K133158 · Beijing Toplaser Technology Co., Ltd. · GEX · Feb 28, 2014 · General, Plastic Surgery
Device Facts
Record ID
K133158
Device Name
ND: YAG LASER SYSTEM
Applicant
Beijing Toplaser Technology Co., Ltd.
Product Code
GEX · General, Plastic Surgery
Decision Date
Feb 28, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4810
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
Tattoo Removal Treatment of Vascular Lesions Treatment of Pigmented Lesions, Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology
Device Story
Qmetrx 1000 (Victory 11) is a Q-switched Nd:YAG laser system. Inputs: electrical power; user-selected parameters (wavelength, energy, spot size, fluence). Operation: flashlamp-pumped Nd:YAG crystal generates 1064 nm beam; frequency-doubling non-linear crystal generates 532 nm beam. Red diode aiming beam reflected via beamsplitter. Combined beams delivered via articulated arm to focusing handpiece. Used in clinical dermatology settings by physicians. Output: therapeutic laser energy for tissue interaction. Clinical benefit: removal of tattoos/lesions and soft tissue surgery. Internal microprocessor controls wavelength selection, power, and cooling.
Clinical Evidence
Bench testing only. No clinical data required. Device validated against design specifications and predicate device performance standards.
Technological Characteristics
Q-switched Nd:YAG laser; 1064 nm and 532 nm (KTP) wavelengths. Flashlamp-pumped. Articulated arm delivery with focusing handpiece. Internal microprocessor control. Standards: IEC 60825-1, IEC 60601-2-22, IEC 60601-1, IEC 60601-1-2, UL 60601-1.
Indications for Use
Indicated for tattoo removal (dark ink at 1064nm; red, sky blue, green at 532nm), treatment of vascular lesions (port wine birthmarks, telangiectasias, spider/cherry angioma, spider nevi at 532nm), treatment of pigmented lesions (Nevus of Ota at 1064nm; café-au-lait, solar/senile lentigines, Becker's nevi, freckles, Nevus spilus at 532nm), and incision, excision, ablation, and vaporization of soft tissue in general dermatology.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
{0}------------------------------------------------
## FEB 2 8 2014
### 510(k) Summary
# Nd: YAG Laser System Beijing Toplaser Technology Co., Ltd. (As required by 21 CFR 807.92) K Number: K133158
1. Date Prepared: Sep. 22, 2013
2. Sponsor Information Beijing Toplaser Technology Co., Ltd. East 30 Floor, Building M7, No.1 Jiuxianqiao East Road, Chaoyang District, Beijing 100015, China Contract Person: Zhang Xiaosong, General Manager Phone: +86-10-64354759 Fax: +86-10-64356591
3. Proposed Device Information Device Common or Usual Name: Nd: YAG Laser Device Trade or Proprietary Name: Qmetrx 1000 (Victory 11) Nd: YAG Laser System Classification Name: Laser instrument, Surgical, Powered Regulation Number: 21 CFR 878.481 0 Product Code: GEX Panel: General and Plastic Surgery Model: QMetrx1000 (Victory-11)
4. Predicate Device Medlite C6 Q-Switched Nd: YAG Laser (K014234) Medlite™ C3 Q-Switched Nd: YAG Laser (K011677) Manufactured by Continuum Electro-Optics, Inc.
#### 5. Device Description
The Nd: YAG Laser System is based on the Q-Switched Nd:YAG (1064 nm) and frequency doubled KTP Nd:YAG (532 nm) laser technology. There is one optical cavity containing the Nd:YAG crystal. The frequency doubled KTP
{1}------------------------------------------------
Nd:YAG wavelength is achieved by directing the Nd:YAG laser beam through a frequency doubling non-linear crystal. The Nd:YAG laser is activated by means of the use of flashlamp. A red diode aiming beam is reflected onto a coaxial beam path using a beamsplitter assembly. The combined therapeutic and aiming beams are guided by articulated arm to a focusing handpiece.
Both laser wavelengths share a common power supply, control system, and cooling system. The internal microprocessor can be directed to select either the Nd:YAG or the KTP Nd:YAG laser wavelength. The physician is able to select the desired wavelength and the related output energy ,spot size ,fluence via control panel.
6. Intended use and Indications of Use
- · Tattoo Removal
- · Treatment of Vascular Lesions
- · Treatment of Pigmented Lesions,
• Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology
| 1064nm | 532nm |
|--------------------------------|------------------------------------------------------------------------------------------------|
| Tattoo Removal | Tattoo Removal |
| Dark ink: blue and black | Light ink: red Light ink: sky blue and green |
| Treatment of Pigmented Lesions | Treatment of Vascular Lesions |
| Nevus of ota | Port wine birthmarks Telangiectasias Spider angioma Cherry angioma Spider nevi |
| | Treatment of Pigmented Lesions |
| | Café-au-lait birthmarks Solar lentiginos Senile lentiginos Becker's nevi Freckles Nevus spilus |
#### 7. Substantial Equivalence
The product Nd:YAG Laser System, model QMetrx1000(Victory-11) shares the similar indications for use, design features, functional features, same safety compliance. Therefore the product Nd:YAG Laser System, model
{2}------------------------------------------------
QMetrx1000(Victory-11) is substantially equivalent to its predicate device cited above without raising new safety and/or effectiveness issues.
#### 8. Testing
The Nd: YAG Laser System is designed, tested and will be manufactured in accordance with both mandatory and voluntary standards, including:
IEC 60825-1: Safety of laser products - Part 1: Equipment classification, . requirements and user's quide.
IEC 60601-2-22: Medical Electrical Equipment - Part 2: Particular . requirements for the safety of diagnostic and therapeutic laser equipment.
IEC 60601-1: Medical Electrical Equipment - Part 1: General requirements . for safety.
IEC 60601-1-2: Medical Electrical Equipment -Part 1: General requirements . for safety-2, Collateral Standard: Electromagnetic compatibility - Requirements and tests.
UL 60601-1:2003 R6.03 .
Non-Clinical Conclusion:
Laboratory testing was conducted to validate and verify' that the proposed device, QMetrx1000(Victory-11) Nd: YAG Laser System met all design specifications and was substantially equivalent to the predicate device. No Clinical Information is required.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles an abstract human figure with three flowing lines representing the body and arms.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 28, 2014
Beijing Toplaser Techonology Company, Ltd. Mr. Zhang Xiaosong General Manager East 3rd Floor, Building M7, No. 1 Jiuxianqiao East Road Chaoyang District, Beijing 100015 CHINA
Re: K133158
Trade/Device Name: Ometrx 1000 (Victory 11) Nd: YAG Laser System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: GEX Dated: November 28, 2013 Received: December 2, 2013
Dear Mr. Xiaosong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{4}------------------------------------------------
Page 2 - Mr. Zhang Xiaosong
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Felipe Aguel
for
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# 4. Indications for Use Statement
### Indications for Use
510(k) Number (if known): K133158
Device Name: Qmetrx 1000 (Victory 11) Nd: YAG Laser System Indications For Use:
- · Tattoo Removal
- · Treatment of Vascular Lesions
- · Treatment of Pigmented Lesions,
- · Incision, Excision, Ablation, Vaporization of Soft Tissue for General Dermatology
| 1064nm | 532nm |
|--------------------------------|---------------------------------|
| Tattoo Removal | Tattoo Removal |
| ● Dark ink: blue and black | ● Light ink: red |
| | ● Light ink: sky blue and green |
| Treatment of Pigmented Lesions | Treatment of Vascular Lesions |
| ● Nevus of ota | ● Port wine birthmarks |
| | ● Telangiectasias |
| | ● Spider angioma |
| | ● Cherry angioma |
| | ● Spider nevi |
| | Treatment of Pigmented Lesions |
| | ● Café-au-lait birthmarks |
| | ● Solar lentiginos |
| | ● Senile lentiginos |
| | ● Becker's nevi |
| | ● Freckles |
| | ● Nevus spilus |
AND/OR Over-The-Counter Use __________________________________________________________________________________________________________________________________________________ Prescription Use YES (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE) Felipe Aguel Date: 2014.02.28
14:20:51 -05'00'
(Division Sign-Off) for BSA Division of Surgical Devices 510(k) Number ________________________________________________________________________________________________________________________________________________________________ Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.