The Milestone™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade I Spondylolisthesis at the involved level. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via transforaminal approach. These implants are to be used with autogenous bone graft. DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies. These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
Device Story
Milestone Spinal System consists of lumbar cages manufactured from INVIBIO PEEK-OPTIMA LT1; designed for interbody fusion in lumbar spine. Cages feature hollow geometry for autogenous bone graft packing; angular teeth for endplate grip; tantalum markers for radiographic visualization. Implanted by surgeons via open, minimally invasive posterior, or transforaminal approaches. Used in conjunction with FDA-cleared supplemental fixation instrumentation. Provides structural support and correction during fusion procedures. Benefits patients by facilitating spinal fusion in DDD cases.
Clinical Evidence
No clinical data provided. Substantial equivalence supported by bench testing including ASTM F2077-03 (mechanical testing for intervertebral body fusion devices), ASTM F88/F88M-09 (seal strength), and USP <85> (bacterial endotoxin limits).
Indicated for skeletally mature patients with degenerative disc disease (DDD) at one or two levels from L2 to S1, with or without Grade I Spondylolisthesis or retrolisthesis, who have failed six months of non-operative treatment. Requires use with autogenous bone graft and supplemental fixation.
Regulatory Classification
Identification
An intervertebral body fusion device is an implanted single or multiple component spinal device made from a variety of materials, including titanium and polymers. The device is inserted into the intervertebral body space of the cervical or lumbosacral spine, and is intended for intervertebral body fusion.
Special Controls
*Classification.* (1) Class II (special controls) for intervertebral body fusion devices that contain bone grafting material. The special control is the FDA guidance document entitled “Class II Special Controls Guidance Document: Intervertebral Body Fusion Device.” See § 888.1(e) for the availability of this guidance document.(2) Class III (premarket approval) for intervertebral body fusion devices that include any therapeutic biologic (e.g., bone morphogenic protein). Intervertebral body fusion devices that contain any therapeutic biologic require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.
{0}------------------------------------------------
K133053 Page 1 of 4
# JUL 1 5 2014
### 510(k) Summary
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
The assigned 510(k) Number: __________________________________________________________________________________________________________________________________________________
- 7.1 Date of Submission
07/14/2014
- 7.2 Sponsor Identification
### Weigao Orthopaedic Device Co., Ltd.
No 26 Xiangjiang Road, Tourist Resorts Weihai, Shandong, 264203, China
Establishment Registration Number: 3006639944
Contact Person: Han Wang Position: Quality & Technique Manager Tel: +86-631-5788966 Fax: +86-631-5660958 Email: wanghan@wegortho.com
7.3 Designated Submission Correspondent
Ms. Diana Hong & Mr. Lee Fu
### Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China
Tel: +86-21-22815850, Fax: 240-238-7587 Email: info@mid-link.net
{1}------------------------------------------------
#### 7.4 Identification of Proposed Device
Trade Name: Milestone™ Spinal System Common Name: Intervertebral Body Fusion Device
Regulatory Information: Classification Name: Intervertebral body fusion device Classification: II; Product Code: MAX: Regulation Number: 21 CFR 888.3080; Review Panel: Orthopedic;
#### Intended Use Statement:
The Milestone™ Spinal System is indicated for interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade I Spondylolisthesis at the involved level. These patients should be skeletally mature and have had six months of non-operative treatment. These implants may be implanted via an open or a minimally invasive posterior approach. Alternatively, these implants may also be implanted via transforaminal approach. These implants are to be used with autogenous bone graft.
DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radiographic studies.
These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
### 7.5 Device Description
The Milestone Spinal System consists of INVIBIO PEEK-OPTIMA LT1 lumbar cages of various lengths and heights, which can be inserted between two lumbosacral vertebral bodies to give support and correction during lumbar interbody fusion surgeries. They are designed with angular teeth to allow the implant to grip the superior and inferior end plates. The hollow geometry of the implants allows them to be packed with autogenous bone graft in lumbar interbody fusion procedures. The cages contain radiographic tantalum markers used for both intra and post-operative positioning and visualization. The cages are provided sterile. The Milestone Spinal System also includes various device specific instruments which are provided non-sterile.
### 7.6 Identification of Predicate Device
510(k) Number: K073291 Product Name: CAPSTONE® Spinal System Manufacturer: Medtronic Sofamor Danek USA, Inc.
{2}------------------------------------------------
### 7.7 Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
- 公 1SO11137-2:2012, Sterilization of healthcare products-Radiation-Part 2: Establishing the sterilization dose:
- ア USP <85> Bacterial Endotoxin Limit;
- ASTM F88/F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials; ア
- 公 ASTM F 2077-03 Test Methods For Intervertebral Body Fusion Devices
- 7.8 Clinical Test Conclusion
No clinical study is included in this submission.
- 7.9 Substantially Equivalent (SE) Comparison
| Item | Proposed Device(s) | Predicate Device(s) |
|-------------------|---------------------------------------|---------------------------------------|
| Product Code | MAX | MAX |
| Regulation Number | 21 CFR 888.3038 | 21 CFR 888.3038 |
| Intended Use | The Milestone™ Spinal System is | The CAPSTONE® Spinal System is |
| | indicated for interbody fusion with | indicated for with autogenous bone |
| | autogenous bone graft in patients | graft in patients with degenerative |
| | with degenerative disc disease | disc disease (DDD) at one or two |
| | (DDD) at one or two levels from | contiguous levels from L2 to S1. |
| | L2 to S1. These DDD patients may | These DDD patients may also have up |
| | also have up to Grade I | to Grade I Spondylolisthesis or |
| | Spondylolisthesis or retrolisthesis | retrolisthesis at the involved level. |
| | at the involved level. These patients | DDD is defined as discogenic back |
| | should be skeletally mature and | pain with degeneration of the disc |
| | have had six months of | confirmed by history and radiographic |
| | non-operative treatment. These | studies. These patients should be |
| | implants may be implanted via an | skeletally mature and have had six |
| | open or a minimally invasive | months of non-operative treatment. |
| | posterior approach. Alternatively, | These implants may be implanted via |
| | these implants may also be | an open or a minimally invasive |
| | implanted via transforaminal | posterior approach. Alternatively, |
| | approach. These implants are to be | these implants may also be implanted |
| | used with autogenous bone graft. | via an anterior and/or |
| | | transformational approach. These |
Table 6-1 Comparison of Technology Characteristics
3
{3}------------------------------------------------
# K133053 Page 4 of 4
4 .
| | | DDD is defined as discogenic back<br>pain with degeneration of the disc<br>confirmed by history and<br>radiographic studies. | implants are intended to be used with<br>supplement fixation instrumentation,<br>which has been cleared by the FDA<br>for use in lumbar spine. |
|--------------------------------|--------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------|
| | | These devices are intended to be<br>used with supplemental fixation<br>instrumentation, which has been<br>cleared by the FDA for use in the<br>lumbar spine. | |
| Sterile | | Radiation Sterilized | Radiation Sterilized |
| Single Use | | Yes | Yes |
| Shelf Life | | 4 years | 4 years |
| Performance | | Comply with ASTM F2077-03 | Comply with ASTM F2077-03 |
| Features | | a convex, bullet nose design | a convex, bullet nose design |
| | | angular teeth | angular teeth |
| | | hollow geometry | hollow geometry |
| Patient<br>contact<br>material | Cage | INVIBIO PEEK-OPTIMA LTI | PEEK |
| | Instrument | 630 Stainless Steel | 630 Stainless Steel |
| | radiographic | tantalum | tantalum |
| | maker | | |
| Physical | Height | 8mm, 10mm, 12mm, 14mm | 8mm, 10mm, 12mm, 14mm |
| specification | Length | 22mm, 26mm, 32mm, 36mm | 22mm, 26mm, 32mm, 36mm |
7.10 Substantially Equivalent (SE) Conclusion
Based on the comparison above, the proposed device is determined to be Substantially Equivalent (SE) to the predicate device.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 15, 2014
Weigao Orthopaedic Device Company, Limited % Ms. Diana Hong Mid-Link Consulting Company, Limited P.O. Box 120-119 Shanghai, 200120. CHINA
Re: K133053
Trade/Device Name: Milestone Spinal System Regulation Number: 21 CFR 888.3080 Regulation Name: Intervertebral body fusion device Regulatory Class: Class II Product Code: MAX Dated: June 13, 2014 Received: June 16, 2014
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical
{5}------------------------------------------------
Page 2 - Ms. Diana Hong
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803). please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours.
# Ronald P. Jean -S for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K133053
Device Name Milestone Spinal System
Indications for Use (Describe)
The Milestone Spinal System is interbody fusion with autogenous bone graft in patients with degenerative disc disease (DDD) at one or two levels from L2 to S1. These DDD patients may also have up to Grade I Spondylolisthesis or retrolisthesis at the involved level. These patients should be skeletally mature and have had six months of non-operative treatment. These implanted via an open or a minimally invasive posterior approach. Alternatively, these implanted via transforaminal approach. These implants are to be used with autogenous bone graft.
DDD is defined as discogenic back pain with degeneration of the disc confirmed by history and radios.
These devices are intended to be used with supplemental fixation instrumentation, which has been cleared by the FDA for use in the lumbar spine.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FOR FOR FDA USE ONLY - CARRET PLATE 11:48:00:00
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Anton E. Dmitriev. PhD
# Division of Orthopedic Devices
This section applies only to requirements of the Paperwork Reduction Act of 1995.
### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.