K133246 · Zimmer, Inc. · HRS · Dec 17, 2013 · Orthopedic
Device Facts
Record ID
K133246
Device Name
ZIMMER DISTAL RADIUS PLATING SYSTEM
Applicant
Zimmer, Inc.
Product Code
HRS · Orthopedic
Decision Date
Dec 17, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 888.3030
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Zimmer Distal Radius Plating System is intended for use in surgical reduction, internal fixation, and stabilization of fractures and osteotomies of the distal radius. Examples of distal radius fractures include but are not limited to compression fractures, intra-articular and extra-articular fractures, displaced fractures, fractures in osteopenic bone, non-unions, and mal-unions.
Device Story
Zimmer Distal Radius Plating System comprises volar, medial, and dorsal locking plates with associated screws; used for internal fixation and stabilization of distal radius fractures and osteotomies. Plates feature screw holes accepting locking or non-locking screws; slots accept non-locking screws only. Locking screws/pegs allow 15° angulation (30° cone). Anatomically contoured design creates fixed or variable angled constructs for simple or multi-fragment fractures. System utilized by surgeons in clinical settings for fracture management; provides mechanical stability to bone segments to facilitate healing.
Clinical Evidence
No clinical data. Substantial equivalence supported by non-clinical bench testing, including sterilization validation (gamma, 20kGy, SAL 10^-6), shelf-life testing (10 years), biocompatibility (ISO 10993-1), and mechanical performance testing (fatigue, torsional strength, pull-out strength).
Technological Characteristics
Materials: Ti-6Al-4V ELI alloy. Type II anodized plates/screws; color anodized non-locking screws. Locking/non-locking screw capability. 15-degree variable angle locking technology. Sterilization: Gamma radiation. Shelf life: 10 years.
Indications for Use
Indicated for surgical reduction, internal fixation, and stabilization of distal radius fractures and osteotomies, including compression, intra-articular, extra-articular, displaced, and osteopenic bone fractures, as well as non-unions and mal-unions.
Regulatory Classification
Identification
Single/multiple component metallic bone fixation appliances and accessories are devices intended to be implanted consisting of one or more metallic components and their metallic fasteners. The devices contain a plate, a nail/plate combination, or a blade/plate combination that are made of alloys, such as cobalt-chromium-molybdenum, stainless steel, and titanium, that are intended to be held in position with fasteners, such as screws and nails, or bolts, nuts, and washers. These devices are used for fixation of fractures of the proximal or distal end of long bones, such as intracapsular, intertrochanteric, intercervical, supracondylar, or condylar fractures of the femur; for fusion of a joint; or for surgical procedures that involve cutting a bone. The devices may be implanted or attached through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
K133246
Page 1 of 3
Image /page/0/Picture/2 description: The image shows the logo for Zimmer. The logo consists of a circle with a stylized letter "Z" inside it. Below the circle, the word "Zimmer" is written in a font that appears to be outlined.
DEC 1 7 2013
P.O. Box 708 Warsaw, IN 46581-0708 574 267-6131
### 510(k) Summary
Sponsor:
Zimmer, Inc. P.O. Box 708 Warsaw, IN 46581-0708
Contact Person:
Romil Sheth Specialist, Regulatory Affairs, Zimmer Inc. Telephone: (574) 371 - 1621 Fax: (574) 371 - 8760
Date Summary Prepared:
10/15/2013
Trade Name:
Zimmer Distal Radius Plating System
Common Name:
Classification Name
Locking Plate System
Plate, Fixation, Bone (21 CFR 888.3030); Screw, Fixation, Bone (21 CFR 888.3040)
Product Code:
and Reference:
HRS, HWC
Orthopedic/87
Classification Panel:
Predicate Device(s):
-Stryker® Leibinger Universal Distal Radius System, Stryker, K040022, cleared on March 12th, 2004 - Zimmer® Periarticular Locking Plate System, Zimmer; K040593, cleared on April 12th, 2004
13
{1}------------------------------------------------
Purpose and Device Description:
Intended Use:
Comparison to Predicate Device:
Performance Data (Nonclinical and/or Clinical):
The Zimmer Distal Radius Plating System is a plate and screw system intended for internal fixation and stabilization of fractures and osteotomies of the distal radius. The Zimmer Distal Radius Plating System consists of Volar, Medial and Dorsal locking plates. All the plates have screw holes which can accept either locking or non-locking screws and slots which can accept only non-locking screws. Locking screws and pegs may be inserted at 15° in any direction (30° cone) from the nominal screw hole trajectory. Thus, anatomically contoured plate design combined with locking screw technology creates fixed or variable angled constructs for use in simple or multi-fragment fractures. All the plates and screws are made from Ti-6AI-4V ELI alloy. All the plates and screws are Type II anodized except the nonlocking screws which are color anodized.
The Zimmer Distal Radius Plating System is intended for use in surgical reduction, internal fixation, and stabilization of fractures and osteotomies of the distal radius. Examples of distal radius fractures include but are not limited to compression fractures, intra-articular and extra-articular fractures, displaced fractures, fractures in osteopenic bone, non-unions, and mal-unions.
The Zimmer Distal Radius Plating System is similar in intended use, materials, sterility, and performance characteristics to the predicate devices.
Non-clinical performance and conclusions: -Sterilization Validation: These devices have been validated to demonstrate that at a minimum gamma
dose of 20kGy they can be terminally sterilized to a SAL of 10-6 or better.
{2}------------------------------------------------
-Shelf Life: Accelerated aging test conducted shows that the sterile devices included in this submission have a shelf life of 10 years
-Biocompatibility: Biocompatibility testing of the material used to manufacture plates and screws included in this submission was conducted per ISO 10993-1 and Good Laboratory Practices (21 CFR § 58). All testing passed.
-Performance testing: Testing performed included; Dynamic/Fatigue construct testing, evaluating the torsional failure strength, driving torque and axial pull out strength for the Zimmer Distal Radius Plating System screws.
The results of non-clinical performance testing demonstrate that the devices are safe and effective and substantially equivalent to the predicate devices.
#### Clinical performance and conclusions:
Clinical data and conclusions were not needed to show substantial equivalence.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol, with its wings forming a protective shape. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.
Food and Drug Administration ... 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
December 17, 2013
Zimmer. Incorporated Mr. Romil Sheth Specialist, Trauma Regulatory Affairs P.O. Box 708 Warsaw, Indiana 46581
Re: K133246
Trade/Device Name: Zimmer Distal Radius Plating System Regulation Number: 21 CFR 888.3030 Regulation Name: Single/multiple component metallic bone fixation appliances and accessories Regulatory Class: Class II Product Code: HRS, HWC Dated: October 21, 2013 Received: October 22, 2013
Dear Mr. Sheth:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (2) CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
#### Page 2 - Mr. Romil Sheth
forth in the quality systems (OS) regulation (21 CFR Part 820): and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If vou desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address. http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
## Ronaldflean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
A * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * * *
#### Indications for Use
#### 510(k) Number (if known): K133246
### Device Name:
ﻣﻦ ﺍﻟﻤﺮﺍﺟﻊ ﺍﻟﻤﻠﻴﻨﻴﺔ
Zimmer Distal Radius Plating System
#### Indications for Use:
The Zimmer Distal Radius Plating System is intended for use in surgical reduction, internal fixation, and stabilization of fractures and osteotomies of the distal radius. Examples of distal radius fractures include but are not limited to compression fractures, intra-articular and extraarticular fractures, displaced fractures in osteopenic bone, non-unions, and malunions.
Prescription Use _ X _ ' (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(Please do not write below this line - Continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
# Elizabeth L. Frank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.