← Product Code [KRD](/productcode/KRD) · K133237

# GEL-BEAD EMBOLIZATION SPHERES (K133237)

_Vascular Solutions, Inc. · KRD · Apr 25, 2014 · Cardiovascular · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133237

## Device Facts

- **Applicant:** Vascular Solutions, Inc.
- **Product Code:** [KRD](/productcode/KRD.md)
- **Decision Date:** Apr 25, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 870.3300
- **Device Class:** Class 2
- **Review Panel:** Cardiovascular
- **Attributes:** Therapeutic

## Indications for Use

Gel-Bead embolization spheres is intended for use in embolization of hypervascular tumors.

## Device Story

Gel-Bead embolization spheres are biodegradable gelatin spheres supplied in a 20 ml syringe with saline. Used by interventional radiologists or physicians in clinical settings to treat hypervascular tumors. Device is delivered via a catheter to the target vessel; spheres create mechanical obstruction to blood flow, confirmed by angiography. Occlusion results in infarction of the target tissue. Biodegradable nature allows for temporary or permanent vessel blockage depending on clinical requirements. Benefits include targeted tumor devascularization.

## Clinical Evidence

No human clinical data provided. Evidence consists of a GLP animal study (n=12 miniature swine) comparing Gel-Bead to Embosphere. Primary endpoints included successful arterial occlusion confirmed by angiography and histopathological assessment of target organs. Results showed consistent, reliable implantation, successful arterial occlusion, and well-demarcated foci of infarction in target organs, with no systemic abnormalities or clinically significant pathology findings. Biocompatibility testing (cytotoxicity, sensitization, irritation, acute systemic toxicity, hematology, genotoxicity) met all acceptance criteria.

## Technological Characteristics

Biodegradable gelatin spheres; sizes 100-300 µm, 300-500 µm, 500-700 µm, 700-1000 µm. Delivered via catheter. Sterilized by Gamma irradiation. Single-use. Mechanical obstruction principle of operation.

## Regulatory Identification

A vascular embolization device is an intravascular implant intended to control hemorrhaging due to aneurysms, certain types of tumors (e.g., nephroma, hepatoma, uterine fibroids), and arteriovenous malformations. This does not include cyanoacrylates and other embolic agents, which act by polymerization or precipitation. Embolization devices used in neurovascular applications are also not included in this classification, see § 882.5950 of this chapter.

## Special Controls

*Classification.* Class II (special controls.) The special control for this device is the FDA guidance document entitled “Class II Special Controls Guidance Document: Vascular and Neurovascular Embolization Devices.” For availability of this guidance document, see § 870.1(e).

## Predicate Devices

- Embosphere Microspheres ([K021397](/device/K021397.md))
- Gel-Block Embolization Pledgets ([K113266](/device/K113266.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

510(k) Summary [As required by 21 CFR 807.92] Date Prepared: April 25, 2014 510(k) Number: K133237

### Submitter's Name / Contact Person

Manufacturer Vascular Solutions, Inc. 6464 Sycamore Court North Minneapolis, MN 55369 USA Establishment Registration # 2134812 Contact Person Ellie Gillespie Sr. Regulatory Product Specialist Tel: 763-656-4300 Fax: 763-656-4253

#### General Information

| Trade Name          | Gel-Bead embolization spheres                                                                                                                                      |
|---------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Common / Usual Name | embolization spheres                                                                                                                                               |
| Classification Name | 870.3300, KRD - Device, vascular, for promoting embolization, Class I                                                                                              |
| Predicate Device(s) | K021397 – Embosphere Microspheres (Biosphere Medical, Inc./Merit<br>Medical Systems, Inc.)<br>K113266 - Gel-Block Embolization Pledgets (Vascular Solutions, Inc.) |

#### Device Description

The Gel-Bead embolization spheres (Gel-Bead) consists of biodegradable gelatin spheres pre-filled in a 20 ml syringe. The syringe contains 1 ml of spheres suspended in 5 ml of saline. Gel-Bead is offered in four size ranges: 100-300 µm, 300-500 µm, 500-700 µm and 700-1000 µm. The spheres are intended to be used with a delivery catheter with an inner diameter that is adequate for sphere delivery (not included). The finished product is sterilized by Gamma irradiation and is intended for single use only.

## Intended Use / Indications

Gel-Bead embolization spheres is intended for use in embolization of hypervascular tumors.

{1}------------------------------------------------

#### Technological Characteristics

11

Gel-Bead has the same intended use as both predicate devices but Embosphere and Gel-Block are also intended for use in arteriovenous malformations and Embosphere is also intended for use in symptomatic uterine fibroids. Gel-Bead is similar in design to the predicate devices as they are gelatincontaining embolization devices delivered through a catheter with the same principle of operation (create a mechanical obstruction to blood flow in the target vessel, as evidenced by disruption in the contrast flow rate). Gel-Bead size ranges are within the range of Embosphere Microspheres (Embosphere). Gel-Bead is biodegradable like Gel-Block embolization pledgets (Gel-Block). The only technological difference between the subject device and the predicate devices is the materials of construction.

#### Substantial Equivalence and Summary of Studies

The subject device design was evaluated through biocompatibility and animal tests to provide evidence of safe and effective use of Gel-Bead. Gel-Bead is substantially equivalent to the specified predicate devices based on comparisons of device functionality, technological characteristics, and indications for use. The subject device design has been verified through an animal study and the following biocompatibility tests performed as recommended by ISO 10993-1:

- · Cytotoxicity
	- o L929 MEM Elution
	- Agar Diffusion o
- · Sensitization
	- Kligman Maximization Sensitization Test (Direct) o
- · Irritation/Intracutaneous Reactivity
	- o Intracutaneous Injection (Direct)
- · Acute Systemic Toxicity
	- o Acute Systemic Injection Test
	- 0 Material Mediated Rabbit Pyrogen (Direct Exposure)
- · Hematology
	- o Hemolysis (Direct)
	- Hemolysis (Indirect) 0
	- Complement Activation (Direct) O
- · Genotoxicity
	- Bacterial Mutagenicity Test Ames Assay O
	- o Mouse Lymphoma Forward Mutation Assay
	- Rodent Bone Marrow Micronucleus Assay O

{2}------------------------------------------------

A GLP animal study was conducted on twelve mature miniature swine, with animals survived up to 12 weeks following implantation. Eight animals were implanted with Gel-Bead and four animals implanted with a control (Embosphere). Gel-Bead spheres were consistently and reliably implanted in the selected target arteries of appropriate size and not in additional non-target tissues or regions. All instances of Gel-Bead delivery resulted in successful arterial occlusion at the time of implant, as confirmed by angiography. There were similar well-demarcated foci of infarction, indicative of successful embolization, observed in the target organs of both test article animals. There were no clinically significant abnormalities identified in the clinical pathology blood results that negatively reflected on either the test or control devices. Examination of the tissues distant to the implantation sites did not identify any systemic abnormalities.

Verification, animal study, and biocompatibility test results met the specified acceptance criteria and did not raise new safety or performance issues. Therefore, Gel-Bead is substantially equivalent to the predicate devices.

{3}------------------------------------------------

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

Image /page/3/Picture/3 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name in a circular arrangement around a symbol. The symbol is a stylized representation of a human figure embracing a bird, which is meant to represent the department's mission of protecting the health of all Americans and providing essential human services.

## April 25, 2014

Vascular Solutions, Inc. Ellie Gillespie Senior Regulatory Product Specialist 6464 Sycamore Court North Minneapolis, MN 55369

K133237 Re:

> Trade/Device Name: Gel-Bead™ Embolization Spheres Regulation Number: 21 CFR 870.3300 Regulation Name: Vascular Embolization Device Regulatory Class: Class II Product Code: KRD Dated: March 18, 2014 Received: March 20, 2014

Dear Ms. Gillespie:

We have reviewed your Section 510(k) premarket notification of intent to market the device we nave rove and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to conimered prices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The I ou may, morely manso of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), II your device is blassified (oos aothols. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act ulal FDA has made a determination that your istered by other Federal agencies. You must comply of of all the Act's requirements, including, but not limited to: registration and listing (21 CFR

{4}------------------------------------------------

## Page 2 - Ellie Gillespie

Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical devicerelated adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

Kenneth J. Cavanaugh -S

for

Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

# Indications for Use

K133237 510(k) Number (if known): _

Device Name: Gel-Bead embolization spheres

Indications for Use:

Gel-Bead embolization spheres is intended for use in embolization of hypervascular tumors.

Prescription Use _ X (Part 21 CFR 801 Subpart D)

Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)

(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Image /page/5/Picture/8 description: The image shows the name "Kenneth J. Cavanaugh -S" in a bold, sans-serif font. The name is written in black ink. There is a logo in the background that is partially obscured by the text. The logo appears to be a stylized design with geometric shapes.

---

**Source:** [https://fda-staging.innolitics.com/device/K133237](https://fda-staging.innolitics.com/device/K133237)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
