K133031 · Industrial Water Solutions, Inc. · MMK · Nov 15, 2013 · General Hospital
Device Facts
Record ID
K133031
Device Name
SHARPSOLOGY 2
Applicant
Industrial Water Solutions, Inc.
Product Code
MMK · General Hospital
Decision Date
Nov 15, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.5570
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The container is intended for use in healthcare settings. The container is designed to safely contain sharps prior to removal from the generating facility, during transportation and until ultimate treatment and disposal of the sharps. The container features a canteen style, vertical drop lid or a mailbox style, vertical drop lid. The container is housed in a wall mount or bracket attached to the wall. The wall mount is made of plastic and the bracket is made of stainless steel. The container is translucent and can be selected in a natural or red color. The lid is a natural color. The container volume is 2.25 gallons. The dimensions of the container are 13.173 inches wide, 11.417 inches high, and 5.993 inches in depth ( front to back ). The lid surface dimensions for the canteen style lid are 6.197 inches wide and 13.031 inches long. The dimensions for- the mailbox style lid are 6.194 inches wide and 13.029 inches long. The diameter of the lid aperture for the canteen style, vertical drop lid is 2.135 inches. The opening of the lid aperture for the mailbox, vertical drop lid is 2.5 inches by 9.5 inches. The dimensions of the wall mount are 12.25 inches wide, 16.625 inches high and 6.25 inches in depth. (front to back). The dimensions of the bracket are 12,75 inches wide, 17.125 inches high and 7 inches in depth. (front to back). The wall mount is available from Industrial Water Solutions, Inc ..
Device Story
Sharpsology 2 is a reusable 2-gallon plastic sharps disposal container; used in healthcare settings to collect needles and medical waste. Device consists of a container base, wall mount/bracket, and two interchangeable lid types (canteen-style or mailbox-style vertical drop). Healthcare personnel deposit sharps through the lid aperture; sharps fall to the bottom. Container features a visible 'Fill Line' for monitoring capacity. When full, the container is removed from the wall mount, sealed with a cap or locking mechanism, and transported to a facility for emptying, cleaning, and sanitization. The device is constructed from high-density polypropylene specialty impact copolymer (container/canteen lid) and 'No Break' polyethylene (mailbox lid). It is puncture-resistant, leak-resistant, and stable. Reusability offers environmental and cost benefits over disposable alternatives. Device is designed for a 5-year lifespan. No needles are broken or sheared; no 'unwinders' are included. Output is the safe containment of sharps during storage and transport.
Clinical Evidence
Bench testing only. Performance testing included: 49 CFR 178.603 (Free Fall Drop), 49 CFR 178.606 (Stacking), 49 CFR 178.608 (Vibration), Health Devices 22 (Leak Resistance, Puncture Resistance, Freestanding Capability). Simulated life-use testing (105 cycles of use, washing, and sanitizing) showed no failure. Puncture force recorded at 10.15 lbf (exceeding 3.37 lbf requirement).
Technological Characteristics
Materials: Polypropylene Specialty Impact Copolymer (container, canteen lid), 'No Break' Polyethylene (mailbox lid), ABS (wall mount). Dimensions: 13.173" x 11.417" x 5.993". Volume: 2.25 gallons. Features: Vertical drop aperture, bayonet-style lid closure, wire tie security. Standards: 49 CFR 178.603/606/608, OSHA 29 CFR 1910-1030, ANSI standards. Reusable; non-sterile.
Indications for Use
Indicated for healthcare professionals in healthcare settings for the safe collection and containment of medical sharps prior to removal, transport, and disposal.
Regulatory Classification
Identification
A hypodermic single lumen needle is a device intended to inject fluids into, or withdraw fluids from, parts of the body below the surface of the skin. The device consists of a metal tube that is sharpened at one end and at the other end joined to a female connector (hub) designed to mate with a male connector (nozzle) of a piston syringe or an intravascular administration set.
{0}------------------------------------------------
# NOV 1 5 2013
K13303
#### 510(k) SUMMARY
Industrial Water Solutions, Inc.
2692 Willowlawn Street Roanoke, Virginia 24018
540/772-6341 540/772-7944
July 14, 2013
Thomas P. Stewart
## 807.92(a)(1)
COMPANY NAME:
COMPANY ADDRESS:
TELEPHONE NUMBER: FACSIMILE NUMBER:
CONTACT PERSON:
SUMMARY PREPARATION DATE:
## 807.92(a)(2)
TRADE or PROPRIETARY NAME: Sharpsology 2 COMMON NAME: Reusable Sharps Disposal Container CLASSIFICATION: Class II - Accessory Device Regulation Number 880 5570 CLASSIFICATION NAME: Product Code
## 807.92(a)(3)
EQUIVALENT DEVICE #2:
EQUIVALENT DEVICE #3
CONTAINER LID .....CANTEEN STYLE
(510(k) #K992626
EQUIVALENT DEVICE #1: Sure-Way Sharps Container lid CONTAINER LID ... ... MAILBOX STYLE The Sure-Way Sharps Container lid ("lid") is used on the top of the Sure-Way Sharps Container and is a Mailbox Style lid.. The Sure-Way Sharps Container lid has been approved under 510(k) #K992626.
MMK
Sharps Away Vertical Drop Lid
The Sharps Away Vertical drop lid is used on top of the Sharps Away Reusable Sharps Container and is a canteen style lid. The Sharps Away Vertical Drop Lid has been approved under 310(k) # 072667
Sure-Way Reusable Sharps Container
The Sure-Way Sharps Container has been approved under 510(k) #K992626. It is described as a reusable plastic container into which used needles and other medical waste are discarded. It is a nonsterile device for nonsterile applications.
807.92(a)(4)
CONTAINER BASE:
DESCRIPTION:
Industrial Water Solutions, Inc. will market and sell the Sharpsology 2 container, wall mount and lids. Industrial Water Solutions, Inc. does not own or operate any dumping, cleaning, processing, or hauling operations for any medical waste, including sharps medical waste.
This premarket notification is for a 2 gallon container, wall mount and (2) two lids only. This is the only container to be manufactured at this time. Only one size and one model.
All wash equipment, scanning software, barcode readers and any other equipment or software is solely the responsibility of the end-user to purchase and use.
510(k) Summary Page 1 of 6
{1}------------------------------------------------
The Sharpsology 2 is a reusable plastic sharps disposal container. The container is a nonsterile device for nonsterile applications. The container is intended to be used in an appropriate wall mount unit, wall bracket or stand alone tray. It features a canteen style, vertical drop lid and a mailbox style, vertical drop lid for safe and easy disposal of sharps. Simply deposit syringes vertically into the opening of the lid, and they are deposited automatically into the container. Container is easily identified as full by looking at both the fill line on the container and biohazard label attached to container. When the canteen style lid is being used and the container is full, the supplied cap is attached to the lid by turning clockwise until it snaps into place for transport. When using the mailbox style lid and the container is full, pull the rotating portion of lid forward and snap lock the rotating portion into the two locking tabs located on the front of the lid. The Sharpsology 2 has a hole in one end of the lid and container that match, a plastic cable wire tie is inserted to look the lid to the container for transportation. Then just replace it with a clean sanitized unit, Using proper cleaning and sterilizing equipment, cleaning facility will empty and sanitize the unit and return it to the user facility to use for the next switch out. The durable high-density polypropylene specialty impact copolymer construction of the Sharpsology 2 makes it leak resistant, puncture resistant and stable. It meets or exceeds all OSHA recommendations for sharps containers, and because it is reusable, it is more environmentally friendly and less expensive than disposable alternatives. The projected life expectancy of the device is at least (5) five years. Other sharps containers using the same material for the same purpose have remained in use over (5) five years without showing failure. The container is intended to be used in healthcare settings. The container is designed to safely contain sharps prior to removal from the generating facility, during transportation and until ultimate treatment and disposal of the sharps.
The Sharpsology 2 container is inserted into a wall mount unit available from Industrial Water Solutions, Inc. that is lockable and allows access to the opening of the lid of the Sharpsology 2 container for needle disposal. Healthcare personnel drop used sharps into the opening of the lid of the container, Sharps drop to the bottom of the container. When the container is full or up to the "Fill Line" of the container, the container is then taken out of the wall mount or bracket on the wall. The label on the container has very visible "Fill Line" imprinted on the outside. The container also has a very visible fill line, as well. A new container or one that has been sanitized for reuse is then put back into the wall mount or bracket for use. The container is then ready for transport to the cleaning facility for sanitation.
A label will be attached to the container specifying the intended use of the container. A sample of this label is attached at the end of this document. (See Attachment 1)
> The Sharpsology 2 container is a reusable two gallon sharps container used for the collection of sharps at the healthcare facility. The container is housed in a wall mount or bracket attached to the wall. Healthcare personnel simply drop used sharps into the vertical drop opening on top of the lid and release. The used sharps deposited into the container and drop to the bottom of the container. The container is used until it is deemed to be full by visually inspecting to see if the sharps level is up to the "Fill Line". Once the container is deemed to be full, it is then removed from the wall mount or bracket and gotten ready for transport. A new container or a used, sanitized container is then placed in the wall mount or bracket for use, The Sharpsology 2 container has no handles, but is easily picked up by placing hands on either side of the container and lifting. The average weight of the container plus sharps inside a full container will average five and seven pounds. The used or full containers are then transported to the cleaning and sanitizing facility for sanitation, inspection, packaging, and made ready for reuse. The container is lockable. Each container has a bayonet style lid closure that horizontally slides into place and is removed mechanically for emptying and cleaning. The lid features a wire tie to further help hold the lid in place.
> > 510(k) Summary Page 2 of 6
{2}------------------------------------------------
The scientific concept for the Sharpsology 2 container was a composite of current containers on the market and what improvements might be made. The Sharpsology 2 container is made of a Polypropylene Specialty Impact Copolymer with a superior impact and puncture resistant qualities. The container itself is very similar to many containers already on the market. The lid, as well, is very similar to current market models already in use. The device is puncture resistant. The device is leak resistant. The device is labeled in accordance with OSHA requirements and ANSI standards. A biohazard label is permanently affixed to the front and back of the container by adhesive backing. The container features (4) four raised legs for stability when used in a free standing mode. The device is capable of maintaining a stable upright position during use in a free standing mode. These four feet are located on the underside of the container at all four corners.
The container does not incorporate any \features to break, bend, or shear needles. The container does not incorporate any "unwinders".
The dimensions of the container are 13.173 inches wide, 11.417 tall and 5.993 inches in depth (front to back).
The dimensions of the wall mount are 16.625 inches high, 12.25 inches wide and 6.75 inches in depth. (Front to Back)
The nominal thickness of the container is 0.122 inches. The tare weight of the container is 1.68 pounds. The volume of the container 2.25 gallons up to the fill line. The color of the container and lid is natural or red . The lids are both vertical drop lids with an opening of 2.135 inches diameter for the canteen style lid and 2.5 inches by 9.5 inches for the mailbox style lid. The dimensions of the lid are 13.029 inches long by 6.194 inches wide (front to back). The nominal thickness of the lid is . 138 inches
If the canteen style lid is being used with the container, the locking cap to be used is 2.135 inches in diameter and is made from a very durable ABS material. The measurements of the base of the predicate device and the Sharpsology 2 container are entirely the same.
The material used for the container and the canteen style lid is Polypropylene Specialty Impact Copolymer. The container and lid color is a natural or red. The material used for the mailbox style lid is a "No Break" Polyethylene. Due to the structural dynamics of the mailbox style, this material gives far more integrity to the lid and prevents breakage.
The wall mount is made of ABS material, making it very durable and impact resistant. The body of the wall mount is natural and the door is clear.
Both the container and lid are natural or red in color, which allows the healthcare personnel to easily see the amount of sharps in the container to determine how full the container is at any time. The cap for the canteen style lid is also a natural color. The container is translucent and either the natural color or red allows healtheare personnel to easily see the amount of sharps in the container at any given time. The red container is commonly used in the industry and allows virtually the same ability to see sharps level as the natural color at any time. The predicate device canteen style lid is manufactured by Solutions, Inc. They have many containers in different sizes that come in the red color. Other manufacturers of sharps containers also have several red containers available.
Both lids have openings and design that prevents anyone from putting a hand into the container, thus preventing needle stick.
510(k) Summary Page 3 of 6
{3}------------------------------------------------
## 807.92(a)(5)
INTENDED USE:
The container is intended for use in healtheare settings. The container is designed to safely contain sharps prior to removal from the generating facility, during transportation and until ultimate treatment and disposal of the sharps. The container features a canteen style, vertical drop lid or a mailbox style, vertical drop lid. The container is housed in a wall mount or bracket attached to the wall. The wall mount is made of plastic and the bracket is made of stainless steel. The container is translucent and can be selected in a natural or red color. The lid is a natural color. The container volume is 2.25 gallons.
The dimensions of the container are 13.173 inches wide, 11.417 inches high, and 5.993 inches in depth ( front to back ). The lid surface dimensions for the canteen style lid are 6.197 inches wide and 13.031 inches long. The dimensions for- the mailbox style lid are 6.194 inches wide and 13.029 inches long. The diameter of the lid aperture for the canteen style, vertical drop lid is 2.135 inches. The opening of the lid aperture for the mailbox, vertical drop lid is 2.5 inches by 9.5 inches.
The dimensions of the wall mount are 12.25 inches wide, 16.625 inches high and 6.25 inches in depth. (front to back) The dimensions of the bracket are 12,75 inches wide, 17.125 inches high and 7 inches in depth. (front to back). The wall mount is available from Industrial Water Solutions, Inc ..
## 807.92(a)(6)
The Sharpsology 2 reusable contained is equivalent to the Sure-Way Reusable Sharps Container and is constructed of a translucent, natural colored Polypropylene Specialty Impact Copolymer.
It is injection molded and varies in thickness on average 0.002". The base of the container is similar in design and ergonomic characteristics to the Sure-Way Reusable Sharps Container. The Sharpsology 2 unit is manufactured with a thick wall (0.122" ± 0.002" ) and stronger Polypropylene Specialty Impact Copolymer to help confer reusable qualities. A side by side comparison linds very little significant differences in the two devices.
There are no significant functional differences between the Sharpslogy 2 container. The predicate lid for the Sure-Way container is a mailbox style lid and has a hole in the center of the lid to vertically drop sharps after use for disposal. The Sharpsology 2 mailbox style lid also has a vertical drop for disposal of sharps after use. The second lid available is a catten style, vertical drop lid. The predicate canteen style lid is the Shaps Away Vettical Drop Lid from Solutions, Inc. The Sharps Away cancen style lid is a vertical drop lid which allows disposal of sharps by pulling a plug that is in the hole atop of the Sharpsology 2 canteen style lid is a vertical drop lid that has a cap that son of the lid to close it and unscrews for safe disposal of sharps after use. Both predicate lids are injected molded. The lid and container base assembly are designed to be compatible with the mechanical dumping and washing equipment. The Sharpsology 2 container has the same intended use, principles of operation, and technological characteristics as it's predicate device. No new issues of safety or effectiveness are raised. The Surc-Way container intended use says that it is to be used for the disposal of contaminated medical sharps in healtheare facilities. The Sharpsology 2 is virtually the same. The intended use for the Sharpsology 2 is said to be in healthcare settings. Pictures of both containers and lids are attached. (See attachment no.2)
| COMPARISON TABLE (lid) | Sure-Way Reusable Sharps Lid<br>......Mailbox Style<br>Sharps Away Reusable Sharps<br>Lid......Canteen Style | Sharpsology 2 Reusable Sharps Lid<br>Mailbox Style<br>Canteen Style |
|------------------------|---------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------|
| Indications for use | Healthcare Sharps | Same |
| Target population | Healthcare Professionals | Same |
| Design | Vertical Drop | Same |
| Materials | Mailbox Style......High density<br>Polyethylene (HDPE)<br>Canteen Style.... High density<br>Polyethylene HDPE | Mailbox Style...."No Break" Polyethylene<br>Canteen Style... Polypropylene Specialty<br>Impact Copolymer |
| Performance | Multiple Use | Same |
| Mechanical safety | Vertical Drop | Same |
| Where used | Healthcare facilities/labs | Same |
| Standards met | 49 CFR/HD 22 (multi-use) | Same |
510(k) Summary Page 4 of 6
{4}------------------------------------------------
The Sharpsology 2 reusable container and lid are substantially equivalent to the Sure-Way Reusable Sharps Container and lid The Sharpsology 2 canteen style lid and the Sharps Away canteen style lid are substantially equivalent.
| COMPARISON TABLE (container) | Sure-Way Reusable Sharps | Sharpsology 2 Reusable Sharps | |
|------------------------------|----------------------------------|---------------------------------------------|--|
| | Container | Container | |
| Indications for use | Healthcare Sharps | Same | |
| Target population | Healthcare Professionals | Same | |
| Design | Injection mold tapered | Same | |
| Materials | High density Polyethviene (HDPE) | Polypropylene Specialty Impact<br>Copolymer | |
| Performance | Multiple Use | Same | |
| Where used | Healthcare facilities/labs | Same | |
| Standards met | 49 CFR/HD 22 (multi-use) | Same | |
## 807.92(b)(1)
-2.7.10 -
Over a two-day period from August 20, 2013, fifteen (15) red colored Sharpsology 2 containers with canteen style lids and fifteen (15) red colored Sharpsology 2 containcrs with mailbox style lids were subjected to a simulated life-use of 105 cycles of closing, filling, depositing wase, opening, washing and santizing. The red Sharpsology 2 containers were from the same lot that was tested by Arama Package and Container Testing Services, Inc. in July, 2013 (Test report dated July 24, 2013). After simulated use, there were no visible signs of failure. Containers were then performance tested and passed. The labels withstood temperatures of 180 degrees, washing with cleaners at 160 degrees rinsing at 180 degrees and wiping without any degrading at all. { See attachment no.3)
## 807.92(b)(2) and (3)
The Sharpsology 2 meets and exceeds the primary design characteristics needed to comply with the OSHA Blood borne Pathogens Standard. Date for the following tests have been provided and are as follows:
| Free Fall Drop | 49 CFR 178.603 | 5 drops from 4 feet | Pass |
|-------------------------|-------------------|----------------------------|------|
| Stacking | 49 CFR 178.606 | 24 Hours under 80 lbs. | Pass |
| Vibration | 49 CFR 178.608 | 1 hour repetitive bounce | Pass |
| Leak Resistance | Health devices 22 | 24 hours filled with water | Pass |
| Puncture | Health Devices 22 | Needle penetration force | Pass |
| Freestanding Capability | Health Devices 22 | Full, ½ Full and 5% Full | Pass |
"The package, as submitted and tested, visually appears to satisfy the test criteria and is capable of preventing the loss or dispersal of the contents for conditions normal to transport" (Container Testing Laboratory, Inc.) (See Attachment 2).
The impact resistance of the Sharpsology 2 container was determined in accordance with the U.S. Deparment of Transportation "Drop Test" (49 CFR 178.603) A laboratory test report is attached.
Mcthod: The packages were filled for testing to 95% capacity with attached needles. Five (5) filled packages, closed for shipment, were immediately subjected to (5) five free fall drops each from a height of 3.9 feet (1.2 meters) onto a solid concrete floor from several and all angles.
Results; The Sharpsology 2 container met with the requirements set forth in 49 CFR 178.603 when using polypropylene wire ties for the lids. The containers did not exhibit any signs of damage from the test. No Ruptures/No leakage.
The puncture resistance of the Sharpsology 2 contained in accordance with proposed "Safe Disposal of Medical Waste Sharps", Health Devices 122, Nos. 8-9, August-September 1993, Pages 379-380.
Method: A sample of several sections of containers were taken, including the thinness areas of the container was subjected to the penetrations and the puncture force was recorded.
510(k) Summary Page 5 of 6
{5}------------------------------------------------
. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Results: All samples net the requirements of the "Safe Disposal of Medical Waste Sharps", Health Devices 122, Nos. 8-9, August-September 1993, Pages 379-380. The minimum puncture force (10.15 lbt) recorded exceeds the required minimum of 3.37 lbf.
The leak resistance was determined in accordance with proposed "Sate Disposal of Medical Waste Sharps", Health Devices 122, nos. 8-9, August-September 1993, Pages 383-384.
Method: Three (3) containers were placed on a level tabletop and filled with water to the highest point that would retain water as it was being poured to them. The water-filled units were left in the upright position undisturbed for 24 hours,
Results: All containers passed with no leakage/no liquid loss.
The vibration test was determined in accordance with 49 CFR Section 178.608.
Method: Three (3) samples filled with disposable hypodermics with needles, closed for shipment, were placed on a vibration platform and vibrated in normal shipping orientation for one (1) hour ay 235 CPM. Immediately thereatler, the containers were removed and examined.
Results: All containers passed with No leakage/no rupture.
The Stacking Test was determined in accordance with 49 CFR Section 178.606.
Method: Three (3) containers were filled with disposable hypodermic with attached needles, closed as for shipment and subjected to a static compression load of 78.48 pounds, equivalent to a 10 feet high stack of identical packages, continuously for 24 hours.
Results: All containers passed with No rupture/No leakage,
The Sharpsology 2 Reusable Sharps Container was found to be very stable of being let in a free standing mode. All containers remain stable without tipping in accordance with OSHA specifications 29 CFR 1910-1030,
#### 807.93 (a)(1)
CONCLUSION: The Sharpsology 2 container met or exceeded all the above tests performed. All of the above less are standard for the FDA requirements to meet 510(k) approval. Therefore, this container has been found to be safe and effective for use as a vessel for the handling and disposal of sharps waste(i.e. needles). The Sharpsology 2 container has been found capable of preventing the loss or dispersal of the contents of sharps waste. The Sharpsology 2 container is sunstantially equivalent to alorementioned predicate device and performs at least as safe and effective as the legally marketed device identified as the predicate device.
510(k) Summary Page 6 of 6
{6}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/6/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features the department's name arranged in a circular fashion around a symbol. The symbol consists of three stylized shapes that resemble human figures or abstract forms, positioned in a way that suggests movement or interaction.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
November 15, 2013
Industrial Water Solutions, Incorporated C/O Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO, MN 55313
Re: K133031
Trade/Device Name: Sharpsology 2 Regulation Number: 21 CFR 880.5570 Regulation Name: Reusable Sharps Disposal Container Regulatory Class: II Product Code: MMK Dated: October 29, 2013 Received: October 31, 2013
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
## Page 2 - Mr. Job
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
Sincerely vours.
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Image /page/7/Picture/7 description: The image shows a logo with the text "Tejashri Purohit-Sheth, M.D. Clinical Deputy Director DAGRID". The logo is in black and white and has a stylized design. The text is arranged in a way that it forms a part of the logo's structure. The word "FOR" is at the bottom right corner of the image.
Kwame Ulmer M.S. Acting Division Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510{k} Number (if known) K133031
ないです。 ・
Device Name
Sharpsology 2 Reusable Sharps Container
### Indications for Use (Describe)
The container is intended for use in healthere settings. The contain sharps prior to removel from the generating facility, during transportation and until ultimate treatment and disposal of the sharps.
The container features a canten style, vertical drop lid or a mallect style, vertical drop lid. The container is housed in a will mount attected to the wall. The container is transform and can be selected in a natural or red color. The wall mount is a natural color with a clear door. The container volume is 2.25 gallors up to the fill line.
The dimensions of the continer are 13.173 inches wide, 11.417 inches bigh (front to back). The lid surface dimensions for the canteen style lid are 6.197 inches wide and 13.031 inches long. The mallow style ild are 6.194 inches wide and 13.029 inches long. The lid sperture for the canteen syle, vertical drop Ild is 2.135 inches. The opening of the lid aperture for the mailbox style, vertical drop lid is 2.5 inches.
The dimensions of the wall mount are 12.25 inches wide, 16.625 inches in depth (front to back).
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Submart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## Please DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY Concurrence of Center for Devices and Redictorical Health (CDRH) (Signature)
Elizabeie-S
.2013.105-05'00'
FORM FDA 3881 (9/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.