← Product Code [FBK](/productcode/FBK) · K133148

# BIOMARC COAXIAL NEEDLE (K133148)

_Carbon Medical Technologies, Inc. · FBK · May 13, 2014 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133148

## Device Facts

- **Applicant:** Carbon Medical Technologies, Inc.
- **Product Code:** [FBK](/productcode/FBK.md)
- **Decision Date:** May 13, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 876.1500
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology

## Indications for Use

The BiomarC Coaxial Needle allows delivery of fiducial markers into soft tissue.

## Device Story

BiomarC Coaxial Needle is a sterile, single-use, disposable device designed to facilitate the delivery of fiducial markers into soft tissue. The device consists of an outer cannula with a female luer lock hub, an inner stylet with a male luer lock hub, and a flexible slip ring depth stop. It is intended for use in conjunction with BiomarC Fiducial Markers, with the outer cannula sized one gauge larger than the corresponding marker needle. The device functions by creating a channel through soft tissue to allow for the placement of the markers. It is a manual, mechanical instrument used by clinicians in a medical setting.

## Clinical Evidence

No clinical data. Substantial equivalence is based on technological characteristics, design, and history of use of similar devices.

## Technological Characteristics

Materials: Stainless steel (type 304). Design: Coaxial needle with outer cannula, inner stylet, and flexible slip ring depth stop. Tip: Trocar. Sterilization: Ethylene Oxide (EO) to SAL 1x10^-6. Energy source: Manual/Mechanical. Connectivity: None.

## Regulatory Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

## Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls:

1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following:
   1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures.
   2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use.
   3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use.
2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes.
3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches
4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error.
5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program.
(6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms.
(7) Labeling must include:
(i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified;
(ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA;
(iii) Identification of compatible devices;
(iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification;
(v) Reprocessing instructions for reusable components;
(vi) A shelf life for any sterile components;
(vii) A description of the device-specific use training program;
(viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and
(ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance.
(8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include:
(i) Device motion accuracy and repeatability;
(ii) System testing;
(iii) Instrument reliability;
(iv) Crosstalk;
(v) Table motion control;
(vi) Thermal effects on tissue;
(vii) User-device interface performance;
(viii) Workspace access testing; and
(ix) Performance testing with compatible devices.
(9) Software verification, validation, and hazard analysis must be performed.
(10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed.
(11) Performance data must demonstrate the sterility of all patient-contacting device components.
(12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life.
(13) Performance data must validate the reprocessing instructions for the reusable components of the device.
(14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible.
(15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic.
(16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information:
(i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and
(ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

## Predicate Devices

- Carbon Medical Technologies Endoscopic Injection Needle ([K042615](/device/K042615.md))
- Coaxial Biopsy Needle ([K936194](/device/K936194.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Page 1 of 2

510(k) Summary 5

### 5. 510(k) Summary

5 -

ं . . .

Andrew Adams Director - Regulatory Affairs & Quality Assurance Carbon Medical Technologies, Inc. 1290 Hammond Road Saint Paul, MN 55110

651-653-8512 Phone: Fax: 651-407-1975

Submitted: September 27, 2013

#### Device Name

| Trade Name:          | BiomarC Coaxial Needle                            |
|----------------------|---------------------------------------------------|
| Classification Name: | Endoscopic Injection Needle, 21 CFR 876.1500, FBK |
|                      | Implantable Clip, 21 CFR 878.4300, NEU            |
| Common/Usual Name:   | Coaxial Needle                                    |

### Predicate Devices

Carbon Medical Technologies Endoscopic Injection Needle (K042615), by Carbon Medical Technologies, Inc. Coaxial Biopsy Needle (K936194), by Bard Gynecology & Radiology

#### Indication for Use

The BiomarC Coaxial Needle allows delivery of fiducial markers into soft tissue.

#### Device Description

The BiomarC Coaxial Needle is a sterile, pyrogen free, single patient use, disposable coaxial needle that consists of an outer cannula with an attached female luer lock hub, an inner stylet with an attached male luer lock hub, and a flexible slip ring style depth stop.

The BiomarC Coaxial Needle is designed for use with BiomarC Fiducial Markers. The outer cannula is only one gauge size larger than the appropriate BiomarC Fiducial Marker needle, e.g., 18g BiomarC Coaxial Needle for a 19g BiomarC Fiducial Marker needle.

> BiomarC Coaxial Needle Premarket Notification 510(k) 5-1 (May 8, 2014)

# MAY 1 3 2014

{1}------------------------------------------------

## Technological Characteristics and Performance

ﺎ

The technological characteristics and performance of the BiomarC Coaxial Needle are substantially equivalent to the predicate devices (Table 5).

| Characteristic                     | Proposed Device                                                                                                                                     | Predicate Device                                                                                                                                                                                              | Predicate Device                                                                                                                                                                                                            |
|------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Trade name                         | BiomarC Coaxial Needle                                                                                                                              | Carbon Medical<br>Technologies Endoscopic<br>Injection Needle                                                                                                                                                 | Coaxial Biopsy Needle                                                                                                                                                                                                       |
| 510(k) number                      |                                                                                                                                                     | K042615                                                                                                                                                                                                       | K936194                                                                                                                                                                                                                     |
| 510(k) holder                      | Carbon Medical<br>Technologies, Inc.                                                                                                                | Carbon Medical<br>Technologies, Inc.                                                                                                                                                                          | Bard Gynecology and<br>Radiology                                                                                                                                                                                            |
| Indications for<br>use             | The BiomarC Coaxial<br>Needle allows delivery of<br>fiducial markers into soft<br>tissue.                                                           | The Carbon Medical<br>Technologies Injection<br>Needle is an accessory for<br>currently marketed<br>endoscopes to allow<br>delivery of injectable<br>materials into tissue during<br>an endoscopic procedure. | The coaxial biopsy needle<br>guide is intended for use as<br>a guiding needle in<br>obtaining core, biopsy<br>samples from soft tissue<br>such as liver, kidney,<br>spleen, lymph nodes and<br>various soft tissue lesions. |
| Use                                | Single use                                                                                                                                          | Single use                                                                                                                                                                                                    | Single use                                                                                                                                                                                                                  |
| Sterility                          | Sterilized by EO with an<br>SAL of 1x10-6                                                                                                           | Sterilized by EO with an<br>SAL of 1x10-6                                                                                                                                                                     | Sterilized by EO                                                                                                                                                                                                            |
| Pyrogens                           | Pyrogen free                                                                                                                                        | Pyrogen free                                                                                                                                                                                                  | Pyrogen free                                                                                                                                                                                                                |
| Cannula/stylet<br>rod material     | Stainless steel, type 304                                                                                                                           | Stainless steel                                                                                                                                                                                               | Stainless steel                                                                                                                                                                                                             |
| Tip style                          | Trocar tip                                                                                                                                          | Spinal tip                                                                                                                                                                                                    | Trocar or blunt tip                                                                                                                                                                                                         |
| Device body<br>contact<br>category | External communicating<br>device, tissue, limited                                                                                                   | External communicating<br>device, tissue, limited                                                                                                                                                             | External communicating<br>device, tissue, limited                                                                                                                                                                           |
| Safety                             | Safety has been established<br>for the identical nature and<br>duration of patient contact<br>through a history of use in<br>other similar devices. | Safety has been established<br>for the identical nature and<br>duration of patient contact<br>through a history of use in<br>other identical and similar<br>devices.                                          | Unknown                                                                                                                                                                                                                     |
|                                    | Able to create a channel<br>through soft tissue                                                                                                     | Able to create a channel<br>through soft tissue                                                                                                                                                               | Able to create a channel<br>through soft tissue                                                                                                                                                                             |
| Performance                        | Able to allow delivery of<br>fiducial markers through<br>channel created                                                                            | Able to allow delivery of<br>injectable materials through<br>channel created                                                                                                                                  | Able to guide another<br>device through channel<br>created                                                                                                                                                                  |

BiomarC Coaxial Needle Premarket Notification 510(k) 5-2 (May 8, 2014)

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized depiction of an eagle or bird-like figure with three curved lines representing its body and wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird symbol.

#### DEPARTMENT OF HEALTH & HUMAN SERVICES

Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

May 13, 2014

Carbon Medical Technologies Incorporated Mr. Andrew J. Adams Director, Regulatory Affairs & Quality Assurance 1290 Hammond Road Saint Paul, Minnesota 55110

Re: K133148

Trade/Device Name: BiomarC Coaxial Needle Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: Class II Product Code: FBK, NEU Dated: March 6, 2014 Received: March 7, 2014

Dear Mr. Adams:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set

{3}------------------------------------------------

Page 2 - Mr. Andrew J. Adams

forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address

http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.

Sincerely yours,

# David Krause -S

- Binita S. Ashar, M.D., M.B.A., F.A.C.S. for . Acting Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure

{4}------------------------------------------------

4. Indications for Use Statement

510(k) Number: K133148

Device Name: BiomarC Coaxial Needle

Indications for Use:

The BiomarC Coaxial Needle allows delivery of fiducial markers into soft tissue.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Joshua C. Nipper -S

Page 1 of _

BiomarC Coaxial Needle Premarket Notification 510(k) 4-1 (May 8, 2014)

---

**Source:** [https://fda-staging.innolitics.com/device/K133148](https://fda-staging.innolitics.com/device/K133148)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
