The Fotona F-22 Handpiece is intended for: - In fractionated mode: - Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece; - In non-fractionated mode: - General Surgery Indications: Surgical incision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary; The Fotona FS-01 Handpiece is intended for: - Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece.
Device Story
The F-22 and FS-01 are laser handpiece accessories for the Fotona Dynamis Er:YAG (2940 nm) laser system. The system uses a flashlamp-pumped solid-state rod to generate laser energy, which is delivered via an articulated arm to the handpiece. A 650 nm red diode aiming beam assists the clinician. The system console, operated by a physician, allows control of laser energy and pulse duration via a footswitch. The F-22 supports both fractionated and non-fractionated modes, while the FS-01 is limited to fractionated dermatological resurfacing. The device is used in clinical settings for soft tissue ablation, coagulation, and surgical procedures. By delivering precise laser energy, the device enables controlled tissue removal and coagulation, benefiting patients through minimally invasive surgical and dermatological outcomes.
Clinical Evidence
Clinical and histological study performed using the Fotona Dynamis SP laser system with FS-01 and F-22 handpieces to demonstrate efficacy in dermatological resurfacing and soft tissue ablation. No specific quantitative metrics (e.g., sensitivity, specificity) provided.
Technological Characteristics
Er:YAG laser (2940 nm); flashlamp-pumped solid-state rod; articulated arm beam delivery; 650 nm diode aiming beam; pulsed output mode; water cooling; footswitch control. Complies with IEC 60601-1, IEC 60601-1-2, IEC 60601-1-4, IEC 60601-1-6, IEC 60601-2-22, and IEC 60825-1.
Indications for Use
Indicated for dermatological soft tissue resurfacing (fractionated mode) and general surgery applications including incision, vaporization, coagulation, tissue dissection, excision of lesions, and partial/complete resection of internal organs (non-fractionated mode).
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
Fotona Dynamis Er:YAG/Nd:YAG Laser System Family (K101306)
{0}------------------------------------------------
# 5. 510(k) Summary
# 1. SUBMITTER'S INFORMATION
Name:
Fotona d.d.
Address:
Stegne 7 1000 Ljubljana, Slovenia Phone: +386 15009100 Fax: + 386 5009 200
Contact Person:
Stojan Trošt, QA&RA Manager Phone: + 386 1 5009 299 E-mail: stojan.trost@fotona.com
Date Prepared: July 8, 2014
# 2. DEVICE INFORMATION
| Trade Name: | Fotona F-22 Laser Handpiece (F-Runner), |
|-------------|-----------------------------------------|
| | Fotona FS-01 Laser Handpiece |
Common Name: Er:YAG Surgical Laser
Classification Name: Powered laser surgical instrument with microbeam\fractional output
Product Code: ONG, GEX
# 3. PREDICATE DEVICES
Fotona Dynamis Er:YAG/Nd:YAG Laser System Family (K101306),
JOULE Multi-Platform System (K101916),
Dermablate Effect, ASCLEPION Laser Technologies GmbH (K081541)
JUL 1 1 2014 K1328DL
{1}------------------------------------------------
#### 4. DEVICE DESCRIPTION
The F-22 Handpiece and FS-01 Handpiece each attach to the Dynamis Er: Y AG/Nd: YAG Laser Systems. The Fotona Dynamis Family is based on Er:YAG (2940 nm) and Nd:YAG (1064 nm) laser technology. The F-22 handpiece and FS-01 handpiece are designed to be used with the Er:YAG (2940 nm) laser wavelength only. The laser system console consists of a flashlamp pumped Er: YAG laser source, power supply, water colling unit, electronics, and a footswitch. Electrical power is supplied to the console by the facility's power source. A red diode aiming beam (650 nm) is combined with the therapeutic Er:YAG laser beam. The combined therapeutic and aiming beams are guided through an articulated arm to an optical manual or scanner hand piece (in the case of the Er:YAG laser). Fotona's power supply and electronics, integrated into the laser system, allows control of the laser energy and the laser pulse duration. The user activates laser emission by means of a footswitch.
#### 4. INTENDED USE
The Fotona F-22 Handpiece is intended for:
- In fractionated mode: -
- Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece;
- In non-fractionated mode: ।
- General Surgery Indications: Surgical incision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary;
The Fotona FS-01 Handpiece is intended for:
- Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece.
### 5. SUMMARY OF TECHNOLOGICAL CHARACTERISTICS
The F-22 Handpiece and FS-01 Handpiece accessories to Fotona Dynamis family have the same technological and design characteristics (design, chemical composition, energy source: wavelength, active medium, cooling system, power supply, beam delivery, controls, housine) as the previously cleared devices. The output characteristics are for the intended use the same as those of the predicate devices. The risk and benefits for the F-22 Handpiece and FS-01 Handpiece accessories to Fotona Dynamis Fotona Dynamis Laser System family are identical to the predicate devices when used for similar clinical applications.
{2}------------------------------------------------
| Er: YAG<br>2940 nm | Sciton JOULE Multi-<br>Platform System Er:YAG<br>(K101916) | Fotona Dynamis Laser<br>System Family<br>(K101306) | Dermablate Effect,<br>ASCLEPION Laser<br>Technologies GmbH<br>(K081541) | Fotona Dynamis Laser<br>System Family<br>(new submission) |
|--------------------|------------------------------------------------------------|----------------------------------------------------|-------------------------------------------------------------------------|-----------------------------------------------------------|
| Wavelength | 2940 mm | 2940 nm | 2940 nm | 2940 nm |
| Laser media | Flashlamp pumped solid<br>state<br>rod | Flashlamp pumped solid<br>state<br>rod | Flashlamp pumped solid<br>state<br>rod | Flashlamp pumped solid<br>state<br>rod |
| Aiming beam | 630 - 680 nm | 650 nm | 650 nm | 650 nm |
| Output mode | Pulsed | Pulsed | Pulsed | Pulsed |
| Pulse energy | 20-8000 mJ | 30 - 3000 mJ | Up to 1500 mJ | 30 - 3000 mJ |
| Pulsewidth | 100 - 500 us | 100 - 1500 us | 400 us | 100 - 300 us |
| Repetition rate | Up to 40 Hz | up to 50 Hz | Up to 20 Hz | up 10 50 Hz |
| Power | 40W | up to 20 W | Up to 12W | up to 20 W |
| Beam Delivery | Articulated arm | Articulated arm | Articulated arm | Articulated arm |
| User interface | LCD Touchscreen | Push button control | LCD Touchscreen | Push button control |
| Handpieces | ProFractional | R08 | Asclepion Dermablate<br>Effect with MicroSpot<br>handpiece | FS-01<br>F22 |
Figurl : Comparison table of the technical characteristics for the predicate laser system and for the host laser system for the newly submitted handpieces
# 6. STATEMENT OF SUBSTANTIAL EQUIVALENCE
The F-22 Handpiece and FS-01 Handpiece are substantially equivalent in terms of indications for use and technology based on technical characteristics to the following predicate devices when used according to its intended use: Fotona Dynamis Er.YAG/Nd:YAG Laser System Family (K101306) and JOULE Multi-Platform System with Profractional handpiece (K101916), Dermablate Effect, ASCLEPION Laser Technologies GmbH (K081541).
# 7. TESTING
#### Clinical testing:
A clinical and histological study with the Fotona Dynamis SP laser system (a model from The Fotona Dynamis Er:YAG/Nd:YAG Laser System Family) using FS-01 stamping fractionated handpiece and F22 scanning fractionated handpiece has been performed to demonstrate the benefits of dermatological procedures requiring resurfacing and ablation of soft tissue with the fractionated handpieces.
{3}------------------------------------------------
#### Non-clinical performance testing
Fotona F-22 Laser Handpiece (F-Runner) and Fotona FS-01 Laser Handpieces are designed. tested and will be manufactured in accordance with both mandatory and voluntary standards including:
- 21 CFR 1040.10 Performance Standards for Light Emitting Products, Laser products .
- 21 CFR 1040.11 Performance Standards for Light Emitting Products, Specific . purpose laser products
- IEC 60601-1 Ed.3.0: 2005 + Corr 1. 2006 + Corr 2. 2007 Medical electrical . equipment. Part 1: General requirements for basic safety and essential performance
- IEC 60601-1-2 Ed.3.0: 2007 Medical electrical equipment. Part 1-2: General . requirements for basic safety and essential performance. Collateral standard: Electromagnetic compatibility-Requirements and tests
- IEC 60601-1-4 Ed. 1.1: 1996 + A1:1999 Medical electrical equipment. Part 1-4: . Collateral standard: Programmable electric medical systems.
- IEC 60601-1-6 Ed.2.0:2007 Medical electrical equipment: Part 1-6: General . requirements for basic safety and essential performance - Collateral standard: Usability
- . IEC 60601-2-22 Ed. 2.0: 1996 - Medical electrical equipment, Part 2: Particular requirements for safety diagnostic and therapeutic laser equipment.
- IEC 60825-1 Ed. 2.0 /2007 Safety of laser products. Part 1: Equipment classification . and requirements.
- EN 62304: 2006 Medical device Software software life-cycle process .
- ISO 17664:2004 Sterilization of medical devices Information to be provided by the . manufacturer for the processing of resterilizable medical devices
Laboratory testing was conducted to validate and verify that the proposed handpieces: Fotona F-22 Laser Handpiece (F-Runner) and Fotona FS-01 Laser Handpieces met all design specifications and was substantially equivalent to the predicate devices.
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three curved lines representing its wings and body. The eagle is enclosed in a circle, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is written around the perimeter of the circle.
#### DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-0002
July 11, 2014
Fotona d.d. Stojan Trost Quality Assurance & Regulatory Affairs Manager Stegne 7, 1000 Ljubljana, Slovenia
Re: K132806 Trade/Device Name: Fotona F-22 Laser Hand piece (F-Runner), Fotona FS-01 Laser Handpiece Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: Class II Product Code: ONG, GEX Dated: June 12, 2014 Received: June 16, 2014
Dear Stojan Trost:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you; however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA 's issuance of a substantial-equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21
{5}------------------------------------------------
Page 2 - Stojan Trost
CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/Resourcesfor You/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Industry and Consumer Education at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# David Krause -S
- for Binita S. Ashar, M.D., M.B.A., F.A.C.S. Director Division of Surgical Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
#### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
# Indications for Use
Form Approved: OMB No, 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
510(k) Number (if known) K132806
Device Name
Fotona F-22 (f-runner), Fotona FS-0) laser handpieces
Indications for Use (Describe)
The Fotona F-22 Handpiece is intended for:
. In fractionated mode:
Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece;
: ` In non-fractionated mode:
General Surgery Indications: Surgical incision, vaporization and coagulation of soft tissue during any general surgery application where skin incision, tissue dissection, excision of lesions, complete or partial resection of internal organs, lesions, tissue ablation and vessel coagulation is necessary;
The Fotona FS-01 Handpiece is intended for:
Dermatological procedures requiring resurfacing of soft tissue with fractionated handpiece.
Type of Use (Select one or both, as applicable)
2 Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON A SEPARATE PAGE IF NEEDED.
FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
Neil R Ogden -S 2014.07.10 16:45:17 -04'00'
{7}------------------------------------------------
This section applies only to requirements of the Paperwork Reduction Act of 1995. "DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per resoonse, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.