ARTISTE SOLUTION WITH SYS_VC10C AND SYNGO RT THERAPIST WORKSPACE, V4.3.1 UPGRADE KIT, ARIA CONNECTIVITY OPTION FOR SIEME

K132935 · Siemens Medical Solutions USA, Inc. · IYE · Feb 21, 2014 · Radiology

Device Facts

Record IDK132935
Device NameARTISTE SOLUTION WITH SYS_VC10C AND SYNGO RT THERAPIST WORKSPACE, V4.3.1 UPGRADE KIT, ARIA CONNECTIVITY OPTION FOR SIEME
ApplicantSiemens Medical Solutions USA, Inc.
Product CodeIYE · Radiology
Decision DateFeb 21, 2014
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.5050
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc therapy (mARC) and precision stereotactic radiation therapy for lesions, turnors and conditions anywhere in the head and body where radiation therapy is indicated. The syngo® RT Therapist workspace is a component of the linear accelerator system and is based on the syngo@ architecture. The syngo@ RT Therapist workspace contains software applications to support patient selection/setup, patient position, treatment delivery/verification. and treatment recording. The syngo® RT Therapist Workspace v4.3.1, can be interfaced with third party devices conforming to the DICOM Standard.

Device Story

syngo® RT Therapist Workspace v4.3.1 is a software component for Siemens ARTISTE, ONCOR, and PRIMUS linear accelerators; supports patient setup, positioning verification, treatment delivery/verification, and recording. Operates within clinical radiation oncology environments; used by radiation therapists/clinicians. Integrates with third-party Oncology Information Systems (OIS), Verification & Recording (V&R) systems, Treatment Planning Systems (TPS), and PACS via DICOM. Facilitates clinical workflow by managing patient data and treatment parameters; ensures accurate delivery of radiation therapy. Benefits patients by enabling precise, verified radiation treatment for cancer lesions and tumors.

Clinical Evidence

Bench testing only. Software verification and regression testing performed at unit, integration, and system levels on production prototype devices. Testing confirmed requirements in System Requirements Specifications (SRS) and Component Requirements (CRS) were met. No clinical data presented.

Technological Characteristics

Software-based workspace utilizing proprietary syngo® architecture. Supports DICOM connectivity for third-party OIS/V&R/TPS/PACS integration. Features include patient setup, position verification, treatment localization, portal imaging, data processing, and image reformatting. Includes Siemens Remote Service (SRS) and virus protection. Operates on Siemens linear accelerator control consoles.

Indications for Use

Indicated for patients requiring X-ray photon and electron radiation therapy for cancer treatment, including 3D conformal, IMRT, mARC, and stereotactic radiation therapy for lesions, tumors, and conditions anywhere in the head and body.

Regulatory Classification

Identification

A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K132935 Page 1 of 5 FEB 2 1 2014 # Section 5 # 510(k) Summary | Date Prepared: | | September 12, 2013 | | | |--------------------------|----------------------------------------------------------------|------------------------------------------------------------------------------------------------------|--|--| | Submitter: | | Siemens Medical Solutions USA, Inc.<br>Radiation Oncology<br>757A Arnold Drive<br>Martinez, CA 94553 | | | | Contact: | | Christine Dunbar<br>Senior Manager, Regulatory Affairs | | | | Phone:<br>Fax:<br>Email: | (925)293-5442<br>(925)602-8008<br>christine dunbar@siemens.com | | | | | Proprietary Name: | | syngo® RT Therapist Workspace, v4.3.1<br>And the optional 30 party OIS Connectivity System | | | | Common Name: | | Accessory To; Medical Charged-Particle Radiation Therapy System | | | | Classification: | | 892.5050 | | | | Product Code: | | IYE | | | | | | | | | Substantial Equivalence Claimed To: | Product | 510(k)<br>Clearance / Date | Claim of Equivalence for: | |----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-----------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | COHERENCE™ RT Therapist<br>Connect update to Sys_32A<br>(RTTC v2.3) with Control<br>Console 9.2.18 and 11.0.212<br>supporting 3rd party OIS, V&R<br>called ARIA and Varian<br>Treatment8 and third party TPS<br>and PACs systems | K123812 /<br>March 01, 2013 | Syngo® RT Therapist with software<br>version 4.3.1, with Control Console v13,<br>an update to Sys_VC10C supporting 3rd<br>party OIS, V&R called ARIA and Varian<br>Treatment9 and third party TPS and<br>PACs systems | | ARTISTE™ Solution with SYS-<br>VC10A for the syngo® RT<br>Therapist (RTT v4.3, CC 13)<br>supporting 3rd party OIS, V&R | K121295 / June<br>12, 2012 | Syngo® RT Therapist with software<br>version 4.3.1, with Control Console v13,<br>an update to Sys_VC10C supporting 3rd<br>party OIS, V&R called ARIA and Varian | 8 ARIA OIS and Varian Treatment V&R systems are manufactured by Varian Medical Systems. ARIA OIS and Varian Treatment Vart Systems are manufactured by Varian Medical Systems. 9 ARIA OIS and Varian Treatment V&R systems are manufactured by Varian Medical Systems. 510(k) for SIEMENS syngo® RT Therapist v4.3.x and 3rd Party OlS Section/ Page - 5/1 {1}------------------------------------------------ | called MOSAIQ and Sequencer<br>V&R10 and third party TPS and<br>PACs systems. | | Treatment11 and third party TPS and<br>PACs systems | |---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | syngo® RT Therapist Connect<br>(RTT v4.2, CC 12)<br>(ARTISTE, ONCOR and<br>PRIMUS Linacs) supporting 3rd<br>party OIS, V&R called MOSAIQ<br>and Sequencer V&R12 and third<br>party TPS and PACs systems | K103606 /<br>April15, 2011 | syngo® RT Therapist with software<br>version 4.3.1, with Control Console v13,<br>an update to Sys_VC10C supporting 3rd<br>party OIS, V&R called ARIA and Varian<br>Treatment13 and third party TPS and<br>PACs systems. | The update to the syngo® RT Therapist Workspace, v4.3.1 as described in this premarket notification has the same intended use and fundamental scientific technical characteristics as the predicate devices listed above. ## Description Summary ## syngo® RT Therapist Workspace, v4.3.1 ## Technological Characteristics: The syngo® RT Therapist Workspace v4.3.1 release is intended to update customers with the currently released syngo® RT Therapist Workspace software with version v4.3 for ARTISTE, ONCOR & PRIMUS Linear Accelerator systems. The technological characteristics and the fundamental technology of the syngo® RT Therapist Workspace v4.3.1 remain unchanged from the currently cleared products. ## The syngo® Software Architecture: The syngo® RT Therapist software utilizes the proprietary syngo® software architecture design that provides a method of delivering customized software applications based on the modality as clinically supporting packages. SIEMENS utilizes, as part of the Oncology clinical focus package, multiple syngo® software applications for patient set-up and position verification, treatment localization, treatment verification, portal imaging as well as data processing, image reformatting, display and printing. The currently cleared syngo® based software products also includes an array of image-oriented software tools, support for DICOM connectivity, Siemens Remote Service [SRS], and virus protection features. Note: syngo® is both a proprietary software architecture and a market brand. This update is intended to be backwards compatible to upgrade the currently cleared ONCOR and PRIMUS family of medical linear accelerators and their Control Consoles (v9.2+ & v11.0, v12), the COHERENCE™ RT Therapist (v2.3) workspace (K123812). Additionally, the ARTISTE 11 ARIA OIS and Varian Treatment V&R systems are manufactured by Varian Medical Systems. 12 MOSAIQ OIS and Sequencer V&R systems are manufactured by IMPAC Medical / Elekta AB. 13 ARIA OIS and Varian Treatment V&R systems are manufactured by Varian Medical Systems. 510(k) for SIEMENS syngo® RT Therapist v4.3.x and 3d Party OlS Section/ Page - 5/2 <sup>10</sup> MOSAIQ OIS and Sequencer V&R systems are manufactured by IMPAC Medical / Elekta AB. {2}------------------------------------------------ Solution with the syngo® RTT v4.1x and v4.2x (K103606) can also be upgraded. These previous versions can be migrated to the current release (syngo® RTT. v4.3) and Control Console 13. #### General Safety and Effectiveness: The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device. Risk management is ensured via a risk analysis, which is used to identify potential hazards and mitigations. These potential hazards are controlled by software means, user instructions, verification of requirements and validation of the clinical workflow to ensure that the product meets its intended uses. To minimize electrical, mechanical and radiation hazards, SIEMENS adheres to recognized and established industry practice and relevant international standards. Refer to Section 21 for the Risk Management documentation. #### Intended Use: The intended use of the SIEMENS branded ARTISTE™ , ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-Ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc therapy (mARC) and precision stereotactic radiation therapy for lesions, turnors and conditions anywhere in the head and body where radiation therapy is indicated. The syngo® RT Therapist workspace is a component of the linear accelerator system and is based on the syngo® architecture. The syngo® RT Therapist workspace contains software applications to support patient selection/setup, patient positioning verification, treatment delivery/verification, and treatment recording. The syngo® RT Therapist Workspace v4.3.1 can be interfaced with third party devices conforming to the DICOM Standard. #### Substantial Equivalence: The Substantial Equivalence comparison chart demonstrates the comparison of the technological characteristics of the synqo® RT Therapist Workspace update to the currently cleared predicate devices. The syngo® RT Therapist Workspace v4.3.1 (project Sys_VC10C) does not change the intended use of the original ARTISTE™ Solution with the syngo® RT Therapist Workspace v4.3 (project Svs VC10A) or the previously cleared ONCOR™ or PRIMUS™ Siemens branded Linear Accelerator Systems. #### Bench Testing: {3}------------------------------------------------ Bench testing in the form of Unit, Integration and System Integration testing was performed to evaluate the performance and functionality of the software update v4.3.1 for the RT Therapist and regression testing the Control Console version 13. All testable requirements in the System Requirements Specifications (SRS) and Component Requirements (CRS) for the Sys_VC10C project, and additionally the specific requirements for the implementation of the third party OIS have been successfully verified and traced in accordance with the Siemens product development (lifecycle) process (PDP). The software verification and regression testing has been performed successfully to meet their previously determined acceptance criteria as stated in the Test Concept. #### Non-Clinical Test Results: Validation of the syngo® RT Therapist Workspace v4.3.1, implementation of the optional third party OIS and regression testing with Control Consoles 13 has been performed at the System test level on production prototype devices by appropriately trained and knowledgeable test personnel. System level validation and regression testing has been performed successfully, demonstrating that the software meets the acceptance criteria as noted in the system test plans. #### Testing to Consensus Standards: The syngo® RT Therapist Workspace v4.3.1 and the Control Console 13 supporting the third party OIS and V&R implementation have been tested to meet the requirements for conformity (where applicable) to multiple industry standards. Refer to Sections 9 and 17 for this content. #### Substantial Equivalence to Predicates: The verification testing to the System requirements for the syngo® RT Therapist workspace, validation of the intended use, and the regression testing to the existing syngo® RT Therapist software and Control Console functional requirements, is intended to support the claim of substantial equivalence to the following predicates: - The COHERENCE™ RT Therapist Connect v2.3 with Control Consoles 9.2.18 and . 11.0.212 supporting 3d party OIS & V&R (ARIA and Varian Treatment™ System) and third party TPS and PACs systems (K123812). - The ARTISTE™ Solution with SYS-VC10A for the syngo® RT Therapist (RTT v4.3, . CC 13) supporting 310 party OIS & V&R (MOSAIQ and Sequencer V&R15) and third party TPS and PACs systems. (K121295). - The syngo® RT Therapist Connect (RTT v4.2, CC 12) (ARTISTE, ONCOR and . PRIMUS Linacs) supporting 30 party OIS & V&R (MOSAIQ and Sequencer V&R10) and third party TPS and PACs systems (K103606). <sup>14</sup> ARIA OIS and Varian Treatment V&R systems are manufactured by Varian Medical Systems. <sup>15</sup> MOSAIQ OIS and Sequencer V&R systems are manufactured by IMPAC Medical / Elekta AB. <sup>16</sup> MOSAIQ OIS and Sequencer V&R systems are manufactured by IMPAC Medical / Elekta AB. <sup>510(</sup>k) for SIEMENS syngo® RT Therapist v4.3.x and 3rd Party OIS Section/ Page - 5/4 {4}------------------------------------------------ K132935 page 5 of 5 ## Summary: In summary, it is SIEMENS' opinion that the syngo® RT Therapist Workspace update to v4.3.1 including the optional 3" party OIS Connectivity system (ARIA OIS), does not introduce any new potential safety risks and is substantially equivalent to, and performs as well devices. .... {5}------------------------------------------------ Image /page/5/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" around the perimeter. Inside the circle is a stylized symbol that resembles a human figure embracing a three-barred cross, which is the symbol of the Public Health Service. ### DEPARTMENT OF HEALTH & HUMAN SERVICES Public Health Service Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002 #### February 21, 2014 SIEMENS MEDICAL SOLUTIONS USA, INC. % CHRISTINE DUNBAR SENIOR MANAGER, REGULATORY AFFAIRS 757A ARNOLD DRIVE MARTINEZ CA 94553 Re: K132935 Trade/Device Name: syngo® RT Therapist Workspace, v4.3.1 & optional 3d party OIS Connectivity System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: IYE Dated: October 4, 2013 Received: October 7, 2013 Dear Ms. Dunbar: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {6}------------------------------------------------ Page 2-Ms. Dunbar and the country of the states a If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Sincerely vours. Jaming Jr. Marge Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration ndications for Use 510(k) Number (if known) K132935 ## Device Name syngo RT Therapist Workspace, v4.3.1 Indications for Use (Describe) The intended use of the SIEMENS branded ARTISTE™, ONCOR™ and PRIMUS™ family of linear accelerator systems is to deliver X-ray photon and electron radiation for the therapeutic treatment of cancer. The linear accelerator systems are high-dose and high-dose rate medical linear accelerators optimized for 3D conformal radiation therapy, intensity-modulated radiation therapy (IMRT), modulated arc therapy (mARC) and precision stereotactic radiation therapy for lesions, turnors and conditions anywhere in the head and body where radiation therapy is indicated. The syngo® RT Therapist workspace is a component of the linear accelerator system and is based on the syngo@ architecture. The syngo@ RT Therapist workspace contains software applications to support patient selection/setup, patient position, treatment delivery/verification. and treatment recording. The syngo® RT Therapist Workspace v4.3.1, can be interfaced with third party devices conforming to the DICOM Standard. Type of Use (Select one or both, as applicable) 2 Prescription Use (Part 21 CFR 801 Subpart D) Over-The-Counter Use (21 CFR 801 Subpart C) Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement on last page. #### PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED. FOR FDA USE ONLY Concurrence of Center for Devices and Radiological Health (CDRH) (Signature) Michael D. O'HaraFORM FDA 3881 (1/14) PSC Publishing Sen sees (30) ) 441-47 an E {8}------------------------------------------------ This section applies only to requirements of the Paperwork Reduction Act of 1995. ### ... . . . . . . . . "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW." The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to: > Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov "An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...