K132802 · Shanghai Motex Healthcare Co., Ltd. · KGO · May 21, 2014 · General, Plastic Surgery
Device Facts
Record ID
K132802
Device Name
MOTEX POWDERED NITRILE SURGICAL GLOVES
Applicant
Shanghai Motex Healthcare Co., Ltd.
Product Code
KGO · General, Plastic Surgery
Decision Date
May 21, 2014
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4460
Device Class
Class 1
Indications for Use
The Motex Powdered Nitrile Surgical Gloves are sterile disposable devices made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. The lubricating or dusting powder used in the glove is excluded.
Device Story
Motex Powdered Nitrile Surgical Gloves are sterile, single-use, synthetic rubber (nitrile) gloves designed for use by operating room personnel. The gloves function as a physical barrier to prevent cross-contamination between the wearer and the surgical wound. They are provided in white and green color variants and various sizes (6 to 8.5). The device is worn manually by clinicians during surgery. The gloves are sterilized via radiation. The device benefits the patient by reducing the risk of surgical site contamination.
Clinical Evidence
No clinical data. Bench testing only, including physical property testing (ASTM D3577-09e1), water leak testing (ASTM D5151-06), residual powder testing (ASTM D6124-06), water extractable protein testing (ASTM D5712-10), and biocompatibility testing (ISO 10993-10:2010, ISO 10993-12:2007).
Indicated for operating room personnel to protect surgical wounds from contamination during surgical procedures. Not for use by patients.
Regulatory Classification
Identification
A non-powdered surgeon's glove is a device intended to be worn on the hands of operating room personnel to protect a surgical wound from contamination. A non-powdered surgeon's glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
{0}------------------------------------------------
## SHANGHAI MOTEX HEALTHCARE CO., LTD.
No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China Telephone: 86-21-5979 9888 Fax: 86-21-23010718
MAY 2 1 2014
## II. 510(K) Summary of Safety and Effectiveness (Per 21 CFR 807.92)
## 2.1. General Information Establishment
- I Manufacturer: Shanghai Motex Healthcare Co., Ltd.
- Address: No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China .
- 그 Owner Number: 9041164 Registration Number: 9615978
- Contact Person: Dr. Jen, Ke-Min E-mail: ceirs.jen@msa.hinet.net Tel: +886-3-5208829 ; Fax: +886-3-5209783
- Date Prepared: May 20, 2014
## Device
- Proprietary Name: Motex Powdered Nitrile Surgical Gloves, Item 6010 White Motex Powdered Nitrile Surgical Gloves, Item 6010 Green
- . Common Name: Surgeon's glove
- . Surgeon's Gloves Classification Name:
- . Product Code: KGO
● Regulation Number: Class I, 878.4460
## 2.2. Safety and Effectiveness Information
### ● Predicate Device:
Claim of Substantial Equivalence (SE) is made to NUZONE Nitrile Powdered Surgical Gloves (K000178)
#### ● Device Description:
Motex Powdered Nitrile Surgical Gloves are made of synthetic rubber. They are sterilized by radiation, and intended to be used in surgery to prevent the cross contamination between patient and user.
#### . Indications for Use:
The Motex Powdered Nitrile Surgical Gloves are sterile disposable devices made of synthetic rubber intended to be worn by operating room personnel to protect a · surgical wound from contamination. The lubricating or dusting powder used in the glove is excluded.
### . Device Characteristics:
Single use only. Not made with natural rubber latex.
{1}------------------------------------------------
No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China Telephone: 86-21-5979 9888 Fax: 86-21-23010718
### ● Technological Characteristics:
Motex Powdered Nitrile Surgical Gloves characteristics are summarized below compared to ASTM and ISO standards to the predicate device:
| Characteristic | Standard |
|-----------------------------------|-------------------|
| Dimensions | ASTM D3577-09e1 |
| Physical Properties | ASTM D3577-09e1 |
| Freedom from Holes | ASTM D3577-09e1 |
| Water leak testing | ASTM D5151-06 |
| Residual powder testing | ASTM D6124-06 |
| Water extractable protein testing | ASTM D5712-10 |
| Biocompatibility | ISO 10993-10:2010 |
| | ISO 10993-12:2007 |
| Sterilization Validation | ISO 11137-1:2006 |
| | ISO 11137-2:2006 |
### ● Powder Residual:
Surgeon's gloves meet powder level requirements for "Powder" designation per ASTM D6124-06, Standard test method for residual powder on medical gloves. The results generated values will below 15mg/dm2 of residual powder per glove.
#### . Biocompatibility Test Reports:
According to ISO 10993 series standards, we completed the biological evaluation and the results of these studies show that the Motex Powdered Nitrile Surgical Gloves safe for their intended use. Besides, we also completed the following reports for each item 6010_White and 6010_Green including:
- Skin Sensitization Test (Maximization test), Sesame oil extract;
- . Skin Sensitization Test (Maximization test), Sodium Chloride extract;
- � Skin Irritation Test, Sesame oil extract;
- Skin Irritation Test, Sodium Chloride extract;
#### . Clinical Data:
Not applicable.
{2}------------------------------------------------
# SHANGHAI MOTEX HEALTHCARE CO., LTD.
No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China
Telephone: 86-21-5979 9888 Fax: 86-21-23010718
## Comparison Table
| Similarity: | | | |
|----------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|
| Comparison Feature | Predicate device | Subject device | |
| Device name | Nuzone Nitrile Surgical Gloves,<br>Powdered, Green<br>(K000178) | Motex Powdered Nitrile Surgical<br>Gloves<br>(K132802) | |
| Indications For Use | The surgical glove is a device made<br>of synthetic rubber intended to be<br>worn by operating room personnel<br>to protect a surgical wound from<br>contamination. | The Motex Powdered Nitrile<br>Surgical Gloves are sterile<br>disposable devices made of<br>synthetic rubber intended to be worn<br>by operating room personnel to<br>protect a surgical wound from<br>contamination. The lubricating or<br>dusting powder used in the glove is<br>excluded. | |
| Prescription/OTC<br>Device | Over-the-Counter Use | Over-the-Counter Use | |
| Product code | KGO, Class I, 878.4460 | KGO, Class I, 878.4460 | |
| Device Design | They are coated with nitrile coating<br>and are offered powdered and<br>sterile. | They are coated with nitrile coating<br>and are offered powdered and<br>sterile. | |
| Major Material<br>Composition | Nitrile and<br>Absorbable dusting powder<br>. - | Nitrile: Acrylonitrile- butadiene<br>rubber latex; i.e. NBR<br>(CASH<br>9003-18-3)<br>Absorbable dusting powder USP<br>(FDA PMA Number: P890070) | |
| Performance Test<br>for Pinhole,<br>Dimensions, and<br>Physical properties | Meets ASTM D3577-09 | Meets ASTM D3577-09 | |
| Biocompatibility | Primary Skin irritation test<br>ASTM F 719-81 and Dermal<br>Sensitization Test ASTM<br>F720-81, indicates "under the<br>conditions of the study, not an<br>irritant" and "under<br>conditions of the study, not a<br>sensitizer. " | Meets ISO 10993-10:2010/<br>ISO 10993-12:2007<br>Gloves are "under the conditions<br>of the study, not an irritant" and<br>"under conditions of the study,<br>not a sensitizer. " | |
| Sterilization Validation | Sterile Nitrile Powdered Surgical<br>Gloves | Meets ISO 11137-1:2006,<br>ISO 11137-2:2006 | |
| Powder Residual | N/A | Meets ASTM D6124-06<br>Results generated values below<br>15mg/dm² of residual powder per<br>glove | |
| Water Extractable<br>Protein Testing | N/A | Meets ASTM D5712-10 | |
{3}------------------------------------------------
## SHANGHAI MOTEX HEALTHCARE CO., LTD.
No.369, Jiasong Zhong Road, Huaxin, Qingpu, Shanghai, 201708, China Telephone: 86-21-5979 9888 Fax: 86-21-23010718
### Difference:
| Product Trade Name | Nuzone Nitrile Surgical Gloves,<br>Powdered, Green | Motex Powdered Nitrile Surgical<br>Gloves<br>6010 White and 6010 Green |
|--------------------|----------------------------------------------------|--------------------------------------------------------------------------|
| Specification size | N/A | 6, 6.5, 7, 7.5, 8, 8.5 |
#### . Substantial Equivalence (SE)
A claim of substantial equivalence is made to NUZONE Nitrile Powdered Surgical Gloves (K000178). Both of the subject device and predicate device are the same Powdered Surgical Gloves which have the similar indications for use. material composition with Nitrile, meet the performance tests by ASTM D3577-09, and also completed the biocompatibility, sterilization validation test reports. The major differences of the two devices are the different colorants and specification. That just means the subject devices have two colorants: white and green, and the predicate device has only one green colorant; and the specifications of the subject device are limited on the 6~8.5 sizes. These differences are not relating to the safety or effectiveness aspects. Thus they are substantially equivalent.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized symbol that resembles a caduceus, a traditional symbol of medicine, but with three abstract shapes instead of snakes.
Public Health Service
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring MD 20993-0002
### May 21, 2014
Shanghai Motex Healthcare Company, Limited C/O Dr. Ke-Min Jen Official Correspondent No. 369, Jiasong Zhong Road Huaxin, Oingpu, Shanghai CHINA 201708
Re: K132802
Trade/Device Name: Motex Powdered Nitrile Surgical Gloves, Item 6010 White Motex Powdered Nitrile Surgical Gloves, Item 6010 Green Regulation Number: 21 CFR 878.4460 Regulation Name: Surgeon's Glove Regulatory Class; 1 Product Code: KGO Dated: April 19, 2014 Received: April 24, 2014
## Dear Dr. Jen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Page 2 - Dr. Jen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers. International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note
the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/Medica|Devices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
## Mary S. Runner -S
Erin I. Keith, M.S. Acting Director Division of Anesthesiology, General Hospital, Respiratory, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### DEPARTMENT OF HEALTH AND HUMAN SERVICES Food and Drug Administration
## Indications for Use
510(k) Number (if known) K132802
Davice Name
Motex Powdered Nitrile Surgical Gloves, Item 6010 White Motex Powdered Nitrile Surgical Gloves, Item 6010 Green
### Indications for Use (Describe)
The Motex Powdered Nitrile Surgical Gloves are sterile disposable devices made of synthetic rubber intended to be worn by operating room personnel to protect a surgical wound from contamination. The lubricating or dusting powder used in the glove is excluded.
Type of Use (Select one or both, as applicable)
_ Prescription Use (Part 21 CFR 801 Subpart D)
Over-The-Counter Use (21 CFR 801 Subpart C)
## PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON A SEPARATE PAGE IF NEEDED.
### FOR FDA USE ONLY
Concurrence of Center for Devices and Radiological Health (CDRH) (Signature)
# Sreekanth Gutala
Digitally signed by Sreekanth Gutala -S
DN: c=US, o=U.S. Government, ou=HHS, ou=FDA, ou=People, 0.9.2342.19200300.100.1.1=2000540490, cn=Sreekanth Gutala -S Date: 2014.05.21 12:28:53 -04'00'
This section applies only to requirements of the Paperwork Reduction Act of 1995.
## *DO NOT SEND YOUR COMPLETED FORM TO THE PRA STAFF EMAIL ADDRESS BELOW."
The burden time for this collection of information is estimated to average 79 hours per response, including the time to review instructions, search existing data sources, gather and maintain the data needed and complete and review the collection of information. Send comments regarding this burden estimate or any other aspect of this information collection, including suggestions for reducing this burden, to:
> Department of Health and Human Services Food and Drug Administration Office of Chief Information Officer Paperwork Reduction Act (PRA) Staff PRAStaff@fda.hhs.gov
"An agency may not conduct or sponsor, and a person is not required to respond to, a collection of information unless it displays a currently valid OMB number."
Form Approved: OMB No. 0910-0120 Expiration Date: January 31, 2017 See PRA Statement below.
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.