FLASH AR; FLASH AR QUICK

K133436 · 3M Deutschland GmbH · ELW · Mar 7, 2014 · Dental

Device Facts

Record IDK133436
Device NameFLASH AR; FLASH AR QUICK
Applicant3M Deutschland GmbH
Product CodeELW · Dental
Decision DateMar 7, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3660
Device ClassClass 2

Indications for Use

Impressions for the production of temporary restorations . All types of preliminary impressions . Impressions of the opposing jaw . Impressions for orthodontic models .

Device Story

Flash AR and Flash AR Quick are medium-bodied (ISO Type 2) vinyl polysiloxane (A-silicone) impression materials. They function as two-component systems (base paste and catalyst) designed to reproduce dental structures. Unlike predicate devices that use automated mixing systems, these materials are delivered in 50 ml Garant cartridges and mixed/extruded via a manual Garant dispenser at a 1:1 volume ratio. Used by dental professionals in clinical settings to capture preliminary impressions, the material provides a physical mold of the patient's teeth and oral anatomy. This output allows clinicians to fabricate temporary restorations, orthodontic models, or study casts, facilitating subsequent dental treatment planning and restorative procedures.

Clinical Evidence

Bench testing only. Performance evaluated per ISO 4823 standards. Results demonstrate substantial equivalence to predicate devices regarding working time, intra-oral setting time, and physical/mechanical properties. Biocompatibility assessments performed per FDA and international guidelines confirm safety for intended use.

Technological Characteristics

Medium-bodied (ISO Type 2) vinyl polysiloxane (A-silicone) impression material. Two-component system (base paste/catalyst). Delivered in 50 ml Garant cartridges for manual 1:1 volume ratio dispensing. Biocompatible composition.

Indications for Use

Indicated for dental patients requiring preliminary impressions, including impressions for temporary restorations, opposing jaw impressions, and orthodontic models.

Regulatory Classification

Identification

Impression material is a device composed of materials such as alginate or polysulfide intended to be placed on a preformed impression tray and used to reproduce the structure of a patient's teeth and gums. The device is intended to provide models for study and for production of restorative prosthetic devices, such as gold inlays and dentures.

Special Controls

*Classification.* Class II (Special Controls).

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K138486 ## 510(k) SUMMARY OF SAFETY AND EFFECTIVENESS MAR - 7 2014 ### Submitter | Company: | 3M Deutschland GmbH | |-----------------------------------|---------------------------------------------------------| | Street: | ESPE Platz | | ZIP-Code, City: | D-82229 Seefeld | | Federal State: | Bavaria | | Country: | Germany | | Establishment Registration Number | 9611385 | | Official Correspondent: | Dr. Desi W. Soegiarto,<br>Regulatory Affairs Specialist | | Phone: | +49-8152-700 1169 | | Fax: | +49-8152-700 1869 | | E-mail: | desi.soegiarto@mmm.com | | Date: | November 05, 2013<br>(revised on December 02, 2013) | ### Name of Devices | Proprietary Name: | Flash AR | |----------------------|----------------------------| | | Flash AR Quick | | Classification Name: | Impression material | | Common Name: | Dental impression material | ### Predicate Devices Flash AR Penta™, Flash AR Penta™ Quick by 3M Deutschland GmbH, Germany .............................................................................................................................................. . Amendment 510(k) Submission {1}------------------------------------------------ ### Description for the Premarket Notification Flash AR and Flash AR Quick are classified as impression materials (21 C.F.R. § 872.3660) because they are devices intended to reproduce the structure of a patient's teeth. As the predicate devices Flash AR Penta™ and Flash AR Penta™ Quick (K131404), Flash AR and Flash AR Quick are medium-bodied (ISO Type 2) consistency A-silicone impression materials for all kinds of preliminary impressions. Like Flash AR Penta™ and Flash AR Penta™ Quick, Flash AR and Flash AR Quick are two component (base paste/catalyst) vinyl polysiloxane impression materials to be used for impressions, where typically alginates are used. Whereas Flash AR Penta™ and Flash AR Penta™ Quick are designed to automatically be mixed and dispensed in all versions of Pentamix™ devices of 3M Deutschland GmbH and the mixing ratio for both materials is base paste:catalyst, 5:1 (by volume), Flash AR and Flash AR Ouick will be delivered in Garant™ cartridges (50 ml plastic cartridges) and can be mixed and extruded using the Garant™ dispenser manufactured by 3M Deutschland GmbH with the mixing ratio for both materials of base paste:catalyst, 1:1 (by volume). In this 510(k) premarket notification Flash AR and Flash AR Quick have been compared to the predicate devices with regard to indications for use, physical and mechanical properties, and chemical composition. Test results according to ISO 4823 showed that Flash AR and Flash AR Quick are substantially equivalent to the predicate devices Flash AR Penta and Flash AR Penta Quick. Flash AR has same working time and intra-oral setting time as its predicate device Flash AR Penta, whereas working time and intra-oral setting time of Flash AR Quick are the same as those of its predicate device Flash AR Penta Quick. The indications for use of Flash AR and Flash AR Quick are the same as those of the predicate devices Flash AR Penta and Flash AR Penta Quick. Comparison for indications for use, performance data, and chemistry shows that Flash AR and Flash AR Quick are substantially equivalent to the predicate devices. {2}------------------------------------------------ Biocompatibility testing was carried out. Biocompatibility evaluations have been performed for Flash AR and Flash AR Quick in consideration of FDA & internationally recognized guidelines. The conclusion of the assessments is that Flash AR and Flash AR Quick materials are biocompatible for its intended use. In summary, it can be concluded that Flash AR and Flash AR Quick are substantially equivalent in safety and effectiveness as the predicate devices Flash AR Penta™ and Flash AR PentaTM Quick by 3M Deutschland GmbH, Germany (K131404). #### Indications for Use: - Impressions for the production of temporary restorations . - . All types of preliminary impressions - . Impressions of the opposing jaw - Impressions for orthodontic models . {3}------------------------------------------------ DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized symbol resembling an eagle or bird in flight, composed of three curved lines. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol. The logo is black and white. #### Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - W066-G609 Silver Spring, MD 20993-0002 March 7, 2014 3M Deutschland GmbH Dental Products Desi Soegiarto, PhD Regulatory Affairs Specialist ESPE Platz D-82229 Seefeld GERMANY Re: K133436 Trade/Device Name: Flash AR and Flash AR Quick Regulation Number: 21 CFR 872.3660 Regulation Name: Impression Material Regulatory Class: II Product Code: ELW Dated: December 3, 2013 Received: December 6, 2013 Dear Dr. Soegiarto: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent ffor the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Dr. Soegiarto Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Sincerely yours, Image /page/4/Picture/7 description: The image contains the text "Erin FReith -S". The text is written in a bold, sans-serif font. The word "FReith" is stylized with a decorative border around the letters. The letters are all capitalized, and the text is left-aligned. Erin I. Keith, M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure. {5}------------------------------------------------ # Indications for Use 510(k) Number (if known): K133434 Device Name: Flash AR Flash AR Quick Indications for Use: - Impressions for the production of temporary restorations . - All types of preliminary impressions . - Impressions of the opposing jaw . - Impressions for orthodontic models . Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runner-S FDA 03/13/07 08/28/05'00 Page 1 of 1 510(k) Submission: Flash AR & Flash AR Quick; Nov. 05, 2013 Page 11 of 240
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...