K133478 · Shijiazhuang Hongzan Plastic Technology Co.,Ltd · LZA · Jun 2, 2014 · General Hospital
Device Facts
Record ID
K133478
Device Name
POWDER FREE POLYETHYLENE EXAMINATION GLOVES
Applicant
Shijiazhuang Hongzan Plastic Technology Co.,Ltd
Product Code
LZA · General Hospital
Decision Date
Jun 2, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
Device Story
Powder Free Polyethylene Examination Gloves are disposable, non-sterile, loose-fitting gloves made of translucent low-density polyethylene. Used in clinical settings by healthcare examiners to provide a flexible barrier film; prevents contamination between patient and examiner. Device is single-use. Performance meets ASTM D5250-06 (2011) standards for tensile strength, elongation, and freedom from pinholes. No clinical testing performed; substantial equivalence established via non-clinical bench testing including biocompatibility (skin irritation and sensitization assays).
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including physical property testing (tensile strength, elongation, dimensions), freedom from pinholes (ASTM D5250-06), residual powder testing (ASTM D6124-06), and biocompatibility testing (primary skin irritation and dermal sensitization per ISO 10993-10).
Technological Characteristics
Material: Low Density Polyethylene. Form factor: Loose-fitting, translucent (clear) disposable glove. Standards: ASTM D5250-06 (2011), ASTM D6124-06. Biocompatibility: ISO 10993-10. Non-sterile. No powder.
Indications for Use
Indicated for use as a disposable, non-sterile patient examination glove worn on the examiner's hand or finger to prevent cross-contamination between patient and examiner.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Predicate Devices
AmerCare Inc. C2 Powder Free Polyethylene Examination Glove (K113639)
Submission Summary (Full Text)
{0}------------------------------------------------
Donggao Industrial Zone, Zanhuang, Hebei, China 050000
Product: Powder Free Polyethylene Examination Gloves
JUN - 2 2014
### 510(K) SUMMARY
This summary of 510(K) safety and effectiveness information is being submitted in accordance with requirements of 21 CFR §807.92.
The assigned 510(K) number is: K133478
### l. Owner's Identification:
Ms. Huizhen Qu Shijiazhuang Hongzan Plastic Technology Co., Ltd Donggao Industrial Zone, Zanhuang, Hebei, China 050000
Tel: 86-311-83601854 Fax: 86-311- 83616934
Contact: Ms. Kathy Liu, Project Manager Address: 3973 Schaefer Ave., Chino, CA 91710 Tel: 909-590-1611 Fax: 909-590-1511 Date Summary Prepared: April 19, 2014
### 2. Name of the Device:
Trade Name: Powder Free Polyethylene Examination Gloves Common Name: Exam Gloves Classification Name: Patient Examination Glove Classification Regulation: 880.6250 Classification Panel: 880 General Hospital and Personal Use Product Code: LZA Device Class: Class I
### 3. Predicate Device Information:
AmerCare Inc. C2 Powder Free Polyethylene Examination Glove (K113639)
### 4. Device Description:
Powder Free Polyethylene Examination Gloves are Patient Examination Gloves, Disposable, single use only and non-sterile. The gloves are made of translucent (clear) Low Density Polyethylene materials and are powder free. The gloves are loose fitting. The physical and performance characteristics of the devices meet all requirements of ASTM standard D-5250-06 (2011) Standard Specification for Poly(vinyl Chloride) Gloves for Medical Application.
{1}------------------------------------------------
Donggao Industrial Zone, Zanhuang, Hebei, China 050000
Product: Powder Free Polyethylene Examination Gloves
### 5. Intended Use of the Device:
A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### 6. Technological Characteristics and Substantial Equivalence:
Shijiazhuang Hongzan Plastic Technology Co., Ltd.'s Powder Free Polyethylene Examination Gloves is substantially equivalent in safety and effectiveness to the AmerCare Inc.'s C2 Powder Free Polyethylene Examination Glove (K113639). The subject device and predicate device use a similar plastic flexible barrier film to achieve a device for the intended use.
And the properties between the subject device and the predicate device are compared in the following table: Device Performance Result of Characteristics Standard Dradicate Device Subject Device omnarison
| Product Code | | Predicate Device | Subject Device | comparison |
|---------------------|---|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|
| Product Code | / | LZA | LZA | Substantial<br>equivalence |
| Intended Use | / | Predicate device is<br>a disposable device<br>intended for<br>medical purpose<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient<br>and examiner. | Subject device is<br>disposable non-<br>sterile device<br>intended for<br>medical purpose<br>that is worn on the<br>examiner's hand or<br>finger to prevent<br>contamination<br>between patient<br>and examiner. | Substantial<br>equivalence |
| Labeling | / | There are no<br>special labeling<br>claims and do not<br>claim gloves as<br>hypoallergenic on<br>labels. | There are no<br>special labeling<br>claims and do not<br>claim gloves as<br>hypoallergenic on<br>labels. | Substantial<br>equivalence |
| Device<br>Materials | / | Polyethylene | Polyethylene | Substantial<br>equivalence |
| Color | / | Translucent (Clear) | Translucent (Clear) | Substantial<br>equivalence |
{2}------------------------------------------------
Donggao Industrial Zone, Zanhuang, Hebei, China 050000
| | Standard | Device Performance | | Result of | |
|--------------------------------------------------------------------------|------------------------------------------|--------------------|--------------------------------------------|--------------------------------------------|----------------------------|
| Characteristics | | Predicate Device | Subject Device | comparison | |
| Device tolerances and specifications & Performance Data: | | | | | |
| Tensile<br>strength:<br>before<br>and<br>after<br>aging | ASTM<br>D5250-06 (2011) | Meets | Meets | Substantial<br>equivalence | |
| Ultimate<br>elongation: before<br>and after aging | ASTM<br>D5250-06 (2011) | Meets | Meets | Substantial<br>equivalence | |
| Freedom<br>from<br>pinholes | ASTM<br>D5250-06 (2011) | Meets | Meets | Substantial<br>equivalence | |
| Dimensions:<br>Overall<br>length,<br>Width, Palm and<br>Finger thickness | ASTM<br>D5250-06 (2011) | Meets | Meets | Substantial<br>equivalence | |
| Residual powder | ASTM<br>D5250-06 (2011)<br>ASTM D6124-06 | Meets | Meets | Substantial<br>equivalence | |
| Biocompatibility | | | | | |
| Primary skin irritation test | | | Passes<br>Not a primary skin<br>irritation | Passes<br>Not a primary skin<br>irritation | Substantial<br>equivalence |
| Dermal sensitization assay | | | Passes<br>Not a dermal<br>sensitization | Passes<br>Not a dermal<br>sensitization | Substantial<br>equivalence |
Product: Powder Free Polyethylene Examination Gloves
Shijiazhuang Hongzan Plastic Technology Co., Ltd's Powder Free Polyethylene Examination Gloves shares the same technology characteristics compared to the predicate device. The subject device performs according to the glove performance standards ASTM D5250-06(2011), biocompatibility requirement and FDA requirements and the labeling claims for the product. It performs as well as the legally marketed predicate device.
### 7. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence are as Follows:
| Characteristics | Applicable<br>Recognized Standards | Performance Results |
|---------------------|------------------------------------------|---------------------|
| Dimensions | ASTM D 5250-06 (2011) | Meets |
| Physical Properties | ASTM D 5250-06 (2011) | Meets |
| Freedom from holes | ASTM D 5250-06 (2011)<br>FDA 21CFR800.20 | Meets |
{3}------------------------------------------------
Donggao Industrial Zone, Zanhuang, Hebei, China 050000
| Characteristics | Applicable FDA-Recognized Standards | Performance Results |
|---------------------------------------------------|-------------------------------------------------------------|---------------------|
| Residual Powder Test | ASTM D 5250-06 (2011)<br>ASTM D6124-06<br>(Reapproved 2011) | Meets |
| Primary Skin Irritation<br>and Skin Sensitization | ISO 10993 Part 10 | Meets |
Product: Powder Free Polyethylene Examination Gloves
#### 8. Discussion of Clinical Tests Performed:
Not Applicable - There is no hypoallergenic Claim. There was no clinical testing required to support the medical device as the indications for use is equivalent to the predicate device. The substantial equivalence of the device is supported by the nonclinical testing.
#### 9. Labeling:
There are no special labeling claims and we do not claim our gloves as hypoallergenic on our labels.
### 10. Conclusions:
Shijiazhuang Hongzan Plastic Technology Co., Ltd.'s Powder Free Polyethylene Examination Gloves conform fully to ASTM D-5250-06(2011) standard as well as applicable 21 CFR references, and meets pinhole FDA requirements, biocompatibility requirements and labeling claims as shown by data discussed above. There are no safety/efficacy issues or new claims from the "substantial equivalence" products cited. Drawn from the complete list of non-clinical tests, the device herein mentioned is as safe, as effective, and performs as well as the predicate device.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized image of an eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
June 2, 2014
Shijiazhuang Hongzan Plastic Technology Company, Limited C/O Ms. Kathy Liu Official Correspondent Hongray USA Medical Products Incorporated 3973 Schaefer Avenue Chino, CA 91710
Re: K133478
Trade/Device Name: Powder Free Polyethylene Examination Gloves Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LZA Dated: April 19, 2014 Received: April 30, 2014
Dear Ms. Liu:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Ms. Liu
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (2) CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.ida.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
, `
Sincerely yours,
Tejashri Purohit-Sheth, M.D.
Tejashri Purohit Sheth, M.D. Clinical Deputy Director
DAGRID/ODE/CDRH FOR
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133478
#### Device Name
Powder Free Polyethylene Examination Gloves
#### Indications for Use (Describe)
A patient examination glove is disposable non-sterile device intended for medical purpose that is worn on the examiner's hand or finger to prevent contamination between patient and examiner.
### Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
THE ON THE FOR MENT FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR FOR CONSTITUTION OF THE CONSTITUTION OF THE CONTRACT THE CONTRACT THE CONTRACT THE CONTRACT THE CONTRACT THE CONTR
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Image /page/6/Picture/13 description: The image shows the name "Sreekanth Gutala" in large, bold font on the left side. To the right of the name is a digital signature, including the text "Digitally signed by Sreekanth Gutala -S". The signature also includes information such as "o=U.S. Government, ou=HHS, ou=FDA" and a date and time stamp of "2014.05.30 16:33:43 -04'00'". The signature appears to be a formal, digital authentication of the name.
FORM FDA 3881 (9/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
# *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete line to review the collections. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
. . . .
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.