Used by surgeons in spinal trauma, reconstructive, and degenerative procedures to secure bone or other implants. Operates as mechanical bone anchor to facilitate solid bony fusion or fracture repair. Surgeon manually installs system to provide structural support during healing. Benefits include temporary stabilization of spinal constructs and bone segments.
Clinical Evidence
Bench testing only. Mechanical testing performed in accordance with ASTM F1798 and FDA guidance for spinal systems, including static band tension, static and dynamic band pull-through, static axial grip, and static torsion grip. No clinical data presented.
Technological Characteristics
Materials: Polyethylene terephthalate (PET) bands with commercially pure titanium tips (ASTM F67); clamps made of titanium alloy, cobalt chromium molybdenum alloy, or stainless steel (ASTM F136, F138, F1295, F1472, F1537). Mechanical fixation via bands and clamps for 4.5mm-6.5mm rods. Manual surgical instrumentation. Galvanic corrosion warning for dissimilar metal combinations.
Indications for Use
Indicated for patients 8 years of age and older requiring temporary bone stabilization during bony fusion or fracture repair. Applications include spinal trauma (sublaminar, interspinous, facet wiring), spinal reconstruction (idiopathic/neuromuscular scoliosis, adult scoliosis, kyphosis, spondylolisthesis), and spinal degenerative surgery as an adjunct to fusion. Contraindicated for use with dissimilar metals due to galvanic corrosion risk.
Regulatory Classification
Identification
A bone fixation cerclage is a device intended to be implanted that is made of alloys, such as cobalt-chromium-molybdenum, and that consists of a metallic ribbon or flat sheet or a wire. The device is wrapped around the shaft of a long bone, anchored to the bone with wire or screws, and used in the fixation of fractures.
Predicate Devices
Zimmer Spine Universal Clamp Spinal Fixation System (K060009)
{0}------------------------------------------------
# 510(k) Summary: SILC™ Fixation System
# MAY 2 1 2014
| Company: | Globus Medical Inc.<br>2560 General Armistead Ave.<br>Audubon, PA 19403 |
|--------------|-------------------------------------------------------------------------|
| 610-930-1800 | |
Christina Kichula Contact: Group Manager, Regulatory Affairs
- Date Prepared: November 12, 2013
SILC™ Fixation System Device Name:
Per 21 CFR as follows: Classification: &888.3010 Bone Fixation Cerclage, Sublaminar Product Codes OWI Regulatory Class II, Panel Code 87
- Zimmer Spine Universal Clamp Spinal Fixation System, Predicate(s): K060009. K110348 Medicrea LigaPASS, K112736 Pioneer Songer Cable System, K922952
# Purpose:
The purpose of this submission is to request clearance of the SILC™ Fixation System.
# DEVICE DESCRIPTION:
The SILC™ Fixation System consists of bands, clamps that mate with 4.5mm-6.5mm diameter rods, and associated manual surgical instruments. Bands are manufactured from polyethylene terephthalate (PET) with commercially pure titanium tips per ASTM F67, and clamps are manufactured from titanium alloy, cobalt chromium molybdenum alloy, or stainless steel, as specified in ASTM F136, F138, F1295, F1472, and F1537. Due to the risk of galvanic corrosion following implantation, stainless steel implants should not be connected to titanium, titanium alloy, or cobalt chromium molybdenum alloy implants.
# INDICATIONS FOR USE:
The SILC™ Fixation System consists of temporary implants for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications:
- Spinal trauma surgery, used in sublaminar, interspinous or facet wiring . techniques:
- Spinal reconstructive surgery, incorporated into constructs for the purpose . of correction of spinal deformities such as idiopathic and neuromuscular
{1}------------------------------------------------
scoliosis in patients 8 years of age and older, adult scoliosis, kyphosis, and spondylolisthesis;
- Spinal degenerative surgery, as an adjunct to spinal fusions. .
The SILC™ Fixation System may also be used in conjunction with other medical implants made of similar metals whenever "wiring" may help secure the attachment of other implants.
### Performance Data:
Mechanical testing (static band tension, static and dynamic band pull-through, static axial grip, and static torsion grip) was conducted in accordance with ASTM F1798 and the "Guidance for Industry and FDA Staff, Guidance for Spinal System 510(k)s," May 3, 2004. Performance data demonstrate substantial equivalence to the predicate devices.
#### Conclusions / Basis of Substantial Equivalence:
The SILC™ Fixation System implants are similar to the predicate implants with respect to technical characteristics, performance, and intended use. The information provided within this premarket notification supports substantial equivalence to the predicate devices. The SILC™ Fixation System performs as well as or better than the predicate devices.
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
May 21, 2014
Globus Medical. Incorporated Ms. Christina Kichula Group Manager, Regulatory Affairs 2560 General Armistead Avenue Audubon, Pennsylvania 19403
Re: K133482
Trade/Device Name: SILC™ Fixation System Regulation Number: 21 CFR 888.3010 Regulation Name: Bone fixation cerclage Regulatory Class: Class II Product Code: OWI Dated: April 21, 2014 Received: April 22, 2014
#### Dear Ms. Kichula:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
# · Page 2 - Ms. Christina Kichula
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthonedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
# Indications for Use Statement
K133482 510(k) Number:
SILC™ Fixation System Device Name:
# INDICATIONS:
The SILC™ Fixation System consists of temporary implants for use in orthopedic surgery. The system is intended to provide temporary stabilization as a bone anchor during the development of solid bony fusion and aid in the repair of bone fractures. The indications for use include the following applications:
- Spinal trauma surgery, used in sublaminar, interspinous or facet wiring . techniques:
- Spinal reconstructive surgery, incorporated into constructs for the purpose ● of correction of spinal deformities such as idiopathic and neuromuscular scollosis in patients 8 years of age and older, adult scoliosis, kyphosis, and spondylolisthesis;
- Spinal degenerative surgery, as an adjunct to spinal fusions. .
OR
The SILC™ Fixation System may also be used in conjunction with other medical implants made of similar metals whenever "wiring" may help secure the attachment of other implants.
x Prescription Use (Per 21 CFR §801.109)
Over-The-Counter Use__________________________________________________________________________________________________________________________________________________________
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
James P. Bertram -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K133482
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.