BEACON TIP TORCON NB ADVANTAGE CATHETER, TORCON NB ADVANTAGE CATHETER
K133130 · Cook, Inc. · DQO · Jan 9, 2015 · Cardiovascular
Device Facts
Record ID
K133130
Device Name
BEACON TIP TORCON NB ADVANTAGE CATHETER, TORCON NB ADVANTAGE CATHETER
Applicant
Cook, Inc.
Product Code
DQO · Cardiovascular
Decision Date
Jan 9, 2015
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1200
Device Class
Class 2
Indications for Use
The catheters are intended for use in the peripheral and coronary vascular system including the carotid arteries in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.
Device Story
Diagnostic intravascular catheters; used for delivery of contrast media to diagnose vascular occlusion or stenosis. Beacon Tip models feature distal radiopaque tip bonded to nylon shaft with braided stainless steel; Torcon NB models feature thinwall radiopaque nylon with stainless steel braiding. Percutaneous introduction into vascular system over wire guide. Used in clinical settings by physicians; provides conduit for contrast media injection; aids visualization of vascular anatomy for diagnostic decision-making.
Clinical Evidence
Bench testing only. Testing included dynamic flow rate, static burst pressure (ISO 10555-2:1996), tensile testing (ISO 10555-1:2009), liquid/air leakage (ISO 10555-1:2013), and biocompatibility (ISO 10993-1:2009) including cytotoxicity, sensitization, irritation, systemic toxicity, hemolysis, implantation, and complement activation.
Technological Characteristics
Materials: Nylon, stainless steel braiding, radiopaque markers. Dimensions: 4.0-8.0 French, 35-170 cm lengths. Accepts 0.025-0.038 inch wire guides. Features: Distal radiopaque tips, optional sideports (0-12). Standards: ISO 10555-1, ISO 10555-2, ISO 10993-1.
Indications for Use
Indicated for use in peripheral and coronary vascular systems, including carotid arteries, for angiographic procedures. For use by physicians trained in angiographic techniques.
Regulatory Classification
Identification
An intravascular diagnostic catheter is a device used to record intracardiac pressures, to sample blood, and to introduce substances into the heart and vessels. Included in this generic device are right-heart catheters, left-heart catheters, and angiographic catheters, among others.
Imager II 5F Selective Angiography Catheter (K121694)
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" written around the perimeter. In the center of the seal is a stylized image of three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
January 9, 2015
Cook, Inc. David Chadwick, PhD, RAC Director, Regulatory Affairs/Regulatory Science 750 Daniels Way Bloomington, Indiana 47404
Re: K133130
> Trade/Device Name: Beacon Tip Torcon NB Advantage Catheter, Torcon NB Advantage Catheter Regulation Number: 21 CFR 870.1200 Regulation Name: Diagnostic Intravascular Catheter Regulatory Class: Class II Product Code: DOO Dated: December 8, 2014 Received: December 9, 2014
Dear David Chadwick, Phd., Rac,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device
{1}------------------------------------------------
related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
Melissa A. Torres -S
For Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number (if known) K133130
#### Device Name
Beacon® Tip Torcon NB® Advantage Catheter and Torcon NB® Advantage Catheter
Indications for Use (Describe)
The catheters are intended for use in the peripheral and coronary vascular system including the carotid arteries in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.
| Type of Use (Select one or both, as applicable) |
|-------------------------------------------------|
|-------------------------------------------------|
| | <span style="font-size: 10pt;">☑ Prescription Use (Part 21 CFR 801 Subpart D)</span> |
|--|--------------------------------------------------------------------------------------|
| | <span style="font-size: 10pt;">☐ Over-The-Counter Use (21 CFR 801 Subpart C)</span> |
#### CONTINUE ON A SEPARATE PAGE IF NEEDED.
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image shows the Cook Medical logo. The logo is set against a red background. The word "COOK" is written in white, block letters at the top, and the word "MEDICAL" is written in white, block letters at the bottom.
COOK INCORPORATED
750 DANIELS WAY, P.O. BOX 489
BLOOMINGTON, IN 47402-0489 U.S.A.
PHONE: 812.339.2235 TOLL FREE: 800.457.4500
WWW.COOKMEDICAL.COM
### 510(k) SUMMARY
#### Submitted By:
| Applicant: | Cook Incorporated |
|-----------------------|------------------------------------------|
| Address: | 750 Daniels Way<br>Bloomington, IN 47404 |
| Contact: | David E. Chadwick, Ph.D., RAC, FRAPS |
| Email: | david.chadwick@cookmedical.com |
| Contact Phone Number: | (812) 335-3575 x102330 |
| Contact Fax Number: | (812) 332-0281 |
| Device Information: | |
Trade Name:
Common Name: Classification Name:
Class: Device Panel: Date Prepared:
#### Indications for Use:
Beacon® Tip Torcon NB® Advantage Catheter and Torcon NB® Advantage Catheter Diagnostic Intravascular Catheter Catheter, Diagnostic Intravascular DQO (21 CFR §870.1200) Class II Cardiovascular January 5. 2015
The catheters are intended for use in the peripheral and coronary vascular system including the carotid arteries in angiographic procedures by physicians trained and experienced in angiographic techniques. Standard techniques for placement of vascular access sheaths, angiographic catheters and wire guides should be employed.
#### Predicate Devices:
Cook Incorporated's Beacon® Tip Torcon NB® Advantage Catheters and Torcon NB® Advantage Catheters have an identical intended use and are similar in terms of principles of operation, materials of construction, and technological characteristics to the predicate devices. The devices, subject of this submission, are substantially equivalent to the Slip-Cath® Beacon Tip Catheters and the Shuttle® Select Slip-Cath® Catheters, and the Slip-Coat™ Catheters manufactured by Cook Incorporated, which are cleared under 510(k) numbers K122937 and K882796. respectively. They are also substantially equivalent to the Imager II 5F Selective Angiography Catheter manufactured by Boston Scientific Corporation, which is cleared under 510(k) number K121694.
{4}------------------------------------------------
## Comparison to Predicate Devices:
It has been demonstrated that the Beacon® Tip Torcon NB® Advantage Catheters and Torcon NB® Advantage Catheters are comparable to the predicate devices in terms of design, intended use, materials, fundamental technology, and principles of operation.
### Device Description:
The Beacon® Tip Torcon NB® Advantage Catheters are visually identified by a distal radiopaque tip bonded onto a nylon catheter shaft with braided stainless steel. Beacon® Tip Torcon NB® Advantage Catheters are available in 4.0, 4.1, 5.0, and 6.0 French with lengths measuring 40 to 170 centimeters. The catheter inner lumen is manufactured to accept a 0.035 or 0.038 inch diameter wire guide. These devices are manufactured without sideports and in a variety of distal tip configurations.
The Torcon NB® Advantage Catheters are selective angiographic catheters manufactured with a thinwall radiopaque nylon material and constructed with stainless steel braiding. The Torcon NB® Advantage Catheters are available in 4.0, 4.1, 5.0, 6.0, 7.0 and 8.0 French with lengths measuring 35 to 150 centimeters. The catheter inner lumen is manufactured to accept a 0.025, 0.035, or 0.038 inch diameter wire guide. These devices are manufactured with zero to twelve sideports, depending on the distal curve radius, and are available in a variety of distal tip configurations. These tips may be tapered at the distal endhole, depending upon device specifications. The catheters are also available in a variety of distal tip configurations.
Beacon® Tip Torcon NB® Advantage Catheters and Torcon NB® Advantage Catheters are for use in angiographic procedures as a conduit for the delivery of contrast media and can help physicians diagnose occlusion or stenosis. The catheters are designed for percutaneous introduction into the vascular system over an appropriately sized wire guide.
### Test Data:
The following tests were performed to demonstrate that the Beacon® Tip Torcon NB® Advantage Catheters and Torcon NB® Advantage Catheters met applicable design and performance requirements and support a determination of substantial equivalence.
- Dynamic Flow Rate and Burst Pressure Evaluation - This study is for characterization purposes; therefore no acceptance criteria apply.
- Static Burst Pressure Testing - Testing verified that, in accordance with ISO 10555-2:1996(E), the test articles withstood a minimum of 1200 psig static pressure for at least 2 seconds without failure. The predetermined acceptance criterion was met.
- Evaluation of Dynamic Pressure During Injection after Three Years Accelerated . Aging – Testing verified that the test articles achieved a flow rate of at least
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image is a logo for Cook Medical. The logo is a red rectangle with the word "COOK" in white, sans-serif font on the top half of the rectangle. There is a registered trademark symbol after the word "COOK". The bottom half of the rectangle has a smaller red rectangle that is angled to the left. The word "MEDICAL" is in white, sans-serif font on the smaller rectangle.
12 mL/sec when pressurized to 1215 psig ± 15 psig without failure. The predetermined acceptance criterion was met.
- Biocompatibility Testing Per ISO 10993-1:2009, the proposed devices were ● classified as external communicating devices in contact with circulating blood for a limited (≤ 24 hours) duration. The following tests were completed and the biocompatibility was deemed acceptable: Cytotoxicity, Sensitization, Irritation/Intracutaneous Reactivity, Systemic Toxicity, Hemolysis, Implantation, Partial Thromboplastin Time, C3a Completement Activation, SC5b-9 Complement Activation, Mutagenicity, Pyrogenicity.
- . Tensile testing - Testing verified that, under proper clinical use, the peak load values would be greater than 10N, in accordance with ISO 10555-1:2009. The predetermined acceptance criterion was met.
- Liquid Leakage Testing Testing verified that, in accordance with ISO 10555-● 1:2013, the catheter maintained a pressure of 300 kPa for 30 seconds without leaking between the catheter shaft and hub. The predetermined acceptance criterion was met.
- Air Leakage Testing Testing verified that, in accordance with ISO 10555-● 1:2013, there was no leakage of air between the catheter shaft and hub when subjected to a negative pressure for 15 seconds. The predetermined acceptance criterion was met.
- Particulate Testing This study is for characterization purposes; therefore no ● acceptance criteria apply.
In conclusion, the results of these tests support a determination of substantial equivalence to the predicate device.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.