The FireSoft-SiHy sphere (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The FireSoft-SiHy toric (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters. The FireSoft-SiHy multifocal (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters. The FireSoft-SiHy multifocal toric (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters. The FireSoft-IC (efrofilcon A) Soft Contact Lenses for daily wear are indicated for and may be prescribed in otherwise non-diseased eves that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting. The FireSoft-55 sphere (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The FireSoft-55 toric (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters. The FireSoft-55 multifocal (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters. The FireSoft-55 multifocal toric (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eves and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters. The FireSoft-49G sphere (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. TheFireSoft-49G toric (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters. TheFireSoft-49G multifocal (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters. The FireSoft-49G multifocal toric (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters. The FireSoft-54G sphere (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity. The FireSoft-54G toric (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters. The FireSoft-54G multifocal (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters. The FireSoft-54G multifocal toric (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
Device Story
Lathe-cut soft contact lenses; fabricated from efrofilcon A, methafilcon A, hioxifilcon B, or hioxifilcon D. Lenses function as refractive media to focus light on retina. Prescribed by eyecare practitioners for daily wear; frequent/planned replacement. Requires cleaning/disinfection via chemical or hydrogen peroxide systems. FireSoft-IC features unique base curve geometry with spheric optic zone, aspheric periphery, and pressure-balancing holes for oxygen transmissibility and tear exchange. Lenses supplied sterile in buffered saline vials. Benefits include visual acuity correction and management of irregular corneal conditions.
Clinical Evidence
No clinical data required. Substantial equivalence established via bench testing, including biocompatibility, toxicology, and physical property verification (water content, oxygen permeability, refractive index) compared to predicate devices.
Technological Characteristics
Lathe-cut soft contact lenses. Materials: efrofilcon A (silicone hydrogel), methafilcon A, hioxifilcon B, hioxifilcon D. Water content: 49%-74%. Oxygen permeability: 15.0-59.8 x 10^-11 (cm2/sec)(ml O2/ml x mm Hg @ 35C). Sterile, vial-packaged. No software/electronics.
Indications for Use
Indicated for aphakic/non-aphakic patients with non-diseased eyes requiring correction of myopia, hyperopia, or astigmatism (up to 12D). Multifocal versions indicated for presbyopia (add power up to +4.00D). FireSoft-IC indicated for irregular corneal conditions (keratoconus, post-graft). Contraindicated for diseased eyes.
Regulatory Classification
Identification
A soft (hydrophilic) contact lens is a device intended to be worn directly against the cornea and adjacent limbal and scleral areas of the eye to correct vision conditions or act as a therapeutic bandage. The device is made of various polymer materials the main polymer molecules of which absorb or attract a certain volume (percentage) of water.
{0}------------------------------------------------
JUN 1 1 2014 Page 1 of 11
# 510 (k) SUMMARY OF SAFETY AND EFFECTIVENESS
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| The assigned 510(k) number is: | K133403 | |
|----------------------------------|-------------------------------------------------------------------|--|
| Applicant information: | | |
| Date Prepared: | June 6, 2014 | |
| Name:<br>Address | Firestone Optics<br>3901 NE 33rd Terrace<br>Kansas City, MO 64117 | |
| Contact Person:<br>Phone number: | David T Rusch<br>President<br>(816) 455-0500 | |
| Consultant: | Martin Dalsing<br>806 Kimball Avenue<br>Grand Junction, CO 81501 | |
| Phone number<br>Email address | (970) 243 5490<br>martindalsing@gmail.com | |
### Device Information:
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
| Device Class: | 2 |
|-------------------------|----------------------------------|
| Regulation Number: | 886.5925 |
| Product Code: | LPL |
| Regulation Description: | Soft (hydrophilic) contact lens. |
{1}------------------------------------------------
Device Trade Name:
FireSoft-SiHy (efrofilcon A) Silicone Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric) FireSoft-IC (efrofilcon A) Silicone Hydrogel (Irregular Cornea; Keratoconus, Post Graft) FireSoft-55 (methafilcon A) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric,) FireSoft-49G (hioxifilcon B) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric) FireSoft-54G (hioxifilcon D) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric)
#### Equivalent Devices:
The FireSoft Daily Wear Soft Contact Lenses are substantially equivalent to the following predicate devices:
Predicate devices:
"Soft K Keratoconus" (efrofilcon A) by Soflex Limited 510(k) number; K122220
"IntelliWave3" (efrofilcon A) by Art Optical Contact Lens, Inc. 510(k) number; K100221
"Contaflex 55" (methafilcon A) by Contamac Ltd. 510(k) number; K023989
"Alden HP 54" (hioxifilcon D) by Alden Optical Labs, Inc. 510(k) number: K091327
"Alden HP 49" (hioxifilcon B) by Alden Optical Labs, Inc. 510(k) number: K981252
{2}------------------------------------------------
#### Device Description:
The FireSoft-SiHy (efrofilcon A) and FireSoft-IC (efrofilcon A) Silicone Hydrogel Soft Contact Lenses are fabricated from efrofilcon A. which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.
The non-ionic lens material, (efrofilcon A) is a group 5, daily wear silicone hydrogel contact lens that is not surface treated and characterized by a high water content. The lens material is composed of silicone monomers cross linked with other monomers and optionally incorporates D&C Green 6 as an integrated handling tint. The lenses are made by lathe-cut for individualized RX. It consists of 26% effofilcon A and 74% water by weight when immersed in a buffered saline solution. The (efrofilcon A) name has been adopted by the United States Adopted Names Council (USAN).
The unique base curve geometry of the FireSoft-IC lens has a spheric optic zone with an aspheric periphery. The front curve has a reinforced optic zone with a center thickness of.36mm (standard), with a special lenticular zone for structural stability. The design incorporates pressure balancing holes to equalize the pressure between the front and back of the lens to improve: Optic stability, Oxygen transmissibility, Tear exchange, and Eliminate air bubbles.
In the hydrated state, the lens conforms to the curvature of the comea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (effofileon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
The Physical properties of the lens are:
| Refractive Index | 1.38 |
|---------------------|------------------------------------------------------------------------|
| Light Transmission | greater than 97% |
| Surface Character | hydrophilic |
| Water Content | 74 % |
| Specific Gravity | 1.048 (hydrated) |
| Oxygen Permeability | 59.8 x 10-11 (cm²/sec) (ml O₂/ml x hPa @ 35°C), (revised Fatt method). |
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 74% water by weight. The lenses will be manufactured in spherical, aspherical, toric, multifocal and irregular comea configurations with the following features and properties.
| · Chord Diameter | 12.0 mm to 16.00 mm |
|---------------------------|----------------------------------|
| · Center Thickness | 0.01 mm to 0.50 mm |
| · Base Curve | 8.0 mm to 9.5 mm |
| · Power Range | -20.00D to +20.00D in 0.25 steps |
| · Cylinder Power (Toric) | -0.25D to -10.00D |
| · Add Power* (Multifocal) | +0.50D to +3.00D |
- Add Power* (Multifocal) . * not applicable for the FireSoft-IC
{3}------------------------------------------------
The FireSoft-55 (methafilcon A) Hydrogel Soft Contact Lenses are fabricated from methafilcon A, which in the dry (unhydrated) state may be machined and polished. The hydrophilic nature of this material allows the lens to become soft and pliable when immersed in an aqueous solution.
The ionic lens material, (methafilcon A) is a co-polymer of 2-Hydroxyethylmethacrylate (2-HEMA) and Methacrylic Acid, cross-linked with ethylene glycol dimethacrylate (EGDMA), plus an initiator. The co-polymer consists of 45% methafilcon A and 55% water by weight when immersed in normal buffered saline solution. The lens is available in clear and with a blue visibility-handling tint, [phthalocyaninato (2-)] copper. The methafilcon A name has been adopted by the United States Adopted Names Council (USAN).
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The (methafilcon A) soft hydrophilic contact lens has a spherical back surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard and appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 55% water by weight. The physical properties of the lens are:
| Refractive Index | 1.405 (hydrated) |
|-----------------------------|--------------------------------------------------------------------------|
| Light Transmission (clear) | greater than 94% |
| Light Transmission (tinted) | greater than 94% |
| Water Content | 55% ± 2% |
| Specific Gravity | 1.067 |
| Oxygen Permeability | 19.6 X 10-11 (cm2/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt method). |
The FireSoft-55 lenses will be manufactured in spherical, toric and multifocal configurations with the following features.
| • Chord Diameter | 12.0 mm to 16.00 mm |
|------------------|---------------------|
|------------------|---------------------|
Center Thickness .
- . Base Curve
- Power Range .
- Cylinder Power (Toric) .
- Add Power (Multifocal) .
0.01 mm to 0.50 mm 8.0 mm to 9.5 mm -20.00D to +20.00D in 0.25 steps -0.25D to -10.00D +0.50D to +3.00D
{4}------------------------------------------------
The FireSoft-49G (hioxifileon B) Hydrogel Soft Contact Lenses are fabricated from (hioxifilcon B) which is a non-ionic, ultra high molecular weight copolymer of 2-hydroxyethyl methacrylate (2-HEMA) and 2,3-Dihydroxypropyl Methacrylate (Glycerol Methacrylate, GMA). It consists of 51% hioxifiloon B and 49% water by weight when immersed in normal buffered saline solution. The lens is available in clear and with a blue visibility handling tint, phthalocyanato (2) - (copper). The hioxifilcon B name has been adopted by the United States Adopted Names Council (USAN).
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly beyond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 49% water by weight. The physical properties of the lens are:
| Refractive Index | 1.425 (hydrated) |
|-----------------------------|--------------------------------------------------------------------------|
| Light Transmission (clear) | greater than 95% |
| Light Transmission (tinted) | greater than 95% |
| Water Content | 49 % ± 2% |
| Specific Gravity | 1.137 |
| Oxygen Permeability | 15.0 X 10-11 (cm2/sec) (ml O2/ml x mm Hg @ 35°C), (revised Fatt method). |
The FireSoft-49G lenses will be manufactured in spherical, toric and multifocal configurations with the following features.
| • Chord Diameter | 12.0 mm to |
|------------------|------------|
|------------------|------------|
- Center Thickness
- Base Curve .
- Power Range .
- Cylinder Power (Toric) ●
- � Add Power (Multifocal)
6.00 mm 0.01 mm to 0.50 mm 8.0 mm to 9.5 mm -20.00D to +20.00D in 0.25 steps -0.25D to -10.00D +0.50D to +3.00D
{5}------------------------------------------------
The FireSoft-54G (hioxifileon D) Hydrogel Soft Contact Lenses are fabricated from (hioxifilcon D) which is a non-ionic, ultra high molecular weight copolymer of 2-hydroxyethyl methacrylate (2-HEMA) and 2,3-Dihydroxypropyl Methacrylate (Glycerol Methacrylate, GMA). It consists of 46% hioxifilcon D and 54% water by weight when immersed in normal buffered saline solution. The lens is available in clear and with a blue visibility handling tint, phthalocyanato (2) - (copper). The hioxifilcon D name has been adopted by the United States Adopted Names Council (USAN).
In the hydrated state, the lens conforms to the curvature of the eye covering the cornea and extending slightly bevond the limbus forming a colorless, transparent optical surface. The hydrophilic properties of the lens require that it be maintained in a fully hydrated state in a solution compatible with the eye. If the lens dries out, it will become hard appear somewhat warped however, it will return to its proper configuration when completely rehydrated in the proper storage solution.
The hydrophilic characteristics allow aqueous solutions to enter the lens and in its fully hydrated state the lens is approximately 54% water by weight. The physical properties of the lens are:
| Refractive Index | 1.408 (hydrated) |
|-----------------------------|-----------------------------------------------------------------|
| Light Transmission (clear) | greater than 95% |
| Light Transmission (tinted) | greater than 95% |
| Water Content | 54 % ± 2% |
| Specific Gravity | 1.137 |
| Oxygen Permeability | 23.0 X 10-11 (cm²/sec) (ml O₂/ml x mm Hg @ 35°C), (revised Fatt |
| | method). |
The FireSoft-54G lenses will be manufactured in spherical, toric and multifocal configurations with the following features.
| • Chord Diameter | 12.0 mm to 16.00 mm |
|--------------------------|----------------------------------|
| • Center Thickness | 0.01 mm to 0.50 mm |
| • Base Curve | 8.0 mm to 9.5 mm |
| • Power Range | -20.00D to +20.00D in 0.25 steps |
| • Cylinder Power (Toric) | -0.25D to -10.00D |
| • Add Power (Multifocal) | +0.50D to +3.00D |
ALL FireSoft Soft Contact Lenses are supplied sterile in vials containing a buffered saline solution. Vial labeling is printed with appropriate name, lot numbering, expiration dating and lens parameter identification. Expiration dating has been established based on studies of product stability, package integrity, and validation of the sterilization process.
{6}------------------------------------------------
#### Intended Use:
The FireSoft-SiHy sphere (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The FireSoft-SiHy toric (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
The FireSoft-SiHy multifocal (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-SiHy multifocal toric (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-IC (efrofilcon A) Soft Contact Lenses for daily wear are indicated for and may be prescribed in otherwise non-diseased eves that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.
The FireSoft-55 sphere (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eves with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The FireSoft-55 toric (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
The FireSoft-55 multifocal (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
)
{7}------------------------------------------------
The FireSoft-55 multifocal toric (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eves and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-49G sphere (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
TheFireSoft-49G toric (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
TheFireSoft-49G multifocal (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-49G multifocal toric (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-54G sphere (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The FireSoft-54G toric (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
The FireSoft-54G multifocal (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-54G multifocal toric (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non-diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
{8}------------------------------------------------
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
#### Testing:
Non-clinical Testing A series of in vivo preclinical toxicology and biocompatibility tests were performed to assess the safety and effectiveness of the FireSoft-SiHy (efrofilcon A), FireSoft-55 (methafilcon A), Firesoft-49G (hioxifilcon B) and Firesoft-54G (hioxifilcon D) Soft Contact Lenses packaged in glass vials. All non-clinical toxicology tests were conducted in accordance with the GLP regulation. All other testing was conducted according to valid scientific protocols.
> Test results of the non-clinical testing on the FireSoft Soft Contact Lenses demonstrate that:
- Lenses supplied in glass vials are sterile for the indicated shelf-life, .
- The packaging material and extracts are not toxic and not irritating, and .
- Lens physical and material properties are consistent with currently marketed . . lenses.
#### Clinical Data
The clinical performance of the (efrofilcon A) (methafilcon A) (hioxifilcon B) (hioxifilcon D) contact lens materials have been previously established, and therefore was not required for this 510(k).
The FireSoft-SiHy and FireSoft-IC (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lenses are identical and have the same manufacturing process (lathe-cut) to the cleared Intelliwave3 (efrofilcon A) Silicone Hydrogel Daily Wear Soft Contact Lens cleared under K100221.
The FireSoft-55, (methafilcon A) Hydrogel Daily Wear Soft Contact Lens is identical and has the same manufacturing process (lathe-cut) to the cleared Contaflex-55 (methafilcon A) Hydrogel Daily Wear Soft Contact Lens cleared under K023989.
The FireSoft-49G, (hioxifilcon B) Hydrogel Daily Wear Soft Contact Lens is identical and has the same manufacturing process (lathe-cut) to the cleared Alden HP49 (hioxifilcon B) Hydrogel Daily Wear Soft Contact Lens cleared under K981252.
The FireSoft-54G, (hioxifilcon D) Hydrogel Daily Wear Soft Contact Lens is identical and has the same manufacturing process (lathe-cut) to the cleared Alden HP54 (hioxifilcon D) Hydrogel Daily Wear Soft Contact Lens cleared under K091327.
{9}------------------------------------------------
#### Conclusions Drawn from Studies
#### Validity of Scientific Data
Several laboratories under Good Laboratory Practice regulations conducted toxicology studies, Microbiology, chemistry, shelf-life stability studies and followed scientific protocols. The data were determined to be scientifically valid under 21 CFR 860.7
#### Substantial Equivalence
Information presented in this Premarket Notification establishes that the FireSoft-SiHy, FireSoft-IC, FireSoft-55, FireSoft-49G and FireSoft-54G Daily Wear Soft Contact Lenses are as safe and effective as the predicate devices when used in accordance with the labeled directions for use and for the requested indication.
#### Risks and Benefits
The risks of the subject devices are the same as those normally attributed to the wearing of Daily Wear Soft Contact Lenses. The benefits to the patient are the same as those for other Daily Wear Soft Contact Lenses.
#### Substantial Equivalence:
The following matrix illustrates the functionality, indication for use, production method, FDA group#, USAN name and material characteristics of the FireSoft-SiHy, FireSoft-IC, FireSoft-55, FireSoft-49G, FireSoft-54G Daily Wear Soft Contact Lenses, as well as the predicate devices.
{10}------------------------------------------------
. . . . . . . . .
ﺮ
:
·
ﮯ
Page 11 of 11
.
.
| | FireSoft-Stly<br>New Device | IntelliWave3<br>(efrofilcon A)<br>predicate device,<br>K100221 | FireSoft-55<br>New Device | Coptaflex-55<br>(methafilcon A)<br>predicate device,<br>K023989 | FireSoft-49G<br>New Device | Alden HP49<br>(bioxifilcon B)<br>predicate device,<br>K981252 | FireSoft-54G<br>New Device | Alden HP54<br>(bioxifileen D)<br>predicate device,<br>K091327 | FireSoft-IC<br>New Device | Soft K<br>(efrofilcon A)<br>predicate device,<br>K122220 |
|---------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Functionality | same as<br>predicate device | The contact lenses act<br>as a refractive medium<br>that focus light rays<br>from near and distant<br>objects on the retina. | same as<br>predicate device | The contact lenses act<br>as a refractive medium<br>that focus light rays<br>from near and distant<br>objects on the retina | same as<br>predicate device | The contact lenses act<br>as a refractive medium<br>that focus light rays<br>from near and distant<br>objects on the retina. | same as<br>predicate device | The contact lenses act<br>as a refractive medium<br>that focus light rays<br>from near and distant<br>objects on the retina. | same as<br>predicate device | The contact lenses act<br>as a refractive medium<br>that focus light rays<br>from near and distant<br>objects on the retina. |
| Indications | same as<br>predicate device | Daily wear, Silicone<br>Hydrogel Soft<br>(hydrophilic) contact<br>lens | same as<br>predicate device | Daily wear, Soft<br>(hydrophilic) contact<br>lens | same as<br>predicate device | Daily wear, Soft<br>(hydrophilic) contact<br>lens | same as<br>predicate device | Daily wear, Soft<br>(hydrophilic) contact<br>lens | same as<br>predicate device | Daily wear, Silicone<br>Hydrogel Soft<br>(hydrophilic) contact<br>lens |
| Production Method | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX | Lathe-Cut,<br>Customized RX |
| FDA Group # | Group # V (Silicone<br>Hydrogel) > 50%<br>Water, Nonionic<br>Polymers | Group # V (Silicone<br>Hydrogel) > 50%<br>Water, Nonionic<br>Polymers | Group # II > 50%<br>Water, Nonionic<br>Polymers | Group # II > 50%<br>Water, Nonionic<br>Polymers | Group # II > 50%<br>Water, Nonionic<br>Polymers | Group # II > 50%<br>Water, Nonionic<br>Polymers | Group # II > 50%<br>Water, Nonionic<br>Polymers | Group # II > 50%<br>Water, Nonionic<br>Polymers | Group # V (Silicone<br>Hydrogel) > 50%<br>Water, Nonionic<br>Polymers | Group # V (Silicone<br>Hydrogel) > 50%<br>Water, Nonionic<br>Polymers |
| USAN name | efrofilcon A | efrofilcon A | methafilcon A | methafilcon A | bioxifilcon B | bioxifilcon B | hioxifileen D | hioxifileen D | efrofilcon A | efrofilcon A |
| Water Content | 74% | 74% | 55% | 55% | 49% | 49% | 54% | 54% | 74% | 74% |
| Oxygen Permeability | $60 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $60 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $19.6 x 10^{-11}$<br>(cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $19.6 x 10^{-11}$<br>(cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $15 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $15 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $23 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method) | $23 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $60 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). | $60 x 10^{-11}$ (cm2/sec)<br>(ml 02/ml x mm Hg<br>@ 35 degrees C),<br>(revised Fatt method). |
| Specific Gravity | 1.139 | 1.139 | 1.067 | 1.067 | 1.136 | 1.136 | 1.140 | 1.140 | 1.139 | 1.139 |
{11}------------------------------------------------
Image /page/11/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized symbol resembling a human figure with outstretched arms, accompanied by the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
June 11, 2014
Firestone Optics, Inc. c/o Mr. Martin Dalsing Official Correspondent 806 Kimball Avenue Grand Junction, CO 81501
Re: K133403
> Trade/Device Name: FireSoft-SiHy (efrofilcon A) Silicone Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric): FireSoft-IC (efrofilcon A) Silicone Hydrogel (Irregular Cornea, Keratoconus, Post Graft); FireSoft-55 (methafilcon A) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric); FireSoft-49G (hioxifilcon B) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric): FireSoft-54G (hioxifilcon D) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric). Regulation Number: 21 CFR 886.5925 Regulation Name: Soft (hydrophilic) contact lens Regulatory Class: Class II Product Code: LPL Dated: May 6, 2014 Received: May 7, 2014
Dear Mr. Dalsing,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
{12}------------------------------------------------
Page 2 - Mr. Martin Dalsing
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm | | 5809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
:
:
Sincerely yours.
Deborah L. Falls -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
#### Indications for Use
510(k) Number (if known) K133403
#### Device Name
FireSoft-54G (hioxifilcon D) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric)
#### Indications for Use (Describe)
The FireSoft-54G sphere (hioxifileon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The FireSoft-54G toric (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
The FireSoft-54G multifical (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-S4G multifocal toric (hioxifilcon D) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
Type of Use (Select one or both, as applicable)
. 11 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2 2
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Leonid Livshitz -S 2014.06.10 12:56:05 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of Information unless it displays a currently valid OMB number."
Form Approvad: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{14}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
Indications for Use
510(k) Number (if known) K133403
Device Name
FireSoft-49G (hioxifilcon B) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric)
Indications for Use (Describe)
The FireSoft-49G sphere (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
TheFireSoft-49G toric (hioxifileon B) Soft Contact Lenses for the correction of visual acuity in aphakic and not aphakic persons with non diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
TheFireSoft-49G multifocal (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-49G multifocal toric (hioxifilcon B) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)# Leonid Livshitz -S 2014.06.09 11:30:02 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other assect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{15}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K133403
#### Device Name
FireSoft IC (efrofilcon A) Silicone Hydrogel (Irregular Cornea, Keratoconus, Post Graph)
Indications for Use (Describe)
The FireSoft-IC (efrofilcon A) Soft Contact Lenses for daily wear are indicated for and may be prescribed in otherwise non-diseased eyes that require a Soft Contact Lens for the management of irregular corneal conditions such as keratoconus and post graft fitting.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
.
# Leonid Livshitz -S 2014.06.09 11:28:14 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@ida.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
FORM FDA 3881 (1/14)
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
{16}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K133403
#### Davice Name
FireSoft SiHy (efrofilcon A) Silicone Hydrogel (Spherical, Toric, Multifocal Toric)
Indications for Use (Describe)
The FireSoft-SiHy sphere (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The FireSoft-SiHy toric (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
The FireSoft-SiHy multifocal (efrofilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-SiHy multifocal toric (efrofilcon A) Soft Conact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent placement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
SED WANT AND A FOR FOR FOR FOR FOR FOR FOR FOR USE ONLY
## Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
## Leonid Livshitz -S 2014.06.09 11:27:31 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{17}------------------------------------------------
Indications for Use
Device Name
FireSoft 55 (methafilcon A) Hydrogel (Spherical, Toric, Multifocal, Multifocal Toric)
Indications for Use (Describe)
The FireSoft-55 sphere (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non-diseased eyes with myopia or hyperopia. The lens may be worn by persons who exhibit refractive astigmatism of .75 diopters or less where the astigmatism does not interfere with visual acuity.
The FireSoft-55 toric (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non diseased eyes with myopia or hyperopia and/or possesses refractive astigmatism not exceeding 12 diopters.
The FireSoft-55 multifocal (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of visual acuity in aphakic and not aphakic persons with non diseased eyes with myopia or hyperopia and are presbyopic requiring add power of up to +4.00 diopters.
The FireSoft-55 multifocal toric (methafilcon A) Soft Contact Lenses for daily wear are indicated for the correction of ametropia (myopia and hyperopia) in aphakic and not aphakic persons with non diseased eyes and/or possesses refractive astigmatism not exceeding 4 diopters and are presbyopic requiring add power of up to +4.00 diopters.
Eyecare practitioners may prescribe the lenses for frequent/planned replacement wear, with cleaning, disinfection and scheduled replacement. When prescribed for frequent/planned replacement wear, the lens may be cleaned and disinfected using a chemical (not heat) lens care system or hydrogen peroxide disinfection systems.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
: .. FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Leonid Livshitz -S 2014.06.09 11:29:05 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.4
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.