K133356 · ETHICON, Inc. · GAW · Dec 6, 2013 · General, Plastic Surgery
Device Facts
Record ID
K133356
Device Name
PROLENE POLYPROPYLENE NONABSORBABLE SUTURE
Applicant
ETHICON, Inc.
Product Code
GAW · General, Plastic Surgery
Decision Date
Dec 6, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.5010
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
PROLENE™ suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Device Story
PROLENE™ Polypropylene Nonabsorbable Suture is a sterile, nonabsorbable surgical suture composed of isotactic crystalline stereoisomer of polypropylene. The device is used by surgeons for soft tissue approximation and ligation. The modified version features a reduced diameter at the suture swage area to better align with finer wire diameter needles, intended to reduce needle hole bleeding. The device is supplied in various configurations (clear or blue-pigmented) and is used in clinical settings such as operating rooms. The suture is attached to a needle; the modification specifically addresses needle attachment force requirements per U.S.P. standards. The device functions as a mechanical fastener for tissue; it does not involve electronic processing or clinical decision-support algorithms.
Clinical Evidence
Bench testing only. Verification testing performed per U.S.P. <871> (Needle Attachment), U.S.P. <881> (Tensile Strength), and U.S.P. <861> (Diameter) demonstrated that the modified device meets all applicable performance specifications.
Technological Characteristics
Material: Isotactic crystalline stereoisomer of polypropylene (synthetic linear polyolefin). Pigmented version uses phthalocyanine blue (Color Index 74160). Sterilization: Ethylene Oxide. Form factor: Suture strand with attached needle. Swaging process adjusted to align suture strand and needle wire diameters. Complies with U.S.P. monographs for diameter, tensile strength, and needle attachment.
Indications for Use
Indicated for general soft tissue approximation and/or ligation in patients requiring cardiovascular, ophthalmic, or neurological surgical procedures.
Regulatory Classification
Identification
Nonabsorbable polypropylene surgical suture is a monofilament, nonabsorbable, sterile, flexible thread prepared from long-chain polyolefin polymer known as polypropylene and is indicated for use in soft tissue approximation. The polypropylene surgical suture meets United States Pharmacopeia (U.S.P.) requirements as described in the U.S.P. Monograph for Nonabsorbable Surgical Sutures; it may be undyed or dyed with an FDA approved color additive; and the suture may be provided with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for this device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
Predicate Devices
Labeling Changes for Generic Suture Types (K946173)
PROLENE Polypropylene Suture (NDA 16-374)
Submission Summary (Full Text)
{0}------------------------------------------------
K133356 Page 1/3
# 510(k) SUMMARY
| Sponsor: | ETHICON, Inc. P.O. Box 151 Route 22 West Somerville, New Jersey 08876 |
|---------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Contact: | Elliott Jang, RAC ETHICON, Inc. P.O. Box 151 Route 22 West Somerville, New Jersey 08876 Phone: 908-541-3817 Fax: 908-218-2595 Email: ejang1@its.jnj.com |
| | DEC 0 6 2013 |
| Date of Submission: | October 28, 2013 |
| Proprietary Name: | PROLENET™ Polypropylene Nonabsorbable Suture |
| Common Name: | Suture, Nonabsorbable, Synthetic, Polypropylene |
| Regulation: | 21 CFR 878.5010 |
| Regulatory Class: | II |
| Product Code: | GAW |
| Predicate Devices: | Labeling Changes for Generic Suture Types, by Ethicon Inc., K946173, cleared January 9, 1995 |
| | PROLENE Polypropylene Suture (Nonabsorbable Surgical Suture, U.S.P., Type B), by Ethicon Inc., NDA 16-374 approved April 16, 1969 followed by multiple supplements (converted from drug to PMA device in 1983 and then reclassified to class II device in 1990) |
| Device Description: | PROLENETM Polypropylene Suture (clear or pigmented) is a nonabsorbable, sterile surgical suture composed of an isotactic crystalline stereoisomer of polypropylene, a synthetic linear polyolefin. The pigmented suture is pigmented blue (with phthalocyanine blue, Color Index Number 74160) to enhance visibility. |
.
ETHICON, Inc.
ﺎ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤ
{1}------------------------------------------------
PROLENE™ Polypropylene Nonabsorbable Suture Special 510(k) Pre-Market Notification
PROLENE™ suture, also available as PROLENE™ HEMO-SEAL™ needle suture, is a needle suture combination in which the diameter of the suture swage area has been reduced to facilitate attachment of finer wire diameter needles. The diameter of the suture strand and the needle wire have been more closely aligned to reduce the degree of needle hole bleeding.
The subject device is identical to the predicate device in all aspects, including design, materials, sterilization, and packaging. The only difference between the proposed and the predicate devices is that the proposed device now meets U.S.P. for needle attachment force and that is now reflected in the labeling.
PROLENE™ suture is indicated for use in general soft tissue Indications for Use approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Summary of Technological Characteristics of Modified Device to Predicate
The principle of operation and fundamental scientific technology of the modified device are equivalent to the predicate device. The performance modification was accomplished via adjustments to the swaging process settings. A comparison between the proposed and the predicate device is given in Table 1 below.
| | Proposed Device | Predicate Device |
|-------------------------|-----------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Number | TBD | K946173 |
| Product Code | Same | GAW |
| Regulation | Same | 21 CFR 878.5010 |
| Absorbable | Same | No |
| Intended Use | Same | PROLENE suture is indicated<br>for use in general soft tissue<br>approximation and/or ligation,<br>including use in cardiovascular,<br>ophthalmic and neurological<br>procedures. |
| How Supplied | Same | The devices are available in<br>one, two, or three dozen boxes. |
| Color | Same | Clear or Pigmented suture<br>strands (pigmented blue with<br>phthalocyanine blue, Color<br>Index Number 74160) |
| Material<br>Composition | Same | Composed of an isotactic<br>crystalline stereoisomer of<br>polypropylene, a synthetic<br>linear polyolefin. |
| Sterilization | Same | Sterilized by Ethylene Oxide |
| Packaging | Same | Plastic tray or paper folder<br>placed within sealed Tyvek - |
Table 1: Device Comparison Table
ETHICON, Inc.
{2}------------------------------------------------
PROLENE™ Polypropylene Nonabsorbable Suture Special 510(k) Pre-Market Notification
| Image: [Blurred image] | copolymer overwrap.<br>Individual overwrap pouch<br>packages placed within<br>paperboard carton. | |
|--------------------------------------------------|--------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| U.S.P.<br>requirements<br>Image: [Blurred image] | Meets U.S.P.<br>Monograph – except<br>for Diameter for 7-0<br>PROLENE | Meets U.S.P. Monograph –<br>except for Diameter for 7-0<br>PROLENE, and except for<br>Needle Attachment for HEMO-<br>SEAL. |
Design Verification Testing (U.S.P. < 871> - Sutures - Needle Performance Data Attachment; U.S.P. < 881> - Tensile Strength; U.S.P. < 861> - Sutures -Diameter) was conducted to demonstrate the performance of the modified PROLENE™ HEMOSEAL™ Polypropylene Nonabsorbable Suture.
Based on the similarities to the predicate device identified in this Conclusions submission as well as the outcome of design verification, and that the devices have the same intended use, the same fundamental technology, and the same principle of operation, we conclude that the modified device is substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three lines representing its wings and body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Frampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
December 6, 2013
Ethicon, Incorporated Mr. Elliott Jang, RAC Manager, Regulatory Affairs Route 22 West, P.O. Box 151 Somerville. New Jersey 08876
Re: K133356
Trade/Device Name: PROLENE™ POLYPROPYLENE NONABSORBABLE SUTURE Regulation Number: 21 CFR 878.5010 Regulation Name: Nonabsorbable polypropylene surgical suture Regulatory Class: Class II Product Code: GAW Dated: November 18, 2013 Received: November 19. 2013
Dear Mr. Jang:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). and obsintherefore, market the device, subject to the general controls provisions of the Act. The . Fou may , increaser , include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must or any rederal station and sequirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
· Page 2 - Mr. Elliott Jang, RAC
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
#### 510{k} Number (if known) K133356
#### Device Name
#### PROLENE™ POLYPROPYLENE NONABSORBABLE SUTURE
#### Indications for Use (Describe)
PROLENE™ suture is indicated for use in general soft tissue approximation and/or ligation, including use in cardiovascular, ophthalmic and neurological procedures.
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.