K132918 · Sense4baby , Inc. · HGM · Nov 22, 2013 · Obstetrics/Gynecology
Device Facts
Record ID
K132918
Device Name
SENSE-4BABY SYSTEM
Applicant
Sense4baby , Inc.
Product Code
HGM · Obstetrics/Gynecology
Decision Date
Nov 22, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.2740
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
The Sense4Baby System Model B+ is indicated for monitoring of maternal and fetal physiologic parameters during the antepartum period. It is to be used by health care professionals in hospitals, clinics, physicians' offices, antepartum rooms, and in the patient's home on the order of a physician.
Device Story
Non-invasive, wireless, external fetal monitor; measures fetal heart rate, maternal heart rate, and uterine contractions during antepartum testing. System transmits data to a gateway device, then to a HIPAA-compliant web-based portal for storage and physician review. Used in hospitals, clinics, physicians' offices, and patient homes under physician order. Healthcare professionals operate the system to review physiologic data; output supports clinical decision-making during antepartum monitoring. Benefits include remote monitoring capability, reducing need for in-person clinic visits.
Clinical Evidence
Bench testing only. Verification and validation activities included simulated use, hardware, software, environmental, mechanical, water ingress, and shipping testing. Compliance with IEC 60601-1, IEC 60601-1-2, IEC 60601-2-37, ISO 10993, and ISO 14971 demonstrated safety and performance equivalent to the predicate.
Technological Characteristics
Wireless, non-invasive fetal/maternal monitor. Components include sensors for heart rate and uterine contractions, gateway device, and web-based portal. Standards: IEC 60601-1 (electrical safety), IEC 60601-1-2 (EMC), IEC 60601-2-37 (ultrasonic monitoring), ISO 10993 (biocompatibility), ASTM D4169 (shipping), IEC 60529 (IP code). Software developed under ISO 13485 and ISO 14971 risk management processes.
Indications for Use
Indicated for monitoring maternal and fetal physiologic parameters (heart rate, uterine contractions) during the antepartum period in patients requiring nonstress testing.
Regulatory Classification
Identification
A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.
{0}------------------------------------------------
Sense 4 Baby. Inc.
## 510(k) Summary
#### Submitters Name:
NOV 222 2013
Jacques Ginestet Sense4Baby. Inc. 3344 N. Torrey Pines Court, Suite 100 La Jolla, CA 92037
## Date of Summary:
This summary was prepared on July 1, 2013
#### Trade Name of the Device:
Sense 4 Babv Svstem, Model B+
#### Common Name:
Fetal Monitor
### Regulatory Classification: Class II (Performance standards)
| Device Panel | Classification | Product Code | Description |
|--------------------------------------------------------|----------------|--------------|------------------------------------------------|
| Obstetrical and<br>Gynecological<br>Monitoring Devices | \$884.2740. | HGM | Perinatal monitoring system<br>and accessories |
## Predicate Device:
Phillips/HP Series 50XM Fetal Monitor 510(k) # K954351.
### Device Description:
The Sense4Baby Model B+ is a non-invasive, wireless, external monitoring system used to measure fetal heart rate, maternal heart rate, and uterine contractions during antepartum (nonstress) testing. The system transmits the measured data to a gateway device, and then a HIPAA-Compliant web based portal for storage and physician review.
#### Indications for Use:
The Sense4Baby System Model B+ is indicated for monitoring of maternal and fetal physiologic parameters during the antepartum period. It is to be used by health care professionals in hospitals, clinics, physicians' offices, antepartum rooms, and in the patients' home on the order of a physician.
#### Technological Characteristics:
The fundamental scientific technology employed in the Sense4Baby, Inc. Model B+ Fetal/Maternal Monitor is substantially equivalent to that of the Phillips/HP Series 50XM Fetal monitor.
Sense4Baby System Model B+ Premarket Notification 510 (k) Vol 005 Page 1 of 3
{1}------------------------------------------------
## Safety Testing (Bench):
The Sense4 Baby System Model B+ complies with applicable voluntary safety standards, in the areas of:
- . Electrical Safety
- . EMC
- . Material safety
- Risk Management .
Safety testing performed in conformance to these voluntary standards was conducted at accredited independent test facilities to demonstrate that the Sense4Baby System Model B+ is as safe and effective as the predicate device. The specific standards employed were:
- IEC 60601-1:2005 + CORR.2 (2007) + Al(2012), Medical electrical equipment Part 1: . General requirements for safety.
- . National deviations of IEC 60601-1: 2005, such as EN 60601-1, Ed. 3 (European Harmonized Standard), AAMI ES 60601-1, Ed. I (US National Standard), CSA - C22.2 No 60601-1:2008 (Canadian National Standard).
- . IEC 60601-1-2:2007, Medical electrical equipment - Part 1-2: General requirements for safety - Collateral standard: Electromagnetic compatibility - Requirements and tests.
- IEC 60601-2-37:2007 Medical Electrical Equipment Part 2-37: Particular requirements . for the basic safety and essential performance of ultrasonic medical diagnostic and monitoring equipment.
- . ISO 10993-1:2009 Biological evaluation of medical devices- Part 1: Evaluation and testing within a risk management process.
- ISO 10993-5:2009: Biological evaluation of medical devices. Part 5- Tests for in vitro . cytotoxicity.
- ISO 10993-10:2010 Biological evaluation of medical devices, Part 10- Tests for irritation t and skin sensitization.
- ASTM D4169:2009 Standard Practice for Performance Testing of Shipping Containers . and Systems.
- IEC 60529:1989 + A 1(1999) Degrees of protection provided by enclosures (IP Code). .
The possibility of hazards arising from hardware and software errors was further minimized in compliance with ISO 14971:2007 and ISO 13485:2003
### Performance Testing (Bench):
Verification and validation activities established the performance, functionality, and reliability characteristics of the device with respect to the predicate. Testing involved sub-system, as well as system level tests. Pass/Fail criteria were based on the specifications cleared for the predicate devices and test results demonstrated substantial equivalence. Performance bench testing included:
- . Simulated Use testing
- . Hardware testing
- . Predicate comparison testing
- Software testing ●
- Environmental testing (operating and storage) .
- Mechanical testing (drop, cleaning, etc.) .
Sense4Baby System Model B+ Premarket Notification 510 (k)
{2}------------------------------------------------
Sense 4 Baby, Inc.
- . Water ingress testing (to IEC 60529)
- . Shipping testing (to ASTM D4169-09)
# Conclusion:
The performance and safety verification and validation test results demonstrate that the Sense4Baby System Model B+ is as safe and effective, and performs as well as the predicate device.
Sense4Baby System Model B+ Premarket Notification 510 (k) Vol 005 Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three wavy lines representing the agency's mission to promote health, well-being, and human services. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle.
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-00002
November 22, 2013
Sense4Baby, Inc. % Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street Buffalo, MN 55313
Re: K132918
> Trade/Device Name: Sense4Baby System model B+ Regulation Number: 21 CFR§ 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM, KNG, HFM Dated: November 7, 2013 Received: November 8, 2013
Dear Mark Job,
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies.
{4}------------------------------------------------
Page 2 - Mark Job
.............................................................................................................................................................................. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements ) as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please go to http://www.fda.gov/AboutFDA/CentersOffices/CDRH/CDRHOffices/ucm115809.htm for the Center for Devices and Radiological Health's (CDRH's) Office of Compliance. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
Benjamin R. Fisher
Benjamin R. Fisher, Ph.D. Director Division of Reproductive, Gastro-Renal, and Urological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): K132918
Device Name: Sense4Baby System Model B+
Indications for Use:
The Sense4Baby System Model B+ is indicated for monitoring of maternal and fetal physiologic parameters during the antepartum period.
It is to be used by health care professionals in hospitals, clinics, physicians' offices, antepartum rooms, and in the patient's home on the order of a physician."
Prescription Use: X (Part 21 CFR 801 Subpart D) Over-The-Counter Use AND/OR (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Benjamin R. Fisher -S 2013.11.22 14:48:52 -05'00'
Page 1 of __ l__
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.