← Product Code [FED](/productcode/FED) · K133359

# ENDORINGS (K133359)

_Endo-Aid, Inc. · FED · Feb 11, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133359

## Device Facts

- **Applicant:** Endo-Aid, Inc.
- **Product Code:** [FED](/productcode/FED.md)
- **Decision Date:** Feb 11, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The EndoRings is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following: Keeping the suitable depth of endoscope's view field

## Device Story

EndoRings is a flexible silicone rubber add-on device for the distal end of an endoscope; features a short tube with three circular layers of flexible wings. During intubation and forward movement, wings fold backward; during withdrawal, wings fold forward to maintain field of view and scope centering. Used in clinical settings by physicians during colonoscopy to facilitate therapeutic interventions (biopsies, polypectomies, APC). Device stabilizes the scope, prevents sudden slippage, and manipulates colonic folds to improve mucosal visualization. Benefits include improved scope stability and maintenance of optimal viewing depth without interfering with the endoscopic image. No electronic or software components; purely mechanical accessory.

## Clinical Evidence

Prospective clinical study of 60 adult patients undergoing standard colonoscopy. Primary endpoint: successful cecum reach (achieved in 100% of cases). Secondary endpoints: safety (no adverse events or colon tissue damage reported), therapeutic intervention success (biopsies/polypectomies rated 'Good' to 'Excellent'), and usability (no malfunctions, high user satisfaction).

## Technological Characteristics

Flexible silicone rubber tube; USP Class VI compliant; biocompatible per ISO 10993. Length 24mm; main body outer diameter 15.7mm; wing open diameter 50mm. Mechanical attachment via friction/fit to endoscope distal tip. Sterilized via Ethylene Oxide.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Arc EndoCuff AEC 110; AEC 120; AEC 140 ([K122565](/device/K122565.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# 510(k) SUMMARY

# EndoAid EndoRings

FEB 1 1 2014

## Applicant Information:

EndoAid, Ltd. ARAN Building P.O. Box 3067 43 Haeshel Street Caesarea 38900 Israel

| Phone:          | (972) 4-6239014  |
|-----------------|------------------|
| Facsimile:      | (972) 4-6273260  |
| Contact Person: | Dan Rottenberg   |
| Date Prepared:  | January 23, 2014 |

#### Device Information:

| Trade Name:           | EndoRings                                        |
|-----------------------|--------------------------------------------------|
| Common or Usual Name: | Endoscopic Distal Attachment                     |
| Classification:       | Class II per 21 CFR 876.1500                     |
| Product Code:         | FED                                              |
| Predicate Device:     | Arc EndoCuff AEC 110; AEC 120; AEC 140 (K122565) |

#### Device Description:

The EndoRings is a flexible silicone rubber add-on device designed to attach to of the designated endoscope. The product features a short tube like shape with three flexible circular layers of wings that easily fold backwards during intubation and scope forward movement, and fold forward during scope withdrawal, to maintain the field of view and scope centering position.

## Intended Use / Indications for Use:

The EndoRings is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following:

* Keeping the suitable depth of endoscope's view field

{1}------------------------------------------------

# Technological Characteristics:

The technological characteristics of EndoAid EndoRings are equivalent to the characteristics of the predicate device.

## Materials

The EndoRings silicone rubber material has been chosen as conforming to USP class VI and tested for biocompatibility in accordance with ISO 10993.

#### Size

The EndoRings length is 24mm, while the predicate Arc EndoCuff (K122565) length is 23.8mm. The main body tube outer diameter is 15.7mm for the EndoRings, and 17.2mm for the EndoCuff. There is a difference in the thin flexible wings /projections maximal free/open diameter which is 50mm for the EndoRings, and 34.8mm for the Arc EndoCuff. The wings fold easily when touching the colon walls, and the diameter of the wings can reduce to 22mm for the EndoRings compared to 17.4mm for the EndoCuff projections.

#### Non-clinical Data

The EndoRings is used to maintain the endoscopes field of view during endoscopy therapy by manipulating colonic folds, maintaining a central position in the lumen and avoiding sudden slippage. The nominated predicate device maximizes the field of view in a similar way.

The bench testing demonstrates that the EndoRings design combines some of the benefits of both being a slim and stabilizing cylindrical device (like the Arc EndoCuff), being able to stabilize and view the mucosa effectively on a greater diameter, without interfering with the standard field of view, since the wings are designed in a way they do not interfere with the endoscopic image.

The design of the device main cylindrical tube made from flexible polymer (silicone rubber) is the means of attachment / retention to the distal end of an endoscope (like the Endoculf).

The EndoAid EndoRings and the Arc EndoCuff both contact the patient in the same way and for the same period of time. Both are made from flexible polymers materials and both are sterilized.

The performance data showed that the sterilization method selected (EndoRings is Ethylene Oxide sterilized and Arc EndoCuff is irradiated K122565) has not introduced any additional risks and the patient contacting materials have been tested for biocompatibility with reports demonstrating no negatives in safety or effectiveness.

The bench testing and clinical study has demonstrated that the device does not introduce any additional risks when undertaking endoscopic therapy and meeting the intended use.

All of these tests demonstrated that the EndoRings meets its intended performance specifications.

{2}------------------------------------------------

# Clinical Data:

The device performance was validated by testing of a final production unit in clinical conditions. The procedures were performed at Elisha Medical Center, Haifa, Israel on a diverse adult population in terms of gender, ethnicity and age. Sixty (60) patients were enrolled in conformance with the device labeling.

The purpose of this prospective clinical study was to establish the feasibility and usability of the ENDORINGSTM when used during standard colonoscopy procedures.

The primary endpoint was reaching the cecum of the colon with the ENDORINGS.

Secondary endpoints were:

- 1. Incidence of complications.
- 2. Ability to complete therapeutic interventions as biopsies, polypectomies, APC etc.
- 3. Procedure time
- 4. Ease of scope insertion, advancement and withdrawal.
- 5. Ability to center the scope inside the gastrointestinal tract.
- 6. Subjective evaluation of the additional area screened by the physician.
- 7. Patient wellbeing

Referring to the criteria above, the study objectives were met:

- . Reaching the cecum: In 100 % of the cases, the primary endpoint, reaching the cecum, was achieved.
- Complications : Regarding secondary safety endpoints, there were no reports of adverse . events, colon tissue damage, or other complications during the procedure or in the 24 hours post procedure follow-up.
- Therapeutic interventions: The ability to insert tool through the working channel and . maneuver the scope distal tip to take specimen were tested: all the applicable answers were scored as "Good" to "Excellent". All of the interventions were successfully completed without any reports of operational difficulties.

{3}------------------------------------------------

- Device performance: The entire procedure was conducted without any interruption. There . was no report of device malfunction or reliability failure.
- . Scoring Analysis: None of the users rated any figure of merit as "unacceptable" or "difficult". All answers varied between "excellent" to "acceptable", while the vast majority of users' responses related to the device were rated as "excellent".
- Patient wellbeing: No adverse events, complications or complaints of patient discomfort . were observed during the procedures or reported in the 24 hours follow up

The above test results and analysis, demonstrated that the EndoRings meets the user needs in clinical environment, under defined conditions. The test pass criteria were met. Thus, the clinical study supports the substantial equivalence of the EndoRings to its predicate device.

## Substantial Equivalence:

After a review of the bench and clinical testing, the company concluded that the EndoRings is as safe and effective as the predicate device for facilitating endoscopic therapy.

The EndoRings and the predicate device have the same indication for use and very similar technological characteristics, and principles of operation. The minor technological differences between the EndoRings and its predicate device raise no new types of safety or effectiveness questions. In vitro verification testing and clinical study demonstrates that the EndoRings performs as intended and meets all design specifications with respect to their mechanical, functional and handling characteristics, and that its materials are biocompatible.

Thus, the EndoAid EndoRings is substantially equivalent to the Arc EndoCuff.

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol resembling an abstract eagle or bird in flight, composed of three curved lines.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

February 11, 2014

EndoAid, Ltd. % Janice M. Hogan Partner Hogan Lovells US LLP 1835 Market Street, 29th Floor Philadelphia, PA 19103

Re: K133359

> Trade/Device Name: EndoRings Regulation Number: 21 CFR& 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: FED Dated: January 23, 2014 Received: January 23, 2014

Dear Janice M. Hogan,

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.

{5}------------------------------------------------

Page 2 - Janice M. Hogan

You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Herbert P. Lerner -S

for

Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

## Indications for Use

510(k) Number (if known) K133359

Device Name EndoRings

Indications for Use (Describe)

The EndoRings is intended to be attached to the distal end of the endoscope to facilitate endoscopic therapy, to be used for the following:

Keeping the suitable depth of endoscope's view field

Type of Use (Select one or both, as applicable)

> Prescription Use (Part 21 CFR 801 Subpart D)

Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

FOR FOR FOR FOR FOR FOR USE ONLY Sandard .. -Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/6/Picture/12 description: The image shows the name "Herbert P. Lerner" and the date and time "2014.02.11 11:22:04 -05:00". The text is in a clear, sans-serif font and is easily readable. The date and time are formatted in a specific way, with the year, month, and day separated by periods, and the time including hours, minutes, seconds, and timezone.

{7}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K133359](https://fda-staging.innolitics.com/device/K133359)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
