K133012 · Path Medical GmbH · EWO · Jun 9, 2014 · Ear, Nose, Throat
Device Facts
Record ID
K133012
Device Name
SENTIERO
Applicant
Path Medical GmbH
Product Code
EWO · Ear, Nose, Throat
Decision Date
Jun 9, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 874.1050
Device Class
Class 2
Attributes
Real-World Evidence, Pediatric
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K133012 · Jun 9, 2014
SENTIERO
Path Medical GmbH
Clinical evaluation reports from external ENT clinics and university medical centers; Patient clinical data from routine diagnostic testing
Clinical evaluations were conducted at multiple external sites to compare the performance and usability of the Sentiero device against predicate devices in a clinical setting. These evaluations involved testing on patients with hearing loss and normal hearing to validate the device's diagnostic modules (ABR, OAE, ASSR, PTA).
Clinical evaluation; Performance validation; Usability testing; Comparative clinical study
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
External clinical evaluations of Sentiero modules; Comparative clinical evaluation; Study Period: 2011 (first clinical evaluation)
Patients with hearing loss and normal hearing; Sample Size: Minimum of 20 participants per evaluation; Number of Sites: Multiple (e.g., University Medical Center Mainz, University of Ferrara, Klinikum rechts der Isar, Kantonsspital Liestal, St.-Elisabeth-Hospital Bochum)
Predicate devices (Cochlea-Scan, Chartr EP 200, AccuScreen, Eclipse)
Morphology and latencies of wave V (ABR), device usability, and diagnostic performance of OAE/ABR/ASSR/PTA modules
Indications for Use
Sentiero is a portable instrument to diagnose all ages or hearing loss. The instrument offers different test methods which can be configured to fit the professional's needs for screening or diagnostic purposes. It offers physiological test methods such as: . Distortion Product Otoacoustic Emissions (DPOAE) . Transient Evoked Otoacoustic Emissions (TEOAE) · Auditory Brainstem Response (ABR) · Auditory Steady State Response (ASSR) · Auditory Impedance and acoustic reflex (TYMP) Additionally if offers standard audiometry (psycho-acoustical). All physiological test methods are especially indicated for use in defining the type and configuration of hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable or to assist in the diagnosis of otologic disorders. Estimation of cochlear hearing thresholds (DPTHRESH) is possible at various frequencies without the need of cooperative interaction with the patient. Acoustic reflex and tympanometry (TYMP) are featured to evaluate the function of the middle and outer ear. For each method, several protocols can be configured. The results are to be used to make further recommendations regarding appropriate intervention strategies. Therefore, Sentiero is intended for use by trained personnel such as audiologists. pediatricians. ENT doctors and other health care professionals in a medical or home environment. In the United States of America. Federal law restricts this device to sale by or on the order of a licensed physician. Available psycho-acoustical methods on Sentiero are especially indicated for use with cooperative patients starting at the age of 2 years or adequate development age, which enables them to do play/interactive audiometry. All other modules are suitable to be used for all ages elder than.infants from 34 weeks (gestational age) that are ready for discharge from the hospital. Sentiero is designed for: 1. Diagnostics, monitoring and follow-up after newborn hearing screening 2. Pre-school, school, and adult hearing screening 3. ENT diagnostics based on measurement of a) Otoacoustic emissions b) Tympanometry and acoustic reflex c) Auditory Brainstem Responses d) Auditory Steady State Responses Sentiero must not be used in cases of external otitis (outer ear canal infection) or in any case which vields to pain when inserting the ear probe or applying any other transducer.
Device Story
Sentiero is a portable/desktop audiometric platform for screening and diagnostics. It accepts inputs from OAE probes, headphones, bone conductors, and electrodes. The device uses configurable modules to perform DPOAE, TEOAE, ABR, ASSR, tympanometry, and pure-tone audiometry. Transducers feature memory chips for automatic calibration and configuration. Operated by trained professionals (audiologists, ENT doctors, pediatricians) in medical or home settings. The device processes signals via internal firmware, displaying waveforms, frequency domain data, and signal statistics on a touch screen. It provides diagnostic information to assist in defining hearing loss type/configuration and otologic disorder diagnosis. Results guide clinical intervention strategies. The device can connect to a PC via USB for data management.
Clinical Evidence
Bench testing validated TEOAE, DPOAE, DPTHRESH, and ABR modules using a simulator device (also used for predicate validation). Clinical evaluation (2011) compared Sentiero to Chartr EP 200 on test subjects, analyzing wave V morphology and latencies; both were deemed reliable. Additional performance data provided by external test sites (e.g., University Medical Center Mainz, University of Ferrara) confirmed equivalence to predicates for various modules (PTA, MAGIC, MATCH, ABR, TEOAE, ASSR). No animal testing conducted.
Technological Characteristics
Handheld/desktop platform with 3.5" or 5.0" touch screen. Biocompatible housing. Transducers (probes, headphones) feature memory chips for calibration/ID. Energy: battery-powered. Connectivity: USB to PC. Standards: IEC 60601-1, IEC 60601-1-2, IEC 60645-1/5/6/7, ISO 10993-1. Software: firmware-based, menu-guided, configurable modules. Tympanometry uses internal pump.
Indications for Use
Indicated for all ages, including infants from 34 weeks gestational age, for hearing screening and diagnostics. Psycho-acoustical methods indicated for cooperative patients aged 2+ years. Contraindicated in cases of external otitis or pain upon transducer insertion.
Regulatory Classification
Identification
An audiometer or automated audiometer is an electroacoustic device that produces controlled levels of test tones and signals intended for use in conducting diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders.
{0}------------------------------------------------
### 510(k) Summary
### SUBMISSION INFORMATION
| Date of preparation: | May 4th, 2014 |
|----------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| 510(k) Submitter: | PATH medical GmbH<br>Landsberger Str. 63<br>82110 Germering<br>Germany<br>Phone: ++49-89-80076502<br>Fax: ++49-89-80076503 |
| Contact Person: | Dr. Johann Oswald, Director<br>Landsberger Str. 63<br>82110 Germering<br>Germany<br>Phone: ++49-89-80076502<br>Fax: ++49-89-80076503<br>Email: oswald@pathme.de |
### DEVICE INFORMATION
| Device Name: | Sentiero |
|------------------------------|----------------------------------------------------------------------------------------------------------------------------------------|
| Device Trade Names: | Sentiero Basic, Sentiero Advanced, Sentiero Desktop |
| Device Identification Codes: | SOH100098, SOH100360, SOD100497 |
| Common Name: | Evoked Response Auditory Stimulator |
| Classification Name: | Evoked Response Auditory Stimulator, Audiometer, Auditory<br>Impedance Tester<br>FDA 21 CFR section 882.1900 and 874.1050 and 874.1090 |
### PREDICATE DEVICES
| Cochlea-Scan | 510(k) number: K061744 |
|---------------|------------------------|
| AccuScreen | 510(k) number: K122067 |
| Chartr EP 200 | 510(k) number: K092373 |
| TITAN IMP440 | 510(k) number: K083861 |
| Eclipse | 510(k) number: K070696 |
### DEVICE DESCRIPTION
Sentiero is an audiometric examination platform which consists of the Sentiero device with a touch screen display together with different accessories such as mains adapter, OAE probes, headphones, bone conductor, electrode cable, ear coupler cable, patient response switch. All connectors and transducers have
a special plug in order to ensure the correct connection to the device. All plugs of the transducers have a memory chip inside which stores the information about the respective transducer (including type of
{1}------------------------------------------------
connector, calibration table). As a result, the Sentiero instrument can be connected flexibly to different kind of transducers while enabling the instrument to 'know' the features of the connected transducer. This information is used within the different modules (test methods which are configured) to guide the user (feedback via display) and help to ensure correct performance of the test methods.
Sentiero is available in two different models: handheld or desktop version. The desktop version is labeled Sentiero (Type Desktop). Both versions are standalone examination platforms and can be connected to a personal computer (PC) via USB for data review and management. The handheld version is portable and is meant to be mainly used as mobile device. The desktop version is portable as well but is meant to be mainly used as stationary device. Both models base on a common hardware platform (printed circuit board, PCB) but with different configurations. Materials in contact with humans are selected to be biocompatible.
Furthermore, each model can be configured to allow different test methods and features (modules) by a license key in the device. Sentiero is based on configurable modules. Sentiero can have one single module or a combination of multiple of modules described in the following intended use.
The measurement application is controlled from a self-contained firmware (software installed on the instrument). The measurement flow is menu guided on a touch screen. Evaluation of test results is based on signal statistics (if available for the test method). Besides that wave forms and result information is displayed for the user's evaluation.
See also "Indications for Use".
The following accessories are available to conduct the different measurement modules:
- · TEOAE probe: PATH EP-TE
- · Clearance by FDA following the submission of a 510(k) K100661 and update on K131141
- · DPOAE probe: PATH EP-DP
- · Clearance by FDA following the submission of a 510(k) K100661 and update on K131141
- In combination with a silicon tube attached to Sentiero SOD and Earprobe EP-EP. the probe . can be used as stimulator for the TYMP module.
- Electrode cable: PATH .
- · shielded, passive cable to connect the instrument to electrodes for ABR or ASSR
- Headphones: Sennheiser HDA 280, Sennheiser HDA 200, Interacoustics DD 45, Holmco PD 81, GN . Otometrics otoInsert. PATH Ear Coupler Cable
- · Only certified standard audiometry headphones are used in conjunction with Sentiero.
- · Bone conductor: RadioEar B71
- · As stated by RadioEar (FDA establishment registration number 2516347): "The Bone Conductor does not come under the FDA Requiations for Medical Devices nor the FDA 510(k) Pre-market Approval provisions."
- · Patient response switch
- Passive switch for the patient to feedback his response to pure tone audiometry (heard or not heard equals pressed or not pressed as indicated in standard ISO 8253-1)
These accessories can be connected to Sentiero based on a special plug, which holds the information about the connected transducer / cable. Therefore the firmware can make use of this information and adapt the measurement procedures accordingly or provide information to the user via its display,
### INTENDED USE
Sentiero is a portable instrument to diagnose all ages or hearing loss. The instrument offers different test methods which can be configured to fit the professional's needs for screening or diagnostic purposes. It offers physiological test methods such as:
- . Distortion Product Otoacoustic Emissions (DPOAE)
- . Transient Evoked Otoacoustic Emissions (TEOAE)
- · Auditory Brainstem Response (ABR)
{2}------------------------------------------------
- · Auditory Steady State Response (ASSR)
- · Auditory Impedance and acoustic reflex (TYMP)
Additionally if offers standard audiometry (psycho-acoustical).
All physiological test methods are especially indicated for use in defining the type and configuration of hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable or to assist in the diagnosis of otologic disorders. Estimation of cochlear hearing thresholds (DPTHRESH) is possible at various frequencies without the need of cooperative interaction with the patient. Acoustic reflex and tympanometry (TYMP) are featured to evaluate the function of the middle and outer ear. For each method, several protocols can be configured. The results are to be used to make further recommendations regarding appropriate intervention strategies. Therefore, Sentiero is intended for use by trained personnel such as audiologists. pediatricians. ENT doctors and other health care professionals in a medical or home environment. In the United States of America. Federal law restricts this device to sale by or on the order of a licensed physician.
Available psycho-acoustical methods on Sentiero are especially indicated for use with cooperative patients starting at the age of 2 years or adequate development age, which enables them to do play/interactive audiometry. All other modules are suitable to be used for all ages elder than.infants from 34 weeks (gestational age) that are ready for discharge from the hospital.
Sentiero is designed for:
- 1. Diagnostics, monitoring and follow-up after newborn hearing screening
- 2. Pre-school, school, and adult hearing screening
- 3. ENT diagnostics based on measurement of
- a) Otoacoustic emissions
- b) Tympanometry and acoustic reflex
- c) Auditory Brainstem Responses
- d) Auditory Steady State Responses
Sentiero must not be used in cases of external otitis (outer ear canal infection) or in any case which vields to pain when inserting the ear probe or applying any other transducer.
### COMPARISON TO PREDICATE DEVICES
In terms of technology, workflow, and accessories, the predicate device AccuScreen shares most of the characteristics of Sentiero. In contrast to AccuScreen, which is a screening device, Sentiero provides diagnostic features based on sophisticated algorithms.
With respect to the intended use, combining psycho-acoustical and physiological test modules in a single handheld device, the Cochlea-Scan is similar.
With respect to diagnostic features on evoked potentials, Chartr EP 200 (ABR and ASSR) and Eclipse (OAE, ABR, and ASSR) can be seen as predicate devices.
With respect to the auditory impedance and acoustic reflex (TYMP), the Titan IMP440 can be seen as predicate device.
In order to facilitate the comparison, first the device setup and technical features and then the modules' features are compared.
1) Device setup and technical features (including accessories)
Design and working principles of Sentiero are very similar to Cochlea-Scan as well as AccuScreen. The similarity can be attributed to the fact that the same engineers at Fischer-Zoth Diagnosesvsteme GmbH (Germany), who designed the Cochlea-Scan at that time, also designed the Sentiero as well as the AccuScreen at PATH medical GmbH. Accessories like probe and cables were designed and are used for both AccuScreen and Sentiero identically. These two systems were also designed in the same period of time using the same technology standards. In contrast thereto, Cochlea-Scan uses older technology standards. Therefore, the comparison to the predicate device with respect to technological characteristics is focused on the comparison between AccuScreen and Sentiero. Non-clinical performance data are also considered substantially equivalent: the development process was conducted by the same engineers and the resulting devices were tested to be compliant to the same standards by independent laboratories.
{3}------------------------------------------------
Both systems, Sentiero and AccuScreen, are battery driven handheld devices with the user interface on a 3.5" touch screen (Sentiero type MOD 100497: 5.0"). Probes and electrode cables are connected to the devices with similar plugs. Probes, electrode cables, ear coupler cable to the PC (USB) are identical.
Sentiero and AccuScreen provide the same measurement methods (DPOAE, TEOAE and ABR) in order to provide statistically sound results. In contrary to the AccuScreen, which provides a pass/refer result as an automated screening result, the signal processing information is used in Sentiero only as an indicator (traffic light or indicator for valid response). In addition, wave forms (ABR) or frequency domain information (TEOAE and DPOAE) are provided together with additional diagnosic information and environmental information of the measurement to inform the user about the quality of the measurement.
The primary mechanical difference between AccuScreen and Sentiero is the housing which contains the printed circuit board (PCB) and the charging of the built-in rechargeable batteries. AccuScreen can be connected to a docking station for charging and data transfer. In contrast thereto, Sentiero must be connected to a custom made plug for connecting to the charger is possible via a USB cable connected directly to the device. Both housings are made of biocompatible material to fulfill the standards accordingly.
Both devices, Sentiero and AccuScreen, show similar performance data and compliance to the following standards:
- · IEC 60601-1:2005
- · IEC 60601-1-2:2007
- · IEC 60601-1-4:1996
- · IEC 62304:2006
- · ISO 10993-1:2009
Safety Electromagnetic compatibility Programmable electrical systems Software lifecycle Biocompatibility
In conclusion, Sentiero is substantially equivalent to the predicate device AccuScreen with respect to technological characteristics and non-clinical performance data.
#### 2) Device measurement features
From a measurement perspective, the methods of predicting a behavioral threshold by means of DPOAE are similar to the predicate device Cochlear-Scan. Also the feature to combine psycho-acoustical test methods like pure tone audiometry and TEOAE or DPOAE on a diagnostic level is similar between both devices. Again, this is supported by the fact, that the same engineers developed both devices. In comparison to Cochlea-Scan, the most significant difference is the technological improvements added to the design of Sentiero. Nevertheless, substantial equivalence is given with respect to the diagnostic features of OAE measurements and hearing threshold predictions.
Additional diagnostic features and measurement techniques were added to Sentiero in comparison to AccuScreen or Cochlea-Scan, which are similar to Chartr EP 200 and Eclipse. Namely the ABR and ASSR features of Sentiero are similar to these predicate devices.
Sentiero uses up to two channels of data recording to run several diagnostic protocols and modalities. The firmware can be configured to run ABR with a traditional click stimulus but also with a tone burst stimulus to predict frequency specific behavioral hearing thresholds. This is similar to the predicate devices Chartr EP 200 and Eclipse. Besides that, Eclipse also provides similar run-time compensated stimuli - similar as Sentiero does (time-delay compensation of the traveling wave on the basilar membrane).
Sentiero can conduct ASSR with a 40 or 80 Hz repetition rate. Several stimuli can be combined together simultaneously and also on both ears to assess responses to various frequencies at the same lime. The evoked response spectrum can yield the information about the presence of the response to stimuli so that ASSR thresholds are supposed to hold a predictable relationship to behavioral thresholds that is dependent on stimulus frequency and the presence or degree of hearing loss. This is similar to the predicate devices Chartr EP 200 and Eclipse.
Besides the difference of dimensions, another difference is that Sentiero can do ABR or ASSR measurements as a standalone system and is therefore not dependent on a PC during the measurement. Chartr EP 200 and Eclipse are all PC-controlled devices.
In a first clinical evaluation in 2011, Sentiero and the predicate device Chartr EP 200 have been tested on the same test persons by different users (trained technicians in audiology). Morphology and latencies of
{4}------------------------------------------------
wave V have been analyzed. Furthermore, the usage of the device was evaluated by the feedback of the users. All results have been evaluated by an expert in audiology. Both systems, Sentiero and Chartr EP 200, were described as reliable clinical ABR devices, which supports the substantial equivalence between both instruments with respect to clinical performance. The run time compensated stimuli and algorithms are similar between Sentiero and Eclipse as they were taken out of the descriptions in published literature.
Sentiero also complies to the relevant standards 60645-1, 60645-7 which displays that TEOAE, DPOAE and evoked potentials can be recorded as safe and as effective as the predicate devices.
Sentiero (Type MOD 100497) can also be configured to allow measurements of auditory impedance and acoustic reflex, which are also present on the predicate device TITAN IMP 440. The other types of Sentiero cannot be configured to use the TYMP module. The stimulus is transmitted via the EarProbe EP-DP in conjunction with a silicon tube in parallel to the probe to deliver the air pressure, which is produced by a pump inside the housing of Sentiero. The TYMP is conform to a IEC 60645-5 / ANSI S3.39 Type 1 acoustic impedance instrument.
In order to facilitate the comparison, the device setup and technical features are compared in the following table to the respective predicate device.
| Parameter for<br>Comparison | Cochlea-Scan | AccuScreen | Chartr EP 200 | Eclipse |
|------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended Use | Same:<br>Hearing<br>screening and<br>diagnostics by<br>means of OAE<br>and pure tone<br>audiometry. | Same:<br>Hearing screening<br>with statistical<br>evaluation of<br>results by means<br>of OAE and ABR.<br>Different:<br>Limited purpose<br>(screening) and<br>age group<br>(newborns). | Same:<br>Hearing<br>diagnostics with<br>ABR and ASSR.<br>Different:<br>No screening, no<br>OAE tests for<br>specific<br>examination of<br>inner ear hearing<br>status. | Same:<br>Hearing<br>diagnostics and<br>neurological<br>screening with<br>OAE, ABR, and<br>ASSR. |
| Patient<br>Population | Same:<br>All ages | Different:<br>Newborns only | Same:<br>All ages | Same:<br>All ages |
| Device<br>Hardware Setup | Same:<br>Standalone,<br>handheld /<br>portable device,<br>battery-<br>powered. | Same:<br>Same touch<br>screen display<br>with same<br>resolution,<br>connectors to<br>cables and<br>transducers.<br>Different:<br>Housing, docking<br>station | Different:<br>Connected to PC,<br>not standalone. | Different:<br>Connected to PC,<br>not standalone. |
| Ear Probe | Same:<br>See K131141. | Same:<br>Same probe,<br>special plugs | | |
| Parameter for<br>Comparison | Cochlea-Scan | AccuScreen | Chartr EP 200 | Eclipse |
| | | including memory. | | |
| Electrode Cable | Same: | Same:<br>Same cable,<br>special plugs<br>including memory. | | |
| Headphones,<br>Insert<br>Earphones | Same:<br>Additional<br>headphones or<br>inserts available<br>- automatic<br>detection<br>through memory<br>in plug | Different:<br>No headphone or<br>insert earphone to<br>be connected. | | |
| Safety,<br>Characteristics,<br>Performance | Same:<br>IEC 10993-1,<br>IEC 60601-1,<br>IEC 60601-1-2,<br>IEC 60601-1-4,<br>EN 1041, ISO<br>980 (ISO<br>15223-1), IEC<br>60645-1, | Same:<br>IEC 10993-1, IEC<br>60601-1, IEC<br>60601-2, 60601-<br>1-4, IEC 60645-6,<br>IEC 606045-7,<br>IEC 62304, IEC<br>62366 (IEC<br>60601-1-6) | Same:<br>IEC 60601-1-2, IEC<br>60601-2-40<br>Different:<br>No<br>encephalography<br>application is<br>mentioned for the<br>predicate device.<br>Nevertheless, the<br>predicate device is<br>compliant with IEC<br>60601-2-26.<br>Sentiero does not<br>comply with EN<br>60601-2-26 as<br>intended use is not<br>for cerebral<br>monitoring or<br>electro-<br>encephalography. | Same:<br>IEC 60601-1, IEC<br>60601-2, IEC<br>60645-1,<br>Different:<br>IEC 60645-3<br>(only relevant to<br>click stimuli -<br>predicate to time<br>compensated<br>stimuli is not<br>affected by this<br>standard)<br>IEC 60601-2-26<br>(see Chartr EP<br>200) |
| Workflow,<br>Menu | Same:<br>General<br>workflow<br>Different:<br>Lower screen<br>resolution,<br>additional keys<br>to enter data | Same:<br>General workflow,<br>online information<br>and device control<br>via header/footer | Different:<br>PC-controlled with<br>separate<br>measurement<br>platform. | Different:<br>PC-controlled with<br>separate<br>measurement<br>platform. |
| Parameter for<br>Comparison | Cochlea-Scan | AccuScreen | Chartr EP 200 | Eclipse |
| Available<br>modules: | Same:<br>DPOAE,<br>TEOAE,<br>DPTHRESHO<br>LD, Standard<br>Audiometry | Same:<br>TEOAE, DPOAE,<br>ABR<br>Different:<br>Sentiero offers<br>diagnostic<br>features instead<br>of screening only | Same:<br>ABR, ASSR | Same:<br>ABR, ASSR |
| Interface to<br>Computer,<br>Software on<br>Computer | Same:<br>USB/RS232<br>connection | Same:<br>USB.connection,<br>data transfer to<br>PC<br>Different: docking<br>station. | Different:<br>Runs on PC | Different:<br>Runs on PC |
5
{5}------------------------------------------------
{6}------------------------------------------------
Table 6-1: Comparison of device setup and technical features (Sentiero against different predicate devices)
1
.
-
| Feature / Comparison | Titan with IMP440 (Predicate) 510(k) No. K083861 | Sentiero | Comments |
|-------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------|----------|
| Indications for use | The Titan Impedance System is an electroacoustic test instrument that produces controlled levels of test tones and signals intended for use in conduction diagnostic hearing evaluations and assisting in the diagnosis of possible otologic disorders. It features tympanometry and acoustic reflexes. | Sentiero features tympanometry and acoustic reflexes. | Same |
| Target population | The devices are suitable for all populations including new-born infants | The devices are suitable for all populations including new-born infants | Same |
| Intended user | The devices are to be used by trained personnel only | The devices are to be used by trained personnel only | Same |
| Safety standard | IEC 60601-1 | IEC 60601-1 | Same |
| EMC standard | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Standards for impedance audiometry | IEC 60645-5/ANSI S3.39, Type 1 | IEC 60645-5/ANSI S3.39, Type 1 | Same |
| Test signal | IEC 60645-1/ANSI S3.6, IEC 60645-3 | IEC 60645-1/ANSI S3.6, IEC 60645-3 | Same |
| Probe tone frequency (tympanometry) | 226 Hz. Optional 678 Hz, 800 Hz and 1000 Hz | 226 Hz. 678 Hz, 800 Hz and 1000 Hz optionally available | Same |
{7}------------------------------------------------
| Probe tone level<br>(tympanometry) | 69 dB HL with AGC | 69 dB HL with AGC | Same |
|------------------------------------|------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Pressure safety limits | -750...+550 daPa | -700...+500 daPa | Equivalent<br>Safety limits need to be<br>within the range of -<br>800...+600 daPa<br>according to IEC<br>60645-5/ANSI S3.39, a<br>smaller range means<br>increased safety |
| Transducers | Ear probe for<br>tympanometry, ear<br>probe and<br>headphone/insert ear<br>phone for reflex<br>measurement | Ear probe for<br>tympanometry, ear<br>probe and<br>headphone/insert ear<br>phone for reflex | Equivalent<br>Different models of ear<br>probe and headphones<br>are used, all of them<br>generating test signals<br>complying to IEC<br>60645-5/ANSI S3.39<br>standards |
| Reflex stimuli | Pure tones at<br>audiometric<br>frequencies, narrow<br>and wide band noise | Pure tones at<br>audiometric<br>frequencies, narrow<br>and wide band noise | Same |
TABLE 6-3: COMPARISON OF TEST MODULE TYMP
## OVERALL CONCLUSION
The clinical tests showed that Sentiero can be used by professionals as efficiently as the respective predicate devices. The nonclinical tests shows that Sentiero is as similar safe, as effective and as performant as the predicate devices.
| Target population | The devices are<br>suitable for all<br>populations including<br>new-born infants | The devices are<br>suitable for all<br>populations including<br>new-born infants | Same |
|------------------------------------------|----------------------------------------------------------------------------------|----------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Intended user | The devices are to be<br>used by trained<br>personnel only | The devices are to be<br>used by trained<br>personnel only | Same |
| Safety standard | IEC 60601-1 | IEC 60601-1 | Same |
| EMC standard | IEC 60601-1-2 | IEC 60601-1-2 | Same |
| Standards for<br>impedance<br>audiometry | IEC 60645-5/ANSI<br>S3.39, Type 1 | IEC 60645-5/ANSI<br>S3.39, Type 1 | Same |
| Test signal | IEC 60645-1/ANSI<br>S3.6, IEC 60645-3 | IEC 60645-1/ANSI<br>S3.6, IEC 60645-3 | Same |
| Probe tone frequency<br>(tympanometry) | 226 Hz. Optional 678<br>Hz, 800 Hz and 1000<br>Hz | 226 Hz. 678 Hz, 800<br>Hz and 1000 Hz<br>optionally available | Same |
| Probe tone level<br>(tympanometry) | 69 dB HL with AGC | 69 dB HL with AGC | Same |
| Pressure safety limits | $-750...+550$ daPa | $-700...+500$ daPa | Equivalent<br>Safety limits need to be<br>within the range of -<br>$-800...+600$ daPa<br>according to IEC<br>60645-5/ANSI S3.39, a<br>smaller range means<br>increased safety |
| Transducers | Ear probe for<br>tympanometry, ear<br>probe and<br>headphone/insert ear | Ear probe for<br>tympanometry, ear<br>probe and<br>headphone/insert ear | Equivalent<br>Different models of ear<br>probe and headphones<br>are used, all of them |
{8}------------------------------------------------
| | phone for reflex<br>measurement | phone for reflex | generating test signals<br>complying to IEC<br>60645-5/ANSI S3.39<br>standards |
|----------------|----------------------------------------------------------------------------|----------------------------------------------------------------------------|--------------------------------------------------------------------------------|
| Reflex stimuli | Pure tones at<br>audiometric<br>frequencies, narrow<br>and wide band noise | Pure tones at<br>audiometric<br>frequencies, narrow<br>and wide band noise | Same |
TABLE 10-3: COMPARISON OF TEST MODULE TYMP
### OVERALL CONCLUSION
Sentiero shows similar safety, effectiveness and performance data as the respective predicate devices.
### CONCISE SUMMARY OF PERFORMANCE TESTING
Sentiero (all types and accessories) was tested by independent external laboratories with respect to the following standards:
- · IEC 60601-1:1988
- · IEC 60601-1:2005
- · · IEC 60601-1-2:2007
- · IEC 60601-2-40:1998
Safety Safety Electromagnetic compatibility Safety of electromyographs and devices for evoked potentials
PATH medical GmbH is certified according to Medical Device Directive 93/42 ECC and holds certificates for compliance with respect to the following standards:
- · ISO 9001:2008 Quality management system
- · ISO 13485:2003 Quality management system (medical devices)
Basing on this, reqular audits and compliance checks are performed by the following notified bodies:
- · Notified body for certification in Europe: DEKRA Certification GmbH (Germany), (see certificate in the Appendix).
- · Notified body for certification in Canada: DEKRA Certification B.V. (Netherlands), (see certificate in the Appendix).
Performance tests were also conducted on a test bench. For this reason a simulator device, which simulates TEOAE, DPOAE or ABR output was used for proof of validity of the derived signal. The same simulator device was distributed by Natus Europe GmbH as a testing device for the predicate devices AccuScreen and Cochlea-Scan.
The device under test (DUT) had to be connected to deliver the stimulus (via insert earphone or ear probe) into a cavity of the simulator. The simulator was connected to the electrode cable of the DUT to record the test result. The stimulator was able to deliver noise or dedicated normative answers derived from literature. The same simulator was used for Sentiero as well as the predicate devices Cochlea-Scan and AccuScreen. The simulator was developed by the same engineers who developed the Cochlea-Scan as well as the AccuScreen. Passing the requirements given by the simulator was part of the acceptance criteria for the verification phase of the device development cycle. Consequently the performance of the test
modules TEOAE, DPOAE, DPTHRES. ABR were validated in comparison to the predicate devices during the in-house validation phase of Sentiero. The test protocol is attached in the Appendix.
Additionally Sentiero was tested against the following standards to show compliance and performance as follows:
- · ISO 10993-1:2009 Biocompatibility
- · IEC 60601-1-4:1996 Programmable electrical system
- · IEC 60601-2-40:1998 Safety of electromyographs and devices for evoked potentials
- · IEC 62304:2006 Software lifecycle
{9}------------------------------------------------
- · IEC 60645-1:2001 Pure-tone audiometer
- · IEC 60645-6:2009 Otoacoustic emissions
- · IEC 60645-7:2009 Acoustically evoked potentials
- · ISO 389-1:1998 Reference levels for pure tones and supra aural headphones
- · ISO 389-2:1994 Reference levels for pure tones and insert earphones
- · ISO 389-3:1994 Reference force levels for pure tones and bone conductors
- · ISO 389-4:1994 Reference levels for narrow band masking noise
- · ISO 389-5:2006 Reference levels for pure tones from 8 to 16 kHz
- · ISO 389-8:2004 Reference levels for pure tones and circumaural headphones
Sentiero successfully passed the tests and checklists of the aforementioned standards. Thus it is shown that Sentiero is as safe, efficient and performing as the predicate devices which also fulfill the standards as mentioned in the comparison table above.
In preparation of the first application for receiving the CE mark for Sentiero, additional test sites were asked to evaluate the performance of relevant test modules and the handling of the instrument. To state clearly: all implemented modules on Sentiero are derived from well-known algorithms, which are described in detail in published literature. The design process of Sentiero included in-house verification tests also with respect to the mentioned literature and and establishment of normative data. In addition, external potential users were asked to validate the measurement results and the usage of the device and test modules (usability tests and performance tests). Comparison measurements had to be done with predicate devices and normal hearing patients with hearing loss. Minimum number of participants per evaluation was 20. Multiple users should be involved to verify usability of the device in comparison to experience level of the user.
No tests were conducted with respect to animal testing.
The test sites, which participated in the evaluation of test modules on Sentiero are shown in Table 10-4. Please note that not all these testers and studies did receive financial support from PATH medical. Most of the evaluation data and feedback was contributed by initiative from the tester. The tester a financial arrangement with PATH medical - except the two sites mentioned in section 9 of this document, where PATH was the sponsor and initiator of the study.
| Sentiero Module | Predicate device | Tester |
|-------------------------------------------------|------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------|
| PTA class 3 | Cochlea-Scan<br>(DPOAE<br>threshold<br>estimation) | Prof. Dr. med. Annerose Keilmann |
| MAGIC (image-<br>based pure tone<br>audiometry) | | Center of Communication Disorders, ENT Clinic,<br>University Medical Center of the Johannes-<br>Guttenberg-University Mainz, Mainz, Germany |
| MATCH (image-<br>based speech test) | Established<br>speech tests<br>(e.g. Mainzer<br>speech test for<br>children) | |
| DPTHRES (DPOAE<br>threshold estimation) | | |
| ABR | Chartr EP 200 | Ph.D. Stavrous Hatzopoulos |
| | | Department of Audiology and Speech Therapy,<br>University of Ferrara, Ferrara, Italy |
| TEOAE | AccuScreen<br>(TEOAE, DPOAE) | Prof. Dr. med. Hans Peter Niedermeyer |
| DPQUICK (DPOAE<br>at fixed levels) | | ENT Clinic, Klinikum rechts der Isar, Technische |
| DPTHRES (DPOAE<br>threshold estimation) | | Universität München, Munich, Germany |
| ASSR | Cochlea-Scan<br>(DPOAE<br>threshold<br>estimation) | Dr. Thomas Rosner |
| | ECLIPSE | ENT Clinic, Klinikum rechts der Isar. Technische |
{10}------------------------------------------------
| Sentiero Module | Predicate device | Tester |
|-------------------------------------------------|----------------------------------------------|----------------------------------------------------------------------------------------------------|
| | | Universität München, Munich, Germany |
| ABR | ABR device,<br>unknown type. | PD Dr. med. Nicolas Schmuziger |
| | | ENT Clinic, Kantonsspital Liestal, Liestal, Switzerland |
| PTA | EchoScreen/Accu<br>Screen and<br>CochleaScan | Prof. Dr. Katrin Neumann |
| MAGIC (image-<br>based pure tone<br>audiometry) | | Department of Phoniatry and Pedaudiology. ENT<br>Clinic, St .- Elisabeth-Hospital, Bochum, Germany |
| TEOAE | | |
| DPQUICK (DPOAE<br>at fixed levels) | | |
| DPTHRES (DPOAE<br>threshold estimation) | | |
| ABR | | |
Table 10-4: List of external tests including tested module and the corresponding tester
Out of these studies reports were generated or published, which indicated that Sentiero is an equivalence to the given predicate devices with respect to clinical application/usage.
{11}------------------------------------------------
Image /page/11/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services (HHS). The logo features a stylized depiction of an eagle or bird in flight, composed of three curved lines that suggest movement and dynamism. The bird is positioned above a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA".
### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring. MD 20993-002
June 9, 2014
PATI Medical GmbH c/o Dr. Johann Oswald Director of PATH Medical GmbH Landsberger Str.63 Germering. Bavaria GM D-82110
Rc: K133012
Trade/Device Name: Sentiero Regulation Number: 21 CFR 874.1050 Regulation Name: Audiometer Regulatory Class: Class II Product Code: EWO. GWJ. ETY Dated: May 1. 2014 Received: May 6. 2014
#### Dear Dr. Oswald:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into cither class II (Special Controls) or class III (PMA). it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition. FDA may publish further announcements concerning your device in the Federal Register.
{12}------------------------------------------------
Page 2 - Dr. Johann Oswald
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Eric A. Mann -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{13}------------------------------------------------
DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration Indications for Use
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known)
Device Name
Sentiero
MICRO
Indications for Use (Describe)
Sentiero is a portable instrument to diages for hearing loss. The instrument offers different test nethods which can be configured to fit the professional's needs for screening or diagnostic purposes. It offers physiological test methods such as:
· Distortion Product Otoacoustic Emissions (DPOAF)
· Transient Evoked Otoacoustic Emissions (TEOAF)
· Auditory Brainstem Response (ABR)
· Auditory Steady State Response (ASSR)
· Auditory Impedance and acousiic relles (TYMP) Additionally if offers standard audiometry (psycho-acousiveal). All physiological test methods are especially indicated for use in defining the type and configuration of hearing loss particularly for individuals whose behavioral audiometric results are deemed unreliable or to assist in the diagnosis of otologic disorders. Estination of cochlear hearing thresholds (DP THRESI) is possible at various frequencies without the need of cooperative interaction with the pairent. Acoussic rellex and tympanomery (TYMP) are featured the functional condition of the middle and outer car. For each method, several protocols can be configured. The results are to be used to make further recommendations regarding appropriate interention strategies. Therefore, Sentiero is intended for use by trained personnel such as audiologiss, pediatricians. ENT doctors and other health care professionals. In the United States of America. Federal law restricts this device to sale by or on the order of a licensed physician.
A vailable psycho-acoustical methods on Sentiero are especially indicated for use with cooperative parting at the age of 2 years or adequate development age, which enables them to do playinters. All other modules are suitable to be used for all ages.
Sentiero is designed for:
1. Diagnostics, monitoring and follow-up after newborn hearing screening
2. Pre-school, school. and adult hearing screening
3. ENT diagnostics based on measurement of
· Otoacoustic emissions
· Tympanometry and acoustic reflex
· Auditory Brainstem Responses
· Auditory Steady State Responses
Sentiero must not be used in cases of external otitis (outer car canal intection) or in any case which yields to pain when inserting the ear probe or applying any other transducer.
Type of Use (Select one or both, as applicable)
Prescription Use (Part 21 CFR 801 Subpan D)
Over-The-Counter Use (21 CFR 801 Subpart C)
{14}------------------------------------------------
## PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
| FOR FDA USE ONLY<br>Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) | |
|--------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------------|
| | |
| | This section applies only to requirements of the Paperwork Reduction Act of 1995. |
*DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response. including the time to review instructions. search existing data sources. gather and mainlain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@ida.hhs.gov
*An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.