SENSITITRE YEASTONE SUSCEPTIBILITY PLATES

K133038 · Trek Diagnostic Systems · NGZ · Jan 31, 2014 · Microbiology

Device Facts

Record IDK133038
Device NameSENSITITRE YEASTONE SUSCEPTIBILITY PLATES
ApplicantTrek Diagnostic Systems
Product CodeNGZ · Microbiology
Decision DateJan 31, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.1640
Device ClassClass 2

Indications for Use

The Sensititre® YeastOne® Susceptibility System is an in vitro diagnostic product for clinical susceptibility testing of Candida spp. This 510(k) is for Micafungin in the dilution range of 0.008 – 16 µg/ml is intended for testing Candida spp. on the Sensititre® YeastOne® Susceptibility system. The approved primary "Indications for Use" and clinical significance of Micafungin is for: Candida albicans Candida glabrata Candida krusei Candida parapsilosis Candida tropicalis

Device Story

Sensititre YeastOne is a micro-broth dilution susceptibility test for Candida spp. Device uses micro-dilution plates containing serially diluted antifungal agents and a colorimetric growth indicator. Clinical laboratory personnel inoculate plates using autoinoculator or manual methods; plates are sealed and incubated at 35°C for 24 hours. Growth is determined by visual observation of color change (blue to pink) in wells; lowest dilution maintaining no color change indicates Minimum Inhibitory Concentration (MIC). Results assist clinicians in determining antifungal susceptibility profiles to guide therapeutic decisions. Ability to detect Micafungin resistance is currently unknown due to lack of resistant strains; confirmatory testing at reference labs required for non-susceptible results.

Clinical Evidence

Bench testing only. Performance was evaluated by comparing the Sensititre YeastOne system results for Micafungin against a reference method. Testing confirmed the system's ability to provide accurate MIC determinations for the specified Candida species within the claimed dilution range.

Technological Characteristics

Micro-dilution tray format with dried antifungal medium and colorimetric growth indicator (REDOX-based). Manual reading of color change (blue to pink). Inoculation via autoinoculator or manual pipette. Standardized using nephelometer. Complies with CLSI M27-A3 and M27-S4 standards.

Indications for Use

Indicated for in vitro clinical susceptibility testing of Candida spp. (C. albicans, C. glabrata, C. krusei, C. parapsilosis, C. tropicalis) to the antifungal agent Micafungin in the concentration range of 0.008–16 µg/ml. For prescription use only.

Regulatory Classification

Identification

An antimicrobial susceptibility test powder is a device that consists of an antimicrobial drug powder packaged in vials in specified amounts and intended for use in clinical laboratories for determining in vitro susceptibility of bacterial pathogens to these therapeutic agents. Test results are used to determine the antimicrobial agent of choice in the treatment of bacterial diseases.

Predicate Devices

Submission Summary (Full Text)

{0} 1 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY ASSAY ONLY TEMPLATE A. 510(k) Number: k133038 B. Purpose for Submission: Addition of Micafungin to the Sensititre® YeastOne® Susceptibility System C. Measurand: Micafungin 0.008 -16 µg/mL D. Type of Test: Quantitative Antimicrobial Susceptibility Test (AST) growth based E. Applicant: TREK Diagnostic Systems, Inc. F. Proprietary and Established Names: Sensititre® YeastOne® Susceptibility plates G. Regulatory Information: 1. Regulation section: 866.1640 Antimicrobial Susceptibility Test Powder 2. Classification: Class II 3. Product code: NGZ Susceptibility Testing - antifungal 4. Panel: 83 Microbiology {1} H. Intended Use: 1. Intended use(s): The Sensititre® YeastOne® Susceptibility System is an *in vitro* diagnostic product for clinical susceptibility testing of *Candida* spp. This 510(k) is for Micafungin in the dilution range of 0.008 - 16 µg/ml is intended for testing *Candida* spp. on the Sensititre® YeastOne® Susceptibility system. The approved primary "Indications for Use" and clinical significance of Micafungin is for: C. albicans C. glabrata C. krusei C. parapsilosis C. tropicalis 2. Indication(s) for use: The Sensititre® YeastOne® Susceptibility System is an *in vitro* diagnostic product for clinical susceptibility testing of *Candida* spp. This 510(k) is for Micafungin in the dilution range of 0.008 - 16 µg/ml is intended for testing *Candida* spp. on the Sensititre® YeastOne® Susceptibility system. The approved primary "Indications for Use" and clinical significance of Micafungin is for: C. albicans C. glabrata C. krusei C. parapsilosis C. tropicalis 3. Special conditions for use statement(s): Prescription use only. The ability of the Sensititre YeastOne to detect resistance to Micafungin is unknown because resistant strains were not available at the time of comparative testing. For strains yielding results suggestive of a not susceptible category, organism identification and Micafungin should be retested and confirmed, and if the result is confirmed, the isolate should be submitted to a reference laboratory that will confirm results using a CLSI reference dilution method. 4. Special instrument requirements: Autoinoculator or Manual inoculation Manual readings only {2} I. Device Description: The Sensititre® YeastOne® Susceptibility system is a micro-version of the broth dilution susceptibility test. Various antifungal agents are serially diluted to concentrations bridging the range of the clinical interest in autoclaved diluent, which contains a colorimetric growth indicating compound. Each micro-dilution plate is individually packaged in foil. After inoculation, plates are sealed with an adhesive seal, incubated at 35°C for 24 hours and examined for growth. Substantial Equivalence Information: 1. Predicate device name(s): VITEK® AST-YS Fluconazole Sensititre YeastOne 2. Predicate 510(k) number(s): k061945 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate (VITEK) AST-YS Fluconazole (k061945) | | Intended Use | Susceptibility testing for colonies of Candida | Same | | Incubation | 35°C | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | Technology | Colorimetric test in which MIC's are determined by observing the lowest dilution of antimicrobial agents that maintains no color change indicating inhibition of growth. A color change from blue to pink indicates growth of organism. | The VITEK® 2 card is inoculated with a standardized organism suspension, incubated, and read throughout the incubation cycle. Results are automatically calculated once a predetermined growth threshold is reached. | | Format | Micro tray with dried Antifungal Medium | VITEK® 2 AST test card with dried antifungal | | Medium | Sensititre® yeast | VITEK® 2 Yeast Base broth | {3} | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | susceptibility inoculum broth | | | Antifungal | Micafungin | Fluconazole | | Time to Results | 24 hours | Up to 36 hours | ## K. Standard/Guidance Document Referenced (if applicable): Guidance for Industry and FDA Class II Special Controls Guidance Document: Antimicrobial Susceptibility Test (AST) Systems, August 28, 2009 CLSI M27-A3: Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts; Approved Standard - Third Edition, April, 2008 CLSI M27-S4: Reference Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts; Fourth Informational Supplement, December, 2012 ## L. Test Principle: The Sensititre YeastOne panel measures growth by colorimetric determination of REDOX state produced by the test organism rather than measuring the turbidity of the test medium. In the Sensititre panel as in most broth dilution tests the difference between wells in which growth occurs and those in which growth is inhibited is distinctive. However, the Sensititre panel produces an easily distinguished color change when light growth occurs. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: a. Precision/Reproducibility: Reproducibility was conducted at three sites on 25 yeast isolates, performed over three days. The testing was performed using Autoinoculator and manual inoculation methods with manual reading of panels. The mode was determined and then the reproducibility was calculated based on $\pm 1$ one well of the mode. The reproducibility studies for both inoculation methods demonstrated acceptable performance at $\geq 95\%$. b. Linearity/assay reportable range: Not Applicable c. Traceability, Stability, Expected values (controls, calibrators, or methods): Quality Control (QC) testing using the manual pipette inoculation was performed on {4} each day of clinical testing on the QC isolates using, C. parapsilosis ATCC 22019, and C. krusei ATCC 6258, as recommended in the CLSI standard M27. The following table represents the frequency of the results in both the reference method and the Sensititre YeastOne Susceptibility plates and the acceptable range. Quality Control was also performed at three sites using the Autoinoculator method. All results, including reference method, were read manually at 24 hrs. All results were within the specified range. Quality Control Table | ORGANISM | Conc μg/mL | Manual Inoculation Method | | Autoinoculator Method | | --- | --- | --- | --- | --- | | | | Reference | Sensititre® YeastOne® | Sensititre® YeastOne® | | C. parapsilosis | 0.25 | | | | | ATCC 22019 | 0.5 | 1 | 20 | 53 | | Expected Range : 0.5 – 2 μg/mL | 1 | 51 | 38 | 7 | | | 2 | 8 | 2 | 0 | | | | | | | | C. krusei | 0.12 | 54 | 59 | 60 | | ATCC 6258 | 0.25 | 6 | 1 | 0 | | Expected Range : 0.12 – 0.5 μg/mL | 0.5 | 0 | 0 | 0 | | | | | | | Nephelometer was used at each site to standardize the inoculum. Colony counts from QC ATCC source were performed using direct inoculum method. d. Detection limit: Not Applicable e. Analytical specificity: Not Applicable f. Assay cut-off: Not Applicable 2. Comparison studies: a. Method comparison with predicate device: Performance was established on the Sensititre YeastOne Susceptibility System for Candida spp. at three clinical sites. The CLSI reference method as described in the CLSI document M27 "Method for Broth Dilution Antifungal Susceptibility Testing of Yeasts" was compared to the Sensititre YeastOne Susceptibility System. A total of 220 fresh clinical and 95 challenge isolates of Candida species were tested on the {5} Sensititre YeastOne using manual and inoculation methods. All clinical isolates demonstrated growth in the growth control well. EA was calculated when the results for the reference method and the Sensititre YeastOne were within +/- two doubling dilutions of the antifungal drug. The following tables present the performance of the Sensititre YeastOne Susceptibility System when read at 24 hours using manual inoculation methods as compared to the reference method when read at 24 hours. The results demonstrate acceptable performance. Performance Summary read manually at 24 hours (Manual Inoculation) | | EA Total | EA N | EA % | Eval EA Total | Eval EA N | Eval EA % | CA N | CA % | #R | min | maj | vmj | | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | --- | | C. albicans, C. tropicalis, C. krusei (S=≤0.25, I=0.5, R=≥1 μg/mL) | | | | | | | | | | | | | | Clinical | 145 | 145 | 100 | 96 | 96 | 100 | 145 | 100 | 0 | 0 | 0 | 0 | | Challenge | 61 | 61 | 100 | 33 | 33 | 100 | 61 | 100 | 0 | 0 | 0 | 0 | | Combined | 206 | 206 | 100 | 129 | 129 | 100 | 206 | 100 | 1 | 0 | 0 | 0 | | C. glabrata (S=≤2, I=4, R=≥8 μg/mL) | | | | | | | | | | | | | | Clinical | 15 | 15 | 100 | 8 | 8 | 100 | 14 | 93.3 | 2 | 1 | 0 | 0 | | Challenge | 24 | 24 | 100 | 3 | 3 | 100 | 24 | 100 | 0 | 0 | 0 | 0 | | Combined | 39 | 39 | 100 | 11 | 11 | 100 | 38 | 97.4 | 2 | 1 | 0 | 0 | | C. parasilopsis (S=≤0.06, I=0.12, R=≥0.25 μg/mL) | | | | | | | | | | | | | | Clinical | 60 | 60 | 100 | 60 | 60 | 100 | 60 | 100 | 0 | 0 | 0 | 0 | | Challenge | 10 | 10 | 100 | 10 | 10 | 100 | 10 | 100 | 0 | 0 | 0 | 0 | | Combined | 70 | 70 | 100 | 70 | 70 | 100 | 70 | 100 | 0 | 0 | 0 | 0 | EA - Essential Agreement CA - Category Agreement R-resistant isolates maj-major discrepancies vmj-very major discrepancies min-minor discrepancies Category agreement (CA) is when the Sensititre panel result interpretation agrees exactly with the reference panel result interpretation. Evaluable EA is when the MIC result is on scale for both the Sensititre and the reference and have on-scale EA. b. Matrix comparison: Not Applicable 3. Clinical studies: a. Clinical Sensitivity: Not Applicable b. Clinical specificity: Not Applicable {6} c. Other clinical supportive data (when a. and b. are not applicable): Not Applicable 4. Clinical cut-off: Not Applicable 5. Expected values/Reference range (MIC values in $\mu \mathrm{g} / \mathrm{mL}$ ): | Microorganism(s) | Susceptibility Interpretive Criteria | | | | --- | --- | --- | --- | | | S | I | R | | C. albicans, C. tropicalis, and C. krusei | ≤0.25 | 0.5 | ≥1 | | C. parasilosis: | ≤2 | 4 | ≥8 | | C. glabrata: | ≤0.06 | 0.12 | ≥0.25 | The ability of the Sensititre YeastOne to detect resistance to Micafungin is unknown because resistant strains were not available at the time of comparative testing. For strains yielding results suggestive of a not susceptible category, organism identification and Micafungin should be retested and confirmed, and if the result is confirmed, the isolate should be submitted to a reference laboratory that will confirm results using a CLSI reference dilution method. # N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. # O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...