The Counter Pain Therapy Device Model P.T.S.-IV is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household work activities.
Device Story
Handheld TENS device for OTC pain relief; features two channels and seven preset programs. User adjusts intensity in 16 steps; treatment time selectable (20, 30, or 40 minutes). Device consists of main controller unit, two pairs of self-adhesive electrodes, and electrode cables. Powered by three internal rechargeable lithium batteries. Used in clinics, hospitals, or home environments by patients for muscle pain relief. Output displayed on backlit LCD screen. Provides electrically generated pulses applied to skin via electrodes to stimulate nerves/muscles. Benefits patient by providing temporary pain relief from muscle strain.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, including electrical safety and EMC testing per IEC 60601-1 and IEC 60601-1-2, and software verification.
Technological Characteristics
TENS device; two-channel output; 16 intensity levels; 7 preset programs. Powered by internal rechargeable lithium batteries. Includes backlit LCD. Electrodes are the only body-contacting components. Conforms to IEC 60601-1 (safety) and IEC 60601-2-10 (nerve/muscle stimulators).
Indications for Use
Indicated for temporary relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household and work activities. Contraindicated for patients with cardiac pacemakers, implanted defibrillators, or other implanted metallic/electronic devices; patients using life-supporting medical electronic devices (e.g., artificial heart/lung); or patients using body-worn medical electronic devices (e.g., ECG).
Regulatory Classification
Identification
A transcutaneous electrical nerve stimulator for pain relief is a device used to apply an electrical current to electrodes on a patient's skin to treat pain.
Predicate Devices
OMRON Electro Therapy Pain Relief, Model PM3030 (K110068)
Submission Summary (Full Text)
{0}------------------------------------------------
K132993
APR 2 4 2014
# 510(k) SUMMARY
Date of Summary Prepared: 01/24/2014
1. Submitter's Name: Counter Scientific Development (GZ) Ltd.
2nd & 3rd Floor, Blg. 6, DaPian Industrial Zone Address: Tangxia Road, Tianhe District, Guangzhou 510665, China
Contact person: Wendi Poon, Manager
Tel: +86-20 -8328 9980
Fax: +86-20-38339500
E-mail Counter_csd@yahoo.com.cn
#### 2. Proposed New Device:
Trade Name: Pain Therapy Device, Model P.T.S .- IV Classification Name: TENS Device Regulation Number: 882.5890 Product Code: NUH Device Class: II
#### 3. Predicate (cleared) Device:
OMRON Electro Therapy Pain Relief, Model PM3030
510(k) Number: K110068
Manufacturer: OMRON Healthcare Co., Ltd., China.
{1}------------------------------------------------
#### 4. Description of Proposed Device:
The Counter Pain Therapy Device, Model P.T.S.-IV features two channels and seven (7) preset programs. The output of both channels can be adjusted individually for intensity in sixteen (16) steps and the treatment time is selectable from 20, 30 or 40 minutes. All features are well indicated by large icons on a backlit LCD screen for easy reading.
The device is powered by three (3) rechargeable Lithium Batteries which are permanently built into the main unit and can be recharged with the charging unit supplied for the device.
The Pain Therapy Device consists of the handheld main unit (controller), two pairs of self-adhesive electrodes, a plug-in charging unit (adaptor 120VAC to 5VDC) and two pairs of electrode cables.
The device has been tested to and meets the requirements of the following recognized consensus standards:
IEC 60601-1:1988 + A1:1991 + A2:1995 "Medical electrical Equipment - Part 1: General requirements for basic safety and essential performance".
IEC 60601-2-10: 1987 (first Edition + amendment 1 (2001) for use in conjunction with IEC 60601-1 (1988). Amendments 1 (1991) and 2 (1995) "Medical electrical equipment -Part 2: "Particular Requirements for safety of nerve and muscle stimulators".
IEC 60601-1-2: 2001 + A1: 2004 & IEC60601-2-10: 1987 + A1: 2001 Clause 36 Medical Electrical Equipment, Part1-2: "General requirements for Basic Safety and Essential Performance" Collateral Standard: Electromagnetic Compatibility.
#### 5. Indication for Use:
The Counter Pain Therapy Device Model P.T.S.-IV is intended for the relief of pain associated with sore or aching muscles of the lower back, arms, or legs due to strain from exercise or normal household work activities.
2
{2}------------------------------------------------
#### 6. Environment of Use:
#### Clinics, Hospitals and home environments
#### 7. Contraindications:
Do not use this device if you have a cardiac pacemaker, implanted defibrillator, or other implanted metallic or electronic device. Such use could cause electrical shock, burns, electrical interference, or death.
Do not use this device together with a life-supporting medical electronic device such as an artificial heart or lung.
Do not use this device together with a body-worn medical electronic device such as an ECG.
#### 8. Technological Characteristics Compared to the Predicate Device
Both, the Counter Pain Therapy Device Model P.T.S.-IV and the predicate device the OMRON TENS Model PM3030 have the same intended use and fundamental technology. Both the Counter unit and the predicate provide electrically generated pulses applied to the skin via electrodes. The main differences between the two devices are that the Counter Pain Therapy Device has 7 program modes versus 3 on the OMRON unit and 2 channels versus 1 channel. However, the Counter device can be viewed as substantially equivalent to the predicate device because: The electrical stimulation provided by the Counter device is substantially equivalent to that commonly employed by TENS devices that have been cleared for marketing without prescription labeling; i.e. for OTC sale. Technological characteristics, features, specifications, materials and intended uses of the Counter device are substantially equivalent to the quoted predicate device. The differences that exist between the devices are insignificant in terms of safety or effectiveness.
# 9. Discussion of Non-Clinical Tests Performed for Determination of Substantial Equivalence is as follows:
Counter did not conduct, nor rely upon, clinical tests to determine substantial equivalence of the Pain Therapy Device Model P.T.S.-IV. vs. the predicate. Non-clinical testing was performed in order to validate the design according to the company's specified design requirements, and to assure conformance with the following voluntary design standards:
IEC 60601-1:1988 + A1:1991 + A2:1995 "Medical electrical Equipment - Part 1: General requirements for basic safety and essential performance".
3
{3}------------------------------------------------
IEC 60601-2-10: 1987 (first Edition + amendment 1 (2001) for use in conjunction with IEC 60601-1 (1988), Amendments 1 (1991) and 2 (1995) "Medical electrical equipment -Part 2: "Particular Requirements for safety of nerve and muscle stimulators".
IEC 60601-1-2: 2001 + A1: 2004 & IEC60601-2-10: 1987 + A1: 2001 Clause 36 Medical Electrical Equipment. Part1-2: "General requirements for Basic Safety and Essential Performance" Collateral Standard: Electromagnetic Compatibility.
## a. EMC and electrical safety
IEC 60601-1 "Medical Electrical Equipment-Part 1: General requirements for basic safety and essential performance ".
IEC 60601-1-2 "Medical Electrical Equipment - Part 1-2: General requirements for basic safety and essential performance-collateral standard: Electromagnetic compatibility-requirements and tests".
#### b. Biocompatibility
The electrodes are the only body contacting parts. They have been tested for biocompatibility as part of their separate FDA 510(k) clearance.
#### c. Software
Based upon the test results it was concluded that the software performs within specifications and is safe to the stated intended use. Since a permanent hazard analysis is implemented in the software development process, and due to the clear software architecture, it is believed that the test protocol sufficiently verifies the software's main functional operation.
#### d. Cleaning
The cleaning instructions as described the Instruction Manual have been tested to be sufficient. Testing involved validation of the manual cleaning method as per the instructions. All testing concluded that that the Pain Therapy Device can be cleaned by the use of the methods described in the Instruction Manual.
## e. EMC and Electrical Safety
Performance testing has established that, with respect to EMC and electrical safety in its intended operational environment, the device conforms to all applicable
4
{4}------------------------------------------------
requirements of IEC 60601-1 and IEC 60601-1-2.
# 8. Conclusions:
The Counter Pain Therapy Device, Model P.T.S.-IV has the same intended use and technological characteristics as the predicate device. Moreover, bench testing and safety report documentation demonstrate that the submitted device could maintain the same safety and effectiveness as that of predicate device. In the other words, the differences do not affect the intended use and do not raise any new questions of safety or effectiveness or alter the fundamental scientific technology of the device. Thus, the Counter Pain Therapy Device, Model P.T.S.-IV is substantially equivalent to the predicate device.
5
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized human figure with outstretched arms, enclosed within a circle. The text "DEPARTMENT OF HUMAN SERVICES - USA" is arranged around the perimeter of the circle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
# April 24, 2014
Counter Scientific Development (GZ) Ltd c/o Guenter Ginsberg, President Media Trade Corporation 11820 Red Hibiscus Drive Bonita Springs, FL 34135
Re: K132993
Trade/Device Name: Counter OTC TENS Device, Model PTS-IV Regulation Number: 21 CFR 882.5890 -Regulation Name: Transcutaneous Electric Nerve Stimulator for Pain Relief Regulatory Class: Class II Product Code: NUH Dated: March 15, 2014 Received: March 18, 2014
Dear Mr. Ginsberg:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28. 1976. the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations. Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{6}------------------------------------------------
CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
:"
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportalProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Image /page/6/Picture/7 description: The image shows a white background with several small, irregularly shaped black spots scattered across it. The spots are concentrated in the upper right and lower right quadrants of the image, with a few isolated spots elsewhere. The spots vary in size and shape, with some appearing as single dots and others as clusters of dots.
Sincerely yours,
# Felipe Aquel -S
Carlos L. Peña, Ph.D., M.S. Director Division of Neurological and Physical Medicine Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known) K132993
Device Name Counter OTC TENS Device, Model PTS-IV
## Indications for Use (Describe)
This TENS Device is to be used for temporary relief of pain associated with sore or aching muscles of lower back, arms, or legs due to strain from exercise or normal household and work activities.
Type of Use (Select one or both, as applicable)
] Prescription Use (Part 21 CFR 801 Subpart D)
> Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
# FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Date: 2014.04.24 11:58:31 -04'00'
FORM FDA 3881 (9/13)
Form Approved: OMB No. 0910-0120 Expiration Date: December 31, 2013 See PRA Statement on last page.
{8}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995.
#### *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW.*
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff(@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
"
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.