The ActiV.A.C., InfoV.A.C., V.A.C. ATS, V.A.C. Freedom, V.A.C. Via, and V.A.C. Simplicity Negative Pressure Wound Therapy Systems are integrated wound management systems for use in acute, extended and home care settings. When used on open wounds, they are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts. When used on closed surgical incisions, they are also intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
Device Story
V.A.C. Therapy Wound Dressings function as part of the V.A.C. Negative Pressure Wound Therapy (NPWT) System. The system applies controlled negative pressure to a wound site via a dressing, which is connected to a powered suction pump. This process removes exudate and infectious material, reduces edema, and promotes granulation tissue formation and perfusion, creating an environment conducive to healing. The system is used in acute, extended, and home care settings by clinicians or patients. The dressing maintains a closed environment over the wound or surgical incision. By managing the wound environment, the device assists in preparing the wound bed for closure and managing post-surgical drainage, potentially accelerating healing and improving patient outcomes.
Clinical Evidence
No clinical study was conducted. Design validation was performed on healthy human volunteers to ensure the modified device meets user requirements, as no reliable bench-top method exists for these specific metrics.
Technological Characteristics
The system consists of wound dressing components used with a powered suction pump. Materials are evaluated for biocompatibility per ISO 10993-1. Key performance characteristics include peel adhesion and moisture vapor transmission rate (MVTR). The system is designed for use in acute, extended, and home care environments.
Indications for Use
Indicated for patients in acute, extended, or home care settings requiring wound management. Applicable to open wounds (chronic, acute, traumatic, subacute, dehisced, partial-thickness burns, ulcers, flaps, grafts) and closed surgical incisions with drainage.
Regulatory Classification
Identification
A powered suction pump is a portable, AC-powered or compressed air-powered device intended to be used to remove infectious materials from wounds or fluids from a patient's airway or respiratory support system. The device may be used during surgery in the operating room or at the patient's bedside. The device may include a microbial filter.
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for KCI. The logo consists of a series of curved lines on the left, followed by the letters "KCI" in a bold, sans-serif font. The letters are outlined in black, and the interior of the letters is white. There is a small circle after the I.
::
...
... ......................................................................................................................................................................
510(k) SUMMARY
- - -- -- -- --
・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・
..
NOV 1 9 2013
-K133276.page 1.of 2
' .
## V.A.C. ® Therapy Wound Dressings
| Submitter Information [21 CFR 807.929(a)(1)] | |
|----------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Name | KCI USA, Inc. (Kinetic Concepts, Inc.) |
| Address | 6203 Farinon Drive<br>San Antonio, TX 78249 |
| Phone number | 210-515-4059 |
| Fax number | 210-255-6727 |
| Establishment<br>Registration Number | 1625774 |
| Name of contact person | Melanie Avila |
| Date prepared | 10/23/2013 |
| Name of the device [21 CFR 807.92(a)(2)] | |
| Trade or proprietary name | V.A.C. ® Negative Pressure Wound Therapy System |
| Common or usual name | Negative Pressure Wound Therapy System |
| Classification name | Negative Pressure Wound Therapy Powered Suction Pump (and components) |
| Regulation | 878.4780 |
| Product Code(s) | OMP |
| Legally marketed device(s) to which<br>equivalence is claimed<br>[21 CFR 807.92(a)(3)] | V.A.C. Therapy Wound Dressings cleared for use under multiple 510(k)s for<br>the KCI V.A.C. Therapy Negative Pressure Wound Therapy Systems. The<br>most recent 510(k) was K120033. |
| Device description<br>[21 CFR 807.92(a)(4)] | V.A.C. Therapy Wound Dressings for use with KCI's V.A.C. NPWT System. |
| Indications for use<br>[21 CFR 807.92(a)(5)] | The ActiV.A.C., InfoV.A.C., V.A.C. ATS, V.A.C. Freedom, V.A.C. Via, and<br>V.A.C. Simplicity Negative Pressure Wound Therapy Systems are integrated<br>wound management systems for use in acute, extended and home care<br>settings.<br><br>When used on open wounds, they are intended to create an environment that<br>promotes wound healing by secondary or tertiary (delayed primary) intention<br>by preparing the wound bed for closure, reducing edema, promoting<br>granulation tissue formation and perfusion, and by removing exudate and<br>infectious material. Open wound types include: chronic, acute, traumatic,<br>subacute and dehisced wounds, partial-thickness burns, ulcers (such as<br>diabetic, pressure or venous insufficiency), flaps and grafts.<br><br>When used on closed surgical incisions, they are also intended to manage the<br>environment of surgical incisions that continue to drain following sutured or<br>stapled closure by maintaining a closed environment and removing exudates<br>via the application of negative pressure wound therapy. |
{1}------------------------------------------------
. ----------------------------------------------------------------------------------------------------------------------------------------------------------------------------
Image /page/1/Picture/1 description: The image shows a logo with the letters "EKCI" in a stylized font. To the left of the letters, there are several curved lines that resemble a stylized "E". The letters "K", "C", and "I" are in a bold, outlined font. A small circle is located to the right of the "I".
- ਸ਼ਾਪੂ
... ...
... ...
#### 510(k) SUMMARY
-----
### V.A.C. ® Therapy Wound Dressings
Summary of the technological characteristics of the device compared to the predicate device [21 CFR 807.92(a)(6)]
| Characteristic | V.A.C. ® Therapy Wound Dressing System (Modified Device) | V.A.C. ® Therapy Wound Dressing System (Predicate) |
|----------------------|----------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Indicated Wound Type | Same as predicate | Chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts |
| Dressing | Same as predicate | Multiple dressing components |
Performance Data [21 CFR 807.92(b)]
Summary of non-clinical tests conducted for determination of substantial equivalence [21 CFR 807.92(b)(1)]
The V.A.C. ® Dressing System was evaluated to ensure conformance to the design specifications.
The following tests were conducted:
- Peel adhesion testing
- Moisture vapor transmission rate (MVTR) testing
- Biocompatibility testing according to ISO10993-1
- Study conducted on healthy human volunteers, under design validation (210 CFR 820.30(g)), to ensure that the proposed modified device meets user requirements
Summary of clinical tests conducted for determination of substantial equivalence or of clinical information [21 CFR 807.92(b)(2)]
There was no clinical study conducted, however, a study was conducted on healthy human volunteers, under design validation (210 CFR 820.30(g)), to ensure that the proposed modified device meets user requirements as there is no reliable method for measuring these requirements on the bench.
Conclusions drawn [21 CFR 807.92(b)(3)]
Testing indicates that the modified V.A.C. ® Therapy Wound Dressing System is substantially equivalent in terms of both indications for use and fundamental scientific technology to the predicate product.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three curved lines representing arms or wings.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
KCI USA, Incorporated Ms. Melanie Avila Regulatory Affairs Project Manager 6203 Fairnon Drive San Antonio, Texas 78249
November 19, 2013
Re: K133276
Trade/Device Name: V.A.C. Negative Pressure Wound Therapy System Regulation Number: 21 CFR 878.4780 Regulation Name: Powered suction pump Regulatory Class: Class II Product Code: OMP Dated: October 23, 2013 Received: October 24, 2013
Dear Ms. Avila:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{3}------------------------------------------------
Page 2 - Ms. Melanie Avila
・・・・・・
ﺍﻟﻤﻮﺿﻮﻋﺎﺕ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﺤﺪﺓ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤﺘﻮﻗﻊ ﺍﻟﻤ
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industrv/default.htm.
Sincerely vours.
# Joshua C. Nipper -S
For
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### INDICATIONS FOR USE
510(k) Number (if known): __ K.133276 Device Name: _ V.A.C. ® Drape Indications for Use:
The ActiV.A.C., InfoV.A.C., V.A.C. ATS, V.A.C. Freedom, V.A.C. Via, and V.A.C. Simplicity Negative Pressure Wound Therapy Systems are Integrated wound management systems for use in acute, extended and home care settings.
When used on open wounds, they are intended to create an environment that promotes wound healing by secondary or tertiary (delayed primary) intention by preparing the wound bed for closure, reducing edema, promoting granulation tissue formation and perfusion, and by removing exudate and infectious material. Open wound types include: chronic, acute, traumatic, subacute and dehisced wounds, partial-thickness burns, ulcers (such as diabetic, pressure or venous insufficiency), flaps and grafts.
When used on closed surgical incisions, they are also intended to manage the environment of surgical incisions that continue to drain following sutured or stapled closure by maintaining a closed environment and removing exudates via the application of negative pressure wound therapy.
Prescription Use Over-The-Counter Use ਮ AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page | of | |
(Posted November 13, 2003)
Image /page/4/Picture/10 description: The image shows the name "Jiyoung Dang -S" in a bold, sans-serif font. The text is black against a white background. There is some noise or artifacts around the letters, possibly due to the image quality or scanning process.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.