K132796 · Huizhou Foryou Medical Devices Co., Ltd. · HOZ · Oct 30, 2013 · Ophthalmic
Device Facts
Record ID
K132796
Device Name
SUNTOUCH EYE SPEAR
Applicant
Huizhou Foryou Medical Devices Co., Ltd.
Product Code
HOZ · Ophthalmic
Decision Date
Oct 30, 2013
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 886.4790
Device Class
Class 2
Indications for Use
The Eye Spears are designed to absorb fluids and remove debris from the operative field or instruments during ophthalmic surgery.
Device Story
SUNTOUCH® Eye Spear is a sterile, single-use ophthalmic sponge device. It consists of an expansive polyvinyl alcohol (PVA) sponge attached to a polypropylene (PP) handle. Used by surgeons during ophthalmic procedures, the device is manually operated to absorb excess fluids and clear debris from the surgical site or instruments. The device functions via passive capillary action of the sponge material. It is provided sterile (SAL 10-6) and is intended for use in an operating room environment.
Clinical Evidence
No clinical data. Substantial equivalence is supported by bench testing, including biocompatibility (ISO 10993-5, ISO 10993-10), package integrity (ASTM F88, F1140, F1929), and physical performance testing (absorbency per BS EN 13726-1, particulate matter per USP <789>, wicking rate, and mechanical strength).
Technological Characteristics
Materials: Polyvinyl alcohol (PVA) sponge, polypropylene (PP) handle. Principle: Passive fluid absorption via capillary action. Dimensions: Various configurations. Sterilization: Radiation (SAL 10-6). Standards: ISO 10993-5/10, ASTM F88/F1140/F1929, BS EN 13726-1, USP <85>/<789>.
Indications for Use
Indicated for use during ophthalmic surgery to absorb fluids and remove debris from the operative field or instruments.
Regulatory Classification
Identification
An ophthalmic sponge is a device that is an absorbant sponge, pad, or spear made of folded gauze, cotton, cellulose, or other material intended to absorb fluids from the operative field in ophthalmic surgery.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.
{0}------------------------------------------------
### 510(k) Summary
ﺎﺭ ﮨﮯ
This 510(k) Summary of 510(k) safety and effectiveness information is being submitted in accordance with requirements of SMDA 1990 and Title 21, CFR Section 807.92.
K132796 The assigned 510(k) Number: _
- Date of Submission: 08/20/2013 1.
## OCT 3 0 2013
- Sponsor Identification 2.
Huizhou Foryou Medical Devices Co., Ltd. North Shangxia Rd., Dongjiang Hi-tech Industry Park, 516005, Huizhou, P. R. China.
Establishment Registration Number: 3007735241
Contact Person: Wei Lu Position: Development Engineer Tel: 86 752 530201 3 Fax: 86 752 5302020 Email: wlu@foryougroup.com
3. Submission Correspondent
Ms. Diana Hong & Mr. Lee Fu Mid-Link Consulting Co., Ltd P.O. Box 120-119 Shanghai, 200120, China Tel: +86-21-22815850 Fax: 240-238-7587 . Email: info@mid-link.net
{1}------------------------------------------------
- 4. Proposed Device Identification
Proposed Device Name: SUNTOUCH® Eye Spear Proposed Device Model: ES1001A-P, ES1001A-P, ES1002A-B, ES1002A-P, ES1002A-P, ES1002B, ES1003B
Proposed Device Common Name: Eye Spear
Regulatory Information: Classification Name: Sponge, Ophthalmic; Classification: II; Product Code: HOZ; Regulation Number: 886.4790; Review Panel: Ophthalmic;
Intended Use Statement: The Eye Spears are designed to absorb fluids and remove debris from the operative field or instruments during ophthalmic surgery.
- 5. Predicate Device Identification
510(k) Number: K002279 Product Name: ML Eye Spear Manufacturer: Med-Logics, Inc
- Device Description 6.
SUNTOUCH® Eye Spears are sterilized, single-use devices designed to absorb fluids and remove debris from the operative field or instruments during ophthalmic surgery. The Eye Spears include seven models which share same material. The models are different in size and configuration of plastic handle.
They are provided sterilized with Sterility Assurance Level (SAL) of 10-6.
- 7. Non-Clinical Test Conclusion
Non clinical tests were conducted to verify that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device. The test results demonstrated that the proposed device complies with the following standards:
ISO 10993-5: 2009, Biological evaluation of medical devices -- Part 5: Tests for In Vitro cytotoxicity. ISO 10993-10: 2010, Biological evaluation of medical devices - Part 10: Tests for irritation and skin 3-2
{2}------------------------------------------------
sensitization.
ASTM F88/F88M-09, Standard Test Method for Seal Strength of Flexible Barrier Materials.
ASTM FI140-07 (Reapproved 2012), Standard Test Methods for Internal Pressurization Failure Resistance of Unrestrained Packages.
ASTM F1929-12, Standard Test Method for Detecting Seal Leaks in Porous Medical Packaging by Dye Penetration.
BS EN 13726-1: 2002, Test methods for primary wound dressings -Part 1: Aspects of absorbency USP 35-NF30:2012, <85> Bacterial Endotoxins Test.
USP 35-NF30:2012, < 789> Particulate Matter in Ophthalmic Solutions
And following physical properties tests were conducted: absorption capacity, release of particulates, wicking rate. pore size, dry density, pH and formaldehyde residues in aqueous extracts, mechanical strength of sponge-handle attachment, and visual inspection for rough edges. The test results demonstrated that the proposed device met all design specifications as was Substantially Equivalent (SE) to the predicate device.
#### Substantially Equivalent (SE) Conclusion 8.
The following table compares the DEVICE to the predicate device with respect to intended use, technological characteristics and principles of operation, etc.
| Item | Proposed Device(s) | Predicate Device(s) |
|--------------------|-----------------------------------------------------------------------------------------------------------------------------------------|---------------------|
| Product Code | HOZ | Same |
| Regulation Number | 21 CFR 886.4790 | Same |
| Intended Use | The Eye Spears are designed to absorb fluids and<br>remove debris from the operative field or<br>instruments during ophthalmic surgery. | Same |
| Configuration | Expansive sponge and plastic handle | Same |
| Performance | Absorbency tested per BS EN 13726-1<br>Microscopic Particle Count tested per USP <789> | Similar |
| Material | Expansive sponge: PVA | Same |
| | Handle: PP | Same |
| Biocompatibility | Comply with ISO 10993-5 and ISO 10993-10. | Same |
| Single Use | Yes | Same |
| Sterilization | Method: Radiation<br>SAL: 10-6 | Same |
| Shelf Life | 5 years | Same |
| Label and Labeling | Meet FDA's Requirements | Same |
Table 3-1 Comparison of Technology Characteristics
{3}------------------------------------------------
The proposed device, SUNTOUCH® Eye Spear, is determined to be Substantially Equivalent (SE) to the predicate device(s), ML Eye Spear (K002279), in respect of safety and effectiveness.
.
.
-
.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of two main elements: a circular text element and a symbol. The text element reads "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" in a circular arrangement. The symbol is a stylized representation of a human figure, with three overlapping profiles suggesting a sense of community and support.
October 30, 2013
Food and Drug Administration 10903 New Hampshire Avenue Document Control Center - WO66-G609 Silver Spring, MD 20993-002
Huizhou Foryou Medical Devices Co., Ltd. c/o Ms. Diana Hong Mid-Link Consulting Co., Ltd. P.O. BOX 120-119. Shanghai, 200120, China
Re: K132796 Trade/Device Name: Suntouch® Eye Spear Regulation Number: 21 CFR 886.4790 Regulation Name: Sponge, Ophthalmic Regulatory Class: Class II Product Code: HOZ Dated: August 28, 2013 Received: September 6, 2013
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. Please note: CDRH does not evaluate information related to contract liability warranties. We remind you, however, that device labeling must be truthful and not misleading.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: Registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); medical device reporting (reporting of medical device-related adverse events) (21 CFR 803); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Ms. Diana Hong
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Division of Small Manufacturers, International and Consumer Assistance at its tollfree number (800) 638-2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding the reporting of adverse events under the MDR regulation (21 CFR Part 803), please go to
http://www.fda.gov/MedicalDevices/Safety/ReportaProblem/default.htm for the CDRH's Office of Surveillance and Biometrics/Division of Postmarket Surveillance.
You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638 2041 or (301) 796-7100 or at its Internet address
http://www.fda.gov/MedicalDevices/ResourcesforYou/Industry/default.htm.
Sincerely yours,
# Deborah L. Falls -S
for Malvina B. Eydelman, M.D. Director Division of Ophthalmic. Ear. Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
### Indications for Use
510(k) Number: K132796
Device Name: Device Name: SUNTOUCH® Eye Spear
Indications for Use:
The Eye Spears are designed to absorb fluids and remove debris from the operative field or instruments during ophthalmic surgery.
Prescription Use_ × (part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use
(21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE- CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of Center of Devices and Radiological Health (CDRH)
| Leonid Livshitz-S |
|-------------------|
| 2013:10:24 |
| '00'04-09:17:34 |
(Division Sign-Off) Division of Ophthalmic and Ear, Nose, and Throat Devices 510(k) Number:________________________________________________________________________________________________________________________________________________________________
page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.