The I.B.S.® Snap-off screws are indicated for bone reconstruction, osteotomy and fracture fixation of bones appropriate for the size of the device. Examples include: Mono-cortical fixation of small bone fragments Weil osteotomy Osteotomies and fractures fixation in the foot and hand.
Device Story
I.B.S. Snap-off screw is a single-piece, self-drilling, self-tapping metallic bone fixation fastener. Device consists of a screw head and a drill shank; shank inserts into a driver for bone insertion. Upon reaching target depth, drill shank twists off and breaks cleanly from screw head. Used in hospital or outpatient surgery centers by surgeons for bone reconstruction, osteotomy, and fracture fixation. Implanted for >30 days. Benefits include stable fixation of small bone fragments in foot and hand procedures. Supplied sterile via gamma irradiation.
Clinical Evidence
Bench testing only. Mechanical testing performed per ASTM F543-07 to assess torque to failure, insertion torque, axial pullout strength, and self-tapping performance. Engineering/dimensional comparison to predicates performed. Results indicated device performed as expected for all tests.
Indicated for bone reconstruction, osteotomy, and fracture fixation of bones appropriate for device size, including mono-cortical fixation of small bone fragments, Weil osteotomy, and foot/hand fracture fixation.
Regulatory Classification
Identification
A smooth or threaded metallic bone fixation fastener is a device intended to be implanted that consists of a stiff wire segment or rod made of alloys, such as cobalt-chromium-molybdenum and stainless steel, and that may be smooth on the outside, fully or partially threaded, straight or U-shaped; and may be either blunt pointed, sharp pointed, or have a formed, slotted head on the end. It may be used for fixation of bone fractures, for bone reconstructions, as a guide pin for insertion of other implants, or it may be implanted through the skin so that a pulling force (traction) may be applied to the skeletal system.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows a logo for "In2Bones". The logo is an oval shape that is split into two parts by a white line. The text "In2Bones" is written in black letters on the right side of the oval. The logo is simple and modern.
ﺔ
23, charren du Va: " 20120 Eculiv "
21 + 7210 - 72 20
outar solar hots
#### K132911 page 1 of 4
### 004_Revised 510(k) Summary for I.B.S. Snap-off Screw (K132911) Response to September 17, 2013 RTA Checklist, Section A4a.pdf
# 510(k) SUMMARY
In2Bones I.B.S." Snap-off screw
| Sponsor identification | In2Bones SAS<br>28 chemin du Petit Bois<br>69130 Ecully - France<br>Phone: +33.4.72.29.26.26<br>Fax: +33.4.72.29.26.29 |
|------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Establishment registration number | New company. Will register following FDA clearance |
| Date of preparation | August 1st, 2013 |
| Contact person | Norman Estrin<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Fax: (301) 294-0126<br>Email: estrin@yourFDAconsultant.com |
| Authorized Agent in the<br>United States | Norman Estrin<br>Estrin Consulting Group LLC<br>9109 Copenhaver Drive<br>Potomac, MD 20854<br>Phone: (301) 279-2899<br>Fax: (301) 294-0126<br>Email: estrin@yourFDAconsultant.com |
| Proprietary Name | I.B.S.® Snap-off screw |
| Common name | Bone fixation screw |
| Device classification regulation | 21 CFR 888.3040: Smooth or threaded metallic bone fixation fastener<br>Class II |
| Device Product<br>Code and Panel | HWC: screw, fixation, bone<br>87 orthopedics |
| Device Description | The I.B.S.® Snap-off screw is a one-piece device made of Titanium<br>Alloy intended to be used as a screw for bone reconstruction,<br>osteotomy and fracture fixation of bones appropriate for the size of the<br>device. |
| | The implant is a self-drilling and self-tapping snap-off screw. It is<br>machined on the end of a drill shank which inserts into a driver. After<br>insertion of the screw into the bone, the drill shank twists off and<br>breaks cleanly from the screw head. |
| | The implant is supplied sterile and is available in various sizes<br>diameters and lengths. |
| | Sizes:<br>The I.B.S.® Snap-off screws are available in various diameters and<br>lengths. |
| | Material:<br>The I.B.S.® Snap-off screws are manufactured from titanium alloy<br>TA6V as per ISO 5832-3 and ASTM F136. They do not have any<br>coating. |
| | Single use:<br>The I.B.S.® Snap-off screws are designed for single use only.<br>Sterilization: |
| | The I.B.S.® Snap-off screws are supplied sterile, using gamma<br>irradiation. |
| | Place of use:<br>The I.B.S.® Snap-off screws are indicated for use in a hospital, or<br>outpatient surgery center where sterile field may be created and<br>maintained. |
| Predicate Devices | Integra / Newdeal Spin screw (K991477)<br>Arthrex Low Profile Screws (K103705)<br>Nexa Orthopedics / Nexa Bone screw (K053394) |
| Indications for use | The I.B.S.® Snap-off screws are indicated for bone reconstruction,<br>osteotomy and fracture fixation of bones appropriate for the size of the<br>device. |
| | Examples include:<br>Mono-cortical fixation of small bone fragments<br>Weil osteotomy<br>Osteotomies and fractures fixation in the foot and hand. |
{1}------------------------------------------------
に
K132911 page 2 of 4
{2}------------------------------------------------
K132911 page 3 of 4
#### COMPARISON OF THE INDICATIONS FOR USE WITH THE PREDICATE DEVICES:
As with the predicate devices, the I.B.S.® Snap-off screws are indicated for surgical implantation longer than 30 days in the fixation of bone fractures or for bone reconstruction.
| 510k number | K132911 | K991477 | K103705 | K053394 |
|------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------|
| manufacturer | In2Bones | Newdeal -- an<br>Integra Lifesciences<br>company | Arthrex | Nexa Orthopedics<br>/ Tornier |
| Name of<br>device | I.B.S.® Snap-off<br>screw | SPIN® screw | Low Profile Screw<br>- QuickFix | Nexa Bone screw<br>system / NexFix |
| Use | Single use | Single use | Single use | Single use |
| Fixation | Bone | Bone | Bone | Bone |
| Indications<br>for use | The I.B.S.® Snap-<br>off screws are<br>indicated for bone<br>reconstruction,<br>osteotomy and<br>fracture fixation of<br>bones appropriate<br>for the size of the<br>device.<br>Examples include:<br>Mono-cortical<br>fixation of small<br>bone fragments<br>Weil osteotomy<br>Osteotomies and<br>fractures fixation in<br>the foot and hand. | The SPIN® Snap-off<br>screw is indicated<br>for fixing the<br>elective osteotomies<br>of the mid-foot<br>bones and the<br>metatarsal and<br>phalanges of the<br>foot only.<br>Examples include:<br>Weil osteotomy<br>Unicortical small<br>bone fixation | The Arthrex Low<br>Profile Screws<br>(2.0-3.0mm solid)<br>are intended to be<br>used as stand-<br>alone bone screws,<br>or in a plate-screw<br>system for internal<br>bone fixation for<br>bone fractures,<br>fusions,<br>osteotomies and<br>non-unions in the<br>ankle, foot, hand,<br>and wrist. | The Nexa bone<br>screw system<br>provides fixation<br>of fractures,<br>fusions, or<br>osteotomies of the<br>hand and foot. |
{3}------------------------------------------------
#### K132911 page 4 of 4
| Comparison of<br>Technological<br>characteristics | The technological characteristics of the I.B.S.® Snap-off screw are the<br>same as the characteristics of predicates devices in terms of intended<br>use and design. All the screws have the following features:<br>- Solid / non-cannulated – the I.B.S.® Snap-off screws and all<br>predicate devices are solid / non-cannulated screws<br>- Self-tapping and self-drilling, with a snap-off design - the<br>I.B.S® Snap-off screws and all predicate devices are self-<br>tapping and self-drilling screws with a snap-off mechanism<br>- Material: made from Titanium alloys, with no new materials<br>being introduced in the product - the I.B.S.® Snap-off screws<br>and all predicate devices are manufactured from Titanium<br>Alloy TA6V<br>- Equivalent size range: the diameters and lengths covered by the<br>predicate devices enable to cover diameters and lengths of the<br>I.B.S.® Snap-off screws<br>- Indicated for surgical implantation longer than 30 days - the<br>I.B.S.® Snap-off screws and all predicate devices are indicated<br>for surgical implantation longer than 30 days |
|---------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Substantial Equivalence<br>Summary | The I.B.S.® Snap-off screw has similar technological characteristics<br>when compared to the predicate devices. |
| Summary Performance<br>Data | Mechanical testing was performed according to ASTM F543-07. This<br>standard describes methods to assess the torque to failure, insertion<br>torque, axial pullout strength, and self-tapping performance of screws.<br>The specification standard provides requirements for materials, finish,<br>marking, care and handling, and the acceptable dimensions and<br>tolerances for metallic bone screws implanted into bone tissue. As the<br>specification standard states, the dimensions and tolerances are<br>applicable only to metallic bone screws described in the specification<br>itself. Additional information must be provided to document that the<br>design of the product will provide adequate mechanical properties for<br>the particular application.<br>In addition, an engineering / dimensional comparison to the predicates<br>was performed to ascertain substantial equivalence.<br>The results of the testing performed by the test lab indicated that the<br>I.B.S.® Snap-off screw performed as expected for each test. |
| CONCLUSION | Based on the evaluations performed, the design and indications of<br>the I.B.S.® Snap-off screw are substantially equivalent to the<br>predicates identified in the 510(k) submission. No new materials or<br>processes are used in the development of this implant.<br><br>In addition, the results of the testing performed by the test lab<br>indicated that the screws performed as expected for each test.<br><br>The I.B.S.® Snap-off screws are acceptable for the application. |
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three lines representing its wings and tail feathers. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
April 4, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MI) 20993-0002
In2Bones SAS % Norman Estrin, Ph.D. Managing Partner Estrin Consulting Group LLC 9109 Copenhaver Drive Potomac, Maryland 20854
Re: K132911
Trade/Device Name: I.B.S.® Snap-off screw Regulation Number: 21 CFR 888.3040 Regulation Name: Smooth or threaded metallic bone fixation fastener Regulatory Class: Class II Product Code: HWC Dated: March 8, 2014 Received: March 11. 2014
Dear Dr. Estrin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments. or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing
{5}------------------------------------------------
Page 2 -Norman Estrin, Ph.D.
(21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers. International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm.
Sincerely yours,
### Mark N. Mellerson -S
Mark N. Melkerson Director Division of Orthopedic Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
K132911 page 1 of 1
#### Indications for Use
510(k) Number (if known): K132911
Device Name: I.B.S snap-off screw
Indications For Use:
The I.B.S snap-off screws are intended for bone reconstruction, osteotomy and fracture fixalinn of bones appropriate for the size of the device. Examples include : Mono-cortical fixation of small bone fragments Weil ostcolomy Osteotomies and fractures fixation in the foot and hand.
Prescription Use _ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
## Elizabet顾迹:rank -S
Division of Orthopedic Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.