← Product Code [FMI](/productcode/FMI) · K133100

# CAREFINE PEN NEEDLE (K133100)

_Facet Technologies · FMI · Jan 23, 2014 · General Hospital · SESE_

**Canonical URL:** https://fda-staging.innolitics.com/device/K133100

## Device Facts

- **Applicant:** Facet Technologies
- **Product Code:** [FMI](/productcode/FMI.md)
- **Decision Date:** Jan 23, 2014
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 880.5570
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The CareFine™ Pen Needle is intended for use with a pen injector device for the subcutaneous injection of insulin.

## Device Story

CareFine™ Pen Needles are single-use, sterile, double-ended cannulas designed for subcutaneous insulin delivery. The device consists of a cannula assembled into an injection-molded hub with internal threads for attachment to pen injectors. The hub allows the cartridge-end of the cannula to penetrate the pen's rubber septum. Both ends of the cannula are lubricated with silicone to facilitate injection and septum penetration. An inner shield protects the patient-end point, and an outer cover provides a sterility barrier. Used by consumers, caregivers, and healthcare professionals in home or clinical settings. The device facilitates insulin administration by connecting the insulin cartridge to the patient's subcutaneous tissue.

## Clinical Evidence

No clinical data required; substantial equivalence supported by non-clinical performance testing, including biocompatibility, sterility, and comparative bench testing against predicate devices.

## Technological Characteristics

Single-lumen, double-ended stainless steel cannula; injection-molded hub and shields; silicone lubricant; EtO terminal sterilization (SAL 10^-6); various gauges (29G-32G) and lengths (4mm-12.7mm).

## Regulatory Identification

A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.

## Predicate Devices

- UltiCare® Disposable Pen ([K100812](/device/K100812.md))
- FlexPen® ([K090111](/device/K090111.md))
- BD Pen Needle ([K100005](/device/K100005.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KI33100

### Facet Technologies, LLC Traditional 510(k) for the CareFine™ Pen Needles Submitted Under 21 CFR § 807.87

JAN 2 3 2014

# Confidential and Proprietary

9 510(k) Summary

55

{1}------------------------------------------------

# 510(k) Summary as Required by 21 CFR §807.92(c)

As required by the Safe Medical Devices Act (SMDA) of 1990 and in accordance with 21 CFR §807.92(a), the 510(k) summary provided below is of sufficient detailed to provide an understanding of the basis for a determination of substantial equivalence.

| Date<br>Prepared          | Summary | September 25, 2013                                                                                                                                 |
|---------------------------|---------|----------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Sponsor<br>Address |         | Facet Technologies, LLC<br>112 Town Park Drive, Suite 300<br>Kennesaw, Georgia 30144                                                               |
| Contact Person            |         | Mary Ann Kinard<br>Senior Regulatory Affairs Manager<br>Phone: (770) 590-6462<br>FAX: (770) 590-6412<br>Email:MaryAnn_Kinard@facettechnologies.com |
| Trade Name                |         | CareFine™ Pen Needles                                                                                                                              |
| Common Name               |         | Insulin Pen Needle                                                                                                                                 |
| Classification Name       |         | Needle, Hypodermic, Single Lumen                                                                                                                   |
| Regulation Number         |         | 21 CFR §880.5570                                                                                                                                   |
| Product Code              |         | FMI                                                                                                                                                |
| Device Classification     |         | II                                                                                                                                                 |
| Review Panel              |         | 80 General Hospital                                                                                                                                |

### 510(k) Summary

CORPORATE HEADQUARTERS 112 Town Park Drive, Suite 300 Kennesaw, GA 30144 Phone: 770-590-6400 Fax: 770-590-6461

MANUFACTURING CENTER 101 Liberty Industrial Parkway McDonough, GA 30253 Phone: 770-898-3223 Fax: 770-898-1535

EUROPEAN OFFICE Unit 7 Mobbs Miller House Christchurch Road Northampton NN1 5LL U.K. Phone: +44 1604 638340 Fax: +44 1604 636 901

Confidential and Proprietary Copyright ® 2013 Facet Technologies, LLC

{2}------------------------------------------------

#### Predicate Marketed Devices:

| [807.92(a)(3)]     |                          |               |
|--------------------|--------------------------|---------------|
| Manufacturer       | Trade Name               | 510(k) Number |
| UltiMed, Inc.      | UltiCare® Disposable Pen | K100812       |
| Novo Nordisk, Inc. | FlexPen®                 | K090111       |
| Becton Dickinson   | BD Pen Needle            | K100005       |

## [(807.92(a)(4)] Description of Device:

CareFine™ Pen Needles are single use, sterile medical devices designed to be used in conjunction with pen injectors and pen cartridges for the subcutaneous injection of insulin. Pen needles are used by consumers, caregivers and health care professionals. They are offered in various gauge sizes (29G, 30G, 31G and 32G) and lengths (4mm, 5mm, 6mm, 8mm, and 12.7mm). CareFine™ Pen Needles are sterile (EtO terminal sterilization), non-toxic and non-pyrogenic.

The pen needle assembly consists of a doubled-ended cannula that is assembled into an injection molded hub using adhesive. The hub has internal threads, which allow it to be screwed onto the pen injector device. This allows the cartridge end of the cannula to penetrate through the rubber septum of the cartridge. The patient-end and the cartridgeend of the cannula are lubricated using a silicone based lubricant for ease of injection and rubber septum penetration.

An injection-molded inner shield is assembled over the patient-end of the cannula to protect the point from damage and accidental needle-sticks. This needle assembly is inserted into a protective injection-molded outer cover and sealed with a peel-away label to provide a sterility barrier and tamper evidence seal. The peel-away label is pre-printed with information, which includes the lot number and needle gauge and length.

The individual needle assemblies are packaged in cartons, and placed into shippers with appropriate labeling. The shipper cases are palletized and sterilized to a SAL of 10° by EtO terminal sterilization.

# [(807.92(a)(5)] Indication for Use:

The CareFine™ Pen Needle is intended for use with a pen injector device for the subcutaneous injection of insulin.

{3}------------------------------------------------

# [(807.92(a)(6)] Technological Characteristics:

The CareFine™ Pen Needles are substantially equivalent to the UltiCare®, Flex Pen® and BD Pen Needles in terms of indications for use, compositions, material and design. Specifically, the following comparisons were made to determine equivalence of the CareFine™M Pen Needle to the referenced predicate pen needles:

- Needle sharpness, .
- . length, gauge,
- needle shield color(s), .
- needle shield dimensions, .
- . needle shield strength,
- hub/needle bond strength, .
- . biocompatibility, materials and
- . sterility

The CareFine™ Pen Needle product family is equivalent to the predicate devices, given that it:

- . Does not modify the indication for use or any labeling change that affects the intended use of the device,
- Does not alter the fundamental scientific technology, .
- Uses the same operating principles, .
- . Is manufactured from the same/similar materials,
- Is sterilized with a resulting SAL of 10-6, and .
- Is packaged using same/similar unit and case materials. .

Based on the comparisons described above to the predicate comparator devices, the CareFineTM Pen Needles do not raise any new issues of safety and effectiveness.

### [(807.92(b)(1)] Non-Clinical Performance Data:

The CareFine™ Pen Needle is manufactured in accordance with the requirements of the current Current Good Manufacturing Practices for Medical Devices and the testing performed which support a determination of substantial equivalence includes:

- . appropriate biocompatibility tests,
- sterility tests. . ●
	- predicate product comparative tests and ●

Confidential and Proprietary Copyright " 2013 Facet Technologies, LLC

{4}------------------------------------------------

### Facet Technologies, LLC Traditional 510(k) for the CareFine™ Pen Needles Submitted Under 21 CFR § 807.87

. performance tests.

The test results are in compliance with existing domestic and international standards. Based on the above test parameter results, the CareFine™ Pen Needles do not raise any new issues of safety and effectiveness.

# · [(807.92(b)(2)] Clinical Performance Data:

Clinical data is not required.

[(807.92(b)(3)] Conclusion:

Based on the design equivalency and the functional testing performed, Facet Technologies has determined that CareFine™ Pen Needles are equivalent to the predicate comparator devices currently cleared for marketing in the United States. Differences between the devices do not raise any significant issues of safety and effectiveness.

# [(807.92(d)] Other Information:

None

End of 510(k) Summary

{5}------------------------------------------------

DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes extending from its wing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the eagle.

#### Public Health Service

Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002

#### January 23, 2014

Facet Technologies, LLC Ms. Mary Ann Kinard Senior Regulatory Affairs Manager 112 Town Park Drive, Suite 300 Kennesaw, GA 30144

Re: K133100

Trade/Device Name: CareFine™ Pen Needles Regulation Number: 21 CFR 880.5570 Regulation Name: Hypodermic Single Lumen Needle Regulatory Class: II Product Code: FMI Dated: December 19, 2013 Received: December 20, 2013

Dear Ms. Kinard:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21

{6}------------------------------------------------

Page 2 - Ms. Kinard

CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to

http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.

You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.

Sincerely yours,

Image /page/6/Picture/6 description: The image contains the text "Kwame Ulmers". The text is arranged vertically, with "Kwame" on top and "Ulmers" below. To the right of the text is a design that includes the letters "EDA" and a triangle shape. The design appears to be stylized or abstract.

for

Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{7}------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration

#### Indications for Use

510(k) Number (if known) K133100

#### Device Name

CareFine™ Pen Needles

#### Indications for Use (Describe)

The CareFine™ Pen Needle is intended for use with a pen injector device for the subcuraneous injection of insulin.

Type of Use (Select one or both, as applicable)

Prescription Use (Part 21 CFR 801 Subpart D)

(2) Over-The-Counter Use (21 CFR 801 Subpart C)

# PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.

THE THE THE PEORIFDAIUSE ONLY ONLY ONLY CLI , "게 대표 프 .

Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)

Image /page/7/Picture/13 description: The image shows the name "Sajjad H. Syed" in large, bold font. To the right of the name is a digital signature. The signature includes the text "Digitally signed by Sajjad H. Syed -S" and other information such as "DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, People, cn=Sajjad H. Syed -S, 2342\19200300.100.1.1=2000601742 0 Date: 2014.01.23 10:40:38 -05'00'".

FORM FDA 3881 (9/13)

PSC Publishing Services (131) 1434722 Fir

Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.

{8}------------------------------------------------

This section applies only to requirements of the Paperwork Reduction Act of 1995.

#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*

The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:

> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov

"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."

---

**Source:** [https://fda-staging.innolitics.com/device/K133100](https://fda-staging.innolitics.com/device/K133100)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).