K133134 · FUJIFILM Sonosite, Inc. · IYN · Nov 4, 2013 · Radiology
Device Facts
Record ID
K133134
Device Name
X-PORTE ULTRASOUND SYSTEM
Applicant
FUJIFILM Sonosite, Inc.
Product Code
IYN · Radiology
Decision Date
Nov 4, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1550
Device Class
Class 2
Attributes
Pediatric, 3rd-Party Reviewed
Indications for Use
The FUJIFILM SonoSite X-Porte Ultrasound System is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include: Ophthalmic Fetal - OB/GYN Abdominal Intra-operative (Abdominal organs and vascular) Pediatric Small Organ (breast, thyroid, testicles, prostate) Trans-vaginal Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Peripheral vessel
Device Story
Mobile, full-featured ultrasound system; acquires/displays high-resolution, real-time ultrasound data. Configurable digital electronic design; interface controlled via backlit touchscreen. Operates in stand-based (AC/battery) or desktop (AC) configurations. Used by physicians in clinical settings. Inputs: ultrasound transducer signals; 3-lead ECG. Outputs: B-mode, M-mode, Color Doppler, PW/CW Doppler, Tissue Doppler Imaging (TDI), and image compounding (SonoMB/MBe). Healthcare providers use visual output for diagnostic evaluation and procedural guidance (needle/catheter placement). Benefits include real-time visualization of anatomy and fluid flow to support clinical decision-making.
Clinical Evidence
Bench testing only. No clinical studies required for substantial equivalence determination. Device evaluated for electrical, thermal, mechanical, and EMC safety; cleaning/disinfection; biocompatibility; and acoustic output per NEMA UD 2/UD 3 standards.
Technological Characteristics
Digital beamformer (128/128 channels using Synthetic Aperture). Transducer frequency 1.0-15.0 MHz. Materials: ABS, Polysulfone UDEL P1700, Polyurethane, PVC, Silicone Rubber. Connectivity: 6 USB 2.0 ports, Wireless 802.11 (a/b/g/n). Interface: 12.1" capacitive touchscreen, 19" LED LCD external monitor. Power: Battery or AC. Software: DICOM 3.0 compliant.
Indications for Use
Indicated for ophthalmic, fetal, OB/GYN, abdominal, intra-operative (abdominal/vascular), pediatric, small organ (breast, thyroid, testicle, prostate), trans-vaginal, musculoskeletal (conventional/superficial), adult/pediatric cardiac, and peripheral vessel imaging. Includes needle/catheter placement guidance and peripheral nerve block procedures.
Regulatory Classification
Identification
An ultrasonic pulsed doppler imaging system is a device that combines the features of continuous wave doppler-effect technology with pulsed-echo effect technology and is intended to determine stationary body tissue characteristics, such as depth or location of tissue interfaces or dynamic tissue characteristics such as velocity of blood or tissue motion. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{0}------------------------------------------------
K133134
Page 1 of 5
# 510(K) Summary
This summary of safety and effectiveness is provided as part of this Premarket Notification in compliance with 21 CFR, Part 807, Subpart E, Section 807.92.
## 1) Submitter's name, address, telephone number, contact person:
FUJIFILM SonoSite, Inc. 21919 30th Drive SE Bothell, WA 98021-3904
Corresponding Official:
E-mail: Telephone: Facsimile: Date prepared: Scott E. Paulson Sr. Manager, Regulatory Affairs Scott.Paulson@sonosite.com (425) 951-6926 (425) 491-8356 September 9, 2013
**NOV 04 2013**
2) Name of the device, including the trade or proprietary name if applicable, the common or usual name, and the classification name, if known:
## Common/ Usual Name
Diagnostic Ultrasound System with Accessories
Proprietary Name
SonoSite X-Porte™ Ultrasound System (subject to change)
## Classification Names
| Name | FR Number | Product Code |
|------------------------------------------|-----------|--------------|
| Ultrasonic Pulsed Doppler Imaging System | 892.1550 | 90-IYN |
| Ultrasonic Pulsed Echo Imaging System | 892.1560 | 90-IYO |
| Diagnostic Ultrasound Transducer | 892.1570 | 90-ITX |
## 3) Identification of the predicate or legally marketed device:
SonoSite Edge Ultrasound System K113156
## 4) Device Description:
The SonoSite X-Porte Ultrasound System is a highly mobile, full featured, general purposic ultrasound system used to acquire and display high-resolution, real-time ultrasound data through multiple imaging modes. X-Porte is a custom fabricated digital electronic design that readly lends itself to be configured for specific ultrasound imaging applications through different system feature selections. The system interface can be customized for the user and controlled using a backlit touchscreen much like what is used in consumer tablet products. X-Porte can be operated in two different configurations, standbased with AC power or battery, and desktop-based with AC power only. In desktop configuration the ultrasound engine can be removed from the stand and used by itself with a single transducer and external monitor.
{1}------------------------------------------------
## 5) Intended Use:
The FUJIFILM SonoSite X-Porte Ultrasound System is a general purpose ultrasound system intended for use by a qualified physician for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
> Ophthalmic Fetal - OB/GYN Abdominal Intra-operative (Abdominal organs and vascular) Pediatric Small Organ (breast, thyroid, testicles, prostate) Trans-vaginal Musculo-skel. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Peripheral vessel
## 6) Technological Characteristics:
SonoSite X-Porte and Edge Ultrasound Systems are both Track 3 devices that employ the same fundamental scientific technology. A comparison table is provided below.
| Feature | SonoSite X-Porte Ultrasound<br>System<br>(this submission) | SonoSite Edge Ultrasound System<br>(K113156) |
|-------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Diagnostic ultrasound imaging or fluid flow<br>analysis of the human body | Diagnostic ultrasound imaging or fluid flow<br>analysis of the human body |
| Indications for Use | Opthalmic<br>Fetal - OB/GYN<br>Abdominal<br>Intraoperative (abdominal organs and<br>vascular)<br>Pediatric<br>Small Organ (breast, thyroid, testicle, prostate)<br>Trans-Vaginal<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Cardiac Adult<br>Cardiac Pediatric | Opthalmic<br>Fetal - OB/GYN<br>Abdominal<br>Intraoperative (abdominal organs and<br>vascular)<br>Intra-operative (Neuro.)<br>Laparoscopic<br>Pediatric<br>Small Organ (breast, thyroid, testicle, prostate)<br>Neonatal Cephalic<br>Adult Cephalic<br>Trans-Rectal<br>Trans-Vaginal<br>Musculo-skeletal (Conventional)<br>Musculo-skeletal (Superficial)<br>Cardiac Adult<br>Cardiac Pediatric<br>Trans-esophageal (cardiac) |
| | Peripheral Vessel<br>Needle guidance | Peripheral Vessel<br>Needle guidance |
| Transducer Types | Linear Array<br>Curved Linear Array<br>Intracavitary<br>Phased Array | Linear Array<br>Curved Linear Array<br>Intracavitary<br>Phased Array<br>Static Probes<br>Trans-esophageal<br>Wobbler Probes |
| Transducer<br>Frequency | 1.0 - 15.0 MHz | 1.0 - 15.0 MHz |
{2}------------------------------------------------
| Feature | SonoSite X-Porte Ultrasound System<br>(this submission) | SonoSite Edge Ultrasound System<br>(K113156) | | | |
|------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--------------|--------------|
| Acoustic Output Display & FDA Limits | Display Feature for Higher Outputs<br>MI Output Display<br>TI Output Display | Display Feature for Higher Outputs<br>MI Output Display<br>TI Output Display | | | |
| Modes of Operation | B-mode Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom<br>Combination Modes<br>Pulsed Wave (PW) Doppler<br>Continuous Wave (CW) Doppler<br>SonoHD2 Noise Reduction<br>SonoMB/MBe Image Compounding<br>Steered CW Doppler<br>Velocity Color Doppler<br>Tissue Doppler Imaging (TDI) | B-mode Grayscale Imaging<br>3D/4D Grayscale Imaging<br>Tissue Harmonic Imaging<br>M-mode<br>Anatomical M-Mode<br>Color M-Mode<br>Color Power Doppler<br>Zoom<br>Combination Modes<br>Pulsed Wave (PW) Doppler<br>Continuous Wave (CW) Doppler<br>SonoHD2 Noise Reduction<br>SonoMB/MBe Image Compounding<br>Steered CW Doppler<br>Velocity Color Doppler<br>Tissue Doppler Imaging (TDI) | | | |
| PW Doppler | Available | Available | | | |
| CW Doppler | Available | Available | | | |
| Velocity Color Doppler | Available | Available | | | |
| Elastography (Strain), and Strain Rate Imaging | Not available | Available | | | |
| ECG Feature | 3-lead ECG input | 3-lead ECG input | | | |
| DICOM | DICOM 3.0 | DICOM 3.0 | | | |
| IMT Measurement | Not available | Available | | | |
| Measurement and Calculations | Obstetrical, cardiac, volume, M-mode, PW and CW Doppler measurement and calculation packages | Obstetrical, cardiac, volume, M-mode, PW and CW Doppler measurement and calculation packages | | | |
| #Transmit Channels | 128 digital channels | 128 digital channels | | | |
| #Receive Channels | 64 digital channels<br>(128 digital channels using Synthetic Aperture) | 64 digital channels<br>(128 digital channels using Synthetic Aperture) | | | |
| Patient Contact | Patient Contact<br>Materials | Transducers: | | Transducers: | Manufacturer |
| | Acrylonitrile-butadien-styrene (ABS)<br>--<br>Dow Medical Adhesive, Type A<br>Epoxy paste adhesive.<br>Polysulfone UDEL P1700 | Acrylonitrile-butadien-styrene (ABS)<br>Cycoloy<br>Dow Medical Adhesive, Type A<br>Epoxy paste adhesive.<br>Polysulfone UDEL P1700 | | | |
:
.
·
{3}------------------------------------------------
| Feature | SonoSite X-Porte Ultrasound<br>System<br>(this submission) | SonoSite Edge Ultrasound System<br>(K113156) |
|---------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Polyurethane<br>Poly-Vinyl-Chloride (PVC)<br>Silicone Rubber<br>Urethane | Polyurethane<br>Poly-Vinyl-Chloride (PVC)<br>Silicone Rubber<br>Urethane |
| | Needle Guides:<br>Acetal copolymer<br>Acrylonitrile-butadien-styrene (ABS) | Needle Guides:<br>Acetal copolymer<br>Acrylonitrile-butadien-styrene (ABS) |
| System<br>Characteristics | X-Porte (stand configuration):<br>Beamformer 128/128 using SA (configurable)<br>12.1" Capacitive touch screen interface<br>19" LED LCD HD external monitor<br>256 gray shades on LED LCD | Edge:<br>Beamformer 128/128 using SA (configurable)<br>Hand held display and control<br>Single 12.1" Liquid Crystal Display (LCD)<br>256 gray shades on LCD |
| | 6 USB 2.0 ports | 2 USB ports |
| | Dimensions: 15.6" (W) x 13.2" (L) x 3.1" (H) | Dimensions: 12.9"(W) x 12.4 (L) x 2.5"(H) |
| | Weight (fully configured with stand, all options<br>and 3 transducers connected): 148.1 lbs | Weight: 8.5 lbs |
| | Stand base Dimensions: 26.4" L x 21.2" W<br>Stand Height (max): 64" (monitor up)<br>Stand Height (min): 42.2" (monitor down) | |
| | System operates via battery or AC power<br>mounted on the stand. | System operates via battery or AC power |
| | Battery life: 1 hour operational - 3 days idle | Battery operated (1.5 - 4 hour operation per<br>charge) |
| | Input: 100 - 240 VAC, 50/60 Hz<br>Output 1: 24VDC output, 275 W max<br>Output 2: 100-240VAC, 50-60 Hz (AC Printer) | 100 - 240V options, 50/60 Hz, 15VDC output |
| | Wireless 802.11 (a/b/g/n) support for image<br>transfer | Wireless 802.11 (a\b\g) support for image<br>transfer |
| | X-Porte (desktop configuration): | |
| | Same software features/capabilities as the<br>stand configuration. Does not have the stand,<br>touch panel interface, DVR, and mobile power<br>unit. | |
| | AC power<br>100 - 240V options, 50/60 Hz | |
| 510(k) Track | Track 3 | Track 3 |
:
{4}------------------------------------------------
K133134
Page 5 of 5
## 7) Determination of Substantial Equivalence:
## Summary of Non-Clinical Tests:
The X-Porte Ultrasound System has been evaluated for electrical, thermal, mechanical and EMC safety. Additionally, cleaning/disinfection, biocompatibility, and acoustic output have been evaluated, and the device has been found to conform to applicable mandatory medical device safety standards. Assurance of quality was established by employing the following elements of product development: Design Phase Reviews, Risk Assessment, Requirements Development, System and Software Verification, Hardware Verification, Safety Compliance Verification, Clinical Validation. All patient contact materials are biocompatible and are materials that are already used in the predicate device. Reports for these elements of product development are referenced in Attachment 6.
The X-Porte Ultrasound System is designed to comply with the following voluntary standards.
| Reference No. | Title | |
|--------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| AAMI/ANSI/ISO<br>10993-1 | Biological evaluation of medical devices -- Part 1: Evaluation and testing within a<br>risk management process | |
| IEC 60601-1 | Medical electrical equipment - Part 1: General requirements for basic safety and<br>essential performance | |
| IEC 60601-1-2 | Medical electrical equipment - Part 1-2: General requirements for basic safety<br>and essential performance - Collateral standard: Electromagnetic compatibility -<br>Requirements and tests | |
| IEC 60601-2-37 | Particular Requirements for the basic safety and essential performance of<br>ultrasonic medical diagnostic and monitoring equipment | |
| ISO 14971 | Medical devices - Application of risk management to medical devices | |
| NEMA UD 2-2004 | Acoustic Output Measurement Standard for Diagnostic Ultrasound Equipment | |
| NEMA UD 3-2004 | Standard for Real-Time Display of Thermal and Mechanical Acoustic Output<br>Indices on Diagnostic Ultrasound Equipment, American Institute of Ultrasound in<br>Medicine | |
| NEMA PS 3.15 | NEMA Ps 3.15:2011, Digital Imaging and Communications in Medicine<br>(DICOM), Part 15: Security and System Management Profiles | |
## Summary of Clinical Tests:
The SonoSite X-Porte Ultrasound System and transducers, subject of this submission, did not reguire clinical studies to support the determination of substantial equivalence.
## 8) Conclusion:
Intended uses and other key features are consistent with traditional clinical practice and FDA quidance. The X-Porte device and predicate both conform to applicable electromedical device safety standards with compliance verified through independent evaluation. The X-Porte device and predicate both meet FDA requirements for Track 3 devices, share indications for use, have biosafety equivalence and are manufactured using the same ISO 13485 quality system. FUJIFILM SonoSite, Inc. believes that the X-Porte Ultrasound System is substantially equivalent with regard to safety and effectiveness to the predicate device(s).
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows a logo with a stylized bird-like symbol consisting of three curved lines, suggesting movement or flight. The symbol is encircled by text, which appears to be part of the organization's name or a related phrase. The logo has a simple, clean design, often used by government agencies or organizations to convey a sense of authority and professionalism.
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-Gold Silver Spring, MI) 20993-0002
November 4, 2013
FUJIFILM SONOSITE, INC. C/O MARK JOB RESPONSIBLE THIRD PARTY OFFICIAL REGULATORY TECHNOLOGY SERVICES LLC 1394 25TH STREET NW BUFFALOMN 55313
Re: K133134
Trade/Device Name: SonoSite X-Porte™ Ultrasound System Regulation Number: 21 CFR 892.1550 Regulation Name: Ultrasonic pulsed doppler imaging system Regulatory Class: II Product Code: IYN, IYO. ITX Dated: October 22, 2013 Received: October 23, 2013
Dear Mr. Job:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract fiability warranties. We remind you, however, that device labeling must be truthful and not misleading.
This determination of substantial equivalence applies to the following transducers intended for use with the SonoSite X-Porte™ Ultrasound System, as described in your premarket notification:
Transducer Model Number
C60xp/5-2 MHz Transducer ICTxp/9-5 MHz Transducer 1.38xp/10-5 MHz Transducer FFFL50xp/15-6 MHz Transducer 1.25xp/13-6 MHz Transducer P21xp/5-1 MHz Transducer
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2-Mr. Job
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that vour device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiution control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21. CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrvidefault.htm. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometries/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely sours.
lor
Janine Morris Director Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
51D(k) Number (if known) K133134
#### Device Name
FUJIFILM SonoSite X-Porte Ultrasound System
#### Indications for Use (Describe)
The FUJIFILM SonoSite X-Porte Ultrasound System is a general purpose ultrasound system intended for use iclan for evaluation by ultrasound imaging or fluid flow analysis of the human body. Specific clinical applications and exam types include:
Ophthalmic Fetal - OB/GYN Abdominal Intra-operative (Abdominal organs and vascular) Pediatric Small Organ (breast, thyroid, testicles, prostate) Trans-vaginal Musculo-skei. (Convent.) Musculo-skel. (Superfic.) Cardiac Adult Cardiac Pediatric Peripheral vessel
Type of Use (Select one or both, as applicable)
> Prescription Use (Part 21 CFR 801 Subpart D)
O Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Smh.7)
FORM FDA 3881 (9/13)
**CRC Publishing Services (201) 344-7440** E
{8}------------------------------------------------
| Table 1.3.1: Diagnostic Ultrasound Indications for Use Form - FUJIFILM SonoSite X | | |
|-----------------------------------------------------------------------------------|--|--|
| Porte Ultrasound System | | |
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | | | | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|----------------------------|------------------|
| Transducer: | N/A | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: | | | | | | |
| Clinical Application | Mode of Operation | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | N | N | N | | N | B+M; B+CD | 1-5 |
| Fetal | N | N | N | | N | B+M; B+PWD; B+CD | 1-3,5 |
| Abdominal | N | N | N | | N | B+M; B+PWD;<br>B+CWD; B+CD | 1-5 |
| Intra-operative (Abdominal<br>organs and vascular) | N | N | N | | N | B+M; B+PWD; B+CD | 1-5 |
| Intra-operative (Neuro.) | N | N | N | | N | | |
| Laparoscopic | N | N | N | | N | | |
| Pediatric | N | N | N | | N | B+M; B+PWD;<br>B+CWD; B+CD | 1-5 |
| Small Organ (breast, thyroid,<br>testicles, prostate) | N | N | N | | N | B+M; B+PWD<br>B+CD | 1-5 |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Trans-rectal | | | | | | | |
| Trans-vaginal | N | N | N | | N | B+M; B+PWD; B+CD | 1,5 |
| Trans-urethral | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | |
| Musculo-skel. (Convent.) | N | N | N | | N | B+M; B+PWD<br>B+CD | 1-5 |
| Musculo-skel. (Superfic.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Intra-luminal | | | | | | | |
| Other (spec.) | | | | | | | |
| Cardiac Adult | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | 1-3,5 |
| Cardiac Pediatric | N | N | N | N | N | B+M; B+PWD;<br>B+CWD; B+CD | 1-3,5 |
| Trans-esophageal (card.) | | | | | | | |
| Other (spec.) | | | | | | | |
| Peripheral vessel | N | N | N | | N | B+M; B+PWD; B+CD | 1-5 |
| Other (spec.) | | | | | | | |
N= new indication; P= previously cleared by FDA; E= added under this appendix
#### Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance.
2: Tissue Harmonic Imaging (THI)
3: **Tissue Doppler Imaging (TDI)**
3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
{9}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | | | | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | C60xp/5-2 MHz Transducer | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: | | | | | | |
| Clinical Application | Mode of Operation | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | | | | | | | |
| Fetal | N | N | N | | N | B+M; B+PWD; B+CD | 1,2,5 |
| Abdominal | N | N | N | | N | B+M; B+PWD; B+CD | 1,2,5 |
| Intra-operative (Abdominal<br>organs and vascular) | N | N | N | | N | B+M; B+PWD; B+CD | 1,2,5 |
| Intra-operative (Neuro.) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | N | B+M; B+PWD; B+CD | 1,2,5 |
| Small Organ (breast, thyroid,<br>testicles, prostate) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Trans-rectal | | | | | | | |
| Trans-vaginal | | | | | | | |
| Trans-urethral | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | |
| Musculo-skel. (Convent.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,2,5 |
| Musculo-skel. (Superfic.) | | | | | | | |
| Intra-luminal | | | | | | | |
| Other (spec.) | | | | | | | |
| Cardiac Adult | | | | | | | |
| Cardiac Pediatric | | | | | | | |
| Trans-esophageal (card.) | | | | | | | |
| Other (spec.) | | | | | | | |
| Peripheral vessel | N | N | N | | N | B+M; B+PWD; B+CD | 1,2,5 |
| Other (spec.) | | | | | | | |
Table 1.3.2: Diagnostic Ultrasound Indications for Use Form – C60xp/5-2 Transducer
N= new indication; P= previously cleared by FDA; E= added under this appendix
## Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance.
2: Tissue Harmonic Imaging (THI)
3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
{10}------------------------------------------------
Table 1.3.3: Diagnostic Ultrasound Indications for Use Form – HFL50xp/15-6 Transducer
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | | | | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | HFL50xp/15-6 MHz Transducer | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: | | | | | | |
| Clinical Application | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | | | | | | | |
| Fetal | | | | | | | |
| Abdominal | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Intra-operative (Abdominal<br>organs and vascular) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Intra-operative (Neuro.) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Small Organ (breast, thyroid,<br>testicles, prostate) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Trans-rectal | | | | | | | |
| Trans-vaginal | | | | | | | |
| Trans-urethral | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | |
| Musculo-skel. (Convent.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Musculo-skel. (Superfic.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Intra-luminal | | | | | | | |
| Other (spec.) | | | | | | | |
| Cardiac Adult | | | | | | | |
| Cardiac Pediatric | | | | | | | |
| Trans-esophageal (card.) | | | | | | | |
| Other (spec.) | | | | | | | |
| Peripheral vessel | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Other (spec.) | | | | | | | |
N= new indication: P= previously cleared by FDA; E= added under this appendix
#### Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and In Heades Pinegally (Color Parisheral new block procedures. Color Doppler includes Power/Velocity/Variance.
Imaging guidance (C maging ( ( HI)
**2:** Tissue Harmonic Imaging (THI)
3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
{11}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | | | | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | ICTxp/9-5 MHz Transducer | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human<br>body as follows: | | | | | | |
| Clinical Application | Mode of Operation | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | | | | | | | |
| Fetal | N | N | N | | N | B+M; B+PWD; B+CD | 1,5 |
| Abdominal | | | | | | | |
| Intra-operative (Abdominal<br>organs and vascular) | | | | | | | |
| Intra-operative (Neuro.) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | | | | | | | |
| Small Organ (breast, thyroid,<br>testicles, prostate) | | | | | | | |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Trans-rectal | | | | | | | |
| Trans-vaginal | N | N | N | | N | B+M; B+PWD; B+CD | 1,5 |
| Trans-urethral | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | |
| Musculo-skel. (Convent.) | | | | | | | |
| Musculo-skel. (Superfic.) | | | | | | | |
| Intra-luminal | | | | | | | |
| Other (spec.) | | | | | | | |
| Cardiac Adult | | | | | | | |
| Cardiac Pediatric | | | | | | | |
| Trans-esophageal (card.) | | | | | | | |
| Other (spec.) | | | | | | | |
| Peripheral vessel | | | | | | | |
| Other (spec.) | | | | | | | |
Table 1.3.4: Diagnostic Ultrasound Indications for Use Form - ICTxp/9-5Transducer
No new indication; P= previously cleared by FDA; E= added under this appendix
# Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other analomical structures and imaging guidance for peripheral nerve block procedures. Color Doppler includes Power/Velocity/Variance.
2: Tissue Harmonic Imaging (THI)
3: Tissue Doppier Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
6: Multi-beam Imaging (SonoMB) in B-Mode
{12}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | | | | | | | |
|-------------------------------------------------------|-------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|--|
| Transducer: | L25xp/13-6 MHz Transducer | | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the human | | | | | | | |
| | body as follows: | | | | | | | |
| Clinical Application | Mode of Operation | | | | | | | |
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) | |
| Ophthalmic | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Fetal | | | | | | | | |
| Abdominal | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Abdominal<br>organs and vascular) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-operative (Neuro.) | | | | | | | | |
| Laparoscopic | | | | | | | | |
| Pediatric | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Small Organ (breast, thyroid,<br>testicles, prostate) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Neonatal Cephalic | | | | | | | | |
| Adult Cephalic | | | | | | | | |
| Trans-rectal | | | | | | | | |
| Trans-vaginal | | | | | | | | |
| Trans-urethral | | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | | |
| Musculo-skel. (Convent.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Musculo-skel. (Superfic.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Intra-luminal | | | | | | | | |
| Other (spec.) | | | | | | | | |
| Cardiac Adult | | | | | | | | |
| Cardiac Pediatric | | | | | | | | |
| Trans-esophageal (card.) | | | | | | | | |
| Other (spec.) | | | | | | | | |
| Peripheral vessel | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 | |
| Other (spec.) | | | | | | | | |
Table 1.3.5: Diagnostic Ultrasound Indications for Use Form – L25xp/13-6 Transducer
N= new indication: P= previously cleared by FDA: E= added under this appendix
## Additional Comments:
1: Includes imaging to assist in the placement of needles and catheters in vascular or other anatomical structures and This Not Southers of Dock procedures. Color Doppler includes Power/Velocity/Variance.
imaging guidance for periores (THI)
2: **Tissue Harmonic Imaging (THI)**
3: Tissue Doppler Imaging (TDI)
4: Steep Needle Profiling (Sono MBe)
5: Multi-beam Imaging (SonoMB) in B-Mode
{13}------------------------------------------------
| System: | FUJIFILM SonoSite X-Porte Ultrasound System | | | | | | |
|-------------------------------------------------------|---------------------------------------------------------------------------------------|---|-----|-----|------------------|---------------------|------------------|
| Transducer: | L38xp/10-5 MHz Transducer | | | | | | |
| Intended Use: | Diagnostic ultrasound imaging or fluid flow analysis of the<br>human body as follows: | | | | | | |
| Clinical Application | | | | | | Mode of Operation | |
| | B | M | PWD | CWD | Color<br>Doppler | Combined<br>(Spec.) | Other<br>(Spec.) |
| Ophthalmic | | | | | | | |
| Fetal | | | | | | | |
| Abdominal | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Intra-operative (Abdominal organs<br>and vascular) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Intra-operative (Neuro.) | | | | | | | |
| Laparoscopic | | | | | | | |
| Pediatric | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Small Organ (breast, thyroid,<br>testicles, prostate) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Neonatal Cephalic | | | | | | | |
| Adult Cephalic | | | | | | | |
| Trans-rectal | | | | | | | |
| Trans-vaginal | | | | | | | |
| Trans-urethral | | | | | | | |
| Trans-esoph. (non-Card.) | | | | | | | |
| Musculo-skel. (Convent.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Musculo-skel. (Superfic.) | N | N | N | | N | B+M; B+PWD; B+CD | 1,4,5 |
| Intra-luminal | | | | | | | |
| Other (spec.) | | | | | | | |
| Cardiac Adult | | | | | | | |
| Cardiac Pediatric | | | | | | | |
| Trans-esophageal (card.) | | | | | |…
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.