K132892 · Bomed, Inc. · DXG · Aug 1, 2014 · Cardiovascular
Device Facts
Record ID
K132892
Device Name
TEBCO-W
Applicant
Bomed, Inc.
Product Code
DXG · Cardiovascular
Decision Date
Aug 1, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1435
Device Class
Class 2
Indications for Use
TEBCO-W is indicated for noninvasive, continuous assessment and monitoring of the left-ventricular performance and global blood flow (Stroke Index, SI and Cardiac Index, CI) in adults, where use of invasive catheters is contraindicated. The device may be used only on the order of a physician and is "By prescription only." The intended use is physician's office and cardiovascular physiology laboratory.
Device Story
TEBCO-W is a wireless thoracic electrical bioimpedance (TEB) monitor for cardiac output assessment. It acquires TEB and ECG signals via 8 thoracic electrodes; ECG serves as a system clock. The device samples TEB at 5ms intervals and transmits digital packets via WiFi (802.11b) to a host PC. The PC reconstructs waveforms and calculates heart rate, stroke index, and cardiac index per heartbeat. Used in physician offices and cardiovascular labs by clinicians. Wireless connectivity improves patient mobility compared to the tethered predicate; battery power enhances safety. Output is displayed on a host PC for clinical decision-making regarding cardiovascular performance.
Clinical Evidence
Clinical testing conducted in an Intensive Care Unit (ICU) environment. A subject was monitored while walking around the ICU while other equipment (ventilators, patient monitors) operated normally. Results showed no observable interference between TEBCO-W and ICU equipment, and no susceptibility of TEBCO-W to noise from ICU equipment. Bench testing also confirmed signal integrity and lack of interference within a private WiFi network.
Technological Characteristics
Thoracic electrical bioimpedance (TEB) sensing; 8 thoracic electrodes. Wireless communication via 802.11b WiFi (2.412-2.462 GHz, max 15 dBm, WPA2 security). Power: 4 rechargeable C-cell batteries. Host system: PC. Software algorithms identical to predicate TEBCO. Form factor: two piggy-back PCBs (4.5" x 2.5").
Indications for Use
Indicated for noninvasive, continuous assessment and monitoring of left-ventricular performance and global blood flow (SI, CI) in adults where invasive catheters are contraindicated. For professional use only; prescription required.
Regulatory Classification
Identification
A single-function, preprogrammed diagnostic computer is a hard-wired computer that calculates a specific physiological or blood-flow parameter based on information obtained from one or more electrodes, transducers, or measuring devices.
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is a stylized symbol of three faces in profile, stacked on top of each other, representing the department's focus on people and health.
August 1, 2014
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center -WO66-G609 Silver Spring, MD 20993-0002
Bomed Inc B. Bo Sramek 59-440 Pili Pl. Kamuela, HI 96743 US
Re: K132892
Trade/Device Name: Tebco-W Regulation Number: 21 CFR 870.1435 Regulation Name: Single-function, preprogrammed diagnostic computer Regulatory Class: Class II Product Code: DXG Dated: June 17, 2014 Received: June 30, 2014
Dear B. Bo Sramek:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{1}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours.
K&l
Skodarek for Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{2}------------------------------------------------
# Indications for Use
510(k) Number:
Device Name: TEBCO-W
Indications For Use: Adults, Hospital, Cardiology, Internal medicine
TEBCO-W is indicated for noninvasive, continuous assessment and monitoring of the left-ventricular performance and global blood flow (Stroke Index, SI and Cardiac Index, CI) in adults, where use of invasive catheters is contraindicated. The device may be used only on the order of a physician and is "By prescription only." The intended use is physician's office and cardiovascular physiology laboratory.
Prescription Use: Yes
AND/OR
Over-The-Counter Use: N/A
Image /page/2/Picture/9 description: The image shows a logo with the letters "K" and "L" in a stylized, handwritten font. The letters are connected, with the "L" appearing to be formed by a single stroke that loops back on itself. Below the logo, the text "Ken Skodacek for" is printed in a smaller, sans-serif font. In the background, there is a faint, stylized image of the FDA logo.
Bram Zuckerman
Page 1 of 1
{3}------------------------------------------------
## 510(k) Summary of Substantial Equivalence BoMed Inc. TEBCO-W: K132892
21 CFR 807.92(a):
#### 21 CFR 807.92(a) (1):
Submitter's name and address:
BoMed Inc. 59-440 Pili Pl. Kamuela, HI 96743
Submitter's telephone number:
808-885-8463
Contact person:
B. Bo Sramek, Ph.D. President BoMed Inc. 59-440 Pili Pl. Kamuela, HI 96743
Date this 510(k) Summary was prepared:
October 17, 2013
### 21 CFR 807.92(a) (2):
| Trade Name of Device: | TEBCO-W; assigned 510(k)#: K132892 |
|----------------------------|------------------------------------------------------------------------------------------------------------------------------|
| Common Name of the Device: | Bioimpedance Cardiac Output, Impedance Cardiograph,<br>Impedance Plethysmograph |
| Regulatory Section: | 21 CFR 870.2770 "Impedance Plethysmograph", and<br>21 CFR 870.1435 "Single-function, preprogrammed diag-<br>nostic computer" |
| Classification: | Class II |
| Product Code: | 74 DXG |
{4}------------------------------------------------
| 21 CFR 807.92(a) (3): | Legally marketed predicate device to which substantial<br>equivalency is claimed |
|-----------------------|----------------------------------------------------------------------------------|
| The Predicate Device: | TEBCO |
| Manufacturer: | HEMO SAPINES INC. |
| Predicate Device K#: | K962128 |
#### 21 CFR 807.92(a) (4): Description of the Device that is the subject of this premarket notification:
TEBCO-W is the substantially equivalent wireless version of the existing tethered predicate device TEBCO (Thoracic Electrical Bioimpedance Cardiac Output, K962128). Both the TEBCO and TEBCO-W are a hardware component of a system for noninvasive measurement of Cardiac Output and other cardiodynamic parameters. The system comprises of TEBCO (or TEBCO-W) and a PC, which acts as a host system. TEBCO (or TEBCO-W) acquire the TEB and data via a patient cable attached to 8 electrodes placed on a patient's thorax. 4 out of 8 electrodes are also used to detect a patient's ECG signal, which is utilized by both devices as a system clock. They sample the instantaneous TEB values @ 5 msec sample rate, and send continuously the binary values of digital samples in packets every 20 msec via a serial communication link to a host system (a PC), where the TEB signals are reconstructed and displayed. Both TEBCO and TEBCO-W also measure the magnitude and timing of the TEB signal in reference to Q-time of the QRS complex, from which the Heart Rate and Stroke Index are calculated, calculate then the value of Cardiac Output and sent the digital values of these parameters via the same serial link once every heartbeat to the host system, where they are displayed.
This table lists the differences between TEBCO and TEBCO-W:
| | TEBCO | TEBCO-W |
|------------------------------|---------------------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------|
| TEB Measurement Current | 7 µA RMS | 7 µA RMS |
| Communication w/ host system | Tethered:<br>Serial RS232 cable<br>4 or 6 bytes packets | Wireless:<br>Serial - WiFi<br>4 or 6 bytes packets<br>WiFi Band:<br>2.412 - 2.462 GHz<br>Modulation: 802.11b<br>Power: max 15 dBm<br>± 1.5 dB<br>Security: WPA2 |
{5}------------------------------------------------
| Connection to patient | 8 thoracic electrodes,<br>8 patient leads,<br>patient cable | 8 thoracic electrodes,<br>8 patient leads,<br>patient cable |
|-----------------------|----------------------------------------------------------------------------------|------------------------------------------------------------------|
| Power | External AC-to-DC<br>5V power source | Built-in, externally re-<br>chargeable four<br>C-cells batteries |
| PC board size | One PCB 9.1" x 5.5" | Two piggy-back PCBs<br>4.5" x 2.5" |
| Patient mobility | Radius of patient<br>movement limited by<br>the length of patient<br>cable (3 m) | Limited by the capabil-<br>ity of the WiFi connec-<br>tion |
#### 21 CFR 807.92(a) (5):
#### Intended use and labeled indications for use
The TEBCO-W (as was TEBCO) device is indicated for noninvasive monitoring and assessment of cardiovascular performance, where use of invasive catheters is contraindicated or expensive. TEBCO-W is labeled "For Professional Use Only" and can be used only on the order of a physician.
#### 21 CFR 807.92(a) (6):
### Technological characteristics
The design, product technology, algorithms used in software, communication protocol, connection to the patient and use of the host system of TEBCO-W to display data are identical to the predicate device TEBCO. The differences are (a) the power source, where the external AC-to-DC converter used with TEBCO is replaced in TEBCO-W by 4 C-cells batteries recharged only externally, and (b) the tethered communication in TEBCO is replaced in TEBCO-W with a wireless communication, utilizing a built-in WiFi feature of today's PC. TEBCO-W will represent a technological improvement to the predicate device TEBCO in two respects: The departure from AC power source improves a patient's safety, while the wireless communication improves his mobility.
### 21 CFR 807.92(b):
### 21 CFR 807.92(b) (1):
#### Brief discussion of nonclinical tests submitted, referenced or relied on in this premarket notification:
The nonclinical test was performed in our laboratory: During the Bench Test, a person representing a patient was being monitored by TEBCO-W carried on subject's chest and his data were continually transmitted to the host system (Acer Aspire
{6}------------------------------------------------
V5-571P-6400 notebook) while the other equipment, representing potential WiFi interference source/recipient, was operating under its normal conditions:
We constructed a private WiFi secured network consisting of three different computers (iMac, Gateway M520X-CTC notebook and MacBook Air) and created a simple program in which all computers continuously communicated via their WiFi with each other. All utilized network computers have a manufacturer built-in WiFi circuit.
The monitored subject than moved around the lab and operator of each of the network computers and the operator of the TEBCO-W host system looked for presence of any disruption in service. We concentrated on TEBCO-W host system: The four digitized analog signals and the digital values of measured parameters are sent out by TEBCO-W every 20 msec in packets (see Attachment 007), while the digitized analog signal content of each packet represents a next dot on the displayed analog signal. Any transmission interference producing missing or erroneous packets would thus be clearly visible as a signal waveform inconsistency. When no interference in this mode could be detected, the monitored subject than placed TEBCO-W adjacent to each of the network computers, also with no interference observable.
### 21 CFR 807.92(b) (2):
#### Brief discussion of clinical tests submitted, referenced or relied on in this premarket notification:
The Clinical Test took place at the North Hawaii Community Hospital, in its Intensive Care Unit. The subject/patient's monitoring was similar to the laboratory Bench Testing: During the Clinical Test, a person representing a patient was being monitored by TEBCO-W carried on subject's chest and his data were continually transmitted via its WiFi link to the host system (Acer Aspire V5-571P-6400 notebook) while all other ICU equipment (patient monitors and ventilators), representing potential interference source, was operating under its normal conditions:
Our subject, monitored by TEBCO-W, walked around the ICU without any observable interference either on the ICU patient monitor (that would indicate TEBCO-W as a source of interference) or on the TEBCO-W host system (which would indicate a TEBCO-W susceptibility to noise emitted by the ICU equipment).
Conclusion 1: TEBCO-W did not produce any interference with the ICU monitors and thus its function could not cause any life-threatening condition to the ICU patient.
Conclusion 2: The ICU setting - the most interference-rich environment - did not produce any observable interference with the TEBCO-W operation.
#### 21 CFR 807.92(b) (3): Conclusions drawn from nonclinical and clinical tests
Based on the results of both nonclinical and clinical studies described in this 510(k) Submission above, we can conclude that TEBCO-W is as safe and effective (therefore substantially equivalent) as the predicate device TEBCO
.... End of 510(k) Summary....
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.