K133265 · Maquet Cardiopulmonary, AG · KFM · Nov 12, 2013 · Cardiovascular
Device Facts
Record ID
K133265
Device Name
QUADROX-IR ADULT AND SMALL ADULT
Applicant
Maquet Cardiopulmonary, AG
Product Code
KFM · Cardiovascular
Decision Date
Nov 12, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 870.4360
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours). The CARDIOHELP System in configuration with the HLS/HIT Set Advanced is intended to be used within the hospital environment and outside the hospital environment (for periods up to six hours), e.g. for intra- and inter-hospital transport. The CARDIOHELP System in configuration with the QUADROX-iR is intended to be used in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery (for periods for up to six hours).
Device Story
The QUADROX-iR is a disposable oxygenator with an integrated centrifugal pump and optional arterial filter, used as part of the CARDIOHELP System for extracorporeal circulation. The device facilitates blood oxygenation and carbon dioxide removal during cardiopulmonary bypass. This modification introduces a reusable stainless steel holding clamp, applied by a perfusionist prior to use, to serve as an interim protective safeguard preventing accidental disconnection of inlet/outlet connectors. The clamp is verified to remain secure under mechanical stress and vibration during operation. The device is intended for use in hospital environments by trained medical professionals. The modification does not alter the fundamental scientific technology or performance specifications of the oxygenator; it enhances mechanical safety by ensuring connector integrity during the six-hour maximum usage period.
Clinical Evidence
No clinical data was required or presented for this submission. Substantial equivalence was established through bench testing and design verification, including mechanical holding force, pull-off force, vibration testing (up to 3G), and usability testing in three hospitals.
Technological Characteristics
Disposable oxygenator with integrated centrifugal pump and optional arterial filter. Modification adds a reusable stainless steel holding clamp. Operates via extracorporeal circulation. No changes to materials, energy source, or fundamental scientific technology. Verification testing confirmed mechanical integrity of the clamp under vibration and load.
Indications for Use
Indicated for patients requiring circulatory and/or pulmonary support via extracorporeal bypass circuit during cardiac surgery or other procedures requiring cardiopulmonary bypass, for durations up to six hours.
Regulatory Classification
Identification
A nonroller-type cardiopulmonary and circulatory bypass blood pump is a prescription device that uses a method other than revolving rollers to pump the blood through an extracorporeal circuit for periods lasting less than 6 hours for the purpose of providing either:(i) Full or partial cardiopulmonary bypass (i.e., circuit includes an oxygenator) during open surgical procedures on the heart or great vessels; or(ii) Temporary circulatory bypass for diversion of flow around a planned disruption of the circulatory pathway necessary for open surgical procedures on the aorta or vena cava.
Special Controls
*Classification* —Class II (special controls). The special controls for this device are:(i) Non-clinical performance testing must perform as intended over the intended duration of use and demonstrate the following: Operating parameters, dynamic blood damage, heat generation, air entrapment, mechanical integrity, and durability/reliability;
(ii) The patient-contacting components of the device must be demonstrated to be biocompatible;
(iii) Sterility and shelf life testing must demonstrate the sterility of patient-contacting components and the shelf life of these components; and
(iv) Labeling must include information regarding the duration of use, and a detailed summary of the device- and procedure-related complications pertinent to use of the device.
(b)
*Nonroller-type temporary ventricular support blood pump* —(1)*Identification.* A nonroller-type temporary ventricular support blood pump is a prescription device that uses any method resulting in blood propulsion to provide the temporary ventricular assistance required for support of the systemic and/or pulmonary circulations during periods when there is ongoing or anticipated hemodynamic instability due to immediately reversible alterations in ventricular myocardial function resulting from mechanical or physiologic causes. Duration of use would be less than 6 hours.(2)
*Classification.* Class III (premarket approval).(c)
*Date premarket approval application (PMA) or notice of completion of product development protocol (PDP) is required.* A PMA or notice of completion of a PDP is required to be filed with FDA on or before September 8, 2015, for any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976, or that has, on or before September 8, 2015, been found to be substantially equivalent to any nonroller-type temporary ventricular support blood pump that was in commercial distribution before May 28, 1976. Any other nonroller-type temporary ventricular support blood pump shall have an approved PMA or declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
#### K133265
NOV 1 2 2013
#### MAQUET Cardiopulmonary AG
Premarket Notification Special 510(k) - Change due to recall/corrective action Modified QUADROX-iR Oxygenators
# 510 (k) Summary [As required by 21 CFR 807.92(c)]
Date:
### October 22, 2013
Submitter:
MAQUET Cardiopulmonary AG Kehler Straße 31 76437 Rastatt Germany
Contact Person:
Whitney Törning Director of Regulatory MAQUET Cardiovascular 45 Barbour Pond Drive Wayne, NJ 07470 Phone: 973-709-7994 Fax: 973-807-9210
Device Trade Names: QUADROX-iR Adult / Small Adult Common/Usual name: Oxygenator with integrated centrifugal pump and with / without integrated Arterial Filter
Classification Name: Oxygenator, cardiopulmonary bypass, heat exchanger, cardiopulmonary bypass, filter, blood cardiopulmonary bypass, arterial line, non-roller type blood pump, cardiopulmonary bypass
{1}------------------------------------------------
Premarket Notification Special 510(k) - Change due to recall/corrective action Modified QUADROX-iR Oxygenators
### Device Modification
The only modification consists of adding a reusable stainless steel holding clamp as an interim protective safeguard / corrective action to prevent the inlet and/or outlet connector from disconnection. The holding clamp is applied by the perfusionist prior to use.
# Legally marketed MAQUET Predicate Devices:
Primary predicate device: QUADROX-iR Adult K102726, QUADROX-iR Small Adult K102726
Secondary predicate devices: QUADROX-i Adult Microporous Membrane Oxygenators K132829. QUADROX-i Small Adult Microporous Membrane Oxygenators K132829, QUADROX-i Pediatric Microporous Membrane Oxygenator K132829, QUADROX-i Neonatal Microporous Membrane Oxygenator K132829, QUADROX-iD Adult Diffusion Membrane Oxygenator K132829 and QUADROX-iD Pediatric Diffusion Membrane Oxygenator K132829.
# Device Description:
# Holding Clamp - interim protective safequard and component of the QUADROX-iR Oxygenator
The holding clamp is applicable to all QUADROX-iR oxygenators which are subject to this submission. The following table provides an overview of the QUADROX-iR oxygenator models and the suitable holding clamp.
| Catalogue number | Holding clamp - picture | suitable for* |
|------------------|-------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| HKH 3572 | Image: Holding clamp | HMO 70100 USA - Quadrox iR Adult w/o filter<br>HMO 71100 USA- Quadrox iR Adult with filter<br>HMO 50100 USA - Quadrox iR Small Adult w/o filter<br>HMO 51100 USA - Quadrox iR Small Adult with filter |
{2}------------------------------------------------
Premarket Notification Special 510(k) - Change due to recall/corrective action Modified QUADROX-iR Oxygenators
* The above listed QUADROX-iR Oxygenator models may also be provided with BIOLINE Coating and may be sold in combination with other devices as described in the original 510(k) (K102726).
Image /page/2/Picture/3 description: The image shows a machine part, possibly from an engine or motor. The part is made of metal and has a cylindrical shape at the top, with a more complex structure below. There are various pipes, tubes, and other components attached to the main body of the part, suggesting it is involved in fluid or gas transfer. The image is a black and white rendering, possibly a technical drawing or a 3D model.
Picture: QUADROX-iR with Holding Clamp
# QUADROX-iR (disposable) as part of the CARDIOHELP System, the indications for use is as follows:
The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours).
The CARDIOHELP System in configuration with the HLS/HIT Set Advanced is intended to be used within the hospital environment and outside the hospital environment (for periods up to six hours), e.g. for intra- and inter-hospital transport".
The CARDIOHELP System in configuration with the QUADROX-iR is intended to be used in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery (for periods for up to six hours).
1 (Note: This section of the indications for use statement does not refer to the HLS/HIT Set Advanced (K102726) and therefore it is not subject to this Special 510(k)).
{3}------------------------------------------------
Premarket Notification Special 510(k) - Change due to recall/corrective action Modified QUADROX-iR Oxygenators
# Technological Characteristics:
The modified device as mentioned above is identical to the originally cleared device, with the exception that an added reusable stainless steel holding clamp is applied by the perfusionist prior to use as an interim protective safequard in the unlikely event that the inlet and/or outlet connector would become loose. Besides this difference, the proposed device is the same in design, intended use, method of operation, components, packaging, and fundamental scientific technology as the predicate device (QUADROX-iR Adult and Small Adult).
### Non-clinical Tests:
The modified device was subjected to design verification tests which are based on the risk assessment. The tests are designed to show that the clamp can be mounted on the Oxygenator, so that the clamp will stay in place during its use, and in the unlikely event that the bonding of the connectors to the housing would fail the connector would not detach from the housing..The tests simulate relevant mechanical conditions in the use of the Oxygenators. The following mechanical tests were performed:
- -Holding Forces - Force applied to the connectors to hold the clamp in place
- Pull off Forces - Force to remove the holding clamp from the housing perpendicular to the clamping force.
- Vibration Test with Shaker Simulate impact forces up to 3G and confirm the secure attachment of the clamp under repeated impact.
- Vibration Test with Cardiohelp Test to demonstrate that the clamp t stays securely fixed over the total application time of up to 6 hours.
- Usability test This test was performed in three hospitals and evaluated as to whether the clamp could be mounted using the Instruction for Use.
The holding clamp as an external component of the cleared QUADROX-iR does not affect the performance specifications related to the intended use.
Based on the tests listed above, the modified device is safe and effective according to the intended use and is substantially equivalent to the originally cleared device.
{4}------------------------------------------------
Premarket Notification Special 510(k) - Change due to recall/corrective action Modified QUADROX-iR Oxygenators
### Clinical Tests:
Clinical results are not required for this submission to support substantial equivalence.
### Conclusion:
Based on the risk analysis, MAQUET Cardiopulmonary AG has conducted the appropriate design verification activities and believes that the modified device is substantially equivalent to the cleared MAQUET primary predicate device.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES USA
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -- WO66-Gb(19 Silver Spring, MD 20993-0002
November 12, 2013
MAQUET Cardiopulmonary AG Whitney Torning Director of Regulatory 45 Barbour Pond Drive Wayne, ND 07470
Re: K133265
Trade/Device Name: CARDIOHELP System with the QUADROX-iR Adult Oxygenator (Models 70100 and 70110 Bioline and Softline Coated) and the QUADROX-iR Small Adult Oxygenator (Models 50100 and 50110 Bioline and Softline Coated) Regulation Number: 21 CFR 870.4360 Regulation Name: Non-roller type cardiopulmonary bypass blood pump
Regulatory Class: Class III
Product Code: KFM, DTZ, DTZ, DTM
Dated: October 22, 2013
Received: October 23, 2013
Dear Ms. Törning:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments. or 10 devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements. including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{6}------------------------------------------------
Page 2 - Ms. Whitney Torning
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
M& Zillehenn
for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known): K133265
# Device Name: CARDIOHELP System (including QUADROX-iR)
Indications for Use The CARDIOHELP System is a blood oxygenation and carbon dioxide removal system used to pump blood through the extracorporeal bypass circuit for circulatory and/or pulmonary support during procedures requiring cardiopulmonary bypass (for periods up to six hours). It is also intended to provide circulatory and/or pulmonary support during procedures not requiring cardiopulmonary bypass (for periods up to six hours).
The CARDIOHELP System in configuration with the HLS/HIT Set Advanced is intended to be used within the hospital environment and outside the hospital environment (for periods up to six hours), e.g. for intra- and inter-hospital transport.
The CARDIOHELP System in configuration with the QUADROX-iR is intended to be used in extracorporeal circulation during cardiopulmonary bypass in cardiac surgery (for periods for up to six hours).
Over-The-Counter Use × Prescription Use Prescription Use _____________________________________________________________________________________________________________________________________________________________ (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Page of of __
M.A. Hillmann
05-2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.