The Inventis Middle Ear Analyzers, Flute and Viola, are intended for use by trained operators in hospitals, nurseries, ENT clinics and audiology offices in conducting diagnostic hearing evaluations and assisting in diagnosis of possible otologic disorders. The Inventis Middle Ear Analyzers are intended for use on individuals six months of age or older. Flute is a tympanometer. Viola is a combination of audiometer and tympanometer, which reduces the amount of necessary equipment.
Device Story
Inventis Flute and Viola are middle ear analyzers; Flute functions as a standalone tympanometer; Viola integrates tympanometry with diagnostic audiometry (air/bone conduction, speech). Devices input acoustic impedance/admittance data via probe tones (226 Hz, 1 kHz) and audiometric stimuli (pure tone, warble, speech). Operated by trained clinicians in hospitals, ENT clinics, and audiology offices. Devices process signals to measure middle ear function and hearing thresholds; output results via graphical color display or optional integrated thermal printer. Clinical utility includes assisting in diagnosis of otologic disorders. Benefits include streamlined diagnostic workflow through combined functionality and improved portability compared to traditional equipment.
Clinical Evidence
Bench testing only. Performance verified against ANSI S3.39 (impedance), ANSI S3.6 (audiometry), and IEC 60645 series standards. Electrical safety and electromagnetic compatibility testing confirmed compliance with IEC 60601-1 and IEC 60601-1-2.
Technological Characteristics
Impedance/admittance measurement via probe tones (226 Hz, 1 kHz). Audiometry (Viola) via air/bone conduction. Graphical color display. Internal flash memory for speech storage. Powered by AC. Safety standards: IEC 60601-1 (Class II, Type BF), IEC 60601-1-2. Performance standards: ANSI S3.39/IEC 60645-5 (Impedance), ANSI S3.6/IEC 60645-1/2 (Audiometry).
Indications for Use
Indicated for diagnostic hearing evaluations and diagnosis of otologic disorders in individuals 6 months of age or older. Used by trained operators in clinical settings.
Regulatory Classification
Identification
An auditory impedance tester is a device that is intended to change the air pressure in the external auditory canal and measure and graph the mobility characteristics of the tympanic membrane to evaluate the functional condition of the middle ear. The device is used to determine abnormalities in the mobility of the tympanic membrane due to stiffness, flaccidity, or the presence of fluid in the middle ear cavity. The device is also used to measure the acoustic reflex threshold from contractions of the stapedial muscle, to monitor healing of tympanic membrane grafts or stapedectomies, or to monitor followup treatment for inflammation of the middle ear.
Special Controls
*Classification.* Class II (special controls). The device, when it is a tympanometer or auditory impedance tester that complies with FDA-recognized consensus standard ANSI S3.39, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 874.9.
Predicate Devices
Interacoustics Audio Traveller, Model AA222 (K022728)
Submission Summary (Full Text)
{0}------------------------------------------------
## INVENTIS MIDDLE EAR ANALYZER TRADITIONAL 510(K) PREMARKET NOTIFICATION
# 510(k) Summary
------------------------------------------------------------------------------------------------------------------------------------------------------------------------------
# 1 - Information
510(k) Notification: K133013
# GENERAL INFORMATION
| Applicant: | Inventis srl |
|------------|-------------------------------------------|
| | Corso Stati Uniti 1/3, 35127 Padova Italy |
| | Tel: +39 049 8962844 |
| | Fax: +39 049 8966 343 |
| | Registration Number: 3008699634 |
#### Massimo Martinelli Contact Person: General Manager E-mail: massimo.martinelli@inventis.it
t
Date: December 18, 2013
## DEVICE INFORMATION
The Inventis Middle Ear Analyzers (Flute and Viola) are conventional Auditory Impedance Testers. Below you can find the information related to the subject device.
Katalogia (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1) (1)
:
And Controller Comments
| Classification: | 21 CFR§874.1090<br>Class II |
|-------------------------|---------------------------------------|
| Product Code: | ETY |
| Trade Name: | Middle Ear Analyzers, Flute and Viola |
| Common name: | Middle Ear Analyzer |
| Classification<br>Name: | Auditory Impedance Tester |
JAN 2 4 2014
・
{1}------------------------------------------------
### PREDICATE DEVICES
Inventis Middle Ear Analyzers are substantially equivalent to the Interacoustics Audio Traveller, Model AA222 for the constructive technological features and the performance.
#### INDICATIONS FOR USE
The Inventis Middle Ear Analyzers, Flute and Viola, are intended for use by trained operators in hospitals, nurseries, ENT clinics and audiology offices in conducting diagnostic hearing evaluations and assisting in diagnosis of possible otologic disorders. The Inventis Middle Ear Analyzers are intended for use on individuals six months of age or older. Flute is a tympanometer. Viola is a combination of audiometer and tympanometer.
#### PRODUCT DESCRIPTION
Flute is stand-alone tympanometry / reflex diagnostic device. Viola is a stand-alone combined unit, integrating a screening middle ear analyzer and a diagnostic audiometer capable of performing air and bone conduction threshold and speech exams. Flute and Viola are available with or without the integrated thermal printer.
# 2 - Substantial Equivalence
Inventis Middle Ear Analyzers (Flute and Viola) represent a modification of the cleared device, Interacoustics Audiotraveller, Model AA222 (510(k) number: K022728). The changes are of minor nature and do not concern the main functions of the predicate device. In particular, the differences regard:
- the pressure range, which is -600 to +400 daPa instead of -600 to +300 daPa and .
- the introduction of an Internal Flash Memory where the speech material is stored. .
In further support of a substantial equivalence determination, Section 12-Substancial Equivalent Discussion provides a comparison table between the submitted device and the predicate device. The main comparison features are reported in the following table:
{2}------------------------------------------------
· ·
.
.
| | Predicate Device | New Device | |
|---------------------------------------------|-------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------|-----------------------------------------------------------------|
| Features | Interacoustics Audiotraveller,<br>Model AA222<br>(510(k) n.: K022728) | Flute | Viola |
| Type | 2 - Impedance<br>(ANSI S3.39)<br>2- Tone Audiometer<br>(ANSI S3.6)<br>B-E Speech Audiometer<br>(ANSI S3.6) | 2-Impedance<br>(ANSI S3.39) | 2-Impedance<br>(ANSI S3.39)<br>3-Tone Audiometer<br>(ANSI S3.6) |
| Probe Tone Frequency | 226 Hz, 678 Hz, 800 Hz, 1 kHz | 226 Hz,<br>1kHz (model HF) | 226 Hz |
| Compliance Range | up to 8.0 ml | Same | |
| Pressure Range | -600 to +300 daPa | -600 to +400 daPa | |
| Safety Limitations for<br>pressuring system | -800 daPa and +600 daPa | Same | |
| Acoustic Reflex<br>outputs | Ipsi (I), Contra (C) | Same | |
| Acoustic Reflex Stimuli | 250Hz (C), 500Hz, 1kHz, 2kHz,<br>3kHz, 4kHz, 6kHz (C), 8kHz (C),<br>BBN, LPN, HPN | 250Hz (C), 500Hz, 1kHz, 2kHz, 4kHz | |
| Audiometry:<br>Available Frequencies | 125Hz, 250Hz, 500Hz, 750Hz,<br>1kHz, 1.5kHz, 2kHz, 3kHz, 4kHz,<br>6kHz, 8kHz | | Same |
| Audiometry Outputs | AC, BC, FF, insert mask | | AC, BC, insert mask |
| Audiometry Inputs | PTA: pure tone, warble | | PTA: pure tone,<br>warble |
| | Speech: EXT1, EXT2, MIC<br>Masking: NBN, WN, SN | | Speech: EXT1, EXT2,<br>INT, MIC<br>Masking: NBN, WN,<br>SN |
| Max Power<br>Consumption | 45VA | 19.2 W | |
| Display Description | Graphic monochrome display | Graphical color display | |
| Size | 16.5 x 12.5 x 4 in<br>(42 x 32 x 10.2 cm) | 12.6 x 12.6 x 3.5 in (32 x 32 x 9 cm);<br>w/printer: 12.6 x 15.4 x 3.5 in (32 x 39 x 9 cm) | |
| Weight | 9 lbs (4,1 kg) | 4.19 lbs (1,9 kg);<br>w/printer: 5.29 lbs (2,4kg) | |
| Safety and<br>Effectiveness | Impedance<br>ANSI S3.39/IEC 60645-5: Type 2 | Impedance<br>ANSI S3.39/IEC 60645-5: Type 2 | |
| | Audiometer<br>ANSI S3.6/IEC 60645-1:<br>Type 2 Tone Audiometer<br>ANSI S3.6/IEC 60645-2:<br>Speech Type B-E | Audiometer (only Viola models)<br>ANSI S3.6/IEC 60645-1:<br>Type 3 Pure-tone Audiometer | |
| | Safety<br>IEC 60601-1, Class I, Type B<br>IEC 60601-1-2 | Safety<br>IEC 60601-1, Class II, Type BF<br>IEC 60601-1-2 | |
-
{3}------------------------------------------------
Inventis Middle Ear Analyzers (Flute and Viola) are substantially equivalent to the predicate device with regard to design, function, safety and technological and performance characteristic, intended use.
Any differences in the technological characteristics between the devices do not raise any new issues of safety or effectiveness. Thus, the proposed Inventis Middle Ear Analyzers are substantially equivalent to the predicate devices.
## TESTING IN SUPPORT OF SUBSTANTIAL EQUIVALENCE DETERMINATION
All necessary bench testing was conducted on the Inventis Middle Ear Analyzers (Flute and Viola) to support a determination of substantial equivalence to the predicate device. The safety and performance testing included the following tests:
- . Electrical and mechanical safety testing
- . Electromagnetic compatibility safety testing
- . Specifications for Instruments to Measure Aural Acoustic Impedance and Admittance (ANSI S3.39)
- . Instruments for the measurement of aural acoustic impedance (IEC 60645-5)
#### CONCLUSION STATEMENT
All the necessary electromagnetic compatibility and electrical safety tests were performed and documented in the Section 17-Electromagnetic compatibility and electrical safety. The results demonstrate that the Inventis Middle Ear Analyzer is in compliance with both the standards IEC 60601-1-2:2007 and IEC 60601-1:2005 and is safe as the predicate device.
All the necessary performance tests in support of substantial equivalence determination were conducted and documented in the Section 18-Performance testing - Bench. The tests demonstrate that the Inventis Middle Ear Analyzer is effective and performs as well as the predicate device.
The minor differences in the technological characteristics between the devices do not raise any new issues of safety or efficacy.
Based on these results, we conclude that the Inventis Middle Ear Analyzer is substantially equivalent to the existing legally marketed device Interacoustics Audiotraveller Model AA222 under Federal Food, Drug and Cosmetic Act.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized depiction of an eagle or bird-like figure with three wing-like shapes. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular fashion around the bird-like figure.
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
#### January 24, 2014
Inventis Srl c/o Mr. Massimo Martinelli General Manager Corso Stati Uniti 1/3 35127, Padova Italy
Re: K133013
Trade/Device Name: Inventis Middle Ear Analyzers, Flute and Viola Regulation Number: 21 CFR 874.1090 Regulation Name: Auditory Impedance Tester Regulatory Class: Class II Product Code: ETY Dated: December 18, 2013 Received: December 23, 2013
Dear Mr. Martinelli:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Mr. Massimo Martinelli
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.goy/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
.. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
# SECTION 04
INDICATIONS FOR USE STATEMENTS
INDICATIONS FOR USE
510(k) Number (if known): K133013
Device Name: Inventis Middle Ear Analyzers, Flute and Viola
Indications for Use: The Inventis Middle Ear Analyzers, Flute and Viola, are intended for use by trained operators in hospitals, nurseries, ENT clinics and audiology offices in conducting diagnostic hearing evaluations and assisting in diagnosis of possible otologic disorders. The Inventis Middle Ear Analyzers are intended for use on individuals six months of age or older. Flute is a tympanometer. Viola is a combination of audiometer and tympanometer, which reduces the amount of necessary equipment.
Prescription Use _ Y (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(Please DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Shuchen Peng -S
Shuchen Peng -S 2014.01.28 13:37:04 -05:00,
SECTION 04
Page 04 - 1 of 1 Pages
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.