DISCOVERY SYSTEM

K133062 · Innovative Visual Systems · HJO · Jun 20, 2014 · Ophthalmic

Device Facts

Record IDK133062
Device NameDISCOVERY SYSTEM
ApplicantInnovative Visual Systems
Product CodeHJO · Ophthalmic
Decision DateJun 20, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1850
Device ClassClass 2

Indications for Use

The Discovery System is intended for use in: -Mapping the lower and higher order aberrations of the eye which includes . measurement of the spherical power, cylindrical power and cylinder axis. - . The measurement and analysis of corneal curvature (corneal refractive power), cylindrical power, and cylinder axis of the cornea. The device also maps the display of the corneal shape. - White-to-white measurements (also called WTW or horizontal corneal . . diameter). - . Retro-illumination imaging of the anterior segment of the eye including intraocular lens imaging. - The measurement of pupil diameter. . - In the automated measurement and analysis of refractive errors of the eye . including hyperopia and myopia from -25.0 to +15.0 diopters spherical, and astigmatism from 0.0 to ± 10.0 diopters.

Device Story

Discovery System combines corneal topographer and ocular wavefront aberrometer; captures anterior corneal topography, ocular wavefront aberrations, retro-illumination images, and iris images. Uses near infrared (NIR) light sources for simultaneous topography and wavefront measurement along same optical axis. Retro-illumination imaging utilizes NIR light reflected off retina; white light/NIR anterior eye light sources optional. Iris images captured to identify eye rotations between exams. Used in clinical settings by eye care professionals. Output provides clinicians with detailed ophthalmic measurements, including refractive error analysis, corneal shape mapping, and anterior segment visualization. Data assists in clinical decision-making regarding refractive status and ocular health. Benefits include efficient, simultaneous capture of multiple diagnostic parameters, reducing exam time and improving measurement alignment.

Clinical Evidence

Bench testing only. Performance evaluated using high-precision contact lens lathe-produced surfaces for system calibration, corneal topography, and ocular wavefront measurements. Light source safety evaluated against ANSI Z136.1, IEC 62471, and ISO 15004-2 standards.

Technological Characteristics

Combines corneal topographer and ocular wavefront aberrometer. Light sources: NIR SLD, NIR LEDs, and White LEDs. Connectivity: Not specified. Sterilization: Not applicable. Software: Automated measurement and analysis of refractive errors and corneal topography.

Indications for Use

Indicated for patients requiring ophthalmic assessment of refractive errors (hyperopia, myopia -25.0 to +15.0 D; astigmatism 0.0 to ± 10.0 D), corneal curvature, ocular wavefront aberrations, pupil diameter, and white-to-white measurements. Used for retro-illumination imaging of the anterior segment and intraocular lenses.

Regulatory Classification

Identification

An AC-powered slitlamp biomicroscope is an AC-powered device that is a microscope intended for use in eye examination that projects into a patient's eye through a control diaphragm a thin, intense beam of light.

Special Controls

*Classification.* Class II (special controls). The device, when it is intended only for the visual examination of the anterior segment of the eye, is classified as Group 1 per FDA-recognized consensus standard ANSI Z80.36, does not provide any quantitative output, and is not intended for screening or automated diagnostic indications, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/1 description: The image shows a logo with the letters "IVS" in a bold, sans-serif font. The letters are arranged horizontally, with "I" on the left, "V" in the middle, and "S" on the right. The entire logo is enclosed within a roughly drawn oval shape. The oval appears to be hand-drawn, with uneven lines and slight imperfections, giving it a somewhat rustic or informal appearance. INNOVATIVE VISUAL SYSTEMS, LLC # 510(k) Summary # Submitter Information | A. Company Name: | Innovative Visual Systems | |-------------------------|---------------------------------------------------------| | B. Company Address: | 386 North York Road<br>Suite 209<br>Elmhurst, IL 60126 | | C. Company Phone / Fax: | (T) 630.530.9700<br>(F) 630.530.1636 | | D. Contact Person: | Donald R. Sanders, MD, PhD<br>Innovative Visual Systems | - E. Date Summary Prepared: May 7, 2014 # Device Identification | A. Regulation: | 21 CFR 886.1850 | |-----------------------------------------------|------------------------------------| | B. Classification: | Class II | | C. Product Codes: | HJO, MMQ, HKO | | D. Common Device Name: | AC-Powered Slit-lamp Biomicroscope | | E. Trade / Proprietary Name: Discovery System | | 386 North York Road •Suite 209 • Elmhurst, Illinois 60126 • Phone: (630) 530-9700 • Fax: (630) 530-1636 . {1}------------------------------------------------ ## Substantial Equivalence: A comparison of the intended use, clinical applications and technological characteristics of the Discovery System and the predicate devices has been performed. The results of this comparison demonstrates that the Discovery System is substantially equivalent to the marketed predicate devices. | Predicate Device | Manufacturer | 510(k) Number | Date Cleared | |------------------|--------------|---------------|--------------| | OPD-Scan™ | Nidek Inc. | K003299 | 3/9/01 | | Slit Lamp BM 900 | Haag-Streit | K100202 | 3/19/2010 | ## Description of the Device: · The Discovery System combines a corneal topographer and an ocular wavefront aberrometer into a single, efficient system. The Discovery System combines several features to provide clinicians with detailed ophthalmic measurements. The Discovery System is designed to quickly and easily capture: - Anterior corneal topography . - Ocular wavefront aberrations . - Retro-illumination images . - Iris images . Using near infrared (NIR) light sources for both the corneal topography and ocular wavefront aberrations permits these measurements to be taken simultaneously along the same optical axis. Retro-illuminated imaging is provided using a NIR light source reflected off the retina and optional white light and/or NIR anterior eye light sources. Iris images are captured during a retro-illumination acquisition and can optionally be captured during all other exam type acquisitions to identify eye rotations between examinations. {2}------------------------------------------------ ## Indications for Use: The Discovery System is intended for use in: - Mapping the lower and higher order aberrations of the eye which includes . measurement of the spherical power, cylindrical power and cylinder axis. - . The measurement and analysis of corneal curvature (corneal refractive power), cylindrical power, and cylinder axis of the cornea. The device also maps the display of the corneal shape. - White-to-white measurements (also called WTW or horizontal corneal . . diameter). - . Retro-illumination imaging of the anterior segment of the eye including intraocular lens imaging. - The measurement of pupil diameter. . - In the automated measurement and analysis of refractive errors of the eye . including hyperopia and myopia from -25.0 to +15.0 diopters spherical, and astigmatism from 0.0 to ± 10.0 diopters. ### Performance Data: The performance data indicate that the Discovery System meets all specified requirements. Light source power measurements were made for all three light sources (at maximum power output by our instrument) at the corneal plane. To evaluate safe light levels for the NIR SLD we used levels specified in ANSI 136.1 and ISO 15004-2. To evaluate safe light levels for the NIR LEDs we used levels specified in IEC 62471 and ISO 15004-2. To evaluate safe light levels for the White LEDs we used levels specified in IEC 62471 and ISO 15004-2. To evaluate safe light levels for combinations of the light sources we used ISO 15004-2. System measurement performance was evaluated using bench testing. This testing consisted of designing and producing surfaces on a high-precision contact lens lathe for system calibration. corneal topography measurements, and ocular wavefront measurements. {3}------------------------------------------------ Image /page/3/Picture/0 description: The image shows the logo for the Department of Health & Human Services USA. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring a symbol that resembles a person embracing or supporting another figure, with flowing lines suggesting movement or support. The emblem is rendered in a dark color, contrasting with the likely lighter background of the seal. Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G6(19 Silver Spring, MI) 20993-0002 June 20, 2014 Innovative Visual Systems Donald R. Sanders. M.D. Ph.D. Manager 386 N. York Road. Suite 209 Elmhurst. IL 60126 Rc: K133062 Trade/Device Name: Discovery System Regulation Number: 21 CFR 886.1850 Regulation Name: AC-powered slit lamp biomicroscope Regulatory Class: Class II Product Code: HJO, MMQ. HKO Dated: May 7, 2014 Received: May 9, 2014 Dear Dr. Sanders: We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading, If vour device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {4}------------------------------------------------ Page 2 - Donald R. Sanders, M.D. Ph.D. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act): 21 CFR 1000-1050. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely yours. # Kesia Y. Alexander -S for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ ### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration # Indications for Use 510(k) Number (if known) K 133062 Device Name Discovery System Indications for Use (Describe) The Discovery System is intended for use in: -Mapping the lower and higher order aberrations of the eye which includes measurement of the spherical power, cylindrical power and cylinder axis. -The measurement and analysis of corneal curvature (comeal refractive power, and cylinder axis of the cornea. The device also maps the display of the corneal shape. - White-to-white measurements (also called WTW or horizontal corneal diameter). -Retro-illumination imaging of the anterior segment of the eye including intraocular lens imaging. - The measurement of pupil diameter. -In the automated measurement and analysis of refractive errors of the eye including hyperopia and myopia from -25.0 to +15.0 diopters spherical, and astigmatism from 0.0 to ± 10.0 diopters. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) ## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. ## FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) # 014.06.17 11:20:14 This section applies only to requirements of the Paperwork Reduction Act of 1995. ## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...