K132813 · Toshibamedical Systems Corporation · JAK · Feb 6, 2014 · Radiology
Device Facts
Record ID
K132813
Device Name
DUAL ENERGY SYSTEM PACKAGE
Applicant
Toshibamedical Systems Corporation
Product Code
JAK · Radiology
Decision Date
Feb 6, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Real-World Evidence
Real-World Evidence
Submission
Device
Sponsor
RWD Sources
RWE Use Summary
Key Tags
K132813 · Feb 6, 2014
DUAL ENERGY SYSTEM PACKAGE
Toshibamedical Systems Corporation
Clinical patient data; Polarized microscopy analysis of surgically extracted/confirmed samples
Clinical data from patients were used to validate the device's ability to identify uric acid in gout cases and to characterize kidney stone composition, with results confirmed by polarized microscopy.
Gout; Kidney stone characterization; Polarized microscopy; Clinical validation
Clinical Evidence
Study Design
Population
Comparator
Key Endpoints
Case study
Patient with suspicion of gout in the right hand; Sample Size: 1
Not applicable for this study
Identification of uric acid confirmed by polarized microscopy
Observational clinical evaluation
Patients with kidney stones; Sample Size: 12
Not applicable for this study
Sensitivity and specificity for detecting pure uric acid stones > 3mm
Indications for Use
The Dual Energy System allows the system to acquire two CT images of the same anatomical location using distinct tube voltages and/or tube currents during two tube rotations. The x-ray dose will be the sum of the dose of each tube rotation at its respective tube voltage and current. Information regarding the material composition of various organs, tissues, and contrast materials may be gained from the differences in x-ray attenuation between these distinct energies. This information may also be used to reconstruct images at multiple energies within the available spectrum, and to reconstruct basis images that allow the visualization and analysis of anatomical and pathological materials. The visualization of the differentiation between uric acid and non-uric acid stones greater than 3mm and the visualization of uric acid presence within surrounding anatomical structures is provided with the Dual Energy system. When used by a qualified physician, a potential application is to determine the course of treatment. Performance of this device may be affected by body size and motion artifacts.
Device Story
Dual Energy System Package (CSDP-001A) is a software suite for Toshiba CT systems; acquires two CT images of identical anatomy using different tube voltages/currents across two rotations. Software processes raw data to: 1) measure CT values/ratios for material identification; 2) generate monochromatic images and contrast-enhanced visualizations; 3) extract/identify uric acid in kidney stones and surrounding tissues. Operated by qualified physicians in clinical settings; output used to inform treatment decisions. Benefits include improved material characterization, beam hardening reduction, and contrast-noise ratio (CNR) enhancement compared to single-energy scans. Performance is subject to patient body size and motion artifacts.
Clinical Evidence
Clinical testing included: 1) Gout case (n=1): confirmed uric acid via polarized microscopy. 2) Kidney stone analysis (n=12): 75% sensitivity and 83.3% specificity for detecting pure uric acid stones >3mm, confirmed by polarized microscopy. 3) Ischemia case (n=1): demonstrated visualization of ischemic bowel region via iodine map. Bench testing included phantom studies for uric acid detection, beam hardening reduction, CNR improvement, and motion artifact assessment.
Technological Characteristics
Software-based dual-energy CT processing package. Operates on Toshiba CT systems. Utilizes two-rotation acquisition at distinct tube voltages/currents. Features include monochromatic image generation, iodine mapping, and material decomposition (uric acid vs. non-uric acid). Designed under ISO 13485 and 21 CFR 820. Software level of concern: Moderate.
Indications for Use
Indicated for patients requiring CT imaging for material composition analysis of organs, tissues, and contrast materials. Specifically indicated for the differentiation of uric acid vs. non-uric acid kidney stones (>3mm) and visualization of uric acid in anatomical structures. Contraindications include limitations due to body size and motion artifacts.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
syngo Dual Energy with extended functionality (K083524)
Submission Summary (Full Text)
{0}------------------------------------------------
K132813
Page 1 of 5
FEB - 6 2014
ТОБНІВА АМЕЙІСА МЕСІСАЦ БУВТЕМБ, І№С. 2441 Michelle Drive, Tustin, CA 92780 Phone: (714) 730-5000
## 510(k) - SUMMARY OF SAFETY AND EFFECTIVENESS
- 1. SUBMITTER'S NAME: Toshiba America Medical Systems, Inc.
- 2. ADDRESS: 2441 Michelle Drive Tustin, CA 92780-2068
- 3. ESTABLISHMENT REGISTRATION: 2020563
- 4. CONTACT PERSON: Paul Biggins Director, Regulatory Affairs (714) 730-5000
# 5. DATE PREPARED:
.
January 9, 2014
## 6. TRADE NAME(S):
Dual Energy System Package, CSDP-001A
# 7. COMMON NAME:
System, X-ray, Computed Tomography
# 8. DEVICE CLASSIFICATION: Class II (per 21 CFR 892.1750)
Page 1 of 5
{1}------------------------------------------------
# 9. PRODUCT CODE / DESCRIPTION:
90JAK - System, Computed Tomography
## 10. PERFORMANCE STANDARD:
None .
## 11. PREDICATE DEVICES:
| Product | Marketed by | 510(k) Number | Clearance Date |
|--------------------------------------------------|----------------------------------------|---------------|----------------|
| Discovery CT 750HD | GE Healthcare | K120833 | June 12, 2012 |
| Brilliance Dual Energy<br>Option | Philips Medical Systems,<br>Inc. | K090462 | June 23, 2009 |
| syngo Dual Energy with<br>extended functionality | Siemens Medical Solutions<br>USA, Inc. | K083524 | April 1, 2009 |
## 12. DEVICE DESCRIPTION:
Dual Energy System Package, CSDP-001A, consists of three software packages intended to be used on Toshiba CT systems which allows for the acquisition of two CT images of the same anatomical location using different tube voltages and/or tube currents during two tube rotations.
CSDP-001A/1 (Dual Energy System) allows the same region to be scanned at two different tube voltages and tube currents and permits the CT values and ratio in the selected region to be measured based on the two image data sets obtained, providing information that is useful for identifying materials.
CSDP-001A/2 (Dual Energy Raw Data Analysis) allows monochromatic images to be generated and permits images in which contrast enhancement is visualized to be generated based on the datasets acquired by scanning the same positions with different tube voltages, making it possible to perform analysis.
CSDP-001A/3 (Dual Energy Composition Analysis) reads data acquired by scanning the same positions with different tube voltages and extracts suspected uric acids from the acquired images.
{2}------------------------------------------------
## 13. INDICATIONS FOR USE:
The Dual Energy System allows the system to acquire two CT images of the same anatomical location using distinct tube voltages and/or tube currents during two tube rotations. The x-ray dose will be the sum of the dose of each tube rotation at its respective tube voltage and current. Information regarding the material composition of various organs, tissues, and contrast materials may be gained from the differences in x-ray attenuation between these distinct energies.
This information may also be used to reconstruct images at multiple energies within the available spectrum, and to reconstruct basis images that allow the visualization and analysis of anatomical and pathological materials.
The visualization of the differentiation between uric acid and non-uric acid stones greater than 3mm and the visualization of uric acid presence within surrounding anatomical structures is provided with the Dual Energy system. When used by a qualified physician, a potential application is to determine the course of treatment.
Performance of this device may be affected by body size and motion artifacts.
## 14. SUBSTANTIAL EQUIVALENCE:
The Dual Energy System Package, CSDP-001A is substantially equivalent to the Discovery CT750HD, the Brilliance Dual Energy Option and the syngo Dual Energy with extended functionality.
Dual Energy System Package, CSDP-001A, performs in a manner similar to the predicate devices in that these systems allow for the acquisition of two CT images of the same anatomical location using different tube voltages and/or currents and when the images are combined they can be used for the visualization and further analysis of anatomical and pathological structures. The subject device accomplishes image acquisition using two rotations with one tube, rather than in one tube rotation as utilized by the predicate devices. The method of kidney stone characterization also differs from that used by Discovery CT750HD in that an image based two material decomposition method is used in the subject device, whereas, the predicate device uses a spectrum of effective atomic number.
#### 15. SAFETY:
This device is designed and manufactured under the Quality System Regulations as outlined in 21 CFR § 820 and ISO 13485 Standards.
{3}------------------------------------------------
## 16. NON-CLINICAL TESTING
Verification and validation testing, hazard analysis and performance testing conducted through bench and clinical testing are included in this submission to support the claims and indications.
## Uric Acid Analysis
Studies using a water phantom with known regions of uric acid adjacent to calcium inserts and a custom anthropomorphic hand phantom containing varying concentrations of monosodium urate were conducted to confirm that the software can provide visualization of uric acid within surrounding structures.
#### Monochromatic Imaging
#### Beam Hardening Reduction
A custom phantom designed to simulate a beam hardening artifact situation was utilized to confirm that a reduction in beam hardening artifacts was possible using the Dual Energy System Package as compared to a single energy scan.
#### Contrast-Noise Ratio (CNR) Improvement
A custom phantom containing tubes filled with varying concentrations of water/iodine was scanned using the Dual Energy System Package to generate monochromatic images. The software selected images with ideal contrast to noise ratio and it was demonstrated that, in comparison single energy scans, the monochromatic images had improved contrast-noise ratio.
## Kidney Stone Analysis
#### Kidnev Stones in Tissue
Kidney stones of known material were inserted into a representative abdominal phantom and scanned with the Dual Energy System Package. Results of the study demonstrated that the software is able to differentiate between uric acid and nonuric acid kidney stones within a representative abdominal phantom.
#### Mixed Stones
Vials consisting of pure and mixed kidney stones were inserted into a phantom and scanned using the Dual Energy System Package. Analysis of the results concluded that the pure uric acid stones within the mixture could be discriminated from the other stones.
#### Body Size
Vials consisting of kidney stones of known composition were inserted into a phantom designed to imitate various human body sizes. The phantom was scanned using the software and analysis of the results concluded that the pure uric acid stones could be discriminated from the other stones.
## Motion
Frequency of patient motion artifacts was evaluated using CT image data obtained from cases with and without breathing. The data was scored on level of motion artifacts and the results of the study concluded that the ratio of motion artifacts that could have a significant effect on the images was nearly non-existent.
{4}------------------------------------------------
K132813
Page 5 of 5
## lodine Map
A water phantom containing tubes filled with diluted contrast medium of varying concentrations was scanned using Dual Energy volume scan. lodine maps were generated by the software and it was concluded that there was no significant difference between the iodine map values and the actual iodine concentrations.
Additionally, software documentation for a Moderate Level of Concern, per the FDA guidance document, "Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices Document" issued on May 11, 2005, is also included as part of this submission.
## 17. CLINICAL TESTING
#### Uric Acid Analysis
A patient with suspicion of gout in the right hand was scanned twice using low and high tube voltage. The volumes were loaded to the Dual Energy analysis software and the resultant images displayed high-lighted regions of suspected uric acid. One of the suspected regions was surgically extracted and confirmed to be uric acid by polarized microscopy analysis.
#### Kidney Stone Analysis
A total of twelve (12) patients were scanned using a Toshiba CT system with the Dual Energy System Package. Stone analysis of the resultant images was performed and it was determined that the system was able to detect pure uric acid stones greater than 3 mm with a 75% sensitivity and an 83.3% specificity. Stone compositions were confirmed using polarized microscopy.
#### lodine Map
One representative clinical data set of a patient with suspicion of ischemia in the small bowel was acquired by Dual Energy software package and an iodine map was output by the software. Analysis of the resultant images demonstrated that visualization of the ischemic region is possible due to the lack of contrast enhancement in the lower perfusion region.
#### 18. CONCLUSION
Dual Energy System Package, CSDP-001A, performs in a manner similar to and is intended for the same use as the predicate devices. Based upon the data presented in this submission including application of design controls and performance data. Toshiba America Medical Systems, believes that Dual Energy System Package, CSDP-001A, is substantially equivalent in safety and effectiveness to the predicate devices.
{5}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an emblem featuring a stylized caduceus, a symbol often associated with healthcare, with a bird-like design incorporated into the staff.
Public Health Service
Food and Drug Administration 0903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
February 6, 2014
Toshiba Medical Systems Corporation % Mr. Paul Biggins Director, Regulatory Affairs Toshiba America Medical Systems, Inc. 2441 Michelle Drive TUSTIN CA 92780
Re: K132813
Trade/Device Name: Dual Energy System Package, CSDP-001A Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: January 9, 2014 Received: January 10, 2014
## Dear Mr. Biggins:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2-Mr. Biggins
. . . . . . . . .
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industrv/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/delault.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
for
Janine M. Morris Director, Division of Radiological Health Office of In Vitro Diagnostics and Radiological Health Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K132813
Device Name:
# Dual Energy System Package, CSDP-001A
Indications for Use:
The Dual Energy System allows the system to acquire two CT images of the same anatomical location using distinct tube voltages and/or tube currents during two tube rotations. The x-ray dose will be the sum of the dose of each tube rotation at its respective tube voltage and current. Information regarding the material composition of various organs, tissues, and contrast materials may be gained from the differences in xray attenuation between these distinct energies.
This information may also be used to reconstruct images at multiple energies within the available spectrum, and to reconstruct basis images that allow the visualization and analysis of anatomical and pathological materials.
The visualization of the differentiation between uric acid and non-uric acid stones greater than 3mm and the visualization of uric acid presence within surrounding anatomical structures is provided with the Dual Energy system. When used by a qualified physician, a potential application is to determine the course of treatment.
Performance of this device may be affected by body size and motion artifacts.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center for Devices and Radiological Health (CDRH)
Smh.7)
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.