Published clinical literature (Bolger et al., 2001)
The sponsor utilized a published clinical study as a literature control to provide comparative safety and effectiveness data for the Vent-Os Sinus Dilation System.
Literature control; Comparative effectiveness
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Bolger et al. (2001) - Safety and outcomes of balloon catheter sinusctomy; Multicenter analysis; Follow-up/Duration: 24-week analysis
Patients undergoing balloon catheter sinusctomy; Sample Size: 142 (at 1 week); 116 (at 3 months); Number of Sites: Multicenter
Not applicable for this study
Ostia patency
Indications for Use
The Vent-Os™ Sinus Dilation System is an instrument intended to provide a means to access the sinus space and to dilate the maxillary sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.
Device Story
Vent-Os Sinus Dilation System is an osmotically driven manual surgical instrument for maxillary sinus ostia dilation. System consists of a dilation device preloaded on a placement instrument. Physician accesses target site via nasal passageway; placement instrument delivers dilation device. Device expands via osmotic pressure, remodeling target tissues; device is subsequently removed. Used in clinical settings by physicians. Output is physical dilation of sinus ostia, facilitating diagnostic or therapeutic access. Benefits include minimally invasive sinus ostia remodeling.
Clinical Evidence
Prospective, non-randomized, single-arm, open-label study (N=34 patients, 57 maxillary ostia). Primary endpoint: ostial patency. Results at 3 months: 93% patency (95% CI: 82.7% - 97.1%), comparable to literature control (Bolger et al., 84% patency). Additional evidence: biocompatibility (ISO 10993), bench testing (weight, profile, force, integrity), and cadaver usability testing.
Technological Characteristics
Osmotically driven dilation device; manual placement instrument. Materials compliant with ISO 10993 (cytotoxicity, sensitization, reactivity). Sterilization via radiation (ANSI/AAMI/ISO 11137-2). No electronic components or software.
Indications for Use
Indicated for adults requiring access to sinus space and dilation of maxillary sinus ostia and associated spaces for diagnostic and therapeutic procedures.
Regulatory Classification
Identification
An ear, nose, and throat manual surgical instrument is one of a variety of devices intended for use in surgical procedures to examine or treat the bronchus, esophagus, trachea, larynx, pharynx, nasal and paranasal sinus, or ear. This generic type of device includes the esophageal dilator; tracheal bistour (a long, narrow surgical knife); tracheal dilator; tracheal hook; laryngeal injection set; laryngeal knife; laryngeal saw; laryngeal trocar; laryngectomy tube; adenoid curette; adenotome; metal tongue depressor; mouth gag; oral screw; salpingeal curette; tonsillectome; tonsil guillotine; tonsil screw; tonsil snare; tonsil suction tube; tonsil suturing hook; antom reforator; ethmoid curette; frontal sinus-rasp; nasal curette; nasal rasp; nasal rongeur; nasal saw; nasal scissors; nasal snare; sinus irrigator; sinus trephine; ear curette; ear excavator; ear rasp; ear scissor, ear snare; ear spoon; ear suction tube; malleous ripper; mastoid gauge; microsurgical ear chisel; myringotomy tube inserter; ossici holding clamp; sacculotomy tack inserter; vein press; wire ear loop; microrule; mirror; mobilizer; ear, nose, and throat punch; ear, nose and throat knife; and ear, nose, and throat trocar.
Bolger W et al. Safety and outcomes of balloon catheter sinusctomy: a multicenter 24-week analysis in 115 patients. Otolaryngol Head and Neck Surg 2001, 137: 10-20
Submission Summary (Full Text)
{0}------------------------------------------------
#### Premarket Notification 510(k) Summary 8
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
## 510(k) Number: K133016
## Applicant Information:
Date Prepared: December 6, 2013 ·
| Name: | SinuSys Corporation |
|----------|----------------------------------------|
| Address: | 4030 Fabian Way<br>Palo Alto, CA 94303 |
| Phone: | 650-213-9988 |
| Fax: | 650-213-9688 |
## Contact Person:
Michael A Daniel, Consultant
| Phone Number: | (415) 407-0223 |
|-------------------|----------------|
| Office: | (925) 254-5228 |
| Facsimile Number: | (925) 254-5187 |
Device Information:
| Device Trade Name: | Vent-Os™ Sinus Dilation System |
|---------------------------|---------------------------------------------------|
| Common Name: | Instrument, ENT Manual Surgical |
| Classification Name(s): | Ear, nose, and throat manual surgical instrument. |
| Product Code/ Regulation: | LRC / 874.4420 |
| Classification: | Class I, Exempt |
# Predicate Devices:
Relieva Spin Sinus Dilation System (K111875) XprESS Multi-Sinus Dilation Tool (K121943) ENTrigue / Ventera (K121351)
## Device Description:
The Vent-OsTM Sinus Dilation System is an osmotically driven device that dilates maxillary sinus ostia and associated spaces. The Vent-Os™ Sinus Dilation System is comprised of a Dilation Device preloaded on a Placement Instrument. The Placement Instrument accesses the target site through the nasal passageway and delivers the Dilation Device. The Vent-Os™ Sinus Dilation Device then expands and remodels the target tissues and then is removed.
### DEC 9 2013
{1}------------------------------------------------
### Premarket Notification 510(k) Summary - Continued 8
# Indications for Use:
The Vent-Os™ Sinus Dilation System is an instrument intended to provide a means to access the sinus space and to dilate the maxillary sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.
# Comparison to Predicate Devices:
The Vent-OsTM Sinus Dilation System has the same intended use and similar indication, mechanism of action and procedural outcomes as the Relieva Spin Sinus Dilation System, the ENTrigue Ventera dilation device and the XprESS Multi-Sinus Dilation Tool (the predicate devices). The SinuSys Dilation System dilates maxillary sinus ostia and associated spaces via pressure applied circumferentially in a substantially equivalent manner to the predicate devices. The technological differences between the Vent-Os™ Sinus Dilation System and its predicates do not raise new questions of safety or efficacy. Performance data demonstrates substantially equivalence in terms of safety and effectiveness to the predicate devices.
# Performance Testing:
Biocompatibility in compliance with EN ISO 10993 requirements was completed. This biocompatibility testing included:
| Item # | Test | Standard Reference | Results |
|--------|----------------------------|--------------------|-----------------|
| 1 | Cytotoxicity (MEM Elution) | ISO 10993-5:2009 | Non-cytotoxic |
| 2 | Sensitization-Maximization | ISO 10993-10: 2010 | Non-sensitizing |
| 3 | Intracutaneous Reactivity | ISO 10993-10: 2010 | Non-irritating |
{2}------------------------------------------------
### Premarket Notification 510(k) Summary - Continued 8
In Vitro bench evaluation including performance testing was completed. This performance testing included the following evaluations:
| Attribute Tested | Pass/Fail |
|-----------------------------------------|-----------|
| Device weight | Passed |
| Placement System Working Length Profile | Passed |
| Placement System Working Length | Passed |
| Crossing Profile | Passed |
| Deployment Force | Passed |
| Dilation Device Diameter @ 1 hour | Passed |
| Dilation Device Working Length @ 1 hour | Passed |
| Dilation Device Integrity | Passed |
| Bond Strength | Passed |
| Proximal Anchoring Force | Passed |
Transportation, sterilization, and shelf-life testing was completed. Sterilization was performed in compliance with all requirements of ANSI/AAMI/ISO 11137-2: Sterilization of Health Care Products - Radiation. Transportation and packaging validation included the following:
| Attribute Tested | Pass/Fail |
|--------------------------------------------|-----------|
| Transit Testing | Passed |
| Sterile Barrier Integrity<br>(Bubble Test) | Passed |
| Packaging & System<br>Visual Inspection | Passed |
| Pouch Seal Testing | Passed |
{3}------------------------------------------------
### Premarket Notification 510(k) Summary - Continued 00
Animal studies confirming safe placement and dilation were conducted. Four (4) Dilation Devices were deployed into four (4) separate sheep maxillary sinuses (MSO).
Cadaver testing was completed by multiple users to validate usability. This cadaver evaluation included use of two (2) adult cadaveric head specimens with the device placed four (4) times into each target site. Investigators confirmed successful endoscopic delivery and placement of the device.
Finally, clinical trials to demonstrate safety and effectiveness were performed by SinuSys Corporation to demonstrate substantial equivalence of the Vent-Os™ Sinus Dilation System to the predicate devices cited. A prospective, non-randomized, single arm, open label, study was conducted at five (5) clinical sites with five (5) investigators. A total of 34 patients and 57 maxillary ostia were treated. Results were comparable to the literature control as summarized in the following table:
| | SinuSys Study | | Bolger Study | |
|---------------|---------------|---------------|--------------|---------------|
| Analysis | Per Protocol | 95% CI* | Per Protocol | 95% CI* |
| | Acute | | 1 week | |
| N | 57 | | 142 | |
| Patent | 54 (95%) | 85.6% - 98.2% | 106 (75%) | 66.9% - 81.1% |
| Non-Patent | 0 | | 0 | |
| Indeterminate | 3 (5%) | 1.8% - 14.4% | 36 (25%) | 18.9% - 33.1% |
| | 3 Months | | 3 Months | |
| N | 55 | | 116 | |
| Patent | 51 (93%) | 82.7% - 97.1% | 98 (84%) | 76.8% - 90.0% |
| Non-Patent | 0 | | 3 (3%) | |
| Indeterminate | 4 (7%) | 2.9% - 17.3% | 15 (13%) | 8.0% - 20.2% |
*Wilson score confidence intervals
Bolger W et al. Safety and outcomes of balloon catheter sinusctomy: a multicenter 24-week analysis in 115 patients. Otolaryngol Head and Neck Surg 2001, 137: 10-20.
### Summary:
Based upon the device description and test data provided in this submission the Vent-Os™ Sinus Dilation System is substantially equivalent to the predicate devices cited.
{4}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows a logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three wavy lines instead of snakes.
December 9, 2013
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
SinuSys Corporation c/o Mr. Michael A. Daniel Regulatory Consultant 8 Snowberry Court Orinda, CA 94563
Re: K133016
Trade/Device Name: Vent-OsTM Sinus Dilation System Regulation Number: 21 CFR 874.4420 Regulation Name: ENT Manual Surgical Instrument Regulatory Class: Class I Product Code: LRC Dated: November 24, 2013 Received: November 26, 2013
Dear Mr. Daniel:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical
{5}------------------------------------------------
Page 2 - Mr. Michael A. Daniel
device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
htto://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
# Denise L. Hampton -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K133016
Device Name: Vent-Os™ Sinus Dilation System
Indications For Use:
The Vent-Os™ Sinus Dilation System is an instrument intended to provide a means to access the sinus space and to dilate the maxillary sinus ostia and associated spaces in adults for diagnostic and therapeutic procedures.
× Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Sunny Park 2013.12.08 20:32:07 -05'00'
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.