The DUAFIT® Interphalangeal implant is intended for fixation of proximal interphalangeal joint arthrodesis of the lesser toes. Examples include: Rigid or semi-rigid deformity of the PIP joint Revision of failed arthroplasty or arthrodesis 2nd toe shortening.
Device Story
DUAFIT® interphalangeal implant is an intramedullary bone fastener for proximal interphalangeal (PIP) joint arthrodesis of lesser toes. Device features proximal taper with barbs and distal blade for stabilization; available in multiple lengths, diameters, and angles (0°, 10°, 17°). Manufactured from PEEK-OPTIMA® polymer. Used in hospital or outpatient surgery centers by surgeons. Implanted into bone to provide fixation for >30 days. Single-use, gamma-sterilized. Provides mechanical stabilization to facilitate arthrodesis, addressing deformities or failed previous procedures.
Clinical Evidence
Bench testing only. Performed static and cyclic four-point bending tests to demonstrate mechanical equivalence to predicate devices. Results met all acceptance criteria.
Technological Characteristics
Material: PEEK-OPTIMA® polymer (ASTM 2026). Design: Intramedullary fastener with proximal tapered barbs and distal blade; cannulated. Energy: None (passive implant). Sterilization: Gamma irradiation. Connectivity: N/A. Form factor: Multiple lengths (11-20mm) and angles (0°, 10°, 17°).
Indications for Use
Indicated for patients requiring proximal interphalangeal (PIP) joint arthrodesis of the lesser toes, including those with rigid or semi-rigid PIP joint deformities, patients undergoing revision of failed arthroplasty or arthrodesis, and patients requiring 2nd toe shortening.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
## JUL 3 0 2014
z8, chemin du Petil Bois
69130 Ecully – FRANCE
1 LL -33 (0)4 72 29 26 25
FA> +33 (0)4 73 29 26 29
contact@in2bones.com www.r. madanes com
## For
| Sponsor identification | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully - France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment registration number | New company. Will register following FDA clearance<br>(Owner/Operator Number: 10046803) |
| Date of preparation | July 23, 2014 |
| Contact person | Norman Estrin<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Fax: (301) 294-0126<br>Email: estrin@yourFDAconsultant.com |
| Authorized Agent in the<br>United States | Norman Estrin<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Fax: (301) 294-0126<br>Email: estrin@yourFDAconsultant.com |
| Proprietary Name | Duafit® interphalangeal implant |
| Common name | Intramedullary bone fastener |
| Device classification regulation | 21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener<br>Class II |
| Device Product<br>Code and Panel | HWC: screw, fixation, bone<br>87 orthopedics |
| Device Description | The DUAFIT® interphalangeal implant is an intramedullary implant,<br>designed to act as a bone fastener for proximal interphalangeal<br>arthrodesis of the lesser rays.<br><br>Design is a combination of a proximal taper with barbs and a distal<br>blade for enhanced stabilization. |
| | It is made available in multiple lengths and diameters, and in 3<br>different angles.<br>The implant is manufactured from PEEK-OPTIMA®, polymer from<br>Invibio®, and is designed for single use only. |
| | <b>Sizes:</b><br>The DUAFIT® interphalangeal implant is available in various angles<br>(straight / 0° - 10° - 17°) and lengths (size 1 to 4, corresponding to 11<br>to 20mm).<br><b>Material:</b> |
| | The DUAFIT® interphalangeal implant is manufactured from<br>Polyetheretherketone PEEK-OPTIMA®, polymer from Invibio®, as per<br>ASTM 2026. It does not have any coating. |
| | <b>Single use:</b><br>The DUAFIT® interphalangeal implant is designed for single use only. |
| | <b>Sterilization:</b><br>The DUAFIT® interphalangeal implant is supplied sterile, using gamma<br>irradiation. |
| | <b>Place of use:</b><br>The DUAFIT® interphalangeal implant is indicated for use in a<br>hospital, or outpatient surgery center where sterile field may be created<br>and maintained. |
| <b>Predicate Devices</b> | Newdeal K-wire (K022599) |
| | Metasurg DigiFuse (K111536) |
| | Parcus PEEK CF push-in suture anchor (K102326) |
| | WMT Pro-Toe (K120645) |
| | Arrowhead Arrow-Lok (K112675) |
| | Smith and Nephew Bioraptor 2.3 PK suture anchor (K071586) |
| | Merete MetaToe (K100414) |
| | Synchro Medical Toegrip (K143477) |
| <b>Indications for use:</b> | The DUAFIT® interphalangeal implant is intended for fixation of<br>proximal interphalangeal joint arthrodesis of the lesser toes. |
| | Examples include: |
| | - rigid or semi-rigid deformity of the PIP joint |
| | - revision of failed arthroplasty or arthrodesis |
| | - 2nd toe shortening |
{1}------------------------------------------------
.
and the commend of the comments of
.
:
:
.
.
and the comments of the comments of
{2}------------------------------------------------
:
| Comparison of | The technological characteristics of the DUAFIT® interphalangeal implant are |
|-----------------|----------------------------------------------------------------------------------------------------------------------------------|
| Technological | the same as the characteristics of predicate devices in terms of intended use and |
| characteristics | design. All these implants have the following features: |
| | Insertion into bone - the DUAFIT® interphalangeal implant and all |
| | predicate devices are intended for surgical implantation into bone for |
| | longer than 30 days. |
| | Tapered barded design in its proximal part - The DUAFIT®<br>1 |
| | interphalangeal implant has similar tapered barded design in its proximal |
| | part, when compared to the Parcus PEEK CF push-in suture anchor |
| | (K102326). Smith and Nephew Bioraptor 2.3 PK suture anchor |
| | (K071586), Metasurg DigiFuse implant (K111536), WMT Pro-Toe |
| | (K120645), Merete MetaToe (K100414) and Synchro Toegrip |
| | (K133477). |
| | Blade design in its distal part - The DUAFIT® interphalangeal implant |
| | has similar blade design in its distal part, when compared to the |
| | Metasurg DigiFuse implant (K111536) and WMT Pro-Toe (K120645). |
| | Its shape is also similar to the distal part of the Synchro Toegrip |
| | (K133477). |
| | Canulated design: The DUAFIT® interphalangeal implant is available<br>- |
| | with a similar canulated design, when compared to the Metasurg |
| | DigiFuse implant (K111536). |
| | Straight / Angled design: The DUAFIT® interphalangeal implant has<br>- |
| | similar straight and angled design, when compared to the Metasurg |
| | DigiFuse implant (K111536), WMT Pro-Toe (K120645) , Arrowhead |
| | Arrow-Lok (K112675) and Synchro Toegrip (K133477) |
| | Made from PEEK: The DUAFIT® interphalangeal implant has exactly<br>- |
| | the same raw material PEEK-OPTIMA®, polymer from Invibio", when |
| | compared to the Smith and Nephew Bioraptor 2.3 PK suture anchor |
| | (K071586). It has similar raw material, when compared to the Parcus |
| | PEEK CF push-in suture anchor (K102326) and Synchro Toegrip |
| | (K133477) . The Newdeal / Integra K-wire (K022599), the WMT Pro- |
| | Toe (K120645) and Arrowhead Arrow-Lok (K112675) are manufactured |
| | from stainless steel 316L, and the Metasurg DigiFuse implant is |
| | manufactured from Titanium Alloy TA6V. The Merete MetaToe is |
| | manufactured from EndoSorb PLGA bioabsorbable material. |
| | Equivalent size range: The DUAFIT® interphalangeal implant has |
| | similar size range, when compared to the Metasurg DigiFuse implant |
| | (K111536), WMT Pro-Toe (K120645), Arrowhead Arrow-Lok |
| | (K112675), Merete MetaToe (K100414) and Synchro Toegrip |
| | (K133477).<br>The DUAFIT® interphalangeal implant has similar intended use and |
| | |
| | mechanical properties when compared to the Merete MetaToe EndoSorb<br>(K100414). Results from testing confirmed that the DUAFIT® |
| | interphalangeal implant is at least equivalent to the predicate device. |
| | |
.
.
.
.
: : : :
.
:
ので、その他の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の人気の
{3}------------------------------------------------
|
| Substantial<br>Equivalence<br>Summary | The DUAFIT® interphalangeal implant has similar indications for use and<br>technological characteristics when compared to the predicate devices. | |
|---------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Summary Performance<br>Data | The following tests were performed to demonstrate that the DUAFIT<br>interphalangeal implant is substantially equivalent to other predicate<br>devices:<br>- Static four-point bending test<br>- Cyclic four-point bending test.<br>The results of these studies showed that the DUAFIT interphalangeal<br>implant met the acceptance criteria. | |
| CONCLUSION | Based on the evaluations performed, the design and indications of<br>the DUAFIT® interphalangeal implant are substantially equivalent<br>to the predicate devices identified in the 510(k) submission. No new<br>materials or processes are used in the development of this implant.<br><br>In addition, the results of the testing performed by the test lab<br>indicated that the implants performed as expected for each test.<br><br>The DUAFIT® interphalangeal implants are acceptable for the<br>application. | |
:
.
:
:
: 上一篇: .
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows a logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized emblem featuring three curved shapes that resemble waves or abstract wings. The emblem is positioned to the right of the text.
Public I lealth Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 30, 2014
In2Bones SAS % Norman F. Estrin, Ph.D. Estrin Consulting Group LLC 9109 Copenhaver Drive Potomac, Maryland 20854
Re: K132912
Trade/Device Name: DUAFIT* Interphalangeal Implant Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: July 3. 2014 Received: July 9, 2014
Dear Dr. Estrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration. Iisting of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA is issuance of a substantíal equivalence determination docs not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-rclated adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Dr. Norman F. Estrin
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours, Lori A. Wiggins
for
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K132912 - page 1 of 1
## Indications for Use
510(k) Number (if known): K132912
Device Name: DUAFIT** Interphalangeal Implant
Indications For Use: The DUAFIT® Interphalangeal implant is intended for fixation of proximal interphalangeal joint arthrodesis of the lesser toes. Examples include: Rigid or semi-rigid deformity of the PIP joint Revision of failed arthroplasty or arthrodesis 2nd toe shortening.
Prescription Use _____________________________________________________________________________________________________________________________________________________________ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use ___________________________________ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Elizabet简介版rank -S
(Division Sign-Off) Division of Orthopedic Devices 510(k) Number: K132912
20000000
.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.