K133180 · Erbe USA, Inc. · GEI · Dec 11, 2013 · General, Plastic Surgery
Device Facts
Record ID
K133180
Device Name
ERBE ESU MODEL VIO DV WITH ACCESSORIES
Applicant
Erbe USA, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Dec 11, 2013
Decision
SESE
Submission Type
Special
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ERBE ESU Model VIO dV with Accessories is intended to deliver High Frequency (HF) electrical current for the cutting and/or coagulation of tissue.
Device Story
The ERBE ESU Model VIO dV is a stand-alone electrosurgical generator that delivers high-frequency (HF) electrical current to hand-held instruments for cutting and coagulating tissue. The device features a 10-inch touchscreen interface for selecting modes (Auto Cut, Dry Cut, Swift Coag, Forced Coag, Bipolar Soft Coag) and effect levels, which determine voltage parameters. It includes four HF receptacles with instrument recognition and a CAN-bus port for remote control. Operated by physicians in surgical settings, the device provides audio and visual error monitoring and an AutoStop function for specific coagulation modes. The output current is delivered via monopolar or bipolar cables to compatible accessories. The device assists surgeons in achieving hemostasis and tissue separation, benefiting patients by providing controlled, flexible surgical energy delivery.
Clinical Evidence
Bench testing only. No clinical or animal performance testing was conducted.
Technological Characteristics
Stand-alone electrosurgical generator; 10" touchscreen interface; 4 HF receptacles (2 monopolar, 2 bipolar) with instrument recognition; CAN-bus connectivity; firewall function. Modes: Auto Cut, Dry Cut, Swift Coag, Forced Coag, Bipolar Soft Coag (with/without AutoStop). Non-sterile, reusable. Design verified/validated per internal quality systems and FDA recognized consensus standards.
Indications for Use
Indicated for the delivery of High Frequency (HF) electrical current for the cutting and/or coagulation of tissue in surgical applications. Intended for prescription use.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
ERBE ESU Model VIO 300 D with Accessories (K083452)
Submission Summary (Full Text)
{0}------------------------------------------------
# ERBE USA Incorporated Special 510(k) for ERBE ESU Model VIO dV with Accessories
Page 1 of 3 October 2013
| Owner/Operator: | ERBE USA Incorporated<br>2225 Northwest Parkway<br>Marietta, GA 30067 |
|---------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Trade (Proprietary) Name: | ERBE ESU Model VIO dV with Accessories |
| Common Name: | Electrosurgical Unit (ESU/Generator) |
| Classification Name, Number, and<br>Product Code: | Electrosurgical, Cutting & Coagulation &<br>Accessories; 21 CFR 878.4400; GEI |
| Legally Marketed Predicate Device: | ERBE ESU Model VIO 300 D with Accessories,<br>510(k) Number K083452<br>Note: VIO stands for Variable Cut and<br>Coagulation. |
# 510(k) SUMMARY [As Required by 21 CFR 807.92(c)]
Description of the Device:
The ERBE ESU Model VIO dV with Accessories is an ElectroSurgical Unit that generates High Frequency (HF) electrical current to cut and/or coagulate tissue. It is a stand-alone generator. The ESU has five clearly defined Cutting and Coagulation Modes with different electrical waveforms and electrical parameters, which are programmed with defined Effect levels. Each Effect level corresponds to a specific voltage. The Modes provide the physician flexibility in interventional applications. Thus the Unit may be used for a broad array of surgical applications.
The ESU user interface consists primarily of a touchscreen surrounded by a small number of physical controls, such as a power switch and connection points for the instruments and accessories with which the generator is compatible.
Various hand instruments and neutral electrodes from ERBE and different manufacturers may be attached to and operated by the generator. The standard Accessories for the ESU consist of reusable Footswitches as well as Monopolar and Bipolar Cables
Intended Use:
The ERBE ESU Model VIO dV with Accessories is intended to deliver High Frequency (HF) current for the cutting and/or coagulation of tissue.
{1}------------------------------------------------
K133180
# ERBE USA Incorporated Special 510(k) for ERBE ESU Model VIO dV with Accessories
Page 2 of 3 October 2013
# Summary of the Technological Characteristics:
## Similarities between modified and predicate devices:
#### Unit
The ERBE ESU Model VIO dV has the same intended use, the same basic technology, protective circuits, and uses the same basic accessories as the predicate ESU (ERBE Model VIO 300 D). Also, the Modes available in the ERBE ESU Model VIO dV ESU [i.e. Auto Cut, Dry Cut, Swift Coag, Forced Coag, and Bipolar Soft Coag with and without Auto Stop] are the same in the predicate. Both generators have user interface displays to select modes, power settings, etc. and have the AutoStop function. Also, both units have audio and visual error monitoring.
The subject ESU and the predicate device both are manufactured by ERBE Elektromedizin GmbH in Germany and are supplied as non-sterile and are reusable. The packaging is also the same for both devices with similar labeling (e.g. Outer Package Label, User Manual, etc.).
# Accessories:
The one pedal and two pedal footswitches are the same for both the subject and the predicate devices. The packaging is also the same with similar labeling.
#### Differences between modified and predicate devices:
The ERBE ESU Model VIO dV differs from the predicate ERBE ESU Model VIO 300 D as follows:
| Subject device | Predicate device |
|--------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------|
| - the unit has a square shape<br>(408 x 163 x 380 mm)<br>with rounded front panel | - the unit is of 'cake slice' shape<br>(425 x 211 x 462 mm)<br>with square front panel |
| - the ESU has a 10" Touchscreen color monitor display | - it has 6" button and LED color monitor display |
| - it has 4 HF receptacles in the front panel<br>(2 monopolar and 2 bipolar) with<br>instrument recognition on both the<br>monopolar and bipolar receptacles. | - it has 3 HF receptacles in the front panel<br>(monopolar, bipolar, multifunction) |
| - it has a receptacle for a CAN-bus for a remote controller | - it has no receptacle for a CAN-bus for a remote controller |
| - it has no receptacle for an APC unit | - it has a receptacle for an APC unit |
| - it has a firewall function | - it has no firewall function |
{2}------------------------------------------------
K133180
Page 3 of 3
#### ERBE USA Incorporated Special 510(k) for ERBE ESU Model VIO dV with Accessories
# October 2013
| - it has no programming capabilities<br>- it ha a recall function<br>- the modes that are present are the<br>principal ones to perform essential<br>cutting and coagulating activities.<br>Auto Cut, Dry Cut, Swift Coag, Forced<br>Coag., Soft Coag, Soft Coag with<br>AutoStop,,<br>No new modes are added. | - it has 99 program possibilities<br>- it has no recall function<br>- the predicate has a wide range of modes<br>such as Auto Cut, High Cut, Dry Cut, Dry<br>Cut °, Precise Cut, Soft Coag, Soft Coag<br>with AutoStop, Swift Coag, Swift Coag °,<br>Classic Coag, Forced Coag, Spray<br>Coag, Precise Coag, Twin Coag, Bipolar<br>Cut, Bipolar Cut +, Bipolar Precise Cut,<br>Endo Cut Q, Endo Cut I, Bipolar Soft<br>Coag +, Bipolar Forced Coag, BiClamp,<br>Bipolar Precise Coag, Precise APC,<br>Forced APC, Pulsed APC, Argon-<br>assisted High Cut, Argon-assisted AUTO<br>Cut, Argon-assisted Dry Cut, Argon-<br>assisted Dry Cut °, Argon-assisted Swift<br>Coag, Argon-assisted Swift Coag °,<br>Argon-assisted Soft Coag, Argon-<br>assisted Forced Coag and Argon-<br>assisted Twin Coag (Note: Only some of<br>these Modes are standard with the<br>predicate device.). |
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
The ERBE ESU Model VIO dV with Accessories has been verified or validated in design control.
ERBE Elektromedizin GmbH, in accordance with established procedures, directs and controls design activities. These procedures involve design and development planning, design input, design review, design verification/design output, design validation, design transfer, as well as design change control. The activities provide the oversight and forum for project approval, formal management/design review, evaluation, and final project approval. This work is documented within design review and the approvals are documented as part of the current change control system.
The ERBE ESU Model VIO dV with Accessories was tested to FDA's "Recognized Consensus Standards". Animal or clinical performance testing was not considered necessarv.
# Conclusion:
The ERBE ESU Model VIO dV with Accessories has the same intended use, principles of operation and technological characteristics as the predicate device in the previously cleared 510(k)s. The subject device has been verified or validated in design control. In conclusion, there are no issues with the subject device that would raise additional safety or efficacy issues, when compared to the predicate devices.
E 3
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows a logo for the Department of Health & Human Services USA. The logo features a stylized emblem with three curved shapes, resembling a person with outstretched arms. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the emblem.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
ERBE USA Incorporated Mr. John Tartal Ouality Assurance/Regulatory Affairs Director 2225 Northwest Parkway Marietta, Georgia 30067
December 11, 2013
Re: K133180
Trade/Device Name: ERBE ESU Model VIO dV with Accessories Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: Class II Product Code: GEI Dated: November 11, 2013 Received: November 12, 2013
Dear Mr. Tartal:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical . device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. John Tartal
··············································································································································································
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Indications for Use
K133180 510(k) Number (if known):
Device Name: ERBE ESU Model VIO dV with Accessories
Indications For Use:
ﺍﻟﻤﺴﻴﺴﻴﺔ ﺍﻟﺘﺤﺘﻮﻱ ﺍﻟﻤﻮﺗﻮﻓﻴﺎ
The ERBE ESU Model VIO dV with Accessories is intended to deliver High Frequency (HF) electrical current for the cutting and/or coagulation of tissue.
Prescription Use 2 (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Image /page/5/Figure/10 description: The image shows a document sign-off for the Division of Surgical Devices. The 510(k) number is K133180. The document is signed by Long H. Chen-A on December 10, 2013. The document is for BSA.
001_4 Indications for Use Statement
004 - 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.