← Product Code [DYB](/productcode/DYB) · K133296

# XCATH (K133296)

_Alseal · DYB · Feb 14, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133296

## Device Facts

- **Applicant:** Alseal
- **Product Code:** [DYB](/productcode/DYB.md)
- **Decision Date:** Feb 14, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.1340
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular

## Indications for Use

XCath (model 2064-XC) is intended to be inserted into the HQS Introducer (model 2064-HQS) to establish multiple and simultaneous conduits through the same vascular access (previously established by the HQS introducer). The whole XCath / HQS allows insertion of several endovascular devices at the same time while minimizing blood loss associated with such insertions.

## Device Story

XCath is a multi-lumen accessory (2, 3, or 4 channels) designed to attach to the proximal end of the HQS Introducer; enables simultaneous insertion of multiple endovascular tools (up to 9.3F) through a single vascular access point. Device consists of a flexible body, silicone valves for each channel to maintain hemostasis, and extension lines with stopcocks and Luer Lock connections. Operated by clinicians in a clinical setting; functions as a conduit extension. By allowing multiple tools to pass through one access site, it reduces the need for additional vascular punctures, potentially decreasing patient trauma and blood loss.

## Clinical Evidence

No clinical data. Evidence consists of bench testing including tensile strength, overpressure resistance, sealing resistance, air leak (vacuum) testing, and compatibility with the HQS introducer. Device validated against ISO 11070:1999 standards.

## Technological Characteristics

Multi-lumen catheter accessory (2, 3, or 4 channels); silicone valves for hemostasis; extension lines with stopcocks and Luer Lock connectors. Compatible with 9.3F (3.1mm) tools. Sterile, single-use. Biocompatibility per ISO 10993-1 for externally communicating devices in contact with circulating blood < 24 hours.

## Regulatory Identification

A catheter introducer is a sheath used to facilitate placing a catheter through the skin into a vein or artery.

## Predicate Devices

- HQS Introducer ([K113849](/device/K113849.md))
- Edwards Percutaneous Sheath Introducer ([K121185](/device/K121185.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K133296

FEB 1 4 2014

# 510(k) Summary 807.92(c)

#### SPONSOR

# 807.92(a)(1)

| Company Name:   | ALSEAL                                          |
|-----------------|-------------------------------------------------|
| Company Address | 2 rue Paul Milleret<br>25000 Besançon<br>France |

| Telephone: | +33(0)3 81 61 69 93 |
|------------|---------------------|
| Fax:       | +33(0)3 81 53 47 65 |

Jean-François DELFORGE Contact Person:

Summary Preparation Date: September 29, 2013

### DEVICE NAME

| DEVICE NAME          |                       | 807.92(a)(2) |
|----------------------|-----------------------|--------------|
| Trade Name:          | XCath (Model 2064-XC) |              |
| Common/Usual Name:   | Catheter Introducer   |              |
| Classification Name: | Introducer, Catheter  |              |
| Regulation Number:   | 870.1340              |              |
| Product Code:        | DYB                   |              |
| Device Class:        | Class II              |              |

## PREDICATE DEVICE

807.92(a)(3)

| K Number | Product Name                              | Manufacturer               |
|----------|-------------------------------------------|----------------------------|
| K113849  | HQS Introducer                            | ALSEAL                     |
| K121185  | Edwards Percutaneous Sheath<br>Introducer | Edwards Life Sciences, LLC |

## DEVICE DESCRIPTION

## 807.92(a){4}

The XCath (model 2064-XC) consists of a multiple ways access (2, 3 or 4 ways) adapted to the HQS Introducer (K113849) to optimize the sealing when several vascular tools up to 9,3F need to be introduced through the same vascular access.

The device features include (1) a flexible multilumen body, with 2, 3 or 4 channels compatible with the open valve of HQS Introducer; (2) a base with 2, 3 or 4 valves, (3) Extension line for each valve with stopcock and Luer Lock connection.

Once the XCath is connected to the HQS Introducer, the device can accommodate vascular tools up to 9,3 French (3,1mm). The XCath (accessory) is fixed in proximal position of the HQS introducer body and allows the simultaneous insertion of several vascular tools.

{1}------------------------------------------------

## DEVICE INDICATIONS FOR USE

# 807.92(a)(5)

XCath (model 2064-XC) is intended to be inserted into the HQS Introducer (model 2064-HQS) to establish multiple and simultaneous conduits through the same vascular access (previously established by the HQS introducer).

The whole XCath / HQS allows insertion of several endovascular devices at the same time while minimizing blood loss associated with such insertions.

|                                                 | Subject Device                                                                                                                                                                                                                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                                  | Predicate Device                                                                                                                                                                                                          |
|-------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Manufacturer                                    | ALSEAL                                                                                                                                                                                                                                                                                                                                                                                                                          | ALSEAL                                                                                                                                                                                                                            | Edwards Life Sciences, LLC                                                                                                                                                                                                |
| Trade Name                                      | XCath (model 2064-XC)                                                                                                                                                                                                                                                                                                                                                                                                           | HQS Introducer                                                                                                                                                                                                                    | Edwards Percutaneous<br>Sheath Introducer<br>AVA 3Xi and<br>AVA High Flow                                                                                                                                                 |
| K number                                        |                                                                                                                                                                                                                                                                                                                                                                                                                                 | K113849                                                                                                                                                                                                                           | K 121185                                                                                                                                                                                                                  |
| Product Code                                    | DYB                                                                                                                                                                                                                                                                                                                                                                                                                             | DYB                                                                                                                                                                                                                               | DYB                                                                                                                                                                                                                       |
| Regulation No.                                  | 870.1340                                                                                                                                                                                                                                                                                                                                                                                                                        | 870.1340                                                                                                                                                                                                                          | 870.1340                                                                                                                                                                                                                  |
| Indications for<br>Use                          | XCath (model 2064-XC) is<br>intended to be inserted<br>into the HQS Introducer<br>(model 2064-HQS) to<br>establish multiple and<br>simultaneous conduits<br>through the same vascular<br>access (previously<br>established by the HQS<br>introducer).<br>The whole XCath / HQS<br>allows insertion of several<br>endovascular devices at<br>the same time while<br>minimizing blood loss<br>associated with such<br>insertions. | The HQS Introducer (Model<br>2064-HQS) is intended to be<br>inserted in the vasculature<br>to provide a conduit for the<br>insertion of endovascular<br>devices while minimizing<br>blood loss associated with<br>such insertions | The Edwards Percutaneous<br>Sheath Introducer is<br>indicated for use in patients<br>requiring access of the<br>venous system and to<br>facilitate catheter insertion<br>(e.g. pulmonary artery or<br>infusion catheter). |
| Principle of<br>Operation                       | Transitions the HQS with<br>a single insertion into a<br>multi lumen device<br>based on the number of<br>access conduits required<br>by the use of multiple<br>silicone valves.                                                                                                                                                                                                                                                 | The tools are introduced in<br>the vasculature by placing a<br>guide through the device<br>and going through a silicone<br>valve minimizing blood loss                                                                            | Transitions from an<br>introducer to a triple lumen<br>device with a fixed or<br>detachable introducer valve<br>as treatment changes<br>accomplished through single<br>percutaneous introducer                            |
| Device Design                                   | Each channel of XCath<br>accessory has an<br>approaching design of HQS<br>introducer (sheath<br>equipped with a silicone<br>valve and a lateral                                                                                                                                                                                                                                                                                 | Sheath equipped with a<br>silicone valve and a lateral<br>extension line with Luer<br>Lock connectors / 3 way<br>stopcock.                                                                                                        | Introducers are composed<br>of valve housing to which a<br>sheath is attached distally<br>and three side<br>arm/extension tubes are<br>connected proximally.                                                              |
|                                                 | extension line with Luer<br>Lock connectors / 3 way<br>stopcock).                                                                                                                                                                                                                                                                                                                                                               | Furnished with a dilator for<br>facilitating the introduction<br>into vasculature.                                                                                                                                                | The valves located in the<br>housing body provide a seal<br>around a catheter when<br>inserted through the<br>introducer and prevent<br>backflow when no catheter<br>is present                                           |
| Number of<br>channels for<br>vasculature access | As an accessory to the HQS<br>Introducer:<br>2 way<br>3 way<br>4 way                                                                                                                                                                                                                                                                                                                                                            | 1 channel Introducer                                                                                                                                                                                                              | Triple lumen with 4th access<br>via the fixed or detachable<br>introducer valve                                                                                                                                           |
| Valve comparison                                | Silicone valve ensures<br>sealing while empty or<br>accessed with guide / max<br>dilator.                                                                                                                                                                                                                                                                                                                                       | Silicone valve ensures<br>sealing while empty or<br>accessed with guide / max<br>dilator                                                                                                                                          | Valve ensures sealing while<br>empty or insertion eased<br>with provided dilator                                                                                                                                          |
| Biocompatibility<br>Testing                     | ISO 10993-1 for External<br>communicating device,<br>direct circulating blood<br>path, duration ≥ 24 hours                                                                                                                                                                                                                                                                                                                      | ISO 10993-1 for External<br>communicating device,<br>direct circulating blood<br>path, duration ≥ 24 hours                                                                                                                        | ISO 10993-1 for External<br>communicating device,<br>direct circulating blood<br>path, duration ≥ 24 hours                                                                                                                |
| Sterilization                                   | Provide sterile;<br>Single use.                                                                                                                                                                                                                                                                                                                                                                                                 | Provide sterile; Single use.                                                                                                                                                                                                      | Provide sterile; Single use.                                                                                                                                                                                              |

# COMPARISON OF TECHNICAL CHARACTERISTICS 807.92(a)(6)

{2}------------------------------------------------

### NON-CLINICAL PERFORMANCE DATA

807.92(b)(1)

#### SAFETY BENCH TESTING

In accordance with ISO 10993-1, XCath (Model 2064-XCath) is categorized as externally communicating devices in contact with circulating blood for less than 24 hours. The following tests are recommended by ISO 10993-1.

#### PERFORMANCE TESTING

The following in vitro testing was performed on the XCath (Model 2064-XCath) in accordance with ISO standards and/or internal procedures to assure reliable design and performance. In vitro design verification testing data demonstrates that the device is in compliance with ISO 11070: 1999 Sterile, single use intravascular catheters introducers and product labeling.

- 1. Tensile strength
- 2. Overpressure resistance
- 3. Dimensions and aspect
- 4. Sealing resistance
- 5. Air leak (vacuum) test
- 6. Compatible with HQS introducer
- 7. Sealing of the valve
- Free passage a guide and 9F vascular tool 8.
- 9. Validation of luer and extension lines
- 10. Ease of assembly
- 11. Traction resistance
- 12. Packaging resistance

{3}------------------------------------------------

- 13. Sterilization tests
- 14. Aging tests

# CONCLUSION

# 807.92(b)(3)

XCath (model 2064-XC) is substantially equivalent to the predicate device in Indications for Use, Materials and design. Safety and performance testing to ISO 11070 and Biocompatibility testing has concluded that the device does not introduce significant questions of safety and is substantially equivalent to the predicate.

{4}------------------------------------------------

Image /page/4/Picture/10 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized representation of three human profiles facing right, stacked on top of each other. The profiles are black and have a flowing, abstract design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the profiles.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002

February 14, 2014

Alseal C/O Yolanda Smith Consultant Smith Associates 1468 Harwell Avenue Crofton, MD 21114 US

Re: K133296

> Trade/Device Name: XCath Model 2064-XC Regulation Number: 21 CFR 870.1340 Regulation Name: Catheter Introducer Accessory Regulatory Class: Class II Product Code: DYB Dated: January 8, 2014 Received: January 10, 2014

Dear Ms. Smith:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

{5}------------------------------------------------

Page 2 - Ms. Yolanda Smith

If your device is classified (see above) into either class II {Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801; medical device reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801) please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours.

M. A. Z. Lilliencron

for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{6}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

# Indications for Use

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

510(k) Number (if known)

K133296

Device Name XCath (Model 2064-XC)

Indications for Use (Describe)

XCath (model 2064-XC) is intended to be inserted into the HQS Introducer (model 2064-HQS) to establish multiple and simultaneous conduits through the same vascular access (previously established by the HQS introducer).

The whole XCath / HQS allows insertion of several endovascular devices at the same time while minimizing blood loss associated with such insertions.

Type of Use (Select one or both, as applicable)

2 Prescription Use (Part 21 CFR 801 Subpart D)

[] Over-The-Counter Use (21 CFR 801 Subpart C)

#### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.

#### FOR FDA USE ONLY

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

M.A. Hillerman

ﺮ

---

**Source:** [https://fda-staging.innolitics.com/device/K133296](https://fda-staging.innolitics.com/device/K133296)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).