K133113 · Kingtech Enterprise, Ltd. · FLL · Dec 26, 2013 · General Hospital
Device Facts
Record ID
K133113
Device Name
KINGTECH FOREHEAD THERMOMETER
Applicant
Kingtech Enterprise, Ltd.
Product Code
FLL · General Hospital
Decision Date
Dec 26, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.2910
Device Class
Class 2
Attributes
Pediatric
Indications for Use
TF1265 and TF1267 Forehead Thermometer is an electronic thermometer using an infraced sensor to detect human body temperature fro the surface of human skin without contact. It is for use by people of all ages in the homecare environment.
Device Story
The KINGTECH Forehead Thermometer (Models TF1265, TF1267) is a non-contact electronic thermometer designed for home use. It utilizes an infrared sensor to detect infrared radiation emitted from the surface of human skin. The device converts this detected infrared energy into a digital temperature value. It is intended for use by individuals of all ages to monitor body temperature. The device provides a temperature reading to the user, which assists in clinical decision-making regarding health status. The TF1267 model features minor differences in exterior dimensions compared to the TF1265.
Clinical Evidence
No clinical data. Bench testing only. Compliance with IEC 60601-1 (safety), EN/IEC 60601-1-2 (EMC), EN12470, ASTM E1965-98, and ASTM E1112-00 standards.
Technological Characteristics
Infrared sensor-based temperature measurement; non-contact operation. Complies with IEC 60601-1, EN/IEC 60601-1-2, EN12470, ASTM E1965-98, and ASTM E1112-00. Handheld form factor.
Indications for Use
Indicated for use by people of all ages for non-contact measurement of human body temperature from the surface of the skin in a homecare environment.
Regulatory Classification
Identification
A clinical electronic thermometer is a device used to measure the body temperature of a patient by means of a transducer coupled with an electronic signal amplification, conditioning, and display unit. The transducer may be in a detachable probe with or without a disposable cover.
Special Controls
(1) Device is not a clinical thermometer with telethermographic functions;
(2) Device is not a clinical thermometer with continuous temperature measurement functions; and
(3) Appropriate analysis and testing (such as that outlined in the currently FDA-recognized editions, as appropriate, of ISO 80601-2-56, “Medical electrical equipment—Part 2-56: Particular requirements for basic safety and essential performance of clinical thermometers for body temperature measurement,” or ASTM E1965, “Standard Specification for Infrared Thermometers for Intermittent Determination of Patient Temperature,” or ASTM E1112, “Standard Specification for Electronic Thermometer for Intermittent Determination of Patient Temperature,” or ASTM E1104, “Standard Specification for Clinical Thermometer Probe Covers and Sheaths”) must validate specifications and performance of the device.
{0}------------------------------------------------
## KINGTECH ENTERPRISES LIMITED
ROOM 2016. 20F., BLOCK B. REGENT CENTER, 70 TA CHUEN PING ST., KWAJ CHUNG. N.T. HONG KONG TEL : 852-24810188 FAX : 852-24253939
## 510(K) SUMMARY
## for KINGTECH Forehead Thermometer, TF1265, TF1267
This summary of 510(K) safety and effectiveness information is being submitted in accordance with the requirements of 21 CFR 807.92.
The assigned 510(K) number is:__K 133113
| Submission Date: | September 22, 2013 | |
|------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Submitter: | Kingtech Enterprises Limited | |
| | Room 2016, 20/F., Block B, Regent Center 70 Ta Chunen Ping<br>Street, Kwai Chung, New Territories, HK | |
| | Tel: +852-24810188 Fax:+852-2423939 | |
| Manufacturer: | Kingtech (Dong Guan) Enterprises Limited | |
| | Farm Village, Da Ling Shan Town, DongGuan, China<br>Tel: +86-76985636260 Fax:+86-76985636350 | |
| Establishment<br>Registration No.: | 3008808166 | |
| Official Contact: | Dr. Jen, Ke-Min<br>Tel: +886-3-5208829 Fax:+886-3-5209783<br>Email: ceirs.jen@msa.hinet.net | |
| Common /<br>Usual Name: | Forehead Thermometer | |
| Trade Name: | KINGTECH Forehead Thermometer. Models: TF1265, TF1267 | |
| Classification<br>Code: | FLL, Class II, 21 CFR 880.2910 | |
| Intended Use: | TF1265, TF1267 Forehead Thermometer is an electronic<br>thermometer using an infrared sensor to detect human body<br>temperature fro the surface of human skin without contact.<br>It is for use by people of all ages in the homecare environment | |
| Predicated<br>Devices: | K122221, Nexus IR30 Thermometer, Model: TD-1265<br>TaiDoc Technology Corporation | |
| Device Description: | The Forehead Thermometer TF1265, TF1267 is characterized by<br>measuring human body temperature from the surface of human<br>skin. It utilizes infrared technology to measure infrared energy<br>emitted from the skin surface when making a temperature<br>measurement. | |
| Test Principle: | The thermometer Measures temperature by reading infrared<br>radiation emitting from the skin and converts it into a temperature<br>value. | |
{1}------------------------------------------------
## KINGTECH ENTERPRISES LIMITED
#### ROOM 2016, 20/P., BLOCK B, REGENT CENTER, 70 TA CHUEN PING ST., KWAI CHUNG, N.T. HONG KONG TEL : 852-24810188 FAX : 852-24253939
Performance Tests: Safety Test: · IEC 60601-1 - Medical electrical equipment Part 1. General requirements for safety, 2005. Electromagnetic Compatibility Test: · EN/IEC 60601-1-2 - Medical electrical equipment, Part 2. Electromagnetic compatibility - Requirements and tests, 2007. In accordance with: EN12470, ASTME 1965-98, and ASTM E1112-00 This 510k submission only need change the application name of the predicate K122221 from "Nexus IR30 Thermometer, Conclusion: TD-1265" to "KINGTECH Forehead Thermometer, TF1265" there is the entire identical specifications and only need to separate into different 510k. And all of the test reports and documentation for this 510k submission were prepared by the applicant of the predicate device: TaiDoc Technology Corporation. We, Kingtech Enterprises Limited, are authorized by BioCare Co., Ltd. who is a branch office of TaiDoc Technology Corporation. We also present the authorized "Statement" as the following page. In addition, another model TF1267 of the subject device, it is only minor difference to the exterior dimension for the TF1265. Thus the new device is substantially equivalent to the predicate devices in this aspect.
#### Clinical Tests:
# Comparison and
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle or bird-like symbol with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular fashion around the bird symbol. The logo is black and white.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G60 Silver Spring, MD 20993-0002
### December 26, 2013
Kingtech Enterprises Limited C/O Dr. Ke-Min Jen Official Correspondent Room 2016, 20/F., Block B, Regent Center 70 Ta Chunen Ping Street Kwai Chung, New Territories HONG KONG
Re: K133113
Trade/Device Name: KINGTECH Forehead Thermometer, Model: TF1265, TF1267 Regulation Number: 21 CFR 880.2910 Regulation Name: Clinical electronic thermometer Regulatory Class: II . Product Code: FLL Dated: September 22, 2013 Received: September 30, 2013
Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 - Dr. Jen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
**Kwame O.**
**Ulmer-S**
for
Erin I. Keith, M.S. Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
#### Indications for Use
510(k) Number (if known) K133113
ﺒﻪ، "ﻧﺎﺀ
#### Device Name
KINGTECH Forehead Thermometer, Models: TF 1265, TF 1267
#### Indications for Use (Describe)
TF1265 and TF1267 Forehead Thermometer is an electronic thermometer using an infraced sensor to detect human body temperature fro the surface of human skin without contact. It is for use by people of all ages in the homecare environment.
Type of Use (Select one or both, as applicable)
[] Prescription Use (Part 21 CFR 801 Subpart D)
χ Over-The-Counter Use (21 CFR 801 Subpart C)
## Please do not write below this line - continue on a separate page if needed.
FOR FDA USE ONLY
| Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | Digitally signed by Richard C. Chapman<br>Date: 2013.12.26 11:50:34 -05'00' |
|------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| FORM FDA 3881 (9/13) | Page 1 of 2 |
{5}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
ﺮﺱ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺎﺣﺔ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟﻤﺴﺘﻮﻯ ﺍﻟ
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.