K132815 · Cytori Therapeutics, Inc. · MUU · Jan 24, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132815
Device Name
PUREGRAFT 50 SYSTEM
Applicant
Cytori Therapeutics, Inc.
Product Code
MUU · General, Plastic Surgery
Decision Date
Jan 24, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cytori Puregraft 50 System is indicated for use in the harvesting, filtering and transferring of autologous fat tissue for reinjecting back into the same patient for aesthetic body contouring.
Device Story
Puregraft 50 System is a sterile, single-use, closed-system bag for processing autologous fat grafts. Input: harvested adipose tissue. Operation: tissue injected into 100mL capacity bag via luer-activated tissue port; bag contains 800-micron and 74-micron filter meshes; manual slider compresses bag to force fluids/debris through filters to waste port; filtered fat retained in bag for delivery. Used in clinical settings by physicians during cosmetic/reconstructive surgery. Output: processed autologous fat graft ready for reinjection. Benefit: provides filtered, concentrated fat tissue for autologous transfer, facilitating body contouring.
Clinical Evidence
No clinical data. Evidence consists of bench-top testing and in vitro performance verification. Comparative testing against predicate devices (K092923) evaluated lipolysis, wetness, lipid content, and RBC/WBC removal, showing equivalent performance across all metrics. Integrity testing also confirmed substantial equivalence.
Technological Characteristics
Sterile, single-use, closed-system bag (184 mm x 96 mm). Materials: medical-grade, DEHP-free polymers (PVC outer shell). Filtration: dual-layer mesh (800 micron and 74 micron). Operation: manual, via luer-activated valves and integrated slider for fluid expression. Sterilization: gamma irradiation. Connectivity: none (standalone).
Indications for Use
Indicated for patients undergoing aesthetic body contouring procedures requiring the harvesting, filtering, and transfer of autologous fat tissue for reinjection into the same patient.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
{0}------------------------------------------------
# Puregraft® 50 System
# ADMINISTRATIVE INFORMATION
Manufacturer Name:
Official Contact:
Cytori Therapeutics, Inc. 3020 Callan Road San Diego, CA 92121. USA
Kenneth K. Kleinhenz Vice President, Global Regulatory Affairs
Telephone (858) 458-0900 Fax (858) 458-0994
#### DEVICE NAME
Classification Name: Trade/Proprietary Name:
Suction Lipoplasty System Puregraft® 50 System
### ESTABLISHMENT REGISTRATION NUMBER
3002642958
#### DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21 CFR 878.5040, Suction Lipoplasty Systems are defined as devices consisting of collection bottles, cannulas, and connecting tubing for use in aesthetic body contouring procedures. Suction Lipoplasty Systems are classified as Class II. They have been assigned Product Code MUU.
#### INDICATIONS FOR USE
The Cytori Puregraft 50 System is indicated for use in the harvesting, filtering and transferring of autologous fat tissue for reinjecting back into the same patient for aesthetic body contouring.
.
JAN 2 4 2014
{1}------------------------------------------------
#### DEVICE DESCRIPTION
#### Design Characteristics
The Puregraft 50 System is a sterile, single use, closed system intended for the preparation and delivery of autologous fat grafts back to the same patient for aesthetic body contouring in cosmetic and reconstructive surgery applications. The dual filtration bag system connects to an off the shelf sterile luer lock syringe through a swabable luer activated valves interface. The attached slider is used to facilitate the movement of liquids through the filter mesh and out of the Puregraft bag through the drain stub as waste. The Puregraft 50 System is composed of the following components:
| Description | Quantity |
|-------------------|----------|
| Outer Membrane | 2 |
| Filter 74 micron | 1 |
| Filter 800 micron | 1 |
| Tissue Port | 1 |
| Drain Port | 1 |
| Attached Slider | 1 |
The Puregraft 50mL System is a sterile, single-use, 100mL capacity mixing bag measuring approximately 184 mm x 96 mm and consists of 2 filters layered between a polyvinyl chloride (PVC) outer shell and 2 ports on the bottom of the bag. Each port is labeled and designed to assure the proper connection is made with a luer fitting. The "tissue" port and the "drain" port contains a female swabable luer fitting designed to connect to a luer syringe. The "tissue" port and the "drain" port are designed as one-way valves to assure that the contents within the Puregraft 50mL Bag are not accidently spilled from the bag. The Puregraft 50mL Bag contains two (2) filters that are continuous within the bag. The first filter is a 800 micron filter mesh and the second filter is a 74 micron filter. The Puregraft 50 Bag has an attached plastic "slider" that is used to compress the Puregraft bag as a means to facilitate the movement of liquids through the filter mesh and out of the Puregraft bag to waste; through the "drain" port. All materials are medical grade and DEHP free.
#### Material Composition
The Puregraft 50 System is fabricated from medical grade polymers and are all DEHP free materials.
#### Sterility
The Puregraft 50 System is sterilized with gamma irradiation.
#### In Vitro Testing
Mechanical testing of the Puregraft 50 System demonstrates that the device is substantially equivalent to the predicate devices.
{2}------------------------------------------------
# EQUIVALENCE TO MARKETED PRODUCT
The Puregraft 50 System shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to a precardaet device: the Cytori Puregraft® 250/PURE System (K092923), and Puregraft 850/PURE System (K113255); both are a Class II medical device that have been cleared for marketing in the United States under K092923 and K113255 respectively.
#### Indications for Use
The Puregraft 50 System and the predicate devices are substantially equivalent as they share identical indications for use language and they are all indicated for the same surgical procedures of harvesting, filtering, and transferring autologous tissues for reinjecting back into the same patient for body contouring. The Puregraft 50 System's indications for use are identical to the cleared indications for use of the predicate devices. Specifically, the Cytori Puregraft 50 System is indicated for:
The Cytori Puregraft 50 System is indicated for use in the harvesting, filtering and transferring of autologous fat tissue for reinjecting back into the same patient for aesthetic body contouring.
#### Design and Materials
The design and materials of the Puregraft 50 System and the Puregraft 250/PURE System (K092923) and Puregraft 850/PURE System (K113255) are substantially equivalent as thy are all single-use, polymer constructed, manually operated systems that receive adipose tissue, filter the adipose tissue, and temporarily hold the adipose tissue until it is removed or placed into a syringe that delivers / re-injects the adipose tissue back into the same patient during the same surgical procedure. The Puregraft 50 System is substantially equivalent to the predicate devices as they are all constructed of a polymeric flexible housing bag with a filter unit within the bag and ports to accept and remove the tissue with a syringe. These predicate devices also share design principles of accepting adipose tissue from the patient and subsequently transport the adipose tissue through a tube into a polymeric collection chamber/bag that contains a filtering mechanism of various pore sizes that restricts the movement of adipose tissue and only allows fluids and small debris to pass through the filter and become deposited into a waste container.
# Mechanical and Nonclinical Testing Summary
As a means to confirm the substantial equivalence between the Puregraft 50 System, the Puregraft 250/PURE System Puregraft 850/PURE System, nonclinical testing was performed on the Puregraft 50 System in addition to the 850/PURE System and Puregraft 250/PURE System predicate device that included biological performance verification and bench-top testing such as tensile strength, pressure, and drop testing. The nonclinical testing demonstrates that the Puregraft 50 System is substantially equivalent to the Puregraft 250/PURE System and Puregraft 850/PURE System predicate devices.
{3}------------------------------------------------
# K132815 Summary
.. :
.. :
# Substantially Equivalent Testing Summary
| | Puregraft 50<br>(Subject Device) | Puregraft 250<br>(K092923) | Correlation |
|------------------------------|----------------------------------|----------------------------|--------------------------|
| Lipolysis (fold change) | PASS | PASS | Substantially Equivalent |
| Wetness (% of control) | PASS | PASS | Substantially Equivalent |
| Lipid Content (% of control) | PASS | PASS | Substantially Equivalent |
| RBC (% of control) | PASS | PASS | Substantially Equivalent |
| WBC (% of control) | PASS | PASS | Substantially Equivalent |
| Integrity Testing | PASS | PASS | Substantially Equivalent |
·
.
·
- --
.
..
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo is circular and contains the department's emblem in the center. The emblem is a stylized human figure embracing a globe. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the emblem.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
January 24, 2014
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
Cytori Therapeutics Incorporated Mr. Kenneth K. Kleinhenz Vice President, Global Regulatory Affairs 3020 Callan Road San Diego. California 92121
Re: K132815
Trade/Device Name: Puregraft 50 System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: Class II Product Code: MUU Dated: December 24, 2013 Received: December 26, 2013
Dear Mr. Kleinhenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Mr. Kenneth K. Kleinhenz
forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Joshua C. Nipper -S
Binita S. Ashar, M.D., M.B.A., F.A.C.S. Acting Director For Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
510(k) Number (if known) KI32815
#### Device Name
Puregraft 50 System
#### Indications for Use (Describe)
The Puregraft 50 System is indicated for use in the harvesting, filtering of autologous fat tissue for reinjecting back into the same patient for aesthetic body contouring.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
Form Approved: OMB No. 0910-0120
Expiration Date: December 31, 2013
See PRA Statement on last page.
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.