The One Plus Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. 3.0mm diameter implants are intental or lateral mandibular incisors. The One Plus Implant System is intended for single use only. It is for delayed loading.
Device Story
One Plus Implant System consists of endosseous dental implants surgically placed into upper or lower jaw bone; intended to support prosthetic artificial teeth. Device fabricated from Titanium 6AL 4V ELI Gr.23 alloy; features RBM-treated fixture body and TiN-coated abutment. Available in multiple diameters (3.00mm–5.25mm) and lengths (11.5mm–14.5mm). Used by dental clinicians in clinical settings for delayed loading protocols. Provides stable foundation for prosthetic restoration; restores masticatory function in partially edentulous patients.
Clinical Evidence
Bench testing only. Evidence includes sterilization validation per ISO 11137-1/2/3, accelerated shelf-life testing per ASTM F1980-07 and ISO 11607-1, and chemical/SEM analysis to verify absence of residuals post-RBM treatment.
Indicated for partially edentulous patients requiring missing tooth replacement to restore mastication. 3.0mm diameter implants are indicated for central or lateral mandibular incisors. Intended for single use and delayed loading.
Regulatory Classification
Identification
An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.
Special Controls
*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2)
*Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use;
(ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads;
(iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system;
(iv) The device must be demonstrated to be biocompatible;
(v) Sterility testing must demonstrate the sterility of the device;
(vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment;
(vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians;
(viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and
(ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.
S-MINI IMPLANT SYSTEM by NEOBIOTECH CO., LTD (K112540)
OSSEOFUSE DENTAL IMPLANT SYSTEM by Dynamic Innovations Inc. (K110577)
Spectra System (ScrewDirect Implant) by Implant Direct LLC (K061319)
Replace One Piece Implant by NOBEL BIOCARE UAS INC. (K023952)
Lifecore PrimaSolo One-Piece Implant System by Lifecore Biomedical, Inc.(K050506)
Submission Summary (Full Text)
{0}------------------------------------------------
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
September 19, 2014
OsseoFuse, Inc. C/O Priscilla Chung LK Consulting Group USA, Inc. 2651 East Chapman Avenue, Suite 110 Fullerton, CA 92831
Re: K133050
Trade/Device Name: One Plus Implant System Regulation Number: 21 CFR 872.3640 Regulation Name: Endossesous Dental Implant Regulatory Class: II Product Code: DZE Dated: August 20, 2014 Received: August 21, 2014
#### Dear Ms. Chung:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act
{1}------------------------------------------------
or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Mary S. Runner -S
Erin I. Keith, M.S Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page
510(k) Number (if known) K133050
Device Name One Plus Implant System
Indications for Use (Describe)
The One Plus Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. 3.0mm diameter implants are intental or lateral mandibular incisors. The One Plus Implant System is intended for single use only. It is for delayed loading.
Type of Use (Select one or both, as applicable)
× Prescription Use (Part 21 CFR 801 Subpart D)
■ Over-The-Counter Use (21 CFR 801 Subpart C)
#### PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON A SEPARATE PAGE IF NEEDED.
#### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
{3}------------------------------------------------
# 510(k) Summary
(K133050)
This summary of 510(K) - safety and effectiveness information is being submitted in accordance with requirements of 21 CFR Part 807.92.
Date: 09/17/2014
# 1. Applicant / Submitter
OsseoFuse, Inc.
5023 North Parkway Calabasas. Calabasas, CA 91302 Tel. 888.446.9995 Fax. 310.356.3183
## 2. Submission Correspondent
LK Consulting Group USA, Inc. 2651 E Chapman Ave. Ste 110, Fullerton CA 92831 Priscilla Chung Phone: 714.202.5789 Fax: 714-409-3357 Email: juhee.c(@lkconsultinggroup.com
## 3. Device
- Trade Name: One Plus Implant System
- · Common Name: Dental Implant
- Classification Name: Endosseous Dental Implant System
- · Product Code: DZE
- · Classification regulation: 21CFR872.3640
# 4. Predicate Device:
- · MS System by OSSTEM IMPLANT CO., LTD (K083067)
- · S-MINI IMPLANT SYSTEM by NEOBIOTECH CO., LTD (K112540)
- · OSSEOFUSE DENTAL IMPLANT SYSTEM by Dynamic Innovations Inc. (K110577)
- · Spectra System (ScrewDirect Implant) by Implant Direct LLC (K061319)
- · Replace One Piece Implant by NOBEL BIOCARE UAS INC. (K023952)
- · Lifecore PrimaSolo One-Piece Implant System by Lifecore Biomedical, Inc.(K050506)
{4}------------------------------------------------
# 5. Description:
The One Plus Implant System is a dental implant system made of Titanium 6AL 4V ELI Gr.23 alloy intended to be surgically placed in the bone of the upper or lower jaw arches. The system is similar to other commercially available products based on the intended use, the technology used, the material composition employed and performance chracteristics, The surface of this system has been treated with R.B.M and the abutment part has TiN coating on it.
The One Plus Implant System is available in the following sizes.
Sizes: 3.00mm (Dia.) x 11.5/13/14.5 3.75mm (Dia.) x 11.5/13/14.5 4.50mm (Dia.) x 11.5/13/14.5 5.25mm (Dia.) x 11.5/13/14.5
## 6. Indication for use:
The One Plus Implant System is intended to use in the treatment of missing teeth to support prosthetic device, such as artificial teeth, in order to restore mastication in partially edentulous patients. 3.0mm diameter implants are intended to be used in central or lateral mandibular incisors. The One Plus Implant System is intended for single use only. It is for delayed loading.
## 7. Basis for Substantial Equivalence
The subject device is substantially equivalent to the noted predicate devices based on tabulated device specifications and properties presented. Based on the comparison analysis, the identical intended use, comparable technological characteristics, and similar general design features, the subject device is substantially equivalent to the predicate devices. There are no significant differences between the One Plus Implant System and other systems currently being marketed which would adversely affect the use of the product. It is substantially equivalent to the predicate devices in design, function, material and intended use.
The comparison chart can be found on the following pages.
{5}------------------------------------------------
#### Comparison Chart
| | Subject Device | Predicate Device 1 | Predicate Device 2 | Predicate Device 3 | Predicate Device 4 |
|--------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(K) Number | K133050 | K083067 | K112540 | K110577 | K061319 |
| Device Name | One Plus Implant System | MS System | S-MINI IMPLANT<br>SYSTEM | OSSEOFUSE DENTAL<br>IMPLANT SYSTEM | Spectra System<br>(ScrewDirect Implants) |
| Applicant | OsseoFuse Co., Ltd. | OSSTEM Implant Co.,<br>Ltd. | NEOBIOTECH CO., LTD. | Dynamic Innovations Inc. | Implant Direct LLC |
| Contract<br>Manufacturer | KJ Meditech Co., Ltd. | - | - | KJ Meditech Co., Ltd. | - |
| Indications for<br>Use | The One Plus Implant<br>System is intended to use<br>in the treatment of<br>missing teeth to support<br>prosthetic device, such as<br>artificial teeth, in order to<br>restore mastication in<br>partially edentulous<br>patients. 3.0mm diameter<br>implants are intended to<br>be used in central or<br>lateral mandibular<br>incisors. The One Plus<br>Implant System is<br>intended for single use<br>only. It is for delayed<br>loading. | The MS System<br>(Denture) is intended to<br>be place in the bone of<br>the upper or lower jaw<br>arches to provide support<br>the prosthetic devices to<br>restore the patient's<br>chewing function,<br>including the denture<br>stabilization. MS System<br>(Denture) is intended for<br>single use only.<br>The MS System (Narrow<br>Ridge) is intended to use<br>in the treatment of<br>missing mandibular<br>central and lateral<br>incisors to support<br>prosthetic device, such<br>as artificial teeth, in<br>order to restore chewing<br>function in partially<br>edentulous patients. MS<br>System (Narrow Ridge)<br>is intended for single use<br>only. It is not for | The Cement type is<br>indicated for use in the<br>treatment of missing<br>maxillary lateral incisors or<br>the mandibular central and<br>lateral incisors to serve as<br>temporary support<br>prosthetic devices during<br>the healing phase of<br>permanent endosseous<br>dental implant, such as<br>artificial teeth, in order to<br>restore chewing function in<br>partially edentulous<br>patients. | The OsseoFuse Dental<br>Implant System is intended<br>for use in partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations including:<br>cemented retained, screw<br>retained, or overdenture<br>restorations, and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework. This system is<br>dedicated for one and two<br>stage surgical procedures<br>and not dedicated for<br>immediate loading. This<br>system is intended for<br>delayed loading. | The Spectra Dental<br>Implant System consists of<br>one-piece or two-piece<br>implants for single-stage or<br>two-stage surgical<br>procedures that are<br>intended for use in<br>partially or fully<br>edentulous mandibles and<br>maxillae, in support of<br>single or multiple-unit<br>restorations and terminal<br>or intermediate abutment<br>support for fixed<br>bridgework. They may be<br>placed in immediate<br>function if initial implant<br>stability can be<br>established.<br>The ScrewDirect 3.0mm<br>implant is indicated for:<br>1. An artificial root<br>structure for single tooth<br>replacement of mandibular<br>central and lateral incisors |
| | | immediate loading. | | | and maxillary lateral<br>incisors.<br>2. Multiple tooth<br>replacements or denture<br>stabilization. |
| Design | One Piece Implant | One Piece Implant | One Piece Implant | Two Piece Implant | One Piece Implant |
| Appearance | Image: One Piece Implant | Image: One Piece Implant | Image: One Piece Implant | Image: Two Piece Implant | Image: One Piece Implant |
| Material | Ti 6Al 4V ELI, Gr.23 | Titanium Alloy | Titanium Gr. 4 | Ti 6Al 4V ELI, Gr.23 | Titanium Alloy |
| Surface<br>Treatment | •RBM Treatment on the<br>fixture body<br>•TiN coating on the<br>abutment | RBM Treatment on the<br>fixture body | RBM Treatment on the<br>fixture body | •RBM Treatment on the<br>fixture body<br>•TiN coating on the<br>abutment | Roughened - HA blasted |
| Implant Sterile | Yes | Yes | Yes | Yes | Yes |
| Sterilization<br>Method | Gamma | Gamma | Gamma | Gamma | Gamma |
| Implant<br>Diameters | 3.00mm, 3.75mm,<br>4.50mm, 5.25mm | 2.50mm 3.00mm | 2.0mm, 2.5mm, 3.0mm,<br>3.5mm | 3.75mm, 4.1mm, 4.5mm,<br>5.25mm | 3.0mm, 3.7mm, 4.7mm,<br>5.7 mm |
| Implant<br>Lengths | 11.5mm, 13mm, 14.5mm | 10.0 mm, 13.0mm,<br>15.0mm | 7.0 - 15.00 mm | 8.5mm - 16.0 mm | 10mm, 13mm, 16mm |
| Product Code | DZE | DZE | DZE | DZE, NHA | DZE |
{6}------------------------------------------------
{7}------------------------------------------------
| | Predicate Device 5 | Predicate Device 6 | |
|---------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| 510(K) Number | K023952 | K050506 | |
| Device Name | Replace One Piece Implant | Lifecore PrimaSolo™ One-Piece Implant System | |
| Applicant | NOBEL BIOCARE UAS INC. | LIFECORE BIOMEDICAL, INC. | |
| Indications for Use | The Replace One Piece Implant is a threaded<br>one-piece implant with an integrated abutment,<br>designed for single-stage surgical procedure<br>and cemented restorations. The Replace One<br>Piece Implant is intended for immediate load<br>on single tooth and multiple tooth applications<br>in good quality bone, to restore chewing<br>function. | Lifecore Biomedical Dental Implant System<br>implants are intended for use in partially or fully<br>edentulous mandibles and maxillae, in support of<br>single or multiple-unit restorations including;<br>cement retained, screw retained, or overdenture<br>restorations, and terminal or intermediate<br>abutment support for fixed bridgework.<br>Specific Intended Uses: The PrimaSolo One-<br>Piece (3.5-5.0mm) Implant is a threaded one-<br>piece implant with integrated abutment designed<br>for single-stage surgical procedure and cemented<br>restorations. The PrimaSolo One-Piece Implant is<br>intended for immediate placement and can be<br>restored with a temporary prosthesis in single<br>tooth and multiple tooth applications with good<br>quality bone.<br>The PrimaSolo One-Piece (3.0mm) Implant is a<br>threaded one-piece implant with an integrated<br>abutment designed for single-stage surgical<br>procedure and is indicated for use in the treatment<br>of missing maxillary lateral incisors or the<br>mandibular central and lateral incisors to support | |
| | | prosthetic devices, such as artificial teeth, in<br>order to restore chewing function in partially | |
| | | edentulous patients. Mandible central and lateral<br>incisors must be splinted if using two or more | |
| | | 3.0mm implants adjacent to one other. | |
| Design | One Piece Implant | One Piece Implant | |
{8}------------------------------------------------
| Appearance | | |
|----------------------|----------------------|-----------------------------------|
| Material | Titanium Gr. 4 | Titanium Alloy |
| Surface Treatment | | RBM Treatment on the fixture body |
| Implant Sterile | Yes | Yes |
| Sterilization Method | | |
| Implant Diameters | 3.5mm, 4.3mm, 5.0mm, | 3.0mm, 3.5mm, 4.1mm, 5.0mm |
| Implant Lengths | 10mm, 13mm, 16mm | 10.0 mm、11.5mm 13.0mm、15.0mm |
{9}------------------------------------------------
# 8. Non-Clinical Testing
- · Sterilization validating testing has been performed in accordance with ISO 11137-1, ISO 11137-2 and ISO 11137-3.
- · Accelerated shelf life testing has been performed in accordance with ASTM1980-07, ISO 11607-1, ISO 11737-1, ISO 11737-2
- Chemical and SEM image analyses have been performed to verify that there is no residual after RBM treatment on the fixture.
There might be differences in sterilization parameters, shelf life and manufacturing processes between the subject device and the predicate devices, however, the test results supported that the subject device is substantially equivalent to the predicate devices.
# 9. Conclusion
The subject device and the predicate devices have intended use and have similar technological characteristics.
Overall, the One Plus Implant System has the following similarities to the predicate devices:
- · has the same intended use,
- · uses the same operating principle,
- · incorporates the same basic design,
- incorporates the same material and the surface treatment.
Based on the similarities, we conclude that the One Plus Implant System is substantially equivalent to the predicate devices.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.