← Product Code [LLZ](/productcode/LLZ) · K133357

# CHARRUAPACS SYSTEM (K133357)

_Charruasoft SA · LLZ · Nov 25, 2013 · Radiology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133357

## Device Facts

- **Applicant:** Charruasoft SA
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Nov 25, 2013
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device

## Indications for Use

CharruaPACS System™ is a software based device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.). Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor.

## Device Story

CharruaPACS System is a software-based PACS for handling, storing, and viewing DICOM-compliant medical images. Inputs include digital images from CT, MR, ultrasound, and radiographic scanners. The system operates as a Windows-based server/workstation architecture, utilizing PostgreSQL for database management and Independent JPEG/OpenJPEG libraries for image compression/decompression. It provides tools for image viewing, including window/level adjustment, zoom, linear/angle measurements, Hounsfield unit measurement, series comparison, and DICOM SR reporting. Used in clinical settings by radiology personnel for workflow management, remote tele-diagnosis, and integration with HIS/RIS/EMR systems. Output is displayed on medical-grade monitors to assist clinicians in diagnostic decision-making. Benefits include replacement of film archives, improved remote access to studies, and streamlined radiology workflow.

## Clinical Evidence

No clinical data. Bench testing only. Verification and validation testing evaluated all input/output functions and operational modes against predetermined acceptance criteria. Testing confirmed the system meets performance requirements and risk analysis specifications.

## Technological Characteristics

Software-based PACS; DICOM 3.x compliant. Supports JPEG, JPEG Lossy, and JPEG 2000 compression. Architecture: Windows Services, PostgreSQL database, Web-based viewer (TIdHTTPServer). Connectivity: Networked via DICOM C-STORE, C-FIND, C-MOVE, C-ECHO. Hardware: Vendor-neutral, recommended Intel i5, 4GB RAM, 400GB HDD. No patient contact; non-life-sustaining.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- INSTARAD ([K080334](/device/K080334.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K133357
Page 1 of 6

This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990.

## Date Prepared:

November 22, 2013

## Submitter's Information: 21 CFR 807.92(a)(1)

Mr. Rafael Sanguinetti Director, CharrauSoft SA Avda. Brasil 2807, Ap. 403 Montevideo, Uruguay, 11300 Tel: +598.270.92850 Email: rafael@charrua.com

#### Trade Name, Common Name and Classification: 21 CFR 807.92(a)(2)

| Product Name:        | CharruaPACS System™                        |
|----------------------|--------------------------------------------|
| Common Name:         | Picture, archive and communications system |
| Classification Name: | System, Image Processing, Radiological     |
| Product Code:        | LLZ                                        |

### Predicate Device: 21 CFR 807. 92(a)(3)

CharruaPACS™ system is substantially equivalent to:

| Device Classification Name        | system, image processing, radiological |
|-----------------------------------|----------------------------------------|
| 510(k) Number                     | K080334                                |
| Device Name                       | INSTARAD                               |
| Original Applicant                | MEDSPHERE TECHNOLOGIES PVT LTD.        |
| Regulation Number                 | 892.2050                               |
| Classification Product Code       | LLZ                                    |
| Date Received                     | 02/07/2008                             |
| Decision Date                     | 02/21/2008                             |
| Decision                          | substantially equivalent (SE)          |
| Classification Advisory Committee | Radiology                              |
| Review Advisory Committee         | Radiology                              |
| summary                           | summary                                |
| Type                              | Traditional                            |
| Reviewed by Third Party           | Yes                                    |
| Expedited Review                  | No                                     |
| Combination Product               | No                                     |
| Device Classification Name        | system, image processing, radiological |

#### Device Description: 21 CFR 807 92(a)(4)

The CharruaPACS System™ consists of the CharruaPACS Workstation/Server and the DICOM Viewer (CWS). The device is a software system to be used to handle and view DICOM compliant studies, which are stored within the device or specified network locations. The System supports DICOM C-STORE SCP for most used SOP classes, with uncompressed Transfer Syntax and JPEG lossless, JPEG lossy and JPEG 2000 Transfer Syntaxes, C-FIND and C-MOVE SCP Study Root model and C-ECHO Verification as SCP.

{1}------------------------------------------------

It is composed of four applications:

- Admin: for the initial configuration. .
- . DICOM Server.
- . WEB Server: WEB server running on port 80, it allows to change CharruaPACS configuration and to access images through a web browser
- HTTP Client: a visualization application that accesses the images in the PACS . through the HTTP port.

The servers run as Windows Services. It also uses the Independent JPEG group library for JPEG compression/decompression and the OpenJPEG group library for JPEG 2000 compression/decompression. The advanced version uses PostgreSQL as database. WebServer is based on Indy Project TIdHTTPServer.

The CharruaPACS System™ is intended for professional use, as a viewing tool for imaging studies, and is a 'Continuous Use' device. This device is also compliant with HIPAA regulations regarding patient privacy (such as restricting access to particular studies, logging access to data). There is no direct patient interaction with the device, therefore, there is no possibility that the CharruaPACS System might lead to a fatal fault or injury to the patient. The CharruaPACS System is intended to work as a standalone service that provides

common PACS functionality.

The System has four main uses:

- . Hard copy media replacement: PACS replaces hard-copy based means of managing medical images, such as film archives.
- . Remote access: It expands on the possibilities of conventional systems by providing capabilities of off-site viewing and reporting (distance education, tele-diagnosis). It enables practitioners in different physical locations to access the same information simultaneously for teleradiology.
- . Electronic image integration platform: PACS provides the electronic platform for radiology images interfacing with other medical automation systems such as Hospital Information System (HIS), Electronic Medical Record (EMR), Practice Management Software, and Radiology Information System (RIS).
- Radiology Workflow Management: PACS is used by radiology personnel to manage . the workflow of patient exams.

## Indications for Use: 21 CFR 807 92(a)(5)

CharruaPACS System™ is a software based device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.). lmages and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation. Mammographic images may only be interpreted using an FDA cleared monitor.

## Technological Characteristics: 21 CFR 807 92(a)(6)

CharruaPACS System™ is a software product that handles and manipulates digital medical images. The device does not contact the patient, nor does it control any life sustaining devices. A physician, providing ample opportunity for competent human intervention interprets images and information being displayed and printed. In general, a PACS (Picture Archiving and Communication System) is a medical imaging technology which provides storage of, and convenient access to, images from multiple modalities. Electronic images and reports are transmitted digitally via PACS: this eliminates the need to manually file, retrieve, or transport film jackets. The universal format for PACS image storage and transfer is DICOM 3.x (Digital Imaging and Communications in Medicine). Non-image data, such as scanned

{2}------------------------------------------------

documents, may be incorporated using consumer industry standard formats like PDF (Portable Document Format), once encapsulated in DICOM. The modified device and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The new device does not raise any new potential safety risks and is equivalent in performance to the existing legally marketed devices. Both systems have been developed to replace traditional film handling in radiology. The 2 devices are substantially equivalent in the areas of design, architecture, general function, application, and intended use.

Any difference between the two devices does not affect safety or efficacy. The predicate device and the new device are compared below:

| Feature/Functions                               |                                  | CharruaPACS<br>Device                                                                         | InstaRad<br>K080334<br>Predicate                                                              | If different, Impact on Safety and or<br>Efficacy                                                                                                                                                                                                                                                                                                                                                                                                                                                        |
|-------------------------------------------------|----------------------------------|-----------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indications for Use                             |                                  | Picture.<br>Archiving,<br>Communications<br>System                                            | Picture.<br>Archiving,<br>Communications<br>System                                            | No difference.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |
| Hardware server<br>(recommended)                | Computer<br>processor            | Intel i5 processor                                                                            | Pentium dual<br>core Intel Xeon<br>2.0+ GHz<br>processor                                      | Yes, differences.<br>The predicate was based upon the latest<br>computer platform at the time of<br>development. Pentium dual core Intel xeon<br>2.0+ GHz processor are older technology<br>and have been basically replaced with a<br>current technology with an Intel i5 processor.<br>The current system configuration has been<br>tested and validated and the results of<br>testing verified that there is no impact on<br>safety or efficacy and that no additional risks<br>have been identified. |
|                                                 | Operating<br>System              | Windows 7<br>Professional                                                                     | Windows 2003                                                                                  | Yes, difference.<br>The predicate was based upon the available<br>Microsoft operating system for that time,<br>Windows 2003 (server OS) which is no<br>longer available or supported by Microsoft.<br>Windows 2003 OS been replaced by<br>Microsoft Windows 7. The system has been<br>tested and validated with Windows 7 and the<br>results of testing verified that there is no<br>impact on safety or efficacy and that no<br>additional risks have been identified.                                  |
|                                                 | Hard disk array                  | Capacity<br>depends on the<br>usage volume.<br>Minimum 400GB                                  | Capacity<br>depends on the<br>usage volume.<br>Minimum 400GB                                  | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Ram                              | Minimum 4 GB                                                                                  | Minimum 1GB                                                                                   | Yes, differences.<br>Microsoft Windows 7 OS and the Intel i5<br>processor require a larger RAM for<br>operating. The system has been tested and<br>validated with Windows 7 and the results of<br>testing verified that there is no impact on<br>safety or efficacy and that no additional risks<br>have been identified.                                                                                                                                                                                |
| Workstation<br>Client Hardware<br>(recommended) | Processor                        | Intel i5 processor                                                                            | Pentium<br>2.0+Ghz. Dual<br>core                                                              | Yes, differences.<br>The predicate was based upon the latest<br>computer platform at the time of<br>development. Pentium dual core Intel Xeon<br>2.0+ GHz processor are older technology<br>and have been replaced with Intel i5<br>processor or equivalent. The current system<br>configuration with the Intel i5 processor has<br>been tested and validated and the results of<br>testing verified that there is no impact on<br>safety or efficacy and that no additional risks                       |
| Feature/Functions                               | CharruaPACS<br>Device            | InstaRad<br>K080334<br>Prodicato                                                              | If different, Impact on Safety and or<br>Efficacy                                             |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                 | Operating<br>System              | Windows 7                                                                                     | Windows XP                                                                                    | Yes, difference.<br>The predicate was based upon the available<br>Microsoft operating system for that time.<br>Windows XP which is no longer available or<br>supported by Microsoft and Windows XP OS<br>been replaced by Microsoft Windows 7. The<br>system has been tested and validated with<br>Windows 7 and the results of testing verified<br>that there is no impact on safety or efficacy<br>and no additional risks have been identified.                                                       |
|                                                 | Display                          | Medical Grade<br>Monitor is<br>recommended.<br>Resolution<br>depends on the<br>modality type. | Medical Grade<br>Monitor is<br>recommended.<br>Resolution<br>depends on the<br>modality type. | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | RAM                              | Minimum 4 GB<br>RAM                                                                           | Minimum 1GB<br>RAM                                                                            | Yes, differences.<br>Microsoft Windows 7 OS requires larger<br>RAM then the old technology of Windows XP<br>(which is no longer sold or supported by<br>Microsoft). The system has been tested and<br>validated with Windows 7 and the results of<br>testing verified that there is no impact on<br>safety or efficacy and that no additional risks<br>have been identified.                                                                                                                             |
|                                                 | Hard Disk                        | 400 GB<br>minimum                                                                             | 400 GB<br>minimum                                                                             | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | System<br>Architecture           | Web based                                                                                     | Web based                                                                                     | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Hardware                         | Vendor Neutral                                                                                | Vendor Neutral                                                                                | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Security                         | Log-on user ID &<br>password                                                                  | Log-on user ID &<br>password                                                                  | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Remote<br>monitoring             | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Server features                  | Database                                                                                      | PostgreSQL                                                                                    | mySQL                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                    |
| Feature/Functions                               | CharruaPACS Device               | InstaRad K080334<br>Prodicato                                                                 | If different, Impact on Safety and or Efficacy                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                                 | Storage                          | RAID                                                                                          | RAID                                                                                          | and the results of testing verified that there is<br>no impact on safety or efficacy and that no<br>additional risks have been identified.                                                                                                                                                                                                                                                                                                                                                               |
|                                                 |                                  |                                                                                               |                                                                                               | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Viewer Features:                                | Image Viewing Layout             | Std formats(up to 2*3)                                                                        | Std formats(up to 4*4)                                                                        | Yes, Differences.<br>In CharruaPACS there is one less "standard"<br>format than the predicate. This difference is<br>a user preference and does not directly<br>affect the image or intended use. The system<br>has been tested and validated and the<br>results of testing verified that there is no<br>impact on safety or efficacy and that no<br>additional risks have been identified.                                                                                                              |
|                                                 | WW/WL                            | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Zoom in/Zoom out                 | yes                                                                                           | yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Hounsfield Measurement           | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Linear and angle<br>measurements | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Series Comparison                | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Scout line display               | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | 3D capabilities                  | No                                                                                            | No                                                                                            | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Stack mode                       | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Gray scale invert                | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Filters                          | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Rotate                           | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Key Image selection              | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | DICOM Print                      | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Windows print                    | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Query/Retrieve                   | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Image compression                | Lossless streaming                                                                            | Lossless streaming                                                                            | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Selection tools                  | Thumbnails                                                                                    | Thumbnails                                                                                    | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
| Reporting<br>module                             | Reporting Interface              | Can be opened<br>from the viewer<br>or from study list                                        | Can be opened<br>from the viewer<br>or from study list                                        | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Report Template Support          | User Defined templates                                                                        | User Defined templates                                                                        | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Digital Signature                | Yes                                                                                           | Yes                                                                                           | No difference                                                                                                                                                                                                                                                                                                                                                                                                                                                                                            |
|                                                 | Report Formats                   | DICOM SR                                                                                      | MS Word                                                                                       | Yes, Differences.<br>MS Word documents are proprietary and<br>only readable by Microsoft Word or<br>compatible applications. DICOM SR is a<br>open standard document readable by any<br>competent DICOM Viewer. DICOM SR is                                                                                                                                                                                                                                                                              |

{3}------------------------------------------------

4 | 17 :

.

{4}------------------------------------------------

{5}------------------------------------------------

| Feature/Functions |                                                     | CharruaPACS<br>Device      | InstaRad<br>K080334<br>Predicate | If different, impact on Safety and or<br>Efficacy                                                                                                                                                                                                                                                                                                 |
|-------------------|-----------------------------------------------------|----------------------------|----------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|                   |                                                     |                            |                                  | Word documents require some kind of<br>reference to keep them linked to the study<br>images and is not automatically linked as are<br>DICOM SR reports. The system has been<br>tested and validated and the results of<br>testing verified that there is no impact on<br>safety or efficacy and that no additional risks<br>have been identified. |
| Other features    | Link to Hospital<br>Information<br>System (HIS)     | Yes                        | Yes                              | No difference                                                                                                                                                                                                                                                                                                                                     |
|                   | Link to<br>Radiology<br>Information<br>System RIIS) | Yes                        | Yes                              | No difference                                                                                                                                                                                                                                                                                                                                     |
|                   | Electronic<br>patient record                        | Through Broker<br>Software | Through Broker<br>Software       | No difference                                                                                                                                                                                                                                                                                                                                     |
|                   | HIPAA                                               | Compliant                  | Compliant                        | No difference                                                                                                                                                                                                                                                                                                                                     |

### Nonclinical Testing:

The complete CharruaPACS System™ configuration has been assessed and tested at the factory and has passed all in-house testing criteria. The Verification & Validation Test Plan was designed to evaluate all input functions, output functions, and actions performed by the CharruaPACS System™ software in each operational mode and followed the process documented in the Validation Test Plan.

Nonclinical testing results are provided in the 510(k). Validation testing indicated that as required by the risk analysis, designated individuals performed all verification and validation activities and that the results demonstrated that the predetermined acceptance criteria were met. If the device is installed by ChurrauSoft SA, integration and installations verification tests are conducted against acceptance criteria prior to release to the client.

## Conclusion: 21 CFR 807 92(b)(1)

The 510(k) Pre-Market Notification for CharruaPACS System™ contains adequate information, data, and nonclinical test results to enable FDA - CDRH to determine substantial equivalence to the predicate device.

The subject device will be manufactured in accordance with the voluntary standards listed in the enclosed voluntary standard survey. The subject and predicate devices are substantially equivalent in the areas of technical characteristics, general function, application, and intended use. The modification to the subject device does not raise any new potential safety risks and is equivalent in performance to existing legally marketed devices.

Nonclinical tests demonstrate that the device is as safe, as effective, and performs as well as the predicate device.

Therefore, CharruaPACS System™ is substantially equivalent to the predicate devices.

{6}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/6/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a symbol. The symbol is a stylized representation of an eagle with three lines extending from its body, possibly representing the department's mission or services.

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

November 25, 2013

CharruaSoft SA % Mr. Carl Alletto Consultant 111 Melanie Drive AUBREY TX 76227

Re: K133357

Trade/Device Name: CharruaPACS System™ Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: October 25, 2013 Received: November 1. 2013 .

Dear Mr. Alletto:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device. subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

{7}------------------------------------------------

Page 2-Mr. Alletto

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYow/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to

http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

![](https://i.imgur.com/99999999999999999999999999999999.png)

for

Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health

Enclosure

{8}------------------------------------------------

# Indications for Use

510(k) Number (if known): K133357

Device Name: CharruaPACS System™

Indications For Use:

CharruaPACS System™ is a software based device that receives digital images and data from various sources (e.g. CT scanners, MR scanners, ultrasound systems, computed & direct radiographic devices, secondary capture devices, scanners, imaging gateways, etc.).

Images and data can be captured, stored, communicated, processed and displayed within the system and or across computer networks at distributed locations. Lossy compressed mammographic images and digitized film screen images must not be reviewed for primary image interpretation.

Mammographic images may only be interpreted using an FDA cleared monitor.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 807 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of Center for Devices and Radiological Health (CDRH)

Smh7)

(Division Sign-Off) Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health 510(k)

---

**Source:** [https://fda-staging.innolitics.com/device/K133357](https://fda-staging.innolitics.com/device/K133357)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
